ECHELON 4000 60mm Compact Stapler (EC3D60C); ECHELON 4000 60mm Standard Stapler (EC3D60S); ECHELON 4000 60mm Long Stapler (EC3D60L); ECHELON 3D 60mm White Reload (ER60W); ECHELON 3D 60mm Blue Reload (ER60B); ECHELON 3D 60mm Green Reload (ER60G); ECHELON 3D 60mm Black Reload (ER60T)
K241630 · Ethicon Endo-Surgery, LLC · GDW · Oct 28, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241630
Device Name
ECHELON 4000 60mm Compact Stapler (EC3D60C); ECHELON 4000 60mm Standard Stapler (EC3D60S); ECHELON 4000 60mm Long Stapler (EC3D60L); ECHELON 3D 60mm White Reload (ER60W); ECHELON 3D 60mm Blue Reload (ER60B); ECHELON 3D 60mm Green Reload (ER60G); ECHELON 3D 60mm Black Reload (ER60T)
Applicant
Ethicon Endo-Surgery, LLC
Product Code
GDW · General, Plastic Surgery
Decision Date
Oct 28, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The ETHICON™ 4000 Staplers and ETHICON™ 3D Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials.
Device Story
Sterile, single-patient-use surgical instrument for simultaneous tissue cutting and stapling. Operates via primary lithium battery pack; features motorized articulation and firing. Input: manual surgeon control of device positioning and activation. Transformation: embedded software controls articulation and firing mechanism; device deploys six staggered rows of staples (3D staples in outer rows, 2D B-formed staples in inner rows). Output: stapled and transected tissue. Used in OR by surgeons for general, urologic, thoracic, and pediatric procedures. Provides permanent tissue closure and apposition; benefits patient through reliable tissue healing and reduced surgical time. Includes lockout feature to prevent firing without reload or with used reload.
Clinical Evidence
No clinical data provided. Evidence based on non-clinical bench testing, including staple performance equivalence, device functional performance, staple line strength characterization, human factors testing, biocompatibility, and pre-clinical hemostasis/tissue healing studies.
Technological Characteristics
Materials: Titanium alloy staples. Energy: Primary lithium battery pack. Form factor: 60mm staple line, 57mm cut line, three shaft lengths (Compact, Standard, Long). Connectivity: None. Sterilization: Not specified. Software: Embedded firmware for articulation and firing control. Biocompatibility: ISO 10993-1 compliant.
Indications for Use
Indicated for transection, resection, and/or creation of anastomoses in open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. No specific age or gender contraindications provided.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 28, 2024
Ethicon Endo-Surgery, LLC Megan Westendorf Regulatory Affairs Specialist 475 Calle C Guaynabo. PR 00969
# Re: K241630
Trade/Device Name: ECHELON 4000 60mm Compact Stapler (EC3D60C); ECHELON 4000 60mm Standard Stapler (EC3D60S); ECHELON 4000 60mm Long Stapler (EC3D60L); ECHELON 3D 60mm White Reload (ER60W); ECHELON 3D 60mm Blue Reload (ER60B); ECHELON 3D 60mm Green Reload (ER60G); ECHELON 3D 60mm Black Reload (ER60T)
Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW, GAG Dated: September 24, 2024 Received: September 24, 2024
Dear Megan Westendorf:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/3 description: The image shows a digitally signed document by Tek N. Lamichhane. The signature includes the date and time of 2024-10-28 17:24:10. The document also lists Tek N. Lamichhane's title as Assistant Director at DHT4B: Division of Plastic and Reconstructive Surgery Devices and OHT4: Office of Surgical and Infection Control Devices, as well as the Office of Product Evaluation and Quality Center for Devices and Radiological Health.
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241630
Device Name
ETHICON 4000 60mm Compact Stapler (EC3D60C); ETHICON 4000 60mm Standard Stapler (EC3D60S); ETHICON 4000 60mm Long Stapler (EC3D60L); ETHICON 3D 60mm White Reload (ER60W); ETHICON 3D 60mm Blue Reload (ER60B); ETHICON 3D 60mm Green Reload (ER60G); ETHICON 3D 60mm Black Reload (ER60T)
Indications for Use (Describe)
The ETHICON™ 4000 Staplers and ETHICON™ 3D Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K241630
# 510(k) Summary K246130
### I. SUBMITTER
| Company: | Ethicon Endo-Surgery, LLC |
|----------|---------------------------|
| | 475 Calle C |
| | Guaynabo, PR 00969 |
- Contact: Megan Westendorf Senior Regulatory Affairs Specialist Ethicon Endo-Surgery, Inc. Phone: 513-479-6951 Email: mwestend(@its.jnj.com
Date Prepared: October 25, 2024
### II. Subject DEVICES
# Trade Names:
Product Codes:
ETHICON™ 4000 60mm Compact Stapler ETHICON™ 4000 60mm Standard Stapler ETHICON™ 4000 60mm Long Stapler ETHICON™ 3D 60mm White Reload ETHICON™ 3D 60mm Blue Reload ETHICON™ 3D 60mm Green Reload ETHICON™ 3D 60mm Black Reload
| EC3D60C |
|---------|
| EC3D60S |
| EC3D60L |
| ER60W |
| ER60B |
| ER60G |
| ER60T |
| Classification Name: | Surgical Stapler |
|----------------------------|---------------------|
| | Staple, Implantable |
| Classification Regulation: | 21 CFR 878.4740 |
| | 21 CFR 878.4750 |
| Device Class: | II |
| Product Code: | GDW, GAG |
### III. Predicate DEVICES
| Predicate Device<br>510(k) Number | Predicate Device Name | Predicate Device Product<br>Codes |
|-----------------------------------|-----------------------------------------------------------------------------------|-----------------------------------|
| K213633 | ECHELON TM 3000 60mm Stapler | ECH60C, ECH60S, ECH60L |
| K183435 | ECHELON Endoscopic Linear Cutter<br>Reload (+Gripping Surface<br>Technology) 60mm | GST60W, GST60B,<br>GST60G, GST60T |
{5}------------------------------------------------
| Reference Device<br>510(k) Number | Reference Device Name | Reference Device Product<br>Codes |
|-----------------------------------|-------------------------------------------|--------------------------------------------------------------|
| K223760 | ECHELON LINEAR™ Cutter<br>Reloads | GLCR60B, GLCR60G,<br>GLCR80B, GLCR80G,<br>GLCR100B, GLCR100G |
| K200420 | ECHELON CONTOUR™ Curved<br>Cutter Reloads | GCR40B, GCR40G |
| K163523 | ECHELON CIRCULAR™ Powered<br>Stapler | CDH23P, CDH25P, CDH29P,<br>CDH31P |
### Reference DEVICES
### IV. Device Description
The ETHICON™ 4000 Stapler and ETHICON™ 3D Reloads are intended for transection, resection and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials.
The ETHICON™ 4000 Stapler and ETHICON™ 3D Reloads are sterile, singlepatient-use instruments that simultaneously cut and staple tissue. There are six staggered rows of staples, three on either side of the cut line. Together, the ETHICON™ 4000 60 mm Staplers and ETHICONTM 3D Reloads deliver 3D staples in the first, second, fifth and sixth rows of staples. The third and fourth rows (nearest the knife) maintain the traditional 2D B-formed staples.
ETHICON™ 4000 60 mm Staplers have a staple line that is approximately 60 mm long and a cut line that is approximately 57 mm long. The Subject stapler device is available in three different shaft lengths - Compact, Standard and Long. The shaft can rotate freely in both directions and an articulation mechanism enables the distal portion of the shaft to pivot to facilitate lateral access to the operative site. The device is packaged with a primary lithium Battery Pack that must be installed prior to use. There is embedded software to articulate and initiate firing of the device.
The instruments are packaged without a reload and must be loaded prior to use. The instrument may be reloaded for a maximum of 12 firings during a single procedure. A Staple Retaining Cap on the reload protects the staple leg points during shipping and transportation. The instruments' lockout feature is designed to prevent a used or improperly installed reload from being re-fired or an instrument from being fired without a reload.
{6}------------------------------------------------
The staples are permanent implants that provide tissue closure and apposition from the time of implant through the critical phases of healing. The staples remain in place for the patient's lifetime, unless in the opinion of the treating physician, they require removal.
The Subject stapler and reloads will not be compatible with previous ECHELON™ staplers and reloads. This is a new platform that will work with 3D reloads only.
#### V. Intended Use/Indications for Use
The ETHICON™ 4000 Staplers and ETHICON™ 3D Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials.
| Characteristic | Subject Device | Predicate Device |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | Staplers: EC3D60C, EC3D60S, EC3D60L<br>Reloads: ER60W, ER60B, ER60G, ER60T | Staplers: ECH60C, ECH60S, ECH60L<br>Reloads: GST60W, GST60B, GST60G, GST60T |
| | The ETHICONTM 4000 Staplers and ETHICONTM 3D Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have applications in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials. | The ECHELON FLEXTM and ECHELONTM 3000 families of staplers and reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney, and spleen. |
| Intended Use | Same | Same |
| Contraindications | Same | Same |
#### VI. Technological Comparison
Confidential
{7}------------------------------------------------
| Characteristic | Subject Device<br>Staplers: EC3D60C, EC3D60S, EC3D60L<br>Reloads: ER60W, ER60B, ER60G, ER60T | Predicate Device<br>Staplers: ECH60C, ECH60S, ECH60L<br>Reloads: GST60W, GST60B, GST60G, GST60T |
|-----------------------------------|----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Sterile, Single<br>Patient Use | Yes | Same |
| Sterilization<br>Method | Same | Same |
| Control Mechanism | Same | Same |
| Shelf Life | Same | Same |
| Biocompatibility of<br>Materials | Meets ISO 10993-1 | Same |
| Packaging<br>Materials | Same | Same |
| Stapler Anvil &<br>End Effector | Compatible with ETHICON 3D<br>Reloads only | Compatible with ETHICON 2D<br>reloads |
| Staple Form | 3D staple form (first, second, fifth<br>and sixth rows)<br>2D staple form (third & fourth<br>rows) | 2D staple form (all rows) |
| Staple Material | Titanium Alloy | Same |
| Staple Line Length | 60 mm | Same |
| Formed Staple<br>Height in Tissue | Same | Same |
| Staple Rows | 6 | Same |
The reference devices demonstrate the same 3D staple design as the Subject device and similar intended use. The same stapling performance tests were conducted on the Subject and reference devices, and all met their respective acceptance criteria. The equivalent staple design characteristics and successful completion of the same stapling performance tests demonstrate safe and effective staple formation with the 3D staple design.
### VII. Non-Clinical and/or Clinical Tests Summary
## Non-Clinical Studies
The nonclinical tests that have been submitted include staple performance equivalence bench testing compared to the predicate; device functional performance bench testing; staple line strength product characterization of Subject and Predicate devices; and the Human Factors testing report. Biocompatibility evaluation was performed according to ISO 10993-1. Pre-clinical data including Hemostasis Performance and Tissue healing response in Abdominal and Thoracic procedures.
{8}------------------------------------------------
## Clinical Studies
The premarket submission did not rely on the assessment of clinical performance data to demonstrate device performance and equivalence.
### Conclusion VIII.
In all cases the subject devices demonstrated either equivalent or improved performance to the predicate devices or passed the functional requirements of the subject device features. The ETHICON™ 4000 and ETHICON™ 3D Reloads therefore meet the acceptance criteria for bench testing and have been demonstrated to be safe and effective.
In conclusion, the performance testing demonstrates that the Subject device performs substantially equivalent to the Predicate devices and does not raise any new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.