K161219 · Maxx Digm, Inc. · EIH · Jan 26, 2017 · Dental
Device Facts
Record ID
K161219
Device Name
Zmaxx T Series
Applicant
Maxx Digm, Inc.
Product Code
EIH · Dental
Decision Date
Jan 26, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6660
Device Class
Class 2
Indications for Use
Zmaxx T Series is used in the manufacture of dental prosthetics.
Device Story
Zmaxx T Series is a zirconia-based ceramic material provided in various shapes (round, square) and sizes; used to manufacture cores of all-ceramic crowns. Input consists of physical ceramic blocks; processed via CAD/CAM milling machines using digital design files. Output is a custom-milled dental prosthetic core. Used in dental laboratories; operated by dental technicians. Healthcare providers use the final prosthetic to restore patient tooth structure and function. Benefits include patient-specific tooth color matching and structural restoration.
Clinical Evidence
Bench testing only. Performance evaluated against ISO 6872 (flexural strength, thermal expansion, chemical solubility, density, shrinkage) and ISO 13356 (monoclinic phase rate). Biocompatibility testing performed per ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization and irritation), and ISO 10993-11 (acute systemic toxicity). Results support substantial equivalence.
Technological Characteristics
Zirconia-based ceramic (ZrO2, HfO2, Y2O3, Al2O3, SiO2, Fe2O3, Er2O3). Classified as ISO 6872 Type 2 Class 4. Form factor: various shapes/sizes for CAD/CAM jigs. Processing: CAD/CAM milling. Biocompatibility: ISO 10993 compliant.
Indications for Use
Indicated for use in the manufacture of dental prosthetics; no specific patient population, age, or gender restrictions provided.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of three human profiles facing to the right, stacked on top of each other. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 26, 2017
Maxx Digm, Inc. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 800 Roosevelt Ste 417 Irvine, California 92620
Re: K161219
Trade/Device Name: Zmaxx T Series Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: April 28, 2016 Received: April 29, 2016
Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
Page 2 - Priscilla Chung
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Susan Russo, DDS, MA
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name
Zmaxx T Series
Indications for Use (Describe)
Zmaxx T Series is used in the manufacture of dental prosthetics.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|__| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
(K161219)
This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: _ December 27, 2016
#### Applicant / Submitter: 1.
Maxx Digm, Inc.
3265 Kifer Road Santa Clara, CA 95051
Tel: 408-462-9608 Fax: 408-542-9871
#### 2. Submission Correspondent:
Priscilla Chung LK Consulting Group USA, Inc. 800 Roosevelt Ste 417, Irvine CA 92620 Fax: 714-409-3357 Phone: 714-202-5789 Email: juhee.c@lkconsultinggroup.com
#### 3. Device:
| Proprietary Name: | Zmaxx T Series |
|------------------------------|---------------------------------------------|
| Common Name: | Dental Frame Material for Dental Prosthesis |
| Classification Name: | Porcelain Powder for Clinical Use |
| Classification: | Class II, 21 CFR 872.6660 |
| Classification Product Code: | EIH |
#### 4. Predicate Device:
Primary Predicate Device: NATURAZ SERIES (K091096) by The Dental Solutions, Inc. Reference Predicate Device: BruxZir™ Anterior (K143330) by Prismatik Dentalcraft, Inc.
{4}------------------------------------------------
#### 5. Device Description:
Zmaxx T Series is zirconia-based ceramic provided in various shapes such as round and square and it is used to manufacture cores of all ceramic crowns, and is classified into ISO 6872 Type 2 Class 4. It is used to manufacture ceramic restorations through cutting process by dental MAD/MAM, computer-assisted design system, or CAD/CAM system. The subject device offers three different shades to meet the needs of different patients tooth colors. It also offers various shape types to be used with various jigs for CAD/CAM milling machine.
Zmaxx T Series offers various types as follows: W-type, T-type, Z-type, A-type, C-type, B-type, K-type, and Q-type. These different types are to accommodate the different types of the jig that is built-in CAD/CAM machine. User can choose a type which fits the jig they have.
| Type | Size Range | Weight Range |
|--------|--------------------------------|--------------|
| W-type | 98.5mm(D) x 10-25mm(T) | 250-580g |
| T-type | 100mm(D) x 10-25mm(T) | 250-615g |
| D-type | 98mm(D) x 10-25mm(T) | 250-595g |
| Z-type | 95mm(D) x 10-30mm(T) | 230-700g |
| A-type | 90mm(W)×73mm(L)×12-25mm(H) | 210-420g |
| C-type | 15.5-22mm(W)×19-25(L)×20-65(H) | 20-125g |
| B-type | 65(W)×35(L)×12-22mm(H) | 85-155g |
| K-type | 75(W)×45(L)×12-22mm(H) | 127-260g |
| Q-type | 100(W)×100(L)×12-22mm(H) | 380-665g |
#### 6. Indications for Use:
Zmaxx T Series is used in the manufacture of dental prosthetics.
#### 7. Performance Data(Non-Clinical):
The following properties were tested based on the referenced standards. All the test results support substantial equivalence to the predicate devices.
- · ISO 6872 Freedom of Extraneous Materials , Radiation Emission , Flexural Strength,
- · Thermal Expansion Coefficient, Chemical Solubility, Density, Shrinkage Factor
- · ISO 13356 Mono Clinic Phrase Rate
- · ISO 10993-5 Cytotoxicity
- · ISO 10993-10 Sensitization & Irritation
- · ISO 10993-11 Acute Systemic Toxicity
- · Other bench testing Visual Inspection, Dimension Accuracy, Package, Uniformity
#### 8. Substantial Equivalence
The Zmaxx T Series is substantially equivalent to the NATURAZ SERIES (K091096) by The Dental Solutions, Inc. and BruxZir™ Anterior (K143330) by Prismatik Dentalcraft, Inc. The following comparison table is presented to demonstrate substantial equivalence. They share the similar
{5}------------------------------------------------
material and similar physical / chemical properties. However, the mixing ratios of the raw materials are different between the devices.
Despite these differences, the performance test results per ISO 6872 shows that the subject device is substantially equivalent to the predicate devices in physical and chemical properties and meets the necessary requirements. In addition, the subject device has been tested for cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), irritation (ISO 10993-10), and acute systemic toxicity (ISO 10993-11) to meet the biocompatibility requirements.
Based on the comparison between the two devices and the performance testing results presented in this 510K, we conclude that the Zmaxx T Series is substantially equivalent to the predicate devices.
| | Subject Device | Primary Predicate Device | Reference Predicate Deice |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K161219 | K091096 | K143330 |
| Device Name | Zmaxx T Series | NATURAZ SERIES | BruxZir™ Anterior |
| Common Name | Porcelain, Powder for<br>clinical use | Porcelain, Powder for clinical<br>use | Porcelain, Powder for clinical<br>use |
| 510k Applicant | Maxx Digm, Inc. | The Dental Solutions, Inc. | Prismatik Dentalcraft, Inc. |
| Contract<br>Manufacturer | DMAX Co. Ltd. | DMAX Co., Ltd. | |
| Indication For Use | Zmaxx T Series is used in<br>the manufacture of dental<br>prosthetics. DMAX CO.,<br>LTD proposes that the<br>materials distributed within<br>the United States be labeled | NaturaZ Series is used in the<br>manufacture of dental<br>prosthetics. The Dental<br>Solution, Inc. proposes that<br>the materials distributed<br>within the United States be<br>labeled:<br>"Caution: Federal (US) law<br>restricts the sale of the device<br>to, or on the order of, licensed<br>professionals." | The device is indicated for use<br>by dental technicians in the<br>construction of custom<br>made all ceramic restorations<br>for anterior and posterior<br>location. |
| CTE(coefficient of<br>thermal expansion | $(10.6±0.5)×10^{-6}K^{-1}$<br>(meeting the ISO 6872<br>requirement) | $(11.5±0.5)×10^{-6}K^{-1}$<br>(meeting the ISO 6872<br>requirement) | $11 ×10^{-6}K^{-1}$<br>(meeting the ISO 6872<br>requirement) |
| Flexural strength | >600MPa<br>(meeting the ISO 6872<br>requirement) | >800MPa<br>(meeting the ISO 6872<br>requirement) | >650MPa<br>(meeting the ISO 6872<br>requirement) |
| Chemical solubility | $<2000 µg/cm²$<br>(meeting the ISO 6872<br>requirement) | $<100 µg/cm²$<br>(meeting the ISO 6872<br>requirement) | $<2000 µg/cm²$<br>(meeting the ISO 6872<br>requirement) |
| Material Component | ZrO2, HfO2, Y2O3, Al2O3,<br>SiO2, Fe2O3, Er2O3 | ZrO2, HfO2, Y2O3, Al2O3, SiO2<br>Fe2O3 | ZrO2, HfO2, Y2O3, Al2O3, SiO2<br>Fe2O3, Er2O3 |
| Shade | A0, A0.5, A2 | A0 | 5 shades |
| Principle of<br>Operation | Install the block on milling<br>machine, upload a designed<br>file and process it | Install the block on milling<br>machine, upload a designed<br>file and process it | Install the block on milling<br>machine, upload a designed file<br>and process it |
{6}------------------------------------------------
| Biocompatibility | Non-toxic and biocompatible<br>(Meeting the ISO 10993-5,<br>10 and<br>10993-11 requirements) | Non-toxic and biocompatible<br>(Meeting the ISO 10993-5, 10<br>and<br>10993-11 requirements) | Non-toxic and biocompatible<br>(Meeting the ISO 10993-3, 5,1<br>and10993-11 requirements) |
|------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
|------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
#### 9. Conclusion:
Based on the testing results, Maxx Digm, Inc. concludes that the Zmaxx T Series are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.