K142503 · Intersection Medical, Inc. · DSB · Jun 5, 2015 · Cardiovascular
Device Facts
Record ID
K142503
Device Name
IMED-Z FLUID STATUS MONITOR
Applicant
Intersection Medical, Inc.
Product Code
DSB · Cardiovascular
Decision Date
Jun 5, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
The IMED-Z Fluid Status Monitor is intended for patients: - With fluid management problems - Taking diuretic medication - Living with Heart Failure ● - Living with End-stage Renal Disease ● - Recovering from Coronary artery Disease related event ● - o Suffering from Recurrent Dehydration This device is intended for use, under the direction of a physician, for the noninvasive monitoring and management of patients with fluid management problems in a variety of medically accepted clinical applications.
Device Story
Non-invasive, battery-powered impedance monitor; determines fluid status changes in patients with fluid management issues. Device applies low-amplitude, high-frequency (2.0 mA at 100 kHz) electrical current to thorax via four-point measurement technique using Ag/AgCl hydrogel electrodes. Base impedance (Zo) decreases with fluid increase, increases with fluid decrease. Used in clinical settings under physician direction. Hardware-controlled measurement timing; mounted on patient strap with short cable to electrode sensor. Provides clinicians with objective impedance data to assist in fluid management and monitoring decisions; aids in early warning of fluid status changes.
Clinical Evidence
Bench testing only. No clinical data. Performance characterized via head-to-head bench testing against predicate ZOE monitor using calibrated precision resistor array to simulate thoracic impedance loads. Results demonstrated matching measurement linearity and accuracy. Biocompatibility (ISO 10993), electrical safety (IEC 60601-1), and EMC (IEC 60601-1-2) testing performed. Usability testing validated human interface design.
Indicated for patients with fluid management problems, including those on diuretics, with heart failure, end-stage renal disease, recovering from coronary artery disease events, or suffering from recurrent dehydration.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 5, 2015
Intersection Medical, Inc. c/o Mr. Warren Craycroft Vice President of Regulatory and Quality 1808 Aston Ave Suite 120 Carlsbad, California 92008-7364
Re: K142503
> Trade/Device Name: IMED-Z Fluid Status Monitor Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance Plethysmograph Regulatory Class: Class II Product Code: DSB Dated: April 30, 2015 Received: May 4, 2015
Dear Mr. Warren Craycroft,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# K142503
# Indications for Use Statement
510(k) Number K142503
Device Name: IMED-Z Fluid Status Monitor
Indications for Use:
The IMED-Z Fluid Status Monitor is intended for patients:
- With fluid management problems
- Taking diuretic medication
- Living with Heart Failure ●
- Living with End-stage Renal Disease ●
- Recovering from Coronary artery Disease related event ●
- o Suffering from Recurrent Dehydration
This device is intended for use, under the direction of a physician, for the noninvasive monitoring and management of patients with fluid management problems in a variety of medically accepted clinical applications.
Prescription Use YES Over-The-Counter Use NO AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## 5. 510(K) SUMMARY
#### SUBMITTER I.
Intersection Medical, Inc. 1808 Aston Avenue, Suite 120 Carlsbad, CA 92008-7364 Tel: (760) 571-3861 FAX: (760)683-6508
Contact Person:
Warren Craycroft Vice President, Regulatory and Quality Phone: (760) 571-3861 Ext 211 Fax: (760) 683-6509 E-mail: wcraycroft(@intersectionmedical.com
Date Prepared: April 28, 2015
## II. DEVICE
| Name of Device: | IMED-Z Fluid Status Monitor |
|-----------------------|--------------------------------------------|
| Common or Usual Name: | Fluid status monitor |
| Classification Name: | Impedance Plethysmograph (21CFR §870.2770) |
| Regulatory Class: | II |
| Product Code: | DSB |
## III. PREDICATE DEVICE
ZOE Fluid Status Monitor, K133301 (January 22, 2014) This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
## IV. DEVICE DESCRIPTION
The IMED-Z Fluid Status Monitor is a non-invasive, battery-powered impedance monitor designed as an "early warning" monitor for determining changes in the fluid status of patients with fluid management problems.
The IMED-Z Fluid Status Monitor works by applying a low-amplitude, high frequency electrical current to the body and measuring the electrical impedance using
{4}------------------------------------------------
a four-point measurement technique through hydrogel (ECG style) electrode contacts. Base Impedance, also known as Zo, decreases when fluid increases and increases when fluid decreases.
The Fluid Status Monitor is designed for use with disposable, self-adhesive silver / silver chloride electrodes that are readily available from Intersection Medical, Inc. IMED-approved electrodes must be used with the IMED-Z Fluid Status Monitor. Zo reading obtained from unapproved electrodes may not be accurate.
The associated components and accessories include the following:
- IMED-Z acquisition module
- Battery pack and cable .
- Battery charger
- Single-use electrode sensor ●
- Single-use electrode interface cables .
- Single-use patient strap
# V. INDICATIONS FOR USE
The IMED-Z Fluid Status Monitor has the identical indications for use as the predicate device ZOE Fluid Status Monitor K133301 (January 22, 2014 ):
The IMED-Z Fluid Status Monitor is intended for patients:
- With fluid management problems ●
- Taking diuretic medication ●
- Living with Heart Failure ●
- Living with End-stage Renal Disease ●
- Recovering from Coronary artery Disease related event ●
- . Suffering from Recurrent Dehvdration
This device is intended for use, under the direction of a physician, for the noninvasive monitoring and management of patients with fluid management problems in a variety of medically accepted clinical applications.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Both the subject and predicate devices are bioimpedance measurement devices that measure thoracic base impedance to assess fluid status. At a high level, the subject and predicate devices are based on the following same technological elements:
- 4-point impedance measurement of the patient's thorax
- Identical electric current output: 2.0 mA at 100 kHz .
- Ag / AgCl hydrogel electrodes ●
- . Type BF degree of protection
- Low-voltage monitor electronics; no mains voltage on-board either monitor ●
{5}------------------------------------------------
The following technological differences exist between the subject and predicate devices:
- . Placement of electrodes: IMED-Z on back; predicate ZOE on chest
- The IMED-Z uses hardware to control measurement timing. The predicate ● ZOE uses a microprocessor and software.
- The IMED-Z is mounted on a patient strap with a short cable connection to ● the electrode sensor. The predicate ZOE is connected to the electrode by a 68 inch cable.
#### VII. PERFORMANCE TESTING:
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
The biocompatibility evaluation for the IMED-Z device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, "" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity .
- Sensitization ●
- Irritation .
The electrode sensor and patient strap are classified as intact skin contact for a duration of less than 24 hours.
# Electrical safety
Electrical safety testing in compliance with IEC 60601-1 Ed 3.1 was completed on April 27, 2015 by Intertek Testing Services NA Inc. The IMED-Z fulfils the requirements of IEC 60601-1. The Report Number is 101945161LAX-001.
# Electromagnetic compatibility (EMC) testing
EMC testing to the IEC 60601-1-2 Ed 3.0 (2007-03) standard was completed on the IMED-Z device on July 28, 2014 by Intertek Testing Services NA Inc. The IMED-Z complies with the requirements of the standard indicated. The Report Number is 101754547BOX-001.
# Design Verification and Validation Activities
Design Verification and validation activities have been performed on the IMED-Z. The IMED-Z meets its design specification requirements, including impedance measurement accuracy.
## Shelf-life testing
Accelerated shelf-life testing has been completed, validating a shelf life of up to 24
{6}------------------------------------------------
### months.
## Usability testing
Usability testing was performed with healthcare workers to validate the human interface design and instructions for use to ensure safe and effective device interactions.
## Performance bench testing
A head-to-head performance bench test with the predicate device was performed to characterize and compare the impedance measurement performance of the IMED-Z Fluid Status Monitor and the predicate ZOE Fluid Status Monitor (K133301). A calibrated precision resistor array was used to present a range of impedance loads to both devices. The IMED-Z device demonstrated substantial equivalence to the predicate ZOE device by matching measurement linearity and measurement accuracy of the ZOE device.
# VIII. CONCLUSIONS
The biocompatibility testing, electrical safety evaluation, and EMC testing demonstrate that the IMED-Z device is as safe as the predicate ZOE Fluid Status Monitor. The hardware verification and validation demonstrate that design requirements, including measurement accuracy requirements, are met. The usability testing ensures that intended users can operate the device safely and effectively. The head-to-head performance bench testing demonstrates that the IMED-Z performs impedance measurements on simulated thoracic impedances as well as the predicate ZOE Fluid Status Monitor that is currently marketed for the same intended use.
Conclusions drawn from performance testing demonstrate that the IMED-Z device is as safe, as effective, and performs as well as the ZOE Fluid Status Monitor K133301, a legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.