ETHICON 4000 60mm Compact Stapler (EC3D60C); ETHICON 4000 60mm Standard Stapler (EC3D60S); ETHICON 4000 60mm Long Stapler (EC3D60L); ETHICON 3D 60mm White Reload (ER60W); ETHICON 3D 60mm Blue Reload (ER60B); ETHICON 3D 60mm Green Reload (ER60G); ETHICON 3D 60mm Black Reload (ER60T)
K250835 · Ethicon Endo-Surgery, LLC · GDW · May 14, 2025 · General, Plastic Surgery
Device Facts
Record ID
K250835
Device Name
ETHICON 4000 60mm Compact Stapler (EC3D60C); ETHICON 4000 60mm Standard Stapler (EC3D60S); ETHICON 4000 60mm Long Stapler (EC3D60L); ETHICON 3D 60mm White Reload (ER60W); ETHICON 3D 60mm Blue Reload (ER60B); ETHICON 3D 60mm Green Reload (ER60G); ETHICON 3D 60mm Black Reload (ER60T)
Applicant
Ethicon Endo-Surgery, LLC
Product Code
GDW · General, Plastic Surgery
Decision Date
May 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The ETHICON™ 4000 and ETHICON™ 3D Reloads are intended for transection, resection and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney, and spleen.
Device Story
Sterile, single-patient-use, battery-powered surgical stapler; delivers 6 staggered rows of staples (3D staples in rows 1, 2, 5, 6; 2D B-formed staples in rows 3, 4). Used in open or minimally invasive surgery by surgeons to simultaneously cut and staple tissue. Features include shaft rotation, articulation mechanism for lateral access, and a lockout to prevent firing used/improperly installed reloads. Powered by a primary lithium battery pack. Embedded software controls articulation and firing. Provides permanent tissue closure/apposition. Benefits include reliable hemostasis and tissue apposition in solid organs (liver, pancreas, kidney, spleen) and general surgical applications.
Clinical Evidence
No clinical data provided. Evidence consists of non-clinical bench testing (vessel sealing, staple height, staple line integrity) and pre-clinical animal testing (porcine model) evaluating hemostasis in liver, spleen, pancreas, and associated vasculature.
Technological Characteristics
Battery-powered, motorized surgical stapler; 60mm staple line; 6-row staple configuration (3D and 2D staples). Materials meet ISO 10993-1. Sterilization: EO (stapler), Gamma (reload). Shelf life: 3 years. Connectivity: None. Software: Embedded for articulation and firing control.
Indications for Use
Indicated for transection, resection, and/or creation of anastomoses in open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures, including liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney, and spleen. Contraindicated for use on ischemic or necrotic tissue, the aorta, or major vessels without proximal/distal control; contraindicated if tissue thickness is outside the specified closed staple height range.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 14, 2025
Ethicon Endo-Surgery, LLC
Megan Westendorf
Senior Regulatory Affairs Specialist
475 Calle C
Guaynabo, PR 00969
Re: K250835
Trade/Device Name: ETHICON 4000 60mm Compact Stapler (EC3D60C); ETHICON 4000 60mm Standard Stapler (EC3D60S); ETHICON 4000 60mm Long Stapler (EC3D60L); ETHICON 3D 60mm White Reload (ER60W); ETHICON 3D 60mm Blue Reload (ER60B); ETHICON 3D 60mm Green Reload (ER60G); ETHICON 3D 60mm Black Reload (ER60T)
Regulation Number: 21 CFR 878.4750
Regulation Name: Implantable staple
Regulatory Class: Class II
Product Code: GDW, GAG
Dated: March 19, 2025
Received: March 20, 2025
Dear Megan Westendorf:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250835 - Megan Westendorf
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K250835 - Megan Westendorf
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TEK N. LAMICHHANE -
S
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
ETHICON 4000 60mm Compact Stapler
Page 13 of 55
| Indications for Use | | |
| --- | --- | --- |
| Submission Number | K250835 | ? |
| Device Name | | ? |
| ETHICON 4000 60mm Compact Stapler (EC3D60C);
ETHICON 4000 60mm Standard Stapler (EC3D60S);
ETHICON 4000 60mm Long Stapler (EC3D60L);
ETHICON 3D 60mm White Reload (ER60W);
ETHICON 3D 60mm Blue Reload (ER60B);
ETHICON 3D 60mm Green Reload (ER60G);
ETHICON 3D 60mm Black Reload (ER60T) | | |
| Indications for Use | | ? |
| The ETHICON™ 4000 and ETHICON™ 3D Reloads are intended for transection, resection and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney, and spleen. | | |
| Type of Use | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K250835
# 510(k) Summary
## I. SUBMITTER
Submission: K250835
Company: Ethicon Endo-Surgery, LLC
475 Calle C
Guaynabo, PR 00969
Contact: Megan Westendorf
Senior Regulatory Affairs Specialist
Ethicon Endo-Surgery, Inc.
Phone: 513-479-6951
Email: mwestend@its.jnj.com
Date Prepared: May 14, 2025
## II. Subject DEVICES
| Trade Names: | Models: |
| --- | --- |
| ETHICON™ 4000 60mm Compact Stapler | EC3D60C |
| ETHICON™ 4000 60mm Standard Stapler | EC3D60S |
| ETHICON™ 4000 60mm Long Stapler | EC3D60L |
| ETHICON™ 3D 60mm White Reload | ER60W |
| ETHICON™ 3D 60mm Blue Reload | ER60B |
| ETHICON™ 3D 60mm Green Reload | ER60G |
| ETHICON™ 3D 60mm Black Reload | ER60T |
Classification Name: Surgical Stapler
Staple, Implantable
Classification Regulation: 21 CFR 878.4740
21 CFR 878.4750
Device Class: II
Product Code: GDW, GAG
{5}
K250835
## III. Predicate DEVICES
| Predicate Device 510(k) Number | Predicate Device Name | Predicate Device Models |
| --- | --- | --- |
| K241629 | ECHELON™ 3000 60mm Stapler | ECH60C, ECH60S, ECH60L |
| K183435 | ECHELON Endoscopic Linear Cutter Reload (+Gripping Surface Technology) 60mm | GST60W, GST60B, GST60G, GST60T |
## Reference DEVICES
| Reference Device 510(k) Number | Reference Device Name | Reference Device Models |
| --- | --- | --- |
| K241630 | ETHICON™ 4000 60mm Stapler | EC3D60C, EC3D60S, EC3D60L |
| K241630 | ETHICON™ 3D Reloads | ER60B, ER60G, ER60T, ER60W |
## IV. Device Description
The ETHICON™ 4000 Stapler and ETHICON™ 3D Reloads are intended for transection, resection and/or creation of anastomoses.
The ETHICON™ 4000 Stapler and ETHICON™ 3D Reloads are sterile, single-patient-use instruments that simultaneously cut and staple tissue. There are six staggered rows of staples, three on either side of the cut line. Together, the ETHICON™ 4000 Staplers and ETHICON™ 3D Reloads deliver 3D staples in the first, second, fifth and sixth rows of staples. The third and fourth rows (nearest the knife) maintain the traditional 2D B-formed staples.
ETHICON™ 4000 60mm Staplers have a staple line that is approximately 60 mm long and a cut line that is approximately 57 mm long. The Subject stapler device is available in three different shaft lengths - Compact, Standard and Long. The shaft can rotate freely in both directions and an articulation mechanism enables the distal portion of the shaft to pivot to facilitate lateral access to the operative site. The device is packaged with a primary lithium Battery Pack that must be installed prior to use. There is embedded software to articulate and initiate firing of the device.
The instruments are packaged without a reload and must be loaded prior to use. The instrument may be reloaded for a maximum of 12 firings during a single procedure. A
{6}
K250835
Staple Retaining Cap on the reload protects the staple leg points during shipping and transportation. The instruments' lockout feature is designed to prevent a used or improperly installed reload from being re-fired or an instrument from being fired without a reload.
The staples are permanent implants that provide tissue closure and apposition from the time of implant through the critical phases of healing. The staples remain in place for the patient's lifetime, unless in the opinion of the treating physician, they require removal.
## V. Intended Use/Indications for Use
This submission is to expand the indications for use to include transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney, and spleen.
The ETHICON™ 4000 Staplers and ETHICON™ 3D Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney, and spleen.
There is no change to the intended use: The ETHICON™ 4000 Stapler and ETHICON™ 3D Reloads are intended for transection, resection and/or creation of anastomoses.
## VI. Technological Comparison
This table provides an overview of the characteristics of the subject device compared to the predicate device. The subject and predicate device share the same Indications for Use.
| Characteristic | Subject Device | Predicate Device |
| --- | --- | --- |
| Indication for Use | The ETHICON™ 4000 Staplers and ETHICON™ 3D Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney, and spleen. | The ECHELON FLEX™, ECHELON™ 3000, and ECHELON ENDOPATH™ families of staplers and reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement |
{7}
K250835
| Characteristic | Subject Device | Predicate Device |
| --- | --- | --- |
| | transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney, and spleen. | materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney, and spleen. |
| Intended Use | Transection, resection, and/or creation of anastomoses. | Same |
| Contraindications | ○ Do not use the instruments on ischemic or necrotic tissue
○ Do not use the instruments on the aorta.
○ Do not use any endocutter on major vessels without making provision for proximal and distal control.
○ Tissue thickness should be carefully evaluated before firing any stapler.
○ Refer to the Reload Product Codes Table below for tissue compression requirement (closed staple height) for each staple size.
○ If tissue cannot comfortably compress to the closed staple height listed in the table, or easily compresses to less than the closed staple height listed in the table, the tissue is contraindicated as it may be too thick or too thin for the selected staple size.
○ These instruments are not intended for use when surgical stapling is contraindicated. | Same |
| Sterile, Single Patient Use | Yes | Same |
| Staple Rows | 6 | Same |
| Staple Form | 3D Staple form (first, second, fifth & sixth rows)
2D Staple form (third & fourth rows) | 2D Staple form |
| Staple Line Length | 60 mm | Same |
| Sterilization Method | Stapler: EO Sterilization
Reload: Gamma Irradiation | Same |
| Shelf Life | 3 Years | Same |
| Biocompatibility of Materials | Meets ISO 10993-1 | Same |
| Packaging Materials | Stapler: PETG Tray with Tyvek Lid
Reload: Flexible C-Film Blister with Tyvek Lid | Same |
4
{8}
K250835
This table provides an overview of the characteristics of the subject device compared to the reference device. The technological characteristics of the subject device are identical to that of the reference device; the only change is the expansion of the indications for use.
| Characteristic | Subject Device | Reference Device |
| --- | --- | --- |
| Indication for Use | The ETHICON™ 4000 Staplers and ETHICON™ 3D Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney, and spleen. | The ETHICON™ 4000 Staplers and ETHICON™ 3D Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials. |
| Intended Use | Transection, resection, and/or creation of anastomoses. | Same |
| Contraindications | ○ Do not use the instruments on ischemic or necrotic tissue ○ Do not use the instruments on the aorta. ○ Do not use any endocutter on major vessels without making provision for proximal and distal control. ○ Tissue thickness should be carefully evaluated before firing any stapler. ○ Refer to the Reload Product Codes Table below for tissue compression requirement (closed staple height) for each staple size. ○ If tissue cannot comfortably compress to the closed staple height listed in the table, or easily compresses to less than the closed staple height listed in the table, the tissue is contraindicated as it may be too thick or too thin for the selected staple size. ○ These instruments are not intended for use when surgical stapling is contraindicated. | Same |
| Sterile, Single Patient Use | Yes | Same |
| Staple Rows | 6 | Same |
{9}
K250835
| Characteristic | Subject Device | Reference Device |
| --- | --- | --- |
| Staple Form | 3D Staple form (first, second, fifth & sixth rows)
2D Staple form (third & fourth rows) | Same |
| Staple Line Length | 60 mm | Same |
| Sterilization Method | Stapler: EO Sterilization
Reload: Gamma Irradiation | Same |
| Shelf Life | 3 Years | Same |
| Biocompatibility of Materials | Meets ISO 10993-1 | Same |
| Packaging Materials | Stapler: PETG Tray with Tyvek Lid
Reload: Flexible C-Film Blister with Tyvek Lid | Same |
## VII. Non-Clinical and/or Clinical Tests
Performance test data (bench and animal) demonstrates that the subject device is substantially equivalent to the predicate device and that the device meets all requirements for the expanded indications for use to include solid organs.
The nonclinical (bench) tests that have been submitted include vessel sealing, formed staple height, staple form quality, and staple line integrity in solid organs. No additional bench testing was needed as the subject device is technologically identical to the reference device.
Pre-clinical (animal) testing includes hemostasis performance in solid organs and vessels. Hemostasis testing of liver, spleen and pancreas using porcine model was conducted to support the expansion of indications to include solid organs. Data was also collected from isolated arteries, veins and bundles/pedicles. Lung and uterine tissues along with their isolated vessels were also included in the testing.
## Clinical Tests
This premarket submission did not rely on the assessment of clinical performance data to demonstrate device performance and equivalence.
## VIII. Conclusion
In all cases the subject devices passed the functional requirements of the device features. The ETHICON™ 4000 and ETHICON™ 3D Reloads have been demonstrated to be safe and effective for the expanded indications for use.
In conclusion, the performance testing demonstrates that the Subject device performs substantially equivalent to the Predicate device and does not raise any new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.