The Multi M Series is used in the manufacture of dental prosthesis.
Device Story
Multi M Series is a zirconia-based ceramic material (ISO 6872 Type 2 Class 5) supplied in various shapes (e.g., round, W-type, Z-type) for use in dental laboratories. The device is processed by dental professionals using CAD/CAM milling machines to manufacture cores for all-ceramic crowns. The clinician or technician selects the appropriate block shape to fit the specific CAD/CAM jig, uploads a digital design file, and mills the ceramic into the final prosthetic shape. The device offers 278 shades to match patient tooth color. By providing a high-strength, biocompatible ceramic substrate, the device enables the creation of durable, aesthetic dental restorations, benefiting patients requiring tooth replacement or repair.
Clinical Evidence
Bench testing only. Performance testing included appearance, dimensions, packaging, uniformity, extraneous materials, flexural strength, linear thermal expansion coefficient, chemical solubility, monoclinic phase fraction, radioactivity, and flexural strength after accelerated aging (all per ISO 6872 or ISO 13356). Biocompatibility testing was conducted in accordance with ISO 10993-3, 5, 6, 10, 11, and 33.
Technological Characteristics
Zirconia-based ceramic (ZrO2, HfO2, Y2O3, Al2O3, SiO2, Fe2O3, Er2O3, CaO). Classified as ISO 6872 Type 2 Class 5. Flexural strength >800MPa; chemical solubility <100 ug/cm2; CTE (10.6±0.5)×10-6K-1. Solid form, non-sterile. Designed for CAD/CAM milling.
Indications for Use
Indicated for use in the manufacture of dental prosthetics for the general population, primarily adults, requiring dental restoration.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 12, 2025
DMAX Co., Ltd.
% Priscilla Chung
Regulatory Affairs Consultant
LK Consulting Group USA, Inc.
18881 Von Karman STE 160
Irvine, California 92612
Re: K250826
Trade/Device Name: Multi M Series
Regulation Number: 21 CFR 872.6660
Regulation Name: Porcelain Powder For Clinical Use
Regulatory Class: Class II
Product Code: EIH,
Dated: February 20, 2025
Received: March 18, 2025
Dear Priscilla Chung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250826 - Priscilla Chung
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250826 - Priscilla Chung
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Multi M Series
Page 7 of 23
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K250826 | ? |
| Please provide the device trade name(s). | | ? |
| Multi M Series | | |
| Please provide your Indications for Use below. | | ? |
| The Multi M Series is used in the manufacture of dental prosthesis. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) summary
1 / 3 page
This summary of 510(K) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: 3/13/2025
1. Submitter
DMAX Co., Ltd.
89, Hosan-ro, Dalseo-gu, Daegu, 42710, Korea
Tel: 82-53-592-3146
2. U.S Agent/Contact Person
Priscilla Chung
LK Consulting Group USA, Inc.
18881 Von Karman Ave STE 160, Irvine CA 92612
Phone: 714.202.5789 Fax: 714-409-3357
Email: juhee.c@lkconsultinggroup.com
3. Device
- Trade Name: Multi M Series
- Common Name: Dental Frame Material for Dental Prosthesis
- Classification Name: Porcelain Powder for Clinical Use
- Product Code: EIH
- Classification regulation: 21 CFR 872.6660
4. Predicate Device:
Zmaxx T series by DMAX Co., Ltd. (K161219)
5. Description:
Multi M Series is zirconia-based ceramic provided in various shapes such as round and it is used to manufacture cores of all ceramic crowns and is classified into ISO 6872 Type 2 Class 5. It is used to manufacture ceramic restorations through cutting process by dental MAD/MAM, computer-assisted design system, or CAD/CAM system. The subject device offers 278 different shades to meet the needs of different patients' tooth colors. It also offers various shape types to be used with various jigs for CAD/CAM milling machine. Multi M Series offers various types as follows: W-type, Z-type. These different types are to accommodate the different types of the jig that is built-in CAD/CAM machine. Users can
{5}
choose a type which fits the jig they have.
# 6. Indication for use:
The Multi M Series is used in the manufacture of dental prosthesis.
# 7. Basis for Substantial Equivalence
# 7.1. Comparison Chart
| | Subject device | Predicate device |
| --- | --- | --- |
| Device Name | Multi M series | Zmaxx T series |
| 510(k) | - | K161219 |
| Product Code | EIH | EIH |
| Manufacturer | DMAX Co., Ltd. | DMAX Co., Ltd. |
| Technology | - | - |
| Indication for Use | Multi M Series is used in the manufacture of dental prosthetics. | Zmaxx T Series is used in the manufacture of dental prosthetics. |
| Principles of Operation | Install the block on milling machine, upload a designed file and process it. | Install the block on milling machine, upload a designed file and process it. |
| Mechanical Testing | ISO 6872:2015 | ISO 6872:2008 |
| Prescription Use | Prescription only | Prescription only |
| Target population | General, mostly adults | General, mostly adults |
| Type of Packaging | Paper box container | Paper box container |
| Packaging Volume(EA) | 1EA | 1EA |
| Shade | 278 shades based on Vita shade guide | 3 colors (A0, A0.5, A2) |
| Storage Conditions | 5-50 °C | 10-35°C |
| General Physical Form | Solid | Solid |
| Sterility | Non-Sterile | Non-Sterile |
| CTE(coefficient of thermal expansion) | (10.6±0.5)×10-6K-1 (meeting the ISO 6872 requirement) | (10.6±0.5)×10-6K-1 (meeting the ISO 6872 requirement) |
| Flexural Strength | >800MPa (meeting the ISO 6872 requirement) | >600MPa (meeting the ISO 6872 requirement) |
| Chemical solubility | <100 ug/cm2 (meeting the ISO 6872 requirement) | <2000 ug/cm2 (meeting the ISO 6872 requirement) |
| Material component | ZrO2.HfO2.Y2O3.Al2O3.SiO2.Fe2O3.Er2O3.CaO | ZrO2.HfO2.Y2O3.Al2O3.SiO2.Fe2O3.Er2O3 |
| Biocompatibility | Non-toxic and biocompatible (Meeting the ISO 10993-5, 10 and 10993-11 requirements) | Non-toxic and biocompatible (Meeting the ISO 10993-5, 10 and 10993-11 requirements) |
510(k) summary
{6}
7.2. Substantial Equivalence
The Multi M series is substantially equivalent to the Zmaxx T series (K161219) made by our company. Although the 510k applicant is Maxx Digm, Inc., we are a contract manufacturer of the device.
The following comparison table is presented to demonstrate substantial equivalence. They share similar materials and similar physical / chemical properties. However, the mixing ratios of the raw materials are different between the devices. Despite these differences, the performance test results per ISO 6872 show that the subject device is substantially equivalent to the predicate devices in physical and chemical properties and meets the necessary requirements. Based on the comparison between the two devices and the performance testing results presented in this 510K, we conclude that the subject device is substantially equivalent to the predicate devices.
8. Non-Clinical Testing
- Performance Tests including Appearance, Dimensions, Packaging, Uniformity (ISO 6872), Extraneous Materials (ISO 6872), Flexural Strength (ISO 6872), Linear thermal Expansion Coefficient (ISO 6872), Chemical Solubility (ISO 6872), Monoclinic Phase Fraction (ISO 13356), Radioactivity (ISO 6872), Flexural strength after accelerated aging (ISO 13356)
- Biocompatibility tests in accordance with ISO 10993-3, 5, 6, 10, 11, and 33
9. Conclusion
The subject device and the predicate device have the same intended use and have the same technological characteristics. Based on the similarities and the test results, we conclude that the subject device is substantially equivalent to the predicate device.
510(k) summary
3 / 3 page
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.