Air compression therapy system FO-3001; Air compression therapy system FO-3008
K201982 · Foshan Hongfeng Co., Ltd. · IRP · Mar 25, 2021 · Physical Medicine
Device Facts
Record ID
K201982
Device Name
Air compression therapy system FO-3001; Air compression therapy system FO-3008
Applicant
Foshan Hongfeng Co., Ltd.
Product Code
IRP · Physical Medicine
Decision Date
Mar 25, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Air compression therapy system Models FO-3001 and FO-3008: Intended for home to temporarily relieve minor muscle aches and/or pains, promote blood circulation in the treated areas.
Device Story
Air compression therapy system (models FO-3001, FO-3008) consists of air pump, air pressure sensor, and inflatable sleeves connected via hoses. Device operates by sequentially inflating and deflating air chambers in sleeves to provide compression massage from limb end to body center. System controlled by microprocessor and sensor; user adjusts pressure settings. Used in home environment by patients for temporary relief of minor muscle aches/pains and circulation promotion. Output is mechanical compression massage. Benefits include improved blood circulation and muscle pain relief. Safety feature includes power button to stop therapy session.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2) and biocompatibility (ISO 10993-5, ISO 10993-10, ISO 10993-12).
Technological Characteristics
Molded ABS enclosure; compressor and valve system for sequential inflation; 4, 6, or 8 chambers; pressure range 30-110mmHg; power 100-240V AC; microprocessor-controlled; standards: IEC 60601-1:2005, IEC 60601-1-11:2015, IEC 60601-1-2:2014, ISO 10993-5:2009, ISO 10993-10:2010, ISO 10993-12:2012.
Indications for Use
Indicated for home use to temporarily relieve minor muscle aches and/or pains and promote blood circulation in treated areas.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
March 25, 2021
Foshan Hongfeng Co., Ltd % Sam Lin Official Correspondent Shanghai Spica Management Consulting Co., Ltd. 609 Room,No.133 Shengang Avenue, Pudong New District Shanghai. 201306 China
Re: K201982
Trade/Device Name: Air compression therapy system FO-3001; Air compression therapy system FO-3008 Regulation Number: 21 CFR 890.5650
Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: July 17, 2020 Received: July 17, 2020
Dear Sam Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jitendra Virani Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K201982
#### Device Name
Air compression therapy system FO-3001; Air compression therapy system FO-3008
#### Indications for Use (Describe)
Air compression therapy system Models FO-3001 and FO-3008: Intended for home to temporarily relieve minor muscle aches and/or pains, promote blood circulation in the treated areas.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Type of Submission | Traditional |
|---------------------------|-----------------------------------------------------------------------------------|
| Date Prepared | February 4, 2021 |
| Submission Sponsor | |
| Manufacturer Name | FOSHAN HONGFENG CO.,LTD. |
| Address | No.4-2 Leqiang Road, Leping Sanshui, Foshan,<br>Guangdong, China |
| Tel | 86-0757-8392028 |
| Email | 573619164@qq.com |
| Contact Person | Dongfeng Cheng |
| Device Identification | |
| Trade Name | Air compression therapy system FO-3001;<br>Air compression therapy system FO-3008 |
| Regulation Number | 21 CFR 890.5650 |
| Regulation Name | Power inflatable tube massager |
| Device Classification | Class II |
| Product Code | IRP |
| Panel | Physical Medicine |
| Previous Submissions | None |
| Application Correspondent | |
| Company Name | Shanghai Spica Management Consulting Co.,Ltd. |
| Address | 609 Room, No.133 Shengang Avenue, Pudong New<br>District, Shanghai, China |
| Tel | 86-15626132181 |
| Email | sam@spicagloble.com |
| Contact Person | Sam Lin |
Air compression therapy system Models FO-3001 and FO-3008: Intended for home to temporarily relieve minor muscle aches and/or pains, promote blood circulation in the treated areas.
#### Device Description
Indications for Use of the Device
Air compression therapy system Models FO-3001 and FO-3008: Air Pressure Therapy System is Section[006] Page 1 of 13
{4}------------------------------------------------
consist of air pressure sensor, air pump, sleeves etc working together as one unit. The air pump is connected to the dedicated sleeves via a series of hoses. The compression massage direction is from limb end to body center by inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient. The sleeve works under the action of sensor and microprocessor.
Air Pressure Therapy System, in medical market, it has a sequential squeezing from distal to proximal, thus help to improve the circulation of blood.
### Performance Testing - Clinical
Not Applicable.
Performance Testing - Animal
Not Applicable.
[Remainder of page left intentionally blank]
{5}------------------------------------------------
## Table 6A: Predicate Device Information
| No. | Regulation<br>Number | 510k<br>Number | Sponsor | Predicate Device<br>Name | Rx or<br>OTC | Treatment<br>Time | Pressure<br>Range |
|-----|----------------------|----------------|----------------------------|---------------------------------------------|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| 1 | 21 CFR<br>890.5650 | K160608 | NormaTec<br>Industries, LP | NormaTec Pulse<br>and NormaTec<br>Pulse Pro | OTC | Stays on until the user turns it off or can be set up to<br>turn off in a range of 10 mins to continuous / User<br>controlled 10 minutes to 175 minutes or continuous –<br>total time over 4 segments | 30-110mmHg |
### Table 6B: Differences between Models
| Characteristic | Air Compression Therapy System: FO-3001 | Air Compression Therapy System: FO-3008 |
|--------------------------------|-----------------------------------------|-----------------------------------------|
| Indications for use | × | × |
| OTC | × | × |
| Device Pressure Range | × | × |
| Treatment Time | × | × |
| Standard | × | × |
| Mode of Compression | × | × |
| Power Source | × | × |
| Power Consumption | × | × |
| Size and appearance of sleeves | × | × |
| Housing Materials | × | × |
Note: *"×" means same.
Section[006]
Page 3 of 13
{6}------------------------------------------------
| Safety Feature | | |
|--------------------|-----------------------------------------|-----------------------------------------|
| Safety Feature | X | X |
| Technology | X | X |
| Dimensions | Different, see the Table 6C for details | Different, see the Table 6D for details |
| Photo | Different, see the Table 6C for details | Different, see the Table 6D for details |
| Number of Chambers | Different, see the Table 6C for details | Different, see the Table 6D for details |
| Work Mode | Different, see the Table 6C for details | Different, see the Table 6D for details |
# Table 6C: Summary of Comparison of FO-3001
| | Subject Device | Predicate Device | Differences Discussion |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Air compression therapy system: FO-3001 | NormaTec Pulse and NormaTec Pulse<br>Pro | N/A |
| 510(k) number | K160608 | K160608 | N/A |
| Manufacturer | FOSHAN HONGFENG CO.,LTD. | NormaTec Industries, LP | N/A |
| Product Regulation | 21 CFR 890.5650 | 21 CFR 890.5650 | Same |
| Classification Name | Massager, Powered Inflatable Tube | Massager, Powered Inflatable Tube | Same |
| Regulation Class | 2 | 2 | Same |
| Product Code | IRP | IRP | Same |
| Indications for<br>use | Intended for home to temporarily relieve<br>minor muscle aches and/or pains, promote<br>blood circulation in treated areas. | The NormaTec Pulse and Pulse Pro is<br>an air pressure massager intended to<br>temporarily relieve minor muscle aches<br>and/or pains, and to temporarily<br>increase circulation to the treated areas. | Same |
| Rx or OTC | OTC | OTC | Same |
| Pressure range | 30-110mmHg | 30-110mmHg | Same |
| Treatment<br>Time | 1-60mins | Stays on until the user turns it off or<br>can be set up to turn off in a range of<br>10 mins to continuous / User controlled<br>10 minutes to 175 minutes or<br>continuous -total time over 4<br>segments. | Similar<br>The treatment time of subject device is<br>smaller than predicate device (K160608)<br>so the difference of treatment time would<br>not raise adversely impact on safety and<br>effectiveness. |
| Standard | IEC 60601-1-11:2015<br>IEC 60601-1-2:2014<br>IEC 60601-1:2005<br>ISO 10993-10:2010<br>ISO 10993-5:2009<br>ISO 10993-12:2012 | ES 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-1-11 | Similar |
| Mode of<br>Compression | Sequential | Sequential Gradient, Peristaltic and<br>Pulsing | Same |
| Power Source | 100-127V/220-240V, 50/60Hz | 12 VDC via an IEC 60601-1 compliant<br>power supply (100-240 VAC input) | Similar |
| | | Optional Integrated rechargeable battery | |
| Power<br>Consumption | 65W | 14W | Similar |
| Dimensions<br>(W*H*D) | 220*190*113MM | 4" x 5" x 9" | Similar |
| Photo | Image: Photo of a black device with buttons and a handle | Image: Photo of a black device with a screen | Similar |
| Size and<br>appearance of<br>sleeves (leg<br>part) | M:91*65cm L:100*74cm<br>XL:110*70cm (Overlapping) | Short: 14" x 43"<br>Standard: 14" x 48"<br>Tall: 14" x 60" | Similar |
| Housing<br>Materials | Molded ABS enclosure | Molded ABS enclosure | Same |
| Number of<br>Chambers | 4,6,8 Chambers for each unit | 5 or less | Similar |
| Work Mode | Image: Diagram of different work modes | Sequential mode:<br>Starting with the distal chamber and<br>progressing up the proximal chamber, each section compresses and the pressure gradually rises to the | Although the subject device provides six<br>kinds of work mode, the Mode A is the<br>same with predicate device (K160608)<br>while the other work modes of subject the device just have difference about the |
| 6 chambers: | | pre-determined air pressure level, then | inflatable order of the different chambers |
| A (Normal Mode): Chamber ① inflating till<br>setup pressure or for 2 seconds, then hold air<br>for 2 seconds, start deflating; chamber ② start<br>like ①. Same way till chamber ⑥, pause for 3<br>seconds, then restart chamber ①②③④⑤⑥<br>again. | | decompresses and the air pressure<br>drops.<br>Once the top section decompresses, the<br>cycle begins again. again. | The treatment pressure range are the same<br>under different work modes, so the<br>difference of pressure range would not raise<br>adversely impact on safety and<br>effectiveness. |
| B (Sequential Squeeze Mode): chamber ①<br>inflating till set up pressure or for 28 seconds,<br>then hold the pressure, chamber ② inflating,<br>till setup pressure or for 28 seconds, then<br>chamber ①② hold pressure in same time,<br>then chamber ③ start inflating, same way till<br>after chamber ⑥. Chamber ①②③④⑤⑥<br>deflating in same time for 3 seconds. Then<br>repeat. | | Normatec Pulse mode:<br>Starting with the distal zone and<br>progressing up the proximal zone, one<br>zone compresses and the pressure<br>gradually rises to the pre-determined<br>air pressure level, holds the air of<br>previous two zone, the other zones do<br>not hold, until the last zone finished,<br>deflate the all last three zone then enter<br>into next cycle. | |
| C (Double Wave Mode: chamber ①②<br>inflating till setup pressure or for 40 seconds,<br>hold air for 2 seconds, then start deflating.<br>Chamber ③④ start inflating till setup pressure<br>or for 40 seconds, hold air for 2 seconds, then<br>deflating, same way for chamber ③④ till<br>chamber ⑤⑥, pause for 3 seconds. Then<br>repeat. | | | |
| D (Whole Squeeze Mode): chamber<br>①②③④⑤⑥ inflating at the same time till<br>setup pressure or for 90 seconds, then<br>deflating in the same time for 3 seconds. Then | | | |
| | Image: Diagram of air pressure | | |
| | Image: Diagram of air pressure | | |
| | | | |
| | repeat.<br>E (Combined B + C): sequential squeeze +<br>double wave<br>F (Combined A + C + D): normal + double<br>wave + whole squeeze.<br>4 and 8 chambers same as 6 chambers above. | | |
| Safety feature | Power button on main unit allows user to stop<br>therapy session at any time | Button on display allows user to stop<br>or pause therapy session at any time | Same |
| Technology | Compressor and valve system which<br>sequentially inflates inflatable chambers | Compressor and valve system which<br>sequentially inflates inflatable chambers | Same |
{7}------------------------------------------------
Section[006]
Page 5 of 13
{8}------------------------------------------------
Section[006]
Page 6 of 13
{9}------------------------------------------------
Section[006]
Page 7 of 13
{10}------------------------------------------------
## Table 6D: Summary of Comparison of FO-3008
| | Subject Device | Predicate Device | Differences Discussion |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Air compression therapy system: FO-3008 | NormaTec Pulse and NormaTec Pulse Pro | N/A |
| 510(k) number | | K160608 | N/A |
| Manufacturer | FOSHAN HONGFENG CO.,LTD. | NormaTec Industries, LP | N/A |
| Product<br>Regulation | 21 CFR 890.5650 | 21 CFR 890.5650 | Same |
| Classification<br>Name | Massager, Powered Inflatable Tube | Massager, Powered Inflatable Tube | Same |
| Regulation<br>Class | 2 | 2 | Same |
| Product Code | IRP | IRP | Same |
| Indications for<br>use | Intended for home to temporarily relieve<br>minor muscle aches and/or pains, promote<br>blood circulation in treated areas. | The NormaTec Pulse and Pulse Pro is an air pressure<br>massager intended to temporarily relieve minor<br>muscle aches and/or pains, and to temporarily increase<br>circulation to the treated areas. | Same |
| Rx or OTC | OTC | OTC | Same |
| Pressure range | 30-110mmHg | 30-110mmHg | Same…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.