K250634 · Fresenius Medical Care Renal Therapies Group, LLC · DSB · Nov 24, 2025 · Cardiovascular
Device Facts
Record ID
K250634
Device Name
BCM2-Body Composition Monitor
Applicant
Fresenius Medical Care Renal Therapies Group, LLC
Product Code
DSB · Cardiovascular
Decision Date
Nov 24, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
The BCM2 is intended for use, under the direction of a physician, for the noninvasive intermittent measurement of fluid status in patients with end-stage kidney disease (ESKD) including those receiving maintenance dialysis. The device does not generate any real-time alarms and outputs from the device should be used in the context of all clinical data to assess a patient's fluid status. The device is intended for use in patients 18 years and older.
Device Story
BCM2 is a multifrequency bioimpedance spectroscopy device for assessing fluid status in ESKD patients. Clinicians (nurses/technicians) apply electrodes to patients in dialysis clinics or hospitals. Device applies low-level signal (50–800 μA) at 50 frequencies (5–1,000 kHz) to measure impedance. System calculates overhydration (OH), urea distribution volume (V), extracellular water (ECW), intracellular water (ICW), and total body water (TBW). Output includes Cole-Cole plot, patient demographics, and fluid parameters. Clinicians interpret outputs alongside other clinical data to guide fluid management decisions. Benefits include objective, noninvasive assessment of hydration status to assist in dialysis treatment planning. Device features touchscreen console, battery/AC power, and calibration/test boxes for verification.
Clinical Evidence
No clinical trials performed. Evidence consists of a literature review supporting device validation, accuracy, and clinical utility for the intended patient population, supplemented by bench testing, software verification/validation, and biocompatibility testing.
Technological Characteristics
Multifrequency bioimpedance spectroscopy device. Operates at 50 frequencies from 5 kHz to 1,000 kHz. Signal intensity 50–800 μA. Components: touchscreen console, disposable electrodes, calibration/test boxes. Power: AC or battery. Biocompatibility per ISO 10993-1. Electrical safety/EMC per IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for patients 18+ with end-stage kidney disease (ESKD), including those receiving maintenance dialysis, for noninvasive intermittent fluid status measurement.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 24, 2025
Fresenius Medical Care Renal Therapies Group, LLC
Christopher Doherty
Regulatory Affairs Fellow
920 Winter St
Waltham, Massachusetts 02451
Re: K250634
Trade/Device Name: BCM2-Body Composition Monitor
Regulation Number: 21 CFR 870.2770
Regulation Name: Impedance Plethysmograph
Regulatory Class: Class II
Product Code: DSB
Dated: October 27, 2025
Received: October 27, 2025
Dear Christopher Doherty:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250634 - Christopher Doherty
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250634 - Christopher Doherty
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Maura Rooney -S
Maura Rooney
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250634
Device Name
BCM2-Body Composition Monitor
Indications for Use (Describe)
The BCM2 is intended for use, under the direction of a physician, for the noninvasive intermittent measurement of fluid status in patients with end-stage kidney disease (ESKD) including those receiving maintenance dialysis.
The device does not generate any real-time alarms and outputs from the device should be used in the context of all clinical data to assess a patient's fluid status.
The device is intended for use in patients 18 years and older.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
FRESENIUS MEDICAL CARE
K250634
510(k) Summary
BCM2 Traditional 510(k)
# 1. 510(K) SUMMARY
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR § 807.92
## 1.1. Submitter’s Information
Name: Fresenius Medical Care Renal Therapies Group, LLC
Address: 920 Winter Street
Waltham, MA 02451-1457
Phone: (781) 996-9103
Contact Person: Christopher Doherty, Regulatory Affairs Fellow
Phone: (781) 296-4663
Preparation Date: 03 March 2025
## 1.2. Device Name
Trade Name: BCM2 – Body Composition Monitor
Common Name: Fluid Status Monitor
Regulation Name: Impedance Plethysmograph
Regulatory Class: Class II per 21 CFR § 870.2770
Product Code: DSB
Product Code Name: Plethysmograph, Impedance
FDA Review Panel: Cardiovascular
## 1.3. Legally Marketed Predicate Device
### 1.3.1. Predicate Device – Intersection Medical Inc. IMED-Z (K142503)
The predicate device for the BCM2 - Body Composition Monitor (hereinafter referred to as the “BCM2”) is the IMED-Z Fluid Status Monitor (K142503). The IMED-Z has not been the subject of a design-related recall.
### 1.3.2. Reference Device – Xitron Hydra 4000
The Xitron Hydra 4000 (K904109) is being used as a reference device to demonstrate the equivalence of technological characteristics (i.e., bioimpedance spectroscopy).
## 1.4. Device Description
The BCM2 is a multifrequency bioimpedance device that can be used by a clinician for the noninvasive intermittent measurement of fluid status in patients with end-stage kidney disease, including those receiving maintenance dialysis. A physician or designated healthcare professional is responsible for interpreting data from the device to determine what action is required as a result of changes in the measurements.
{5}
FRESENIUS MEDICAL CARE
K250634
510(k) Summary
BCM2 Traditional 510(k)
To assess a patient’s hydration status, the BCM2 measures impedance by applying a low-level signal (50–800 μA) for less than 10 seconds at 50 different frequencies from 5 kHz to 1,000 kHz. Measurements are performed by dialysis clinicians (e.g., dialysis nurses, patient care technicians) in a healthcare environment such as a dialysis clinic or hospital.
In the normal clinical workflow, the BCM2 provides the following output parameters:
- Overhydration (OH)
- Urea distribution volume (V)
The following calculated and derived parameters are also available:
- Extracellular water (ECW)
- Intracellular water (ICW)
- Total body water (TBW)
The BCM2 system consists of the BCM2 touchscreen console and the electrode set. The BCM2 console powers and measures the bioimpedance spectroscopy frequencies to assess fluid parameters.
The device can also be operated in battery mode. Battery mode provides flexibility when moving between patients. The battery charge status is shown in the upper corner of the display regardless of which power source is being used. The power supply connection is located on the rear of the console.
The Calibration Box employs different resistors to calibrate the entire range of the BCM2’s measurement (5 kHz to 1,000 kHz). Impedance measurements are performed for each resistor for all frequencies and the data is verified against an expected tolerance range. After the BCM2 has been calibrated, the Test Box is used to verify that the device is functioning properly.
The BCM2 will be available in one (1) configuration for sale within the U.S.
## 1.5. Intended Use
The BCM2 device provides objective intermittent measurement of parameters that assist the healthcare professional (HCP) in the assessment and monitoring of the patient’s hydration and body composition.
In the normal clinical workflow, the device uses the following output parameters:
- Overhydration (OH)
- Urea distribution volume (V urea)
Other calculated and derived parameters are also made available.
The device will also display the Cole-Cole plot and the subject’s weight, height, age, and sex.
## 1.6. Indications for Use
The BCM2 device is intended for use, under the direction of a physician, for the noninvasive intermittent measurement of fluid status in patients with end-stage kidney disease (ESKD) including those receiving maintenance dialysis.
Page 2
{6}
FRESENIUS MEDICAL CARE
K250634
510(k) Summary
BCM2 Traditional 510(k)
The device does not generate any real-time alarms, and outputs from the device should be used in the context of all clinical data to assess a patient's fluid status.
The device is intended for use in patients 18 years and older.
## 1.7. Comparison of Technological Characteristics with the Predicate Device
The following technological characteristics of the BCM2 are substantially equivalent to the predicate device IMED-Z Fluid Status Monitor (K142503):
- Intended Use
- Indications for Use
- Fundamental Scientific Technology/Operating Principle
- Technological Characteristics
The following technological characteristic of the BCM2 is substantially equivalent to the predicate device Xitron Hydra 4000 (K904109):
- Bioimpedance Spectroscopy
## 1.8. Performance Data
The following performance tests were conducted on the BCM2 to support the determination of substantial equivalence:
- Software Verification and Validation
- Functional Design Verification
- Electrical Safety and Electromagnetic Compatibility (EMC)
- Simulated Shipping and Distribution
- Biological Safety (Biocompatibility)
- Cytotoxicity
- Sensitization
- Irritation
- Toxicological Review
All testing met predetermined acceptance criteria.
The BCM2's performance characteristics were demonstrated through bench testing, mechanical integrity testing, electrical safety and electromagnetic compatibility testing, and software testing.
Testing demonstrated that the integrity, durability, and reliability of the disposable electrodes were maintained over their intended shelf life.
The BCM2 is safe and effective for its intended use.
{7}
FRESENIUS MEDICAL CARE
K250634
510(k) Summary
BCM2 Traditional 510(k)
# 1.8.1. Biocompatibility Testing
In accordance with FDA guidance document Use of International Standard ISO 10993, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (September 4, 2020), the BCM2 electrodes are classified as a surface, intact skin, and limited exposure (≤ 24 hours) device.
The BCM2 electrodes were evaluated for biocompatibility in accordance with the requirements of:
- ISO 10993-1:2018, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (September 4, 2020)
Based on the results of the biocompatibility tests performed on the BCM2 electrodes, FMCRTG has concluded that the electrodes are biologically safe for their intended use.
The following tests were performed to support the biological safety of the BCM2 electrodes (Table 1).
Table 1: Biocompatibility Performance Tests
| Biological Safety Endpoint | Test | Acceptance Criteria | Standard | Result |
| --- | --- | --- | --- | --- |
| Sensitization | ISO Closed-patch (Buehler) Test (GLP) | Non-Sensitizer
Passes when numerical grade ≤ 1 (If grade ≤ 1 is observed for controls) | ISO 10993-10:2010 [Rec. Number 2-174] | Passed |
| Irritation | ISO Primary Irritation Study (Animal Irritation) (GLP) | Non-Irritant
Passes when difference between test extract overall mean and corresponding control overall mean is ≤ 1 | | Passed |
| | ISO Primary Irritation Study (Animal Irritation) (GLP) | Non-Irritant
Passes when difference between test extract overall mean and corresponding control overall mean is ≤ 1 | | Passed |
| | ISO Primary Irritation Study (Animal Irritation) (GLP) | Non-Irritant
Passes when difference between test extract overall mean and corresponding control overall mean is ≤ 1 | | Passed |
Page 4
{8}
K250634
510(k) Summary
BCM2 Traditional 510(k)
| Biological Safety Endpoint | Test | Acceptance Criteria | Standard | Result |
| --- | --- | --- | --- | --- |
| | ISO Primary Irritation Study (Animal Irritation) (GLP) | Non-Irritant
Passes when difference between test extract overall mean and corresponding control overall mean is ≤ 1 | | Passed |
## 1.8.1.1. In Vitro Cytotoxicity Testing
The electrode was cytotoxic in *in vitro* cytotoxicity testing.
Additional tests were conducted to fully characterize potential biocompatibility risk to the patient. These tests included the Animal Irritation Test and Closed Patch (Buehler) Test in accordance with ISO 10993-10, both of which had passing results with no response. As described in ISO 10993-5, a positive response does not necessarily indicate toxicity. This result shows that when the electrodes are used on intact skin in a live animal model there were no adverse responses.
## 1.8.2. Human Factors Validation Testing
The BCM2 was validated for its safe and effective use in accordance with FDA guidance document *Applying Human Factors and Usability Engineering to Medical Devices* (03 February 2016).
## 1.8.3. Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical Safety testing was conducted in accordance with IEC 60601-1-2:2014, and IEC 60601-1-2:2014/AMD1:2020. The BCM2 met acceptance criteria for this testing.
The BCM2 was evaluated for electromagnetic compatibility (EMC) in accordance with IEC 60601-1-1:2005, IEC 60601-1:2005/AMD1:2012, and IEC 60601-1:2005/AMD2:2020. The BCM2 met the EMC acceptance criteria.
## 1.8.4. Software Verification and Validation Testing
Unit, integration, regression (system verification), functional, and system testing were performed to demonstrate the effectiveness of the software and to confirm operation of the device. Software verification information within this submission is provided in accordance with the following:
- Content of Premarket Submissions for Device Software Functions (14 June 2023)
- Off-The-Shelf Software Use in Medical Devices (11 August 2023)
- Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (27 September 2023)
- Select Updates for the Premarket Cybersecurity Guidance: Section 524B of the FD&C Act (13 March 2024)
- Applying Human Factors and Usability Engineering to Medical Devices (3 February 2016)
All testing met predetermined acceptance criteria.
The BCM2 is safe and effective for its intended use.
Page 5
{9}
FRESENIUS MEDICAL CARE
K250634
510(k) Summary
BCM2 Traditional 510(k)
## 1.8.6 Animal Studies
No animal studies were performed for the BCM2.
## 1.8.7 Clinical Studies
A literature review was provided in support of the validation and accuracy of the BCM2 device, as well as clinical evidence for the determination of substantial equivalence, the proposed Indications for Use, and the Intended Use in the relevant patient populations.
## 1.9 Conclusion
The Indications for Use, design, principle of operation, and technological characteristics of the BCM2 and its components are substantially equivalent to those of the predicate and secondary predicate devices. Test results demonstrate that the differences between the BCM2, the predicate, and secondary predicate do not raise new concerns with regard to safety or efficacy. FMCRTG concludes that within the meaning of the Medical Device Amendments Act of 1976, the BCM2 is safe and effective for its intended use.
Page 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.