K251088 · Otsuka America Pharmaceutical, Inc. · OZW · Jun 30, 2025 · General Hospital
Device Facts
Record ID
K251088
Device Name
Otsuka Digital Feedback Device
Applicant
Otsuka America Pharmaceutical, Inc.
Product Code
OZW · General Hospital
Decision Date
Jun 30, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6305
Device Class
Class 2
Indications for Use
The Otsuka Digital Feedback Device consists of a miniaturized, wearable sensor for ambulatory recording of physiological and behavioral metrics such as heart rate, activity, body angle relative to gravity (body position), and time-stamped patient-logged events, including events signaled by co-incidence with, or co ingestion with, the ingestible sensor accessory. When the ingestible sensor is ingested, the Otsuka Digital Feedback Device is intended to log, track, and trend intake times. When co-ingested with medication, the tracking and trending of intake times may be used as an aid to measure medication adherence. The Otsuka Digital Feedback Device may be used in any instance where quantifiable analysis of event-associated physiological and behavioral metrics is desirable and enables unattended data collection for clinical and research applications.
Device Story
System comprises wearable patch (D-Tect), ingestible sensor, and software. Ingestible sensor activates in stomach, transmitting unique ID to patch. Patch collects HR, activity, body angle, and event logs. Data transmitted via Bluetooth to general computing device for processing, display, and storage. Used in home healthcare; operated by patient. Healthcare providers review logged intake times and physiological trends to assess medication adherence. Benefits include objective, unattended monitoring of medication intake and behavioral metrics. Device does not provide ECG waveforms, diagnostic functions, or alarms.
Clinical Evidence
Noninterventional validation study (n=55 healthy subjects) evaluated wear duration, comfort, and safety. 51 subjects completed study. Results: 79.0% adherence for ≥5 days; 61.5% adherence for ≥7 days. Average comfort rating 4.7/5.0. Safety: 12.7% (7/55) experienced Grade 1 or 2 skin irritation; no new safety concerns identified.
Technological Characteristics
Wearable patch with 3-axis accelerometer, thermistor, and low-noise AFE. Single-microprocessor architecture. Dimensions: 6.4mm x 113.2mm x 45.2mm. Materials: MED5741 skin adhesive, Axelgaard AG625 hydrogel. Connectivity: Bluetooth. Power: Internal battery. Software: Firmware for data acquisition/encryption; mobile app for display/storage. Biocompatibility: ISO 10993 compliant. Electrical safety: IEC 60601-1/1-2/1-11/1-6 compliant. IP27 ingress protection.
Indications for Use
Indicated for ambulatory recording of physiological/behavioral metrics (HR, activity, body position, patient-logged events) and tracking/trending of ingestible sensor intake times to aid medication adherence measurement in clinical/research settings.
Regulatory Classification
Identification
An ingestible event marker is a prescription device used to record time-stamped, patient-logged events. The ingestible component links wirelessly through intrabody communication to an external recorder which records the date and time of ingestion as well as the unique serial number of the ingestible device.
Special Controls
In combination with the general controls of the FD&C Act, the Proteus Personal Monitor including Ingestion Event Marker is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible and non-toxic;
(2) Nonclinical, animal, and clinical testing must provide a reasonable assurance of safety and effectiveness, including device performance, durability, compatibility, usability (human factors testing), event recording, and proper excretion of the device;
(3) Appropriate analysis and nonclinical testing must validate electromagnetic compatibility performance, wireless performance, and electrical safety; and
(4) Labeling must include a detailed summary of the nonclinical and clinical testing pertinent to use of the device and the maximum number of daily device ingestions.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 30, 2025
Otsuka America Pharmaceutical, Inc.
% Nancy Teague
Senior Director, Global Regulatory Affairs
Otsuka Product Development & Commercialization, Inc.
2440 Research Blvd
Rockville, Maryland 20850
Re: K251088
Trade/Device Name: Otsuka Digital Feedback Device
Regulation Number: 21 CFR 880.6305
Regulation Name: Ingestible Event Marker
Regulatory Class: Class II
Product Code: OZW, DXH
Dated: April 3, 2025
Received: April 9, 2025
Dear Nancy Teague:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251088 - Nancy Teague
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251088 - Nancy Teague
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer W. Shih -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Otsuka Digital Feedback Device
Page 11 of 47
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251088 | ? |
| Please provide the device trade name(s). | | ? |
| Otsuka Digital Feedback Device | | |
| Please provide your Indications for Use below. | | ? |
| The Otsuka Digital Feedback Device consists of a miniaturized, wearable sensor for ambulatory recording of physiological and behavioral metrics such as heart rate, activity, body angle relative to gravity (body position), and time-stamped patient-logged events, including events signaled by co-incidence with, or co ingestion with, the ingestible sensor accessory. When the ingestible sensor is ingested, the Otsuka Digital Feedback Device is intended to log, track, and trend intake times. When co-ingested with medication, the tracking and trending of intake times may be used as an aid to measure medication adherence. The Otsuka Digital Feedback Device may be used in any instance where quantifiable analysis of event-associated physiological and behavioral metrics is desirable and enables unattended data collection for clinical and research applications. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K251088
## 12.5 510(k) Summary
Submitted by: Otsuka America Pharmaceutical, Inc.
Address: 3956 Point Eden Way
Hayward, CA 94545
Telephone: 240-683-3560
Contact Name: Nancy F. Teague
Senior Director, Global Regulatory Affairs
nancy.teague@otsuka-us.com
Name of Device: Otsuka Digital Feedback Device
Tradename: Otsuka Digital Feedback Device
Common Name: Ingestible Event Marker with wearable patch
Classification Name: Ingestible Event Marker, 21 CFR 880.6305
Product Code: OZW
Predicate Device: Otsuka Digital Feedback Device-RW, K223463
Page 1 of 12
{5}
K251088
## 12.5.1 General Device Description
The Otsuka Digital Feedback Device consists of 3 components: a wearable sensor, an ingestible sensor accessory, and software that aggregates, processes, and enables display of data collected by the sensors.
The wearable sensor in the Otsuka Digital Feedback Device is a body-worn sensor that consists of a single-use patch known as the D-Tect wearable sensor or D-Tect Patch. The D-Tect Patch collects physiological and behavioral metrics such as heart rate, activity, body angle, and time stamped patient-logged events, including events signaled by the co-incidence with, or co-ingestion with, the ingestible sensor accessory. Note: While the device includes automated heart rate (HR) measurement, it does not provide an ECG waveform recording for display or analysis. The device is not intended to diagnose heart-related conditions and does not include alarms. HR measurement is not intended to be used in alarm system. HR data may not be accurate for patients with pacemakers.
The ingestible sensor is embedded inside an inactive tablet (the pill or sensor-enabled pill) for ease of handling and swallowing. After the ingestible sensor reaches the stomach, it activates and communicates its presence with a unique identifier to the wearable sensor. When the ingestible sensor is co-ingested with medication, the Otsuka Digital Feedback Device is intended to log, track, and trend medicine intake times to measure medication adherence.
The software on the general computing device receives the data from the wearable sensor for further processing and analysis of metrics. The processed data is then sent to the user interface for display, as well as being saved in a local record database for storage. The software component includes firmware running on the wearable sensor that collects and records the data from the sensors and a software that receives data from the wearable sensor for further analysis, processing, storage, and display to the user.
For purposes of this 510(k), the changes from the predicate device (Otsuka Digital Feedback Device-RW, K223463) to the Otsuka Digital Feedback Device with D-Tect Patch (the device subject of this 510[k]) are the wearable sensor and the software for the user interface. The ingestible sensor remains the same as the predicate device.
## 12.5.2 Indications for Use
Page 2 of 12
{6}
K251088
The Otsuka Digital Feedback Device consists of a miniaturized, wearable sensor for ambulatory recording of physiological and behavioral metrics such as heart rate, activity, body angle relative to gravity (body position), and time-stamped patient-logged events, including events signaled by co-incidence with, or co-ingestion with, the ingestible sensor accessory. When the ingestible sensor is ingested, the Otsuka Digital Feedback Device is intended to log, track, and trend intake times. When co-ingested with medication, the tracking and trending of intake times may be used as an aid to measure medication adherence. The Otsuka Digital Feedback Device may be used in any instance where quantifiable analysis of event-associated physiological and behavioral metrics is desirable and enables unattended data collection for clinical and research applications.
Page 3 of 12
{7}
K251088
# 12.5.3 Technological Characteristics and Predicate Comparison
Technical characteristics and predicate comparison are shown in Table 12.5.3-1.
| Table 12.5.3-1 Technological Characteristics and Predicate Comparison | | |
| --- | --- | --- |
| | Subject of this 510(k) | Predicate Device |
| Trade Name | Otsuka Digital Feedback Device | Otsuka Digital Feedback Device-RW |
| Common Name | Otsuka Digital Feedback Device (with D-Tect Patch) | Otsuka Digital Feedback Device-RW with RW2 Patch |
| 510(k) Number | K251088 | K223463 |
| Indications for Use | The Otsuka Digital Feedback Device consists of a miniaturized, wearable sensor for ambulatory recording of physiological and behavioral metrics such as heart rate, activity, body angle relative to gravity (body position), and time-stamped patient-logged events, including events signaled by the co-incidence with, or co-ingestion with, the ingestible sensor accessory. When the ingestible sensor is ingested, the Otsuka Digital Feedback Device is intended to log, track, and trend intake times. When co-ingested with medication, the tracking and trending of intake times may be used as an aid to measure medication adherence. The Otsuka Digital Feedback Device may be used in any instance where quantifiable analysis of event-associated physiological and behavioral metrics is desirable and enables unattended data collection for clinical and research applications. | The Otsuka Digital Feedback Device-RW consists of a miniaturized, wearable sensor for ambulatory recording of physiological and behavioral metrics such as heart rate, activity, body angle relative to gravity (body position), and time-stamped patient-logged events, including events signaled by the co-incidence with, or co-ingestion with, the ingestible sensor accessory. When the ingestible sensor is ingested, the Otsuka Digital Feedback Device-RW is intended to log, track, and trend intake times. When co-ingested with medication, the tracking and trending of intake times may be used as an aid to measure medication adherence. The Otsuka Digital Feedback Device-RW may be used in any instance where quantifiable analysis of event-associated physiological and behavioral metrics is desirable and enables unattended data collection for clinical and research applications. |
| Common Name | IEM | IEM |
| Classification Name | IEM | IEM |
| Product Code | OZW | OZW |
| Subsequent Product Codes | DXH | DXH |
| Event Marker, Co-ingestion with: | IEM | IEM |
| Product Name | OZW | OZW |
| Product Code | OZW | OZW |
| Subsequent Product Codes | DXH | DXH |
{8}
K251088
| Table 12.5.3-1 Technological Characteristics and Predicate Comparison | | |
| --- | --- | --- |
| | Subject of this 510(k) | Predicate Device |
| Trade Name | Otsuka Digital Feedback Device | Otsuka Digital Feedback Device-RW |
| Common Name | Otsuka Digital Feedback Device (with D-Tect Patch) | Otsuka Digital Feedback Device-RW with RW2 Patch |
| 510(k) Number | K251088 | K223463 |
| Temperature Sensor | Thermistor | Thermistor |
| Accelerometer | 3-axis accelerometer | 3-axis accelerometer |
| Hardware (Ingestible Sensor) | Ingestible event marker | Ingestible event marker |
| Hardware (Wearable Sensor) | Wearable, physiologic sensor (1-component wearable sensor or patch). There is no reuse of this patch. Single Microprocessor Architecture Low noise AFE | Wearable, physiologic sensor (2-component wearable sensor or patch). Dual Microprocessor Architecture Low noise AFE |
| Wear Location | Front torso left and right sides | Front torso left and right sides |
| Mechanical (Wearable Sensor) Device dimensions | Smaller, more comfortable design with a curved membrane connecting the 2 electrodes. 6.4 mm 113.2 mm 45.2 mm | 8.2 mm 110 mm 53 mm |
| Mechanical (Wearable Sensor) Layer Construction | Comparable skin adhesive from the same manufacturer and identical hydrogel material. Skin adhesive MED5741 Hydrogel Axelgaard AG625 | Skin adhesive MED5750A Hydrogel Axelgaard AG625 |
| Firmware (Wearable Sensor) | Collects data from sensor, performs data processing, stores data in the encrypted format until wireless connection is available, sends data to general computing device. | Collects data from sensor, performs data processing, stores data until wireless connection is available, sends data to general computing device. |
Page 5 of 12
{9}
K251088
| Table 12.5.3-1 Technological Characteristics and Predicate Comparison | | |
| --- | --- | --- |
| | Subject of this 510(k) | Predicate Device |
| Trade Name | Otsuka Digital Feedback Device | Otsuka Digital Feedback Device-RW |
| Common Name | Otsuka Digital Feedback Device (with D-Tect Patch) | Otsuka Digital Feedback Device-RW with RW2 Patch |
| 510(k) Number | K251088 | K223463 |
| Improved Pill Detection Capability | Enhanced low-noise analog front end improves sensitivity of the IEM detection by >50%, increasing the likelihood of detection. | IEM detection with the ability to detect signals with lower (nV) amplitudes. |
| Improved Step Counting | Improved step metric based on passive fully continuous step counting. | Periodic step counting using raw motion data sampled briefly each minute to estimate steps per minute. |
| Improved Heart Rate Measurements | Verified against real ECG signals and 100% measurement accuracy for simulated ECG signal. | Verified against simulated ECG signals with >90% measurement accuracy. |
| Software | Aggregates, decrypts, processes, and enables display of data collected by sensors | Aggregates, processes, and enables display of data collected by sensors |
| Cybersecurity | Improved cybersecurity and data privacy. Data encrypted while at rest and in flight. Increased security during Bluetooth pairing, over-the-air upgrade with image signatures and rollback protection. | No encryption of data at rest on the patch, with basic Bluetooth pairing security. |
| Data Telemetry | Bluetooth Technology | Bluetooth Technology |
| Biocompatibility | ISO 10993 compliant | ISO 10993 compliant |
| Electrical Safety/EMC | IEC 60601-1/IEC 60601-1-2 compliant | IEC 60601-1/IEC 60601-1-2 compliant |
AFE = analog front end; D-Tect = Otsuka Digital Feedback Device body-worn sensor;
DW5 = Disposable Wearable 5, earlier generation of wearable sensor (predicate);
EMC = electromagnetic compatibility; IEC = International Electrotechnical Commission;
IEM = ingestible event marker; ISO = International Organization for Standardization; RW = Reusable Wearable; RW2 = Reusable Wearable 2, earlier generation of wearable sensor (predicate); TBD = to be determined
# 12.5.4 Summary of Non-Clinical Performance Data
Performance testing was conducted to demonstrate that the patch meets all safety and effectiveness requirements. Testing focused on features that differ from the predicate
{10}
K251088
device, Otsuka Digital Feedback Device-RW (K223463), as well as comprehensive verification and validation of the complete system.
The testing program successfully demonstrated that acceptance criteria were met for the following performance characteristics:
## Biocompatibility
Biocompatibility testing was performed, and the patch has been evaluated and deemed compliant with all the applicable requirements of ISO 10993-23-2021 (Irritation), ISO 10993-5-2009 (Cytotoxicity), and ISO 10993-10-2021 (Sensitization). Based on Annex A of ISO 10993-1, the patch is considered as a surface device with skin (uncompromised) contact and C - Long term contact duration (>30 d).
## Shelf Life
Material stability and functionality testing was performed, and the patch has a verified shelf life of 3-years from the date of manufacturing.
## Mechanical and Electrical
Formal verification testing was performed to confirm all mechanical and electrical requirements and specifications with passing results.
Examples of the mechanical testing performed include:
LED functionality; on button functionality; structural adhesive and mechanical enclosure integrity; dimensional and physical characteristics; water ingress protection; vibration, impact, push and drop testing, and markings durability.
Examples of electrical testing performed include:
- Static Testing, where conformance with the requirements and specifications is verified by inspection of the hardware design and documentation, such as schematics, BOM, and the hardware assembly.
- Requirement verification testing, to ensure that the design meets all the functional requirements.
- Environmental testing, to ensure functionality over operational, storage, and transportation temperature conditions.
## Patient Electrical Safety and Electromagnetic Compatibility (EMC) Testing
- Electrical Safety Standards Compliance
Page 7 of 12
{11}
K251088
The patch has been evaluated and deemed compliant with all the applicable requirements of the following standards:
- IEC 60601-1: 2005 + A1:2012 + A2:2020
- IEC 60601-1-2: 2014 + A1:2020
- IEC 60601-1-11: 2015 + A1:2020
- IEC 60601-1-6: 2010 + A1:2013 + A2:2020
- Device Safety Profile:
| Degree of Protection: | Type BF applied Part |
| --- | --- |
| Protection against Electrical Shock: | Internally powered medical equipment |
| Mode of Operation: | Continuous |
| Enclosure degree of Ingress Protection: | IP27 (waterproof to 1m – 3.3 feet) |
| Essential Performance: | This device has no essential performance (if the device is compromised, there is no unacceptable risk to the user) |
| Use Environment: | Home Healthcare environment |
- EMC Testing
The patch meets the following EMC classification:
- RF Emissions: CISPR 11 Group 1
- RF Environment: CISPR 11 Class B
- RF Interference Immunity: IEC 61000-4-3 Level 3
- ESD Discharge Immunity: IEC 61000-4-2 Level 4
The patch, mobile device, and software underwent electrical safety and EMC evaluation. The complete system meets the requirements of IEC 60601-1 safety standards and complies to both the third and fourth edition versions of the IEC 60601-1-2 EMC standard.
- Device software
The embedded firmware and mobile application software were developed, verified, and validated in accordance with:
- FDA guidance “Content of Premarket Submissions for Device Software Functions” (2023)
Page 8 of 12
{12}
K251088
- IEC 62304: Medical Device Software – Software Life Cycle Processes
- ISO 14971: Application of Risk Management to Medical Devices
All identified software requirements were tested and successfully passed. No anomalies remain that would impact performance, safety, or effectiveness of the device.
## Embedded Firmware
The embedded firmware controls core device functions including signal acquisition, processing, encrypted data storage, wireless communication, and power management. Verification and validation included functional testing, performance testing, simulated-use and bench testing under representative and boundary conditions.
All firmware testing was completed on the final hardware design. Test results met predefined acceptance criteria, and no unresolved issues remain.
## Mobile Application Software
The mobile application enables secure user interaction and data exchange with the patch. Functional verification and validation included testing for user authentication, access control, secure Bluetooth pairing, connection stability, and data integrity. It also covered time synchronization, encryption and cross-platform compatibility on supported iOS and Android versions and simulated-use scenarios to confirm usability and responsiveness.
Test results met predefined acceptance criteria, and no unresolved issues remain.
## Cybersecurity
Cybersecurity measures have been verified to ensure compliance with Section 524B of the FD&C Act and the FDA guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions” (issued September 27, 2023). In addition, applicable security tests were conducted in accordance with AAMI/Underwriters Laboratories (UL) 2900-1:2017, Clauses 13 to 19. The device incorporates security controls designed to protect patient safety, maintain data integrity, and prevent unauthorized access.
Cybersecurity risk management activities were performed in alignment with the above FDA guidance, as well as to the relevant standards like AAMI TIR-57:2016/2023, and NIST SP 800-30 including ISO 14971:2019 (Medical devices - Application of Risk
Page 9 of 12
{13}
K251088
Management to Medical Devices) for risk management, ensuring a risk-based, comprehensive approach to identifying and mitigating potential cybersecurity threats.
Third party security and penetration testing were also completed, and no critical vulnerabilities were found.
## 12.5.5 Summary of Clinical Performance Data
A noninterventional validation study was conducted in healthy subjects with the primary objective to evaluate wear duration, a specification of the wearable patch, while other specifications were also evaluated (e.g., wear comfort and safety assessment). A total of 55 subjects were tested at 1 site in the United States. All 55 subjects were analyzed for safety, and 51 subjects completed the study. Subjects were asked to wear 2 patches, one at a time, for up to 2 consecutive wear periods that each consisted of up to 7 days. Patches were placed on the torso where the skin was free of hair, clean, dry, and healthy, and the second patch was placed in a location that did not overlap the same area of skin as the first patch.
Efficacy results for wear duration showed a minimum of 79.0% adherence (with 95% confidence using a Wilson Score Interval) for a minimum of 5 days, which passed the 50% patch adherence specification (ie, adherence for ≥ 5 days in ≥ 50% of subjects). Furthermore, a minimum of 61.5% adherence (with 95% confidence) was observed for a minimum of 7 days. Overall, efficacy findings demonstrated that the tradeoff of patch flexibility and adhesive area was successful.
Comfort results showed an average comfort rating of 4.7 (very comfortable) on a scale of 1 to 5 with 1 = very uncomfortable and 5 = very comfortable, which passed the >3.5 minimum comfort specification.
Safety assessment resulting in an AE of Grade 1 (minimal erythema, barely perceptible), or Grade 2 (definite erythema, readily visible; minimal edema or minimal papular response) score occurred for 7/55 (12.7%) subjects, with 6/55 (10.9%) experiencing Grade 1 and 2/55 (3.6%) experiencing Grade 2 (considered clinically relevant). All other subjects had skin irritation scores assessed as Grade 0 (no evidence of irritation). The new patch design has a similar safety profile to the predicate device and no new safety concerns have been identified.
Page 10 of 12
{14}
K251088
# 12.5.6 Testing to Consensus Standards
The following Standards were used for the development and testing of the patch:
| Standards Developing Organization | Designation Number and Edition Date | Standard Title | FDA Recognition Number |
| --- | --- | --- | --- |
| AAMI | TIR57:2016 | Principles for medical device security – Risk management | 13-83 |
| IEC | 60601-1 Edition 3.2 2020-08 Consolidated Version | Medical electrical equipment-Part 1 : General requirements for basic safety and essential performance | 19-49 |
| IEC | 60601-1-6 Edition 3.2 2020-07 Consolidated Version | Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability | 5-132 |
| IEC | 60601-1-11 Edition 2.1 2020-07 Consolidated Version | Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment | 19-38 |
| IEC | TR 60601-4-2 Edition 1.0 2016-05 | Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems | 19-19 |
| IEC | 60601-2-27 Edition 3.0 2011-03 | Medical electrical equipment - Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment [Including: Corrigendum 1 (2012)] | 3-126 |
| IEC | 60601-1-2 Edition 4.1 2020-09 Consolidated Version | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | 19-36 |
{15}
K251088
| Standards Developing Organization | Designation Number and Edition Date | Standard Title | FDA Recognition Number |
| --- | --- | --- | --- |
| ISO | 10993-1 Fifth edition 2018-08 | Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process | 2-258 |
| ISO | 10993-5 Third edition 2009-06-01 | Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity | 2-245 |
| ISO | 10993-10 Fourth edition 2021-11 | Biological evaluation of medical devices - Part 10: Tests for skin sensitization | 2-296 |
| ISO | 10993-23 First edition 2021-01 | Biological evaluation of medical devices - Part 23: Tests for irritation | 2-291 |
| ISO | 14971 Third Edition 2019-12 | Medical devices - Application of risk management to medical devices | 5-125 |
| UL ANSI | 2900-1 First Edition 2017 | Standard for Safety Standard for Software Cybersecurity Network-Connectable Products Part 1: General Requirements | 13-96 |
| IEC | 60086-4 Edition 5.0 2019-04 | Primary batteries - Part 4: Safety of lithium batteries [Including: Corrigendum 1 (2019) and Corrigendum 2 (2020)] | 19-40 |
| ASTM | D4169-22 | Standard Practice for Performance Testing of Shipping Containers and Systems | 14-576 |
## 12.5.7 Conclusion
Based on the same intended use statement, technological characteristics, risk evaluations, device verification and validation testing, and clinical evaluation, the Otsuka Digital Feedback Device does not raise new safety or effectiveness questions and is equivalent to the predicate device.
Page 12 of 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.