K213841 · Premier North America, Inc. · OHS · Jan 6, 2022 · General, Plastic Surgery
Device Facts
Record ID
K213841
Device Name
ENEO TOTALE (Model: ENEOT947)
Applicant
Premier North America, Inc.
Product Code
OHS · General, Plastic Surgery
Decision Date
Jan 6, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ENEO TOTALE is an over the counter device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles.
Device Story
Battery-operated, handheld LED device for home-use; emits red (633nm) and infrared (830nm) light to treat periorbital wrinkles. Device features 24 LEDs (12 red, 12 IR) providing uniform optical power distribution. User operates device by applying to periorbital area for 15-20 minutes per session. Includes LCD display for battery level and treatment time monitoring. Device utilizes constant waveform; therapeutic temperature range 41±2°C. Output helps reduce periorbital wrinkles through light therapy. Charging via cradle.
Clinical Evidence
Bench testing only. Conformance to ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 62471. Biocompatibility testing per ISO 10993-1, ISO 10993-5, and ISO 10993-10. Software validation performed per FDA guidance.
Technological Characteristics
Handheld LED device; ABS and stainless steel construction. Wavelengths: 633±5nm, 830±5nm. Energy source: 3.75Vdc 3000mAh Lithium battery. Output: 24 LEDs (12 red, 12 IR) over 15cm². Constant waveform. Includes LCD display. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62471.
Indications for Use
Indicated for individuals with periorbital lines and wrinkles for use in dermatology.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 6, 2022
Premier North America Inc. % Doris Dong Manager Shanghai CV Technology Co., Ltd. Room 903, No.19 Dongbao Road, Songjiang Area Shanghai, Shanghai 201613 China
Re: K213841
Trade/Device Name: ENEO TOTALE (Model: ENEOT947) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: October 11, 2021 Received: December 9, 2021
Dear Doris Dong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213841
Device Name ENEO TOTALE
Indications for Use (Describe)
ENEO TOTALE is an over the counter device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary [As required by 21 CFR 807.92]
### 1. Submission Information:
| 510(k) Number: | K213841 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | November 24, 2021 |
| Type of 510(k) Submission: | Special |
| Basis for 510(k) Submission: | Change to existing device |
| Owner | Premier North America Inc.<br>3301 SW 42ND ST., FORT LAUDERDALE, FL 33312-6828, USA<br>Tel: +1-855-360-0650<br>Email: customerservice@premierna.com<br>Web: www.premierdeadsea-usa.com |
| Contact: | Doris Dong<br>[Consultant, from Shanghai CV Technology Co., Ltd.]<br>Add: Room 903, No. 19 Dongbao Road, Songjiang Area, Shanghai, 201613 China<br>E-mail: doris.d@ceve.org.cn<br>Tel: 86 21-31261348 / Fax: 86 21-57712250 |
## 2. Device Description:
| Proprietary Name: | ENEO TOTALE |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model: | ENEOT947 |
| Common Name: | Red/IR Light Therapy Device |
| Classification Name: | Light Based Over The Counter Wrinkle Reduction |
| Regulation Number: | 21 CFR 878.4810 |
| Product Code: | OHS |
| Device Class: | II |
| Review Panel: | General & Plastic Surgery |
| Device Description: | ENEO TOTALE is a battery-operated device that uses low power light<br>spectrum at red and infrared LED, at wavelength of 633 ±5nm, 830 ±5nm<br>emitting optical power in a uniform distribution with no hot spots.<br>It is a hand held light emitting diode(LED) device for the treatment of<br>periorbital wrinkles designed for home-use. |
| Indications for use: | ENEO TOTALE is an over the counter device indicated to emit energy in<br>the red and IR region of the spectrum for use in dermatology for the<br>treatment of periorbital wrinkles. |
## 3. Predicate Device:
Avologi ENEO K181659
{4}------------------------------------------------
## 4. Substantial Equivalence to Predicate device:
Detailed comparison data is included in "Section 9 - Substantial Equivalence Discussion" of this 510(k) submission.
Table 1-
| | New Device | Predicate Device | Remark |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| 510(k)<br>Number | K213841 | K181659 | --- |
| Proprietary<br>Name | ENEO TOTALE | Avologi ENEO | --- |
| Product Code | OHS | OHS | Same |
| Class | 2 | 2 | Same |
| Applicant | Premier North America Inc | Premier North America Inc | Same |
| Indications<br>for use | ENEO TOTALE is an over the counter<br>device indicated to emit energy in the<br>red and IR region of the spectrum for<br>use in dermatology for the treatment of<br>periorbital wrinkles. | Avologi ENEO is an over the counter<br>device indicated to emit energy in the<br>red and IR region of the spectrum for<br>use in dermatology for the treatment<br>of periorbital wrinkles. | Same |
| Handheld | Yes | Yes | Same |
| Target<br>Population | Individuals with periorbital lines and<br>wrinkles | Individuals with periorbital lines and<br>wrinkles | Same |
| Location for<br>use | OTC | OTC | Same |
| Materials | ABS and stainless steel | ABS and stainless steel | Same |
| Visible LCD | Yes | No | Same<br>Note 1 |
| Wavelengths | 633 ±5nm, 830 ±5nm | 633 ±5nm, 830 ±5nm | Same |
| Waveform | Constant | Constant | Same |
| Light source | Light emitting diode(LED) | Light emitting diode(LED) | Same |
| Visible light<br>LEDs | Yes | Yes | Same |
| Energy<br>Source | 24 LEDs(12pcs 633nm LEDs+ 12pcs<br>830nm LEDs) over 15cm² | 24 LEDs(12pcs 633nm LEDs+ 12pcs<br>830nm LEDs) over 15cm² | Same |
| Energy Level | 69mW/cm² for 633nm<br>56mW/cm² for 830nm | 69mW/cm² for 633nm<br>55mW/cm² for 830nm | Same<br>Note 2 |
| Therapeutic<br>temperature<br>range | 41±2°C / 105.8±35.6°F | 41±2°C / 105.8±35.6°F | Same |
| Power supply | Adaptor: 100~240V AC 50/60Hz 2.4A<br>Lithium battery: 3.75Vdc, 3000 mAh | Adaptor: 100~240V AC 50/60Hz 2.4A<br>Lithium battery: 3.75Vdc, 3000 mAh | Same |
| Handpiece<br>charging<br>method | In Charging cradle | In Charging cradle | Same |
| Initial<br>treatment<br>course | For the first month (4 weeks),<br>treatment should be performed 3 times<br>a week for 15-20 minutes each time | For the first month (4 weeks),<br>treatment should be performed 3<br>times a week for 15-20 minutes each | Same |
| | (5-7 minutes on each treatment zone). | time (5-7 minutes on each treatment<br>zone). | |
| Maintenance<br>regime | Once a week for 15-20 minutes | Once a week for 15-20 minutes | Same |
| Anatomical<br>Sites | Periorbital Area | Periorbital Area | Same |
| Standard meet | IEC60601-1<br>IEC60601-1-2<br>IEC60601-1-11<br>IEC62471 | IEC60601-1<br>IEC60601-1-2<br>IEC60601-1-11<br>IEC62471 | Same |
| Label and<br>Labeling | Meet FDA's Requirements | Meet FDA's Requirements | Same |
{5}------------------------------------------------
Note 1:
The proposed device has a small LCD display while predicate device do not have. The LCD display of the proposed device is used for improved battery level indication and to display treatment time. Based on this change, the proposed device has been re-evaluated for IEC 60601-1-2, IEC 60601-1-1,IEC 62471 and software validation report, and the test results indicate that this difference does not raise any new safety and effectiveness issues.
Note 2:
The energy density of the proposed device for 633nm is approximate to the predicate device. Subtle differences will not raise any new issues of safety or effectiveness.
## 5. Non-clinical Testing:
The conclusions drawn from the non-clinical testing below demonstrate that the ENEO TOTALE is substantially equivalent to the predicate devices K181659. The ENEO TOTALE has been tested and conforms to international consensus standards:
Electrical safety:
- ANSI AAMI ES60601-1: 2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD);
EMC:
- IEC 60601-1-2:2014, Medical Electrical Equipment -- Part 1-2: General Requirements For Basic Safety ● And Essential Performance -- Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests;
Additional safety testing:
- IEC60601-1-11 Edition 2.0 2015-01 Medical Electrical Equipment -- Part 1-11: General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment;
- IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems Biocompatibility testing:
- ISO 10993-1:2009/(R) 2013, Biological Evaluation Of Medical Devices -- Part 1: Evaluation And Testing Within A Risk Management Process. (Biocompatibility)
- ISO 10993-5:2009/(R) 2014, Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity. (Biocompatibility)
- ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices - Part 10: Tests For
{6}------------------------------------------------
Irritation And Skin Sensitization. (Biocompatibility)
The ENEO TOTALE has been tested to ensure the device meets specifications:
Performance testing
● Software Validation Testing
The ENEO TOTALE's software was tested and validated in accordance with FDA's "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices."
### 6. Conclusion
The conclusion drawn from the substantial equivalence table is that the new device is substantially equivalent to the predicate device and presents no new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.