Veloxion Controller Kit, Veloxion Fluid Control Set, Veloxion Resectoscope, Veloxion Video Control Unit, Veloxion Roll Stand, Waste Management Tubing, Tissue Catch, Waste Management Bags

K191341 · Corinth Medtech, Inc. · FJL · Jun 13, 2019 · Gastroenterology, Urology

Device Facts

Record IDK191341
Device NameVeloxion Controller Kit, Veloxion Fluid Control Set, Veloxion Resectoscope, Veloxion Video Control Unit, Veloxion Roll Stand, Waste Management Tubing, Tissue Catch, Waste Management Bags
ApplicantCorinth Medtech, Inc.
Product CodeFJL · Gastroenterology, Urology
Decision DateJun 13, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Veloxion System is intended for use by trained urologists for endoscopically controlled tissue chip resection and coagulation, and removal of prostate adenomas and bladder tumors (TURBT) via suction channel under continuous flow conditions following resection using a bipolar resectoscope.

Device Story

Veloxion System is an endoscopic surgical device for urological procedures; used by trained urologists in clinical settings. System components: controller with integrated fluid control, footswitch, resectoscope, fluid control set, video control unit, and roll stand. Principle of operation: bipolar radiofrequency (RF) energy for tissue resection and coagulation; peristaltic pumps manage fluid inflow and aspiration of resected tissue chips. Controller monitors intracavitary pressure via dual independent sensors connected to irrigation lumen; software manages pressure limits and notifies user. Video control unit manages camera/light and image storage. System facilitates continuous flow conditions for visualization and tissue removal. Healthcare providers use the system to perform TURBT and prostate adenoma removal; output includes visual feedback on monitor and physical removal of tissue chips. Benefits include controlled resection, hemostasis, and efficient tissue removal via suction.

Clinical Evidence

No new clinical data provided. Substantial equivalence is supported by bench testing and software verification protocols previously reviewed under K190099 and K162979.

Technological Characteristics

Bipolar radiofrequency energy for cut/coagulation. System includes peristaltic pumps for fluid/pressure control, dual independent pressure sensors, and sterile single-use resectoscope and fluid control set. Connectivity includes USB for media download. Software-controlled pressure monitoring and fluid deficit tracking. Dimensions: 220mm working length, 25Fr outer sheath. Electrode movement: linear oscillation at 3000cpm.

Indications for Use

Indicated for trained urologists performing endoscopic tissue chip resection, coagulation, and removal of prostate adenomas and bladder tumors (TURBT) via suction channel under continuous flow conditions using a bipolar resectoscope.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format. June 13, 2019 Corinth MedTech, Inc. Sandeep Saboo Vice President. Ouality Assurance & Regulatory Affairs 1601 S. De Anza Blvd., Suite 200 Cupertino, CA 95014 K191341 Re: Trade/Device Name: Veloxion System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FJL, KQT, GEI Dated: May 15, 2019 Received: May 20, 2019 Dear Sandeep Saboo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or (301) 796-7100). Sincerely, for Glenn B. Bell. Ph.D. Assistant Division Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K191341 Device Name Veloxion System Indications for Use (Describe) The Veloxion System is intended for use by trained urologists for endoscopically controlled tissue chip resection and coagulation, and removal of prostate adenomas and bladder tumors (TURBT) via suction channel under continuous flow conditions following resection using a bipolar resectoscope. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------| | <div> <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## K191341: 510(k) Summary I. ### Submitter Information | Submitter name: | Corinth MedTech, Inc.<br>1601 S. De Anza Blvd, Suite 200<br>Cupertino, CA 95014 | |-----------------|---------------------------------------------------------------------------------------------------------------------------| | Contact person: | Sandeep Saboo<br>Vice President, Regulatory Affairs and Quality Assurance<br>Phone: (408) 996-2517<br>Fax: (408) 996-0621 | | Date Prepared: | 14 May 2019 | #### II. Product Classification | Device Name: | Veloxion System | | |---------------------------|---------------------------|----------------| | Common Name: | Resectoscope | Subject Device | | Regulation: | 21 CFR 876.1500 | | | Regulation Name: | Endoscope and accessories | | | Class: | II | | | Product Code: | FJL | | | Additional Product Codes: | KQT, GEI | | ### III. Predicate Devices The predicate device is the system comprised of the following legally marketed devices as used in combination to which substantial equivalence is claimed: | Predicate | Manufacturer | Predicate Device Names | 510(k)# | Clearance Date | |------------------------------------|----------------------------------------------------|-------------------------------------------|---------|----------------| | Primary Predicate<br>Predicate #1 | Corinth MedTech, Inc. | Veloxion System | K190099 | March 15, 2019 | | Predicate #2<br>(predicate system) | Richard Wolf Medical<br>Instruments<br>Corporation | S-Line Bipolar Resectoscope<br>& ERBE ESU | K062720 | March 22, 2007 | | Predicate #3<br>(predicate system) | Richard Wolf Medical<br>Instruments<br>Corporation | Resection Master & ERBE<br>ESU | K042523 | March 18, 2005 | | Predicate #4<br>(predicate system) | Corinth MedTech, Inc. | Veloxion System | K162979 | March 24, 2017 | Predicate devices have not been the subject of a design related recall. #### IV. Device Description The Veloxion System consists of the following components: - · Veloxion Controller (with Integrated Fluid Control) - o Footswitch - · Veloxion Resectoscope - · Veloxion Fluid Control Set - Veloxion Video Control Unit - Veloxion Roll Stand {4}------------------------------------------------ K191341 The Veloxion System also includes the following Class I accessories for handling of waste collected from the patient (these items only handle waste after it is already outside the patient), which includes: - · Waste Management Tubing: To provide a conduit for transfer of aspirated fluids and tissue from the patient and from under the patient's buttocks drape - · Tissue Catch: For collection of gross resected tissue pieces for pathology. - · Waste Management Bags: To provide bags for final collection of outflow. The Veloxion System provides bipolar resection and coagulation of intrauterine tissue, it distends the uterus by filling with saline, it provides pressure control of the intrauterine cavity for insufflation to facilitate viewing with the integrated hysteroscope and it monitors the fluid deficit (potential fluid absorbed by the patient's body) to the physician established limit. The components of Veloxion System perform the following functions: - · The Veloxion Controller provides bipolar radiofrequency outputs (for cut and coagulation) and fluid/pressure control through the use of two integrated peristaltic pumps, provides the user interface to establish the desired set pressure, monitors intracavitary pressure using dual independent pressure sensors mechanically connected to the Veloxion Fluid Control Set irrigation lumen. The software then monitors, controls and notifies the user when the limits are reached or when specific conditions are met. - The Veloxion Resectoscope is a sterile single use hand held bipolar radiofrequency device configured to provide camera and light for visualization of the anatomy for the resection of tissue and aspiration of resected chips. Fluid inflow and aspiration of the resected chips are controlled by the Controller's peristaltic pumps. - · The Veloxion Fluid Control Set is a sterile single use device that provides conduits for fluid inflow, aspiration of resected tissue and fluids and a diaphragm (pressure membrane) that provides mechanism for the Controller to measure cavity pressure (through the irrigation lumen) during the procedure thereby facilitating the insufflation function. - · The Veloxion Video Control Unit provide control of the camera, light, display image to a commercially available monitor, stores image and video selected by the user from a session, and provides a USB connection for a USB Stick download of stored media by the user. - The Veloxion Roll Stand enables monitoring of saline remaining in the saline bag and facilitates the fluid deficit function. ### V. Indications for Use There is no difference in the indications for use for the modified Veloxion System (subject device) when compared to the predicate devices. | Comparison of Indications for Use | | |-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device | Indications For Use | | Modified<br>Veloxion System<br>(Subject Device) | The Veloxion System is intended for use by trained urologists for endoscopically<br>controlled tissue chip resection and coagulation, and removal of prostate adenomas and<br>bladder tumors (TURBT) via suction channel under continuous flow conditions following<br>resection using a bipolar resectoscope. | | Veloxion System<br>K190099<br>(Primary Predicate)<br>Predicate #1 | The Veloxion System is intended for use by trained urologists for endoscopically<br>controlled tissue chip resection and coagulation, and removal of prostate adenomas via<br>suction channel under continuous flow conditions following resection using a bipolar<br>resecting device. | # Comparison of Indications for Use {5}------------------------------------------------ # Veloxion System SPECIAL 510(k) Premarket Notification | Device | Indications For Use | |-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Predicate #2<br>K062720<br>(Predicate system) | S(a)line Resectoscopes are used for endoscopically controlled removal (ablation) of tissue<br>using 0.9% NaCl solution (saline) as the irrigation medium.<br>For Urology:<br>Urological surgical procedures involving the ablation or removal of soft tissue and where<br>associated hemostasis is required.<br>Transurethral resection of the prostate (TURP) and bladder neck.<br>Transurethral resection of the bladder tumors (TURBT)<br>Transurethral incision of the prostate.<br>Coagulation of bleeding in the lower urinary tract.<br>For Gynecology:<br>Tissue cutting, removal, and desiccation as required or encountered in gynecologic<br>hysteroscopic electrosurgical procedures for the treatment of intrauterine myomas, polyps,<br>adhesions, septa, and benign conditions requiring endometrial ablation.<br>Excision of intrauterine myomas.<br>Excision of intrauterine polyps.<br>Lysis of intrauterine septa.<br>Endometrial ablation. | | Predicate #3<br>K042523<br>(Predicate system) | The Resection Pump 2228 with Resectoscope 8659.xxx is used for endoscopically<br>controlled tissue chip resection and removal of intrauterine polyps, intrauterine myomas or<br>prostate adenomas via suction channel under continuous flow conditions following<br>resection using a high-frequency electrode with a Resectoscope. | | Predicate #4<br>K162979<br>(Predicate system) | The Veloxion System is intended for endoscopically controlled tissue chip resection and<br>coagulation and removal of prostate adenomas via suction channel under continuous flow<br>conditions following resection using a bipolar resecting device. | ## VI. Comparison of Technological Characteristics with the Predicate Device The subject Veloxion System and the previously cleared predicate devices have the same or similar technological characteristics (see Table 1) in terms of basic operating principle and basic design features. {6}------------------------------------------------ ## Veloxion System SPECIAL 510(k) Premarket Notification | | | | | | | Table 1: Comparison of subject Veloxion System and predicate: | |--|--|--|--|--|--|---------------------------------------------------------------| |--|--|--|--|--|--|---------------------------------------------------------------| | Technological Characteristics | VELOXION SYSTEM<br>(Subject) | VELOXION SYSTEM<br>(Predicate System)<br>K190099 & K162979 | Richard Wolf S-Line<br>Resectoscope System<br>(Predicate System)<br>K062720 | Richard Wolf Resection Master<br>System<br>(Predicate System)<br>K042523 | |--------------------------------------------|-------------------------------|----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|--------------------------------------------------------------------------| | Energy Type: | Radiofrequency, bipolar | Radiofrequency, bipolar | Radiofrequency, bipolar | Radiofrequency, bipolar | | RF Functions: | Cut, Coagulation | Cut, Coagulation | Cut, Coagulation | Cut, Coagulation | | Able to Set & Monitor Cavity<br>Pressure? | YES | YES | YES | YES | | Irrigation fluid: | Saline | Saline | Saline | Saline | | Outer Sheath Working Length: | 220mm | K190099: 220mm<br>K162979: 193mm | 186mm | 186mm | | Outer Sheath OD: | 25Fr (8.3mm) | K190099: 25Fr (8.3mm)<br>K162979: 26Fr (8.6mm) | 27Fr (9.0mm) | 27Fr (9.0mm) | | Inner Sheath OD: | 19.5 Fr (6.5mm) | K190099: 19.5 Fr (6.5mm)<br>K162979: N/A | 21 Fr (7mm) | N/A | | Max Electrode Extension outside<br>Sheath: | 20mm | K190099: 20mm<br>K162979: 23mm | 21mm | 21mm | | Electrode Wire OD: | 0.019" | K190099: 0.019"<br>K162979: 0.015" | 0.010inch to 0.015inch | 0.010inch to 0.015inch | | Electrode Movement: | Linear Oscillation at 3000cpm | K190099:<br>Linear Oscillation at 3000cpm<br>K162979:<br>Linear Oscillation at 1000cpm | Stationary on the distal end | Stationary on the distal end | {7}------------------------------------------------ # Corinth MedTech, Inc. There are minor differences in the technological characteristics between the subject Veloxion System and the predicates. The minor differences do not raise new or different questions of safety and effectiveness for the intended use of the device as it pertains to the expansion of the indications for use. ### VII. Performance Data No new performance testing was required to establish substantially equivalence. Results of tests in reviewed as part of clearance of K190099 and K162979 with comparison testing to the predicates (predicate Veloxion System and Richard Wolf Resectoscope) support substantial equivalence to the identified predicates. - . Software Verification Testing was performed using the same protocol reviewed as part of K 190099 for all steps affected by the change and documentation updated per FDA 's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." ### VIII. Conclusions The subject Veloxion System is considered substantially equivalent and as safe and effective as the identified predicate devices (predicate systems).
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...