K250522 · Multi4 Medical AB · FAS · Jun 27, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K250522
Device Name
Multi4 System
Applicant
Multi4 Medical AB
Product Code
FAS · Gastroenterology, Urology
Decision Date
Jun 27, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Multi4 System is intended for use by trained urologists for endoscopically controlled tissue resection and coagulation, and removal of bladder tumors (TURBT) via suction channel under controlled flow conditions following resection using a monopolar resectoscope. The Multi4 System is also intended to deliver injectable materials into the urinary bladder wall during transurethral endoscopic procedures.
Device Story
Multi4 System is an electrosurgical system for adult urethral resection procedures. It comprises a single-use handheld instrument (Multi4 B) and a reusable Multi4 Pump. The system connects to an external electrosurgical unit to deliver RF energy for cutting and coagulation. The Multi4 B instrument is inserted through a standard cystoscope working channel. The pump manages fluid transport, irrigation, and aspiration, facilitating the collection of resected tissue samples into a 'Simple4tainer' for pathology. The device is operated by trained urologists in professional healthcare facilities. The system provides controlled flow conditions during TURBT and enables targeted delivery of injectable materials into the bladder wall. By integrating resection, suction, and injection capabilities into a single system, it streamlines the surgical workflow, potentially improving procedural efficiency and tissue sample retrieval for diagnostic evaluation.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity, pyrogenicity), sterility, packaging, shelf-life, electrical safety/EMC (IEC 60601-1, 60601-1-2, 60601-1-6, 60601-2-2), software verification and validation, and functional performance testing (cut, coagulation, aspiration, irrigation, injection, and simulated use).
Technological Characteristics
Electrosurgical system using monopolar RF energy for cutting and coagulation. Components include a reusable pump and single-use instrument (Multi4 B) with a 27G lancet-point needle (11mm max length). Connectivity includes footswitch control. Sterilization via Gamma radiation. Software developed per IEC 62304. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and IEC 60601-2-2 (electrosurgical equipment).
Indications for Use
Indicated for adult patients undergoing transurethral endoscopic procedures, including tissue resection, coagulation, bladder tumor removal (TURBT), and delivery of injectable materials into the urinary bladder wall.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 27, 2025
Multi4 Medical AB
% Erin Gontang
Senior Consultant
RQM+
2790 Mosside Blvd, Suite #800
Monroeville, Pennsylvania 15146
Re: K250522
Trade/Device Name: Multi4 System
Regulation Number: 21 CFR 876.4300
Regulation Name: Endoscopic Electrosurgical Unit and accessories
Regulatory Class: II
Product Code: FAS, GEI, KQT, FJL, FBK
Dated: February 21, 2025
Received: June 4, 2025
Dear Erin Gontang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250522 - Erin Gontang
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K250522 - Erin Gontang
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark R. Kreitz -S
for Mark J. Antonino, M.S.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250522
Device Name
Multi4 System
Indications for Use (Describe)
The Multi4 System is intended for use by trained urologists for endoscopically controlled tissue resection and coagulation, and removal of bladder tumors (TURBT) via suction channel under controlled flow conditions following resection using a monopolar resectoscope. The Multi4 System is also intended to deliver injectable materials into the urinary bladder wall during transurethral endoscopic procedures.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K250522
Page 1 of 5
^{}[]
510(k) Summary
K250522
DATE PREPARED
June 4, 2025
MANUFACTURER AND 510(k) OWNER
Multi4 Medical AB
Lasarettsgatan 2
SE-553 18 Jönköping
Sweden
Telephone: +46 736 929 000
Official Contact: Miden Melle-Hannah, MD, PhD, CEO
REPRESENTATIVE/CONSULTANT
Erin Gontang, PhD
Allison Komiyama, PhD, RAC
RQM+
Telephone: +1 (412) 899-7422
Email: egontang@rqmplus.com
DEVICE INFORMATION
Trade/Proprietary Name: Multi4 System
Classification Name: Electrode, Electrosurgical, Active, Urological
Regulation Number: 21 CFR 876.4300
Common Name: Endoscopic electrosurgical unit and accessories; Electrosurgical cutting and coagulation device; Endoscope and accessories
Device Class: Class II
Product Code: FAS, KQT, GEI, FBK, FJL
Classification Panel: Gastroenterology/Urology
Premarket Review: Reproductive, Gynecology and Urology Devices (DHT3B)
PREDICATE DEVICE IDENTIFICATION
The Multi4 System is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Predicate |
| --- | --- | --- |
| K191341 | Veloxion System / Corinth MedTech, Inc. | Primary |
| K192498 | Single Use Injection / Olympus Medical Systems | Secondary |
The predicate devices have not been subject to a design related recall.
{5}
K250522
Page 2 of 5
MULTI
# DEVICE DESCRIPTION
The Multi4 System is an electrosurgical system used during urethral resection procedures of adult patients. The system, intended to remove and collect tissue from the bladder, includes a single-use electrosurgical instrument, known as the Multi4 B, as well as the reusable Multi4 Pump. The Multi4 Pump administers energy from an external electrosurgical unit to the handheld Multi4 B instrument, which has electrosurgical functions and allows for fluid transport and tissue sample collection. The Multi4 B instrument accesses the bladder through the working channel of a commercially available cystoscope. The Multi4 System is a prescription device intended for use in professional healthcare facilities by healthcare professionals (HCPs).
# INDICATIONS FOR USE STATEMENT
The Multi4 System is intended for use by trained urologists for endoscopically controlled tissue resection and coagulation, and removal of bladder tumors (TURBT) via suction channel under controlled flow conditions following resection using a monopolar resectoscope. The Multi4 System is also intended to deliver injectable materials into the urinary bladder wall during transurethral endoscopic procedures.
# SUMMARY OF NON-CLINICAL TESTING
The following testing was performed to evaluate the safety and effectiveness of the Multi4 System.
- Software Verification and Validation Testing performed per IEC 62304, and documentation provided per FDA's Guidance for Industry and FDA Staff: Content of Premarket Submissions for Device Software Functions (Issued June 14, 2023).
- Sterility Testing
- Packaging Testing
- Shelf-life Testing
- Biocompatibility Testing
- Cytotoxicity, Irritation, Sensitization, Acute Systemic Toxicity, Pyrogenicity
- Electrical Safety & EMC:
- IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, and IEC 60601-2-2
- Integrity Testing
- Functional Testing
- Cut and coagulation, aspiration, irrigation, injection
- Dimensional Inspection and Testing
- Simulated Use Testing
# EQUIVALENCE TO PREDICATE DEVICE
The subject device has a similar design and dimensions and uses similar or identical materials as the devices cleared in K191341 and K192498. The subject device also has the same intended use and similar technological characteristics as the devices cleared in K191341 and K192498.
Based on the testing performed, including biocompatibility, sterilization validation, packaging validation, shelf-life validation, EMC and electrical safety and EMC testing, software verification and validation testing, and non-clinical performance bench testing, it can be concluded that the subject device does not raise different questions of safety or effectiveness compared to the predicate devices. The testing performed demonstrates that the subject device is as safe and effective as the
{6}
K250522
Page 3 of 5
^{}[]
predicate devices, and the similar indications for use, technological characteristics, and performance characteristics for the proposed Multi4 System are assessed to be substantially equivalent to the predicate devices. The Multi4 System is substantially equivalent to the predicate devices based on the information summarized here:
| | Subject Device | Predicate Device | Predicate Device |
| --- | --- | --- | --- |
| Device Name | Multi4 System | Veloxion System | Single Use Injector |
| Manufacturer Name | Multi4 Medical AB | Corinth MedTech, Inc. | Olympus Medical Systems Corp. |
| 510(k) Number | K250522 | K191341 | K192498 |
| Regulation | 21 CFR 876.1500 - Endoscope and accessories
21 CFR 876.4370 - Gastroenterology-urology evacuator
21 CFR 878.4400 - Electrosurgical cutting and coagulation device and accessories
21 CFR 876.4300 - Endoscopic electrosurgical unit and accessories | 21 CFR 876.1500 - Endoscope and accessories
21 CFR 876.4370 - Gastroenterology-urology evacuator
21 CFR 878.4400 - Electrosurgical cutting and coagulation device and accessories | 21 CFR 876.1500 - Endoscope and accessories |
| Regulation | Endoscope and accessories | Endoscope and accessories | Endoscope and accessories |
| Product Code(s) | FJL - Resectoscope
KQT - Evacuator, Gastro-Urology
GEI - Electrosurgical, Cutting & Coagulation & Accessories
FAS - Electrode, Electrosurgical, Active, Urological
FBK - Endoscopic Injection Needle, Gastroenterology-Urology | FJL - Resectoscope
KQT - Evacuator, Gastro-Urology
GEI - Electrosurgical, Cutting & Coagulation & Accessories | FBK - Endoscopic Injection Needle, Gastroenterology-Urology |
| Classification | Class II | Class II | Class II |
| Indications for Use Statement | The Multi4 System is intended for use by trained urologists for endoscopically controlled tissue resection and coagulation, and removal of bladder tumors (TURBT) via suction channel under controlled flow conditions following resection using a monopolar resectoscope. The Multi4 System is also intended to deliver injectable materials into the urinary bladder wall during transurethral endoscopic procedures. | The Veloxion System is intended for use by trained urologists for endoscopically controlled tissue chip resection and coagulation, and removal of prostate adenomas and bladder tumors (TURBT) via suction channel under continuous flow conditions following resection using a bipolar resectoscope. | This instrument has been designed to be used with an Olympus endoscope to deliver injectable materials into the urinary bladder wall during the transurethral endoscopic procedures. |
{7}
K250522
Page 4 of 5
MAFF
| Device Description | The Multi4 System consists of the following components:
• Multi4 Pump (with Integrated Fluid Control) o Footswitch
• Multi4 B o Resectoscope & Needle o Simple4tainer (For collection of gross resected tissue pieces for pathology) | The Veloxion System consists of the following components:
• Veloxion Controller (with Integrated Fluid Control) o Footswitch
• Veloxion Resectoscope
• Veloxion Fluid Set
• Waste Management Tubing
• Tissue Catch (For collection of gross resected tissue pieces for pathology) | The Single Use Injector has been designed to be used with an Olympus endoscope to deliver injectable materials into the urinary bladder wall during the transurethral endoscopic procedures. The Single Use Injector device is provided in a sterile package. |
| --- | --- | --- | --- |
| Target Population | Adults | Adults | Adults |
| Sterile/ Non-sterile | Sterile; Gamma | Sterile; Unknown | Sterile; Unknown |
| Use Environment | Healthcare Facilities | Healthcare Facilities | Healthcare Facilities |
| Energy Type; Monopolar or Bipolar | Radiofrequency; Monopolar | Radiofrequency; Bipolar | N/A |
| RF Function | Cut, Coagulation | Cut, Coagulation | N/A |
| Control of cavity pressure? | The urologist manages the irrigation and fluid control during the procedure. | The device uses a pressure monitor to evaluate cavity pressure during the procedure. | N/A |
| Irrigation Fluid | Water | Saline | N/A |
| Pump | Dual Action (Irrigation, Aspiration) | Dual Action (Irrigation, Aspiration) | N/A |
| Testing | Biocompatibility Testing
Sterility Testing
Packaging Testing
Shelf-life Testing
EMC/Electrical Safety Testing
Software V&V Testing
Performance Testing – Bench and Simulated Use | Biocompatibility Testing
Sterility Testing
Packaging Testing
Shelf-life Testing
EMC/Electrical Safety Testing
Software V&V Testing
Performance Testing – Bench and Simulated Use | Biocompatibility Testing
Sterility Testing
Packaging Testing
Shelf-life Testing
Performance Testing – Bench |
| Functional Testing | Cut & Coagulation
Aspiration
Irrigation
Injection | Cut & Coagulation
Aspiration
Irrigation
Pressure Control | Injection |
| Needle Width | 27G | N/A | 27G |
| Needle Working Length | 716 mm | N/A | 971 mm |
| Needle Length | 11 mm (maximum) | N/A | 4 mm |
| Tip Shape of Needle | Lancet Point | N/A | Lancet Point |
{8}
K250522
Page 5 of 5
Multi4
## CONCLUSION
Based on the testing performed, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate devices. The Multi4 System is considered safe and effective for its intended use. The similar technological characteristics and performance characteristics for the Multi4 System are assessed to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.