The Coretech Compression System RHB3003 is a sequential compression device which is intended for treatment of patients with the following conditions: - Lymphedema - Venous stasis ulcers - Venous insufficiency - Peripheral edema The device is safe for both home and hospital use.
Device Story
Coretech RHB3003 is a pneumatic compression system for treating lymphedema, venous stasis ulcers, venous insufficiency, and peripheral edema. The system consists of a controller, garments, hoses, and power adapter. The controller inflates and deflates garment chambers in a pre-programmed sequence from distal to proximal, creating a pressure gradient to move excess fluid from impaired lymphatic regions to healthy areas for natural absorption. It operates in sequential gradient pressure or peristaltic modes. Used in hospital, clinical, or home settings, the device is operated by patients or healthcare providers. The provider sets treatment parameters; the device provides automated compression cycles to improve blood flow and microcirculation. Benefits include reduced edema and improved lymphatic drainage. The controller features an LED touch interface and allows pressure adjustments to ensure patient comfort.
Clinical Evidence
Bench testing only. Testing included electrical safety and EMC (IEC 60601-1, IEC 60601-1-11, IEC 60601-1-6), software verification/validation (IEC 62304), and mechanical bench testing (inflation/deflation performance, cycle time accuracy, pressure calibration, burst pressure, garment integrity, and noise performance).
Technological Characteristics
Pneumatic compression system; controller enclosure is plastic. Operates via microprocessor-controlled analog/digital electronics. Pressure range 20-80 mmHg. Garments made of nylon fabric with 4 or 8 chambers. Power: 100-127V/220-240V, 50/60 Hz. Connectivity: None (standalone). Software: Moderate level of concern. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-6, IEC 62304.
Indications for Use
Indicated for patients with lymphedema, venous stasis ulcers, venous insufficiency, or peripheral edema. Safe for home and hospital use.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 8, 2025
Vive Health LLC
% Prithul Bom
Most Responsible Person
Regulatory Technology Services, LLC
1000 Westgate Drive, Suite 510k
Saint Paul, Minnesota 55114
Re: K250817
Trade/Device Name: Coretech Compression System (Coretech RHB3003)
Regulation Number: 21 CFR 870.5800
Regulation Name: Compressible Limb Sleeve
Regulatory Class: Class II
Product Code: JOW
Dated: June 13, 2025
Received: June 13, 2025
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250817 - Prithul Bom
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250817 - Prithul Bom
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Nicole M. Gillette -S
Nicole Gillette
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250817
Device Name
Coretech Compression System RHB3003 (Coretech RHB3003)
Indications for Use (Describe)
The Coretech Compression System RHB3003 is a sequential compression device which is intended for treatment of patients with the following conditions:
- Lymphedema
- Venous stasis ulcers
- Venous insufficiency
- Peripheral edema
The device is safe for both home and hospital use.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
Coretech Compression System RHB3003 (Vive Health)
510(K) Summary
# 510(K) Summary K250817
# Vive Health Coretech Compression System RHB3003
Prepared July 31, 2025
## Submitter
Manufacturer: Vive Health LLC
8955 Fontana Del Sol Way
Naples FL 34109
Contact Person: Joe Fleming - President
Joe.fleming@igbrands.com
239-248-2026
## Device
Device Name: Coretech Compression System (model RHB3003)
Common / Usual Name: Pneumatic compression device
Classification Name: Compressible limb sleeve (21 CFR 870.5800)
Product Code: Sleeve, limb, compressible (JOW)
Device Regulatory Class: Class 2
## Predicate
Device Name: CircuFlow 5150
Premarket Notification Number: K123959
## Reference
Device Name: Airos 6P
Air Compression Therapy System
Premarket Notification Number: K223195
K201982
## Device Description
The Coretech Compression System RHB3003 (Coretech RHB3003) is a compression device that delivers sequential compression to the affected upper and lower extremity of patients with lymphedema, chronic edema, peripheral edema, venous insufficiency and venous stasis ulcers.
This device helps direct and move excess fluid from an impaired lymphatic region to healthy regions, where fluid can be absorbed and processed naturally by the body. The Coretech RHB3003 controller is used to inflate the connected garment chambers. The air chambers inflate and deflate repeatedly and orderly to create circular pressure on the affected limb. The air inflates from the distal end to the proximal end of the patient. This helps to promote the flow of blood and lymph and improves microcirculation. The pressure can be adjusted to avoid discomfort to the patient and deflates as one cycle.
Page 1 of 7
{5}
Coretech Compression System RHB3003 (Vive Health)
510(K) Summary
# Indications for Use
The Coretech RHB3003 is a sequential compression device which is intended for treatment of patients with the following conditions:
- Lymphedema
- Venous stasis ulcers
- Venous insufficiency
- Peripheral edema
The device is safe for both home and hospital use.
The subject and predicate devices have the same intended use and are both prescription devices.
# Comparison of Technological Characteristics with the Predicate Device
The Coretech RHB3003 device and the predicate Airos 6P are the same device type and utilize the same fundamental technology, mode of action, and principles of operation. They are based on the same basic design, software, hardware, and materials, and device features are similar. They both require a power source for operation. A high-level comparison is provided below.
{6}
Coretech Compression System RHB3003 (Vive Health)
510(K) Summary
| | Coretech RHB3003
(subject device) | CircuFlow 5150
(predicate device) | Airos 6P
(reference device) | Comparison |
| --- | --- | --- | --- | --- |
| Product Type | Pneumatic compression device | Pneumatic compression device | Pneumatic compression device | Same |
| Components | Controller
Garments
Hoses
Power adapter | Controller
Garments
Hoses
Power adapter | Controller
Garments
Hoses
Power adapter | Same |
| Mode of Action | Helps direct and move excess fluid from impaired lymphatic regions to healthy regions, where fluid can be absorbed and processed naturally by the body. | Helps direct and move excess fluid from impaired lymphatic regions to healthy regions, where fluid can be absorbed and processed naturally by the body. | Helps direct and move excess fluid from impaired lymphatic regions to healthy regions, where fluid can be absorbed and processed naturally by the body. | Same |
| Principles of Operation | The controller inflates the connected garment chambers in a pre-programmed sequence from distal to proximal with a pressure gradient that provides high distal pressures. The garment chambers are sequentially inflated, pressure is held, and then all chambers are deflated simultaneously. | The controller inflates the connected garment chambers in a pre-programmed sequence from distal to proximal with a pressure gradient that provides high distal pressures. The garment chambers are sequentially inflated, pressure is held, and then all chambers are deflated simultaneously. | The controller inflates the connected garment chambers in a pre-programmed sequence from distal to proximal with a pressure gradient that provides high distal pressures. The garment chambers are sequentially inflated, pressure is held, and then all chambers are deflated simultaneously. | Same |
| Environment of Use | Hospital, clinical, and home settings | Hospital, clinical, and home settings | Hospital, clinical, and home settings | Same |
| Controller Size and Weight | 9" W x 4" D x 8" H
10 lb | 11.0" x 6.0" x 8.0"
4 lb | 10.2" W x 10.5" D x 4.3" H
6.4 lb | Similar; the subject device has a slightly smaller but heavier controller than the |
{7}
Coretech Compression System RHB3003 (Vive Health)
| | Coretech RHB3003
(subject device) | CircuFlow 5150
(predicate device) | Airos 6P
(reference device) | Comparison |
| --- | --- | --- | --- | --- |
| | | | | predicate. |
| Electrical Requirements | 100-127V/220-240V, 50/60 Hz | 100-240V, 47/63 Hz | 100-240VAC, 50/60 Hz | Similar to predicate and reference device |
| Controller Enclosure Material | Plastic | Plastic | Plastic | Same |
| User Interface | Push buttons, LED touch control | Push button controls | Push buttons and Digital LCD screen | Similar; both devices utilize push buttons. The subject device is LED touch control while the predicate is only push-button control. |
| Wireless Functions | None | None | None | Same |
| Software / Hardware Type | Analog and digital electronics with microprocessor | Analog and digital electronics with microprocessor | Analog and digital electronics with microprocessor | Same |
| Software-Concern Level | Moderate level of concern | Moderate level of concern | Moderate level of concern | Same |
| Pressure Range | 20-80 mmHg | 20-80 mmHg | 30-80 mmHg | Same as predicate and similar to the reference device. |
| Output | - Sequential gradient pressure mode
- Peristaltic mode | - Sequential gradient pressure mode | - Sequential gradient pressure mode
- Peristaltic mode | Same as the reference device. The predicate device only includes sequential gradient pressure while the subject and reference device include a |
Page 4 of 7
{8}
Coretech Compression System RHB3003 (Vive Health)
| | Coretech RHB3003
(subject device) | CircuFlow 5150
(predicate device) | Airos 6P
(reference device) | Comparison |
| --- | --- | --- | --- | --- |
| | | | | peristaltic mode. |
| Cycle Time | 82 seconds for both modes | Unknown | 40-60 seconds for gradient Manufacturer identifies "none" for peristaltic | The predicate device cycle time is unknown. The subject device cycle time is slightly longer than the reference device gradient cycle mode. |
| Dwell Time | 15 seconds for gradient mode and 15 seconds for peristaltic mode | 25 seconds | 12 seconds for gradient mode; 5 seconds for peristaltic mode | Different; the subject device dwell time is slightly shorter than the predicate and longer than the reference device. |
| Treatment Time | 30, 45 or 60 minutes | 1 hour maximum | 15, 30, 45, or 60 minutes | Similar; the subject device, predicate and reference device have treatment durations up to 60 minutes |
| Garment Chamber Pressure Control | Pressure based | Pressure based | Pressure based | Same |
| Garment-Controller Connections | Fixed hose connection | Fixed hose connection | Fixed hose connection | Same |
| Garment Port | Standard port | Standard port | Standard port | Same |
| Garment Materials | Nylon fabric | Nylon fabric | Nylon fabric | Same |
Page 5 of 7
{9}
Coretech Compression System RHB3003 (Vive Health)
| | Coretech RHB3003
(subject device) | CircuFlow 5150
(predicate device) | Airos 6P
(reference device) | Comparison |
| --- | --- | --- | --- | --- |
| Garment Chambers | 4 or 8 | 8 | 6 | Different; the subject device may be used utilizing a 4 or 8 chamber garment while the predicate is an 8-chamber garment. The reference device utilizes a 6-chamber garment. |
| Device Cleanability | Cleanable using commonly available mild household products and disinfectants. | Cleanable using commonly available mild household products and disinfectants. | Cleanable using commonly available mild household products and disinfectants. | Same |
Page 6 of 7
{10}
Coretech Compression System RHB3003 (Vive Health)
510(K) Summary
Technological differences between the Coretech RHB3003 device and the predicate device, include:
- Smaller but heavier controller
- LED touch control user interface
- Inclusion of peristaltic mode
- Electrical requirements
- Shorter dwell time
- Treatment time
- Number of garment chambers
These differences do not raise any different questions of safety or effectiveness for the Coretech RHB3003 system compared to the predicate device.
## Performance Data
Based on the differences compared to the predicate device, the following types of testing have been completed for the Coretech RHB3003 device to support the substantial equivalence determination:
- Electrical safety and electromagnetic compatibility (EMC) testing, following IEC 60601-1 Medical electrical equipment – General requirements for basic safety and essential performance, IEC 60601-1-11 - General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment and IEC 60601-1-6 General requirements for basic safety and essential performance – Collateral Standard: Usability
- Software verification and validation testing, following IEC 62304 Medical device software – Software life cycle processes
- Mechanical bench testing, including:
- Start / stop function performance
- Inflation and Deflation function performance
- Cycle time accuracy and performance
- Operating modes, pressure gradient performance
- Pressure calibration performance and accuracy
- Operating / treatment time / therapeutic performance
- Pressure performance and accuracy
- Burst pressure performance / garment Integrity / seam performance / leak test
- Complete firmware / software Integration performance
- Packaging validation
- Noise performance
- Packaging Validation
The technological comparison and performance testing together demonstrate that Coretech RHB3003 has a similar safety and effectiveness profile as the predicate device.
## Conclusion
The Coretech RHB3003 device is substantially equivalent to the legally marketed predicate device.
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.