K223463 · Otsuka America Pharmaceutical, Inc. · OZW · Aug 11, 2023 · General Hospital
Device Facts
Record ID
K223463
Device Name
Otsuka Digital Feedback Device-RW
Applicant
Otsuka America Pharmaceutical, Inc.
Product Code
OZW · General Hospital
Decision Date
Aug 11, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6305
Device Class
Class 2
Indications for Use
The Otsuka Digital Feedback Device-RW consists of a miniaturized, wearable sensor for ambulatory recording of physiological and behavioral metrics such as heart rate, activity, body angle relative to gravity (body position), and timestamped patient-logged events, including events signaled by the co-incidence with, or co-ingestible sensor accessory. When the ingestible sensor is ingested, the Otsuka Digital Feedback Device-RW is intended to log, track and trend intake times. When co-ingested with medication, the tracking of intake times may be used as an aid to measure medication adherence. The Otsuka Digital Feedback Device-RW may instance where quantifiable analysis of event-associated physiological and behavioral metrics is desirable, and enables unattended data collection for clinical and research applications.
Device Story
System comprises wearable 2-component patch (RW2), ingestible sensor, and mobile software. Ingestible sensor activates in stomach, transmitting unique ID to patch. Patch records heart rate, activity, body angle (3-axis accelerometer), and patient-logged events. Data transmitted via Bluetooth to mobile device for processing, display, and storage. Used in ambulatory settings for clinical/research data collection and medication adherence monitoring. Healthcare providers review tracked intake times and physiological trends to inform clinical decisions. Benefits include objective adherence data and unattended physiological monitoring.
Clinical Evidence
No new clinical studies performed for this submission. Previous clinical studies support expanded wear location (entire front abdomen). Bench testing included firmware, mechanical, electrical, and biocompatibility (ISO 10993) verification.
Technological Characteristics
Wearable 2-component patch (RW2); ingestible sensor; 3-axis accelerometer. Materials: skin adhesive MED5750A, Hydrogel Axelgaard AG625 (ISO 10993 compliant). Dimensions: 110x53x8.2 mm. 7-day battery life. Bluetooth connectivity. IEC 60601-1/IEC 60601-1-2 compliant. Software aggregates/processes sensor data.
Indications for Use
Indicated for patients requiring ambulatory monitoring of physiological/behavioral metrics (heart rate, activity, body position) and medication adherence tracking via ingestible sensor co-ingestion. Suitable for clinical and research applications requiring unattended data collection.
Regulatory Classification
Identification
An ingestible event marker is a prescription device used to record time-stamped, patient-logged events. The ingestible component links wirelessly through intrabody communication to an external recorder which records the date and time of ingestion as well as the unique serial number of the ingestible device.
Special Controls
In combination with the general controls of the FD&C Act, the Proteus Personal Monitor including Ingestion Event Marker is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible and non-toxic;
(2) Nonclinical, animal, and clinical testing must provide a reasonable assurance of safety and effectiveness, including device performance, durability, compatibility, usability (human factors testing), event recording, and proper excretion of the device;
(3) Appropriate analysis and nonclinical testing must validate electromagnetic compatibility performance, wireless performance, and electrical safety; and
(4) Labeling must include a detailed summary of the nonclinical and clinical testing pertinent to use of the device and the maximum number of daily device ingestions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
August 11, 2023
Otsuka America Pharmaceutical, Inc. Nancy Teague Senior Director, Global Regulatory Affairs 2440 Research Boulevard Rockville, Maryland 20850
Re: K223463
Trade/Device Name: Otsuka Digital Feedback Device-RW Regulation Number: 21 CFR 880.6305 Regulation Name: Ingestible Event Marker Regulatory Class: Class II Product Code: OZW Dated: July 10, 2023 Received: July 10, 2023
Dear Nancy Teague:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shruti N. Mistry -S
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223463
Device Name Otsuka Digital Feedback Device-RW
#### Indications for Use (Describe)
The Otsuka Digital Feedback Device-RW consists of a miniaturized, wearable sensor for ambulatory recording of physiological and behavioral metrics such as heart rate, activity, body angle relative to gravity (body position), and timestamped patient-logged events, including events signaled by the co-incidence with, or co-ingestible sensor accessory. When the ingestible sensor is ingested, the Otsuka Digital Feedback Device-RW is intended to log, track and trend intake times. When co-ingested with medication, the tracking of intake times may be used as an aid to measure medication adherence. The Otsuka Digital Feedback Device-RW may instance where quantifiable analysis of event-associated physiological and behavioral metrics is desirable, and enables unattended data collection for clinical and research applications.
| Type of Use (Select one or both, as applicable) | | | | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|-----------------------------|-----------------------------|
| <table border="0"><tr><td><table border="0"><tr><td><input checked="true" type="checkbox"/></td><td> For activities under 30 CFR 901 Subpart J </td></tr></table></td><td><table border="0"><tr><td><input type="checkbox"/></td><td> For Construction Activities </td></tr></table></td></tr></table> | <table border="0"><tr><td><input checked="true" type="checkbox"/></td><td> For activities under 30 CFR 901 Subpart J </td></tr></table> | <input checked="true" type="checkbox"/> | For activities under 30 CFR 901 Subpart J | <table border="0"><tr><td><input type="checkbox"/></td><td> For Construction Activities </td></tr></table> | <input type="checkbox"/> | For Construction Activities |
| <table border="0"><tr><td><input checked="true" type="checkbox"/></td><td> For activities under 30 CFR 901 Subpart J </td></tr></table> | <input checked="true" type="checkbox"/> | For activities under 30 CFR 901 Subpart J | <table border="0"><tr><td><input type="checkbox"/></td><td> For Construction Activities </td></tr></table> | <input type="checkbox"/> | For Construction Activities | |
| <input checked="true" type="checkbox"/> | For activities under 30 CFR 901 Subpart J | | | | | |
| <input type="checkbox"/> | For Construction Activities | | | | | |
|X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Submitted by: | Otsuka America Pharmaceutical, Inc. |
|----------------------|------------------------------------------------------------------------------------------------------|
| Address: | 2440 Research Blvd.<br>Rockville, MD 20850 |
| Telephone: | 410-353-8778 |
| Contact Name: | Nancy F. Teague<br>Senior Director, Global Regulatory Affairs<br>nancy.teague@otsuka-us.com |
| Date Submitted: | 18 Nov 2022 |
| Name of Device: | Otsuka Digital Feedback Device-RW |
| Tradename: | Otsuka Digital Feedback Device-RW |
| Common Name: | Ingestible Event Marker with wearable patch |
| Classification Name: | Ingestible Event Marker, 21 CFR 880.6305 |
| Product Code: | OZW |
| Predicate Device: | Proteus Digital Health Feedback Device, K150494<br>Proteus Patch, Proteus Ingestible Sensor, K133263 |
{4}------------------------------------------------
#### General Device Description:
The Otsuka Digital Feedback Device-RW consists of three components: a wearable sensor, an ingestible sensor accessory, and software that aggregates, processes and enables display of data collected by the sensors.
The wearable sensor is a body-worn sensor (also called the patch) that collects physiological and behavioral metrics such as heart rate, activity, body angle, and time-stamped patient-logged events, including events signaled by the co-incidence with, or co-ingestion with, the ingestible sensor accessory. The wearable sensor in the Otsuka Digital Feedback Device-RW is a 2-component patch known as the RW2 wearable sensor or RW2 patch.
The ingestible sensor is embedded inside an inactive tablet (the pill or sensor-enabled pill) for ease of handling and swallowing. After the ingestible sensor reaches the stomach, it activates and communicates its presence with a unique identifier (ID) to the wearable sensor. When the ingestible sensor is co-ingested with medication, the Otsuka Digital Feedback Device-RW is intended to log, track and trend medicine intake times to measure medication adherence.
The software on a general computing device (eg, mobile device) receives the data from the body-worn sensor or patch for further processing and analysis of the behavioral and physiological metrics. The processed data is then sent to the user interface (UI) for display as well as being saved in a local record database for storage.
For purposes of this 510(k), the changes from the predicate device to the device subject of this 510(k) are the wearable sensor and the software for the UI display. The ingestible sensor remains the same from the predicate device.
#### Indications for Use:
The Otsuka Digital Feedback Device-RW consists of a miniaturized, wearable sensor for ambulatory recording of physiological and behavioral metrics such as heart rate, activity, body angle relative to gravity (body position), and time-stamped patient-logged events, including events signaled by the co-incidence with, or co-ingestion with, the ingestible sensor accessory. When the ingestible sensor is ingested, the Otsuka Digital Feedback Device-RW is intended to log, track and trend intake times. When co-ingested with medication, the tracking and trending times may be used as an aid to measure medication adherence. The Otsuka Digital Feedback Device-RW may be used in any instance where
{5}------------------------------------------------
quantifiable analysis of event-associated physiological and behavioral metrics is desirable, and enables unattended data collection for clinical and research applications.
| Product Name | Otsuka Digital Feedback<br>Device-RW with RW2 Patch<br>(Subject Device of this<br>510(k)) | Proteus Digital Health Feedback<br>Device with DW5 Patch | Proteus RW1 Patch<br>Including Ingestible Sensor |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K223463 | K150494 | K133263 |
| Patch Detection<br>Capability | Improved algorithm from<br>DW5 patch which allows for<br>expanded wear location to<br>either side of the torso. | Algorithm allows for patch to be<br>worn on the left side of the torso | Algorithm allows for patch to<br>be worn on the left side of the<br>torso |
| Materials/Layer<br>Construction | RW2 uses skin adhesive<br>MED5750A and Hydrogel<br>Axelgaard AG625. Skin<br>contacting materials pass<br>ISO 10993 testing for<br>irritation, cytotoxicity and<br>sensitization, and/or has been<br>used in another 510(k)-cleared<br>product for similar skin contact<br>use. | DW5 uses skin adhesive<br>MED5750A, Hydrogel Covidien<br>RG-63B and a hydrocolloid foam<br>top. All skin contacting materials<br>pass ISO 10993 testing for<br>irritation, cytotoxicity and<br>sensitization | RW1 uses skin adhesive<br>MED5750A and Hydrogel<br>Covidien RG-63B. All skin<br>contacting materials pass ISO<br>10993 testing for irritation,<br>cytotoxicity and sensitization |
| Battery Life | 7-day battery life | 7-day battery life | 30-day battery life |
| Dimension: | Thickness - 8.2 mm<br>Length - 110 mm<br>Width - 53 mm | Thickness - 6.3 mm<br>Length - 102 mm<br>Width - 60 mm | Thickness - 11 mm<br>Length - 97.5 mm<br>Width - 42 mm |
| Shelf Life | 3 years | 3 years | 1 year |
| Event Marker | Ingestible event marker | Ingestible event marker | Ingestible event marker |
| Accelerometer | 3-axis accelerometry | 3-axis accelerometry | 3-axis accelerometry |
| Hardware<br>(Ingestible<br>Sensor) | Ingestible event marker | Ingestible event marker | Ingestible event marker |
| Hardware<br>(Wearable<br>Sensor) | Wearable, physiologic sensor;<br>2-component patch | Wearable, physiologic sensor;<br>1-component patch | Wearable, physiologic sensor;<br>2-component patch |
| Firmware<br>(Wearable<br>Sensor) | Collects data from sensor,<br>performs data processing,<br>stores data until wireless<br>connection is available, sends<br>data to general computing<br>device | Collects data from sensor,<br>performs data processing, stores<br>data until wireless connection is<br>available, sends data to mobile<br>device | Collects data from sensor,<br>performs data processing,<br>stores data until wireless<br>connection is available, sends<br>data to mobile device |
| Software | Aggregates, processes and<br>enables display of data<br>collected by sensors | Aggregates, processes and<br>displays data collected by sensors | Aggregates, processes and<br>displays data collected by<br>sensors |
| Data Telemetry | Bluetooth Technology | Bluetooth Technology | Bluetooth Technology |
| Biocompatibility | ISO 10993 compliant | ISO 10993 compliant | ISO 10993 compliant |
| Product Name | Otsuka Digital Feedback<br>Device-RW with RW2 Patch<br>(Subject Device of this<br>510(k)) | Proteus Digital Health Feedback<br>Device with DW5 Patch | Proteus RW1 Patch<br>Including Ingestible Sensor |
| 510(k) Number | K223463 | K150494 | K133263 |
| Electrical<br>Safety/EMC | IEC 60601-1/IEC 60601-1-2<br>compliant | IEC 60601-1/IEC 60601-1-2<br>compliant | IEC 60601-1/IEC 60601-1-2<br>compliant |
## Technological Characteristics:
{6}------------------------------------------------
Abbreviations: EMC = electromagnetic compatibility; TBD = to be determined
#### Summary of Non-Clinical Performance Data:
Risk evaluations and verification and validation testing were conducted for each feature that differed from the predicate devices, Proteus Digital Health Feedback Device (K150494) and Proteus Patch, Proteus Ingestible Sensor (K133263), to the Otsuka Digital Feedback Device-RW. The results demonstrate that the acceptance criteria for firmware verification, mechanical verification, electrical verification, biocompatibility evaluation, and system verification testing were met.
#### Summary of Clinical Performance Data:
Clinical studies were conducted to support the expanded wear location of the patch to include the entire front abdomen. However, new clinical studies were not performed to support the changes proposed in this submission.
### Conclusion:
Based on the indication for use statement, technological characteristics, risk evaluations, and device verification and validation testing, the Otsuka Digital Feedback Device-RW does not raise new questions of safety or effectiveness and is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.