K250501 · This AG · HQC · Nov 14, 2025 · Ophthalmic
Device Facts
Record ID
K250501
Device Name
System Sophi
Applicant
This AG
Product Code
HQC · Ophthalmic
Decision Date
Nov 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4670
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The System Sophi is an integrated phacoemulsification lens removal device designed for use in intraocular ophthalmic surgery in the anterior chamber. System Sophi is intended for liquid irrigation and aspiration, lens fragmentation and ultrasonic phacoemulsification, anterior vitrectomy, and bipolar diathermy (high frequency surgical equipment).
Device Story
System Sophi is an integrated phacoemulsification system for anterior chamber ophthalmic surgery. It processes inputs from a surgical handpiece (ultrasonic energy for lens fragmentation) and a cassette interface (regulating balanced saline solution irrigation and aspiration of lens fragments/fluids). The system is operated by a surgeon via a foot pedal and an assistant via a graphical user interface (GUI). It provides ultrasonic phacoemulsification, irrigation/aspiration, anterior vitrectomy, and bipolar diathermy. The device is AC or battery-powered. Output is controlled by the surgeon to facilitate cataract removal and other ophthalmic procedures. The system benefits patients by enabling precise, automated control of fluidics and ultrasonic energy during lens removal, reducing surgical complexity and improving procedural efficiency.
Clinical Evidence
No clinical data was necessary to determine substantial equivalence. Evidence consists of bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), wireless coexistence (IEEE ANSI C63.27), performance testing for lens removal/vitrectomy (IEC 80601-2-58), high-frequency surgical equipment safety (IEC 60601-2-2), biocompatibility (ISO 10993), sterilization validation (ISO 11135, ISO 17665), and software/cybersecurity verification.
Technological Characteristics
Console-based system with 15"/19" touchscreen GUI and foot pedal control. Features peristaltic or venturi pump for fluidics. Phaco module uses 40kHz auto-tuning piezoelectric handpiece. Diathermy module operates at 0.5 MHz. Connectivity includes wired/wireless footswitch. Powered by AC or battery. Software runs on Linux. Sterile accessories are EO sterilized.
Indications for Use
Indicated for intraocular ophthalmic surgery in the anterior chamber, specifically for lens removal, irrigation/aspiration, phacoemulsification, anterior vitrectomy, and bipolar diathermy.
Regulatory Classification
Identification
A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 14, 2025
This AG
Anne-Marie Ripley
Clinical & Regulatory Affairs Consultant
Regulatory Pathways Group, Inc.
440 N. Barranca Ave., #2471
Covina, CA 91723
Re: K250501
Trade/Device Name: System Sophi
Regulation Number: 21 CFR 886.4670
Regulation Name: Phacofragmentation System
Regulatory Class: Class II
Product Code: HQC, HQE
Dated: October 17, 2025
Received: October 17, 2025
Dear Anne-Marie Ripley:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250501 – Anne-Marie Ripley
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K250501 – Anne-Marie Ripley
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CLAUDINE H. KRAWCZYK -S
Claudine Krawczyk
Assistant Director
DHT1A: Division of Ophthalmic Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250501
Device Name
System Sophi
Indications for Use (Describe)
The System Sophi is an integrated phacoemulsification lens removal device designed for use in intraocular ophthalmic surgery in the anterior chamber. System Sophi is intended for liquid irrigation and aspiration, lens fragmentation and ultrasonic phacoemulsification, anterior vitrectomy, and bipolar diathermy (high frequency surgical equipment).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
This AG
Traditional 510(k) Premarket Submission K250501
System Sophi
This
Think In System
# 510(k) Summary
## 1 Submission Sponsor
This AG
Widnauerstrasse 1
Heerbrugg
SG, 9435
Switzerland
Contact: Patrick Vortanz
Email: PatrickVortanz@rayner.com
Phone: +41-071-510-0791
## 2 Official Correspondent
Regulatory Pathways Group, Inc.
440 N. Barranca Ave., #2471
Covina, CA 91723
Contact: Anne-Marie Ripley
Email: aripley@regulatorypathways.com
Phone: 661-645-8546
## 3 Date Prepared
November 4, 2025
## 4 Device Identification
Trade/Proprietary Name: System Sophi
Classification Name: Phacofragmentation system
Regulation Number(s): 886.4670
Product Code(s): HQC, HQE
Class: 2
Classification Panel: Ophthalmic
## 5 Legally Marketed Predicate Device
Device name: EVA NEXUS Ophthalmic Surgical System
510(k) number: K213467
Manufacturer: D.O.R.C. Dutch Ophthalmic Research Center (International) B.V
## 6 Predicate Device Description
The EVA NEXUS Ophthalmic surgical system is used for both anterior segment (i.e. phacoemulsification and removal of cataracts) and posterior segment (i.e. vitreoretinal) ophthalmic surgery, including subretinal microinjection of gaseous or liquids, and laser surgery.
p1 of 7
{5}
This AG
Traditional 510(k) Premarket Submission K250501
System Sophi
This
Think In System
# 7 Indication for Use Statement
The System Sophi is an integrated phacoemulsification lens removal device designed for use in intraocular ophthalmic surgery in the anterior chamber. System Sophi is intended for liquid irrigation and aspiration, lens fragmentation and ultrasonic phacoemulsification, anterior vitrectomy, and bipolar diathermy (high frequency surgical equipment)
# 8 Device Description
The System Sophi is an integrated phacoemulsification lens removal device designed for use in intraocular ophthalmic surgery in the anterior chamber. System Sophi is intended for lens fragmentation and ultrasonic phacoemulsification, liquid irrigation and aspiration, anterior vitrectomy, and bipolar diathermy (high frequency surgical equipment).
A cassette is the interface between the System Sophi device and the surgical handpiece. It is used to regulate the balanced saline solution (BSS) irrigating fluid to the handpiece, to aspirate lens fragments and liquids from the handpiece, to monitor irrigation and aspiration flow and pressure and deposition of lens fragments and fluid into the waste bag for disposal. The cassette ports and tube connectors are coded to ensure proper connections at any time.
The system is controlled with a foot pedal (surgeon) and a graphical user interface (assistant).
System Sophi is powered by mains voltage (AC-powered) or through a battery.
Only company-validated and recommended disposable sterile equipment and designated instruments and hand pieces are allowed.
The Sophi system is offered in three configurations: Sophi Premium, Sophi Eco and Sophi A, differing in the included options and accessories.
# 9 Substantial Equivalence Discussion
The following table compares the System Sophi to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.
# 10 Comparison of Characteristics
| Attribute | Predicate Device EVA NEXUS Ophthalmic Surgical System | Subject Device System Sophi | Device Comparison |
| --- | --- | --- | --- |
| Manufacturer | D.O.R.C. Dutch Ophthalmic Research Center | This AG | N/A |
| US Premarket Notification | K213467 | Pending | N/A |
| Product Codes | HQC HQE HQF FMF | HQC HQE | Different, differences not affecting safety or performance |
| Regulation Number | §886.4670 Phacofragmentation System §886.4150 Vitreous aspiration and cutting instrument §886.4390 Ophthalmic Laser §886.5860 Piston Syringe | §886.4670 Phacofragmentation System §886.4150 Vitreous aspiration and cutting instrument | As above |
| Device Class | II | II | Same |
| Intended Use | Ophthalmic surgery | Ophthalmic surgery | Same |
{6}
This AG
Traditional 510(k) Premarket Submission K250501
System Sophi
This
Think In System
| Attribute | Predicate Device
EVA NEXUS Ophthalmic Surgical System | Subject Device
System Sophi | Device Comparison |
| --- | --- | --- | --- |
| Indications for Use | The EVA NEXUS Ophthalmic surgical system is indicated for both anterior segment (i.e. phacoemulsification and removal of cataracts) and posterior segment (i.e. vitreoretinal) ophthalmic surgery, including subretinal microinjection of gaseous or liquids. | The System Sophi is an integrated phacoemulsification lens removal device designed for use in intraocular ophthalmic surgery in the anterior chamber. System Sophi is intended for liquid irrigation and aspiration, lens fragmentation and ultrasonic phacoemulsification, anterior vitrectomy, and bipolar diathermy (high frequency surgical equipment). | Similar, differences not affecting safety or performance |
| | In addition, the optional laser is indicated for the following:
Condition
Treatment
Diabetic Retinopathy
Proliferative Diabetic Retinopathy
Panretinal Photocoagulation
Clinically Significant Macular Edema
Reticular Teat and Detachments
Lattice Degeneration
Sub-retinal (choroidal) Neovascularization
Reticular Occlusion
Neovascularization secondary to Brand or Central retinal vein occlusion
Chronic macular edema secondary to Branch or Central retinal vein occlusion
Glaucoma
Primary Open-angle Closed Angle
Trapeculoplast Iridotomy or Iridoplasty | N/A | The Subject Device does not include a laser |
| Design functions | - Phacoemulsification
- Vitrectomy
- Diathermy
- Irrigation / Aspiration
- Illumination
- Fluid/Air exchange
- Viscous fluid control (injection, extraction, Micro Injection)
- Laser
- Visualization (Digital overlay) | - Phacoemulsification
- Vitrectomy
- Diathermy
- Irrigation / Aspiration
- Fluid/Air exchange | Different, differences not affecting safety or performance |
| Device Components | Console, footswitch, power cable, accessories for different purposes including phacoemulsification, vitrectomy, illumination, Inspiration/Aspiration, diathermy, and laser. Procedure packs. | Console, footswitch, power cable, accessories for different purposes including phacoemulsification, vitrectomy, Inspiration/Aspiration, diathermy. Procedure packs. | Similar, differences not affecting safety or performance |
p3 of 7
{7}
This AG
Traditional 510(k) Premarket Submission K250501
System Sophi
This
Think In System
| Attribute | Predicate Device EVA NEXUS Ophthalmic Surgical System | Subject Device System Sophi | Device Comparison |
| --- | --- | --- | --- |
| Accessories | Console Plastic dust cover and protective flight case or carton
Disposable - Primary EO sterilization package and protective box Reusable - Protective packaging | Console Plastic dust cover and protective flight case or carton
Disposable - Primary EO sterilization package and protective box Protective packaging | Same |
| Software Operating System | Windows embedded OS | Linux | Different, differences not affecting safety or performance |
| Power Source | AC: 110 – 240 V, 50Hz/60Hz | AC: 110 – 240 V, 50Hz/60Hz | Same |
| Electrical Safety, Electromagnetic Compatibility standards | IEC 60601-1
IEC 60601-1-2 | IEC 60601-1
IEC 60601-1-2 | Same |
| User Interfaces | The user selects settings for the various functions on the Graphical User Interface by the touch the screen. The user starts, stops, and regulates the various functions with the foot switch. | The user selects settings for the various functions on the Graphical User Interface by the touch screen. The user starts, stops, and regulates the various functions with the foot switch. | Same |
| Irrigation (Infusion) modes | Modes: Fixed/AIC/SMART IOP | Modes: Active IOP or gravitation | Similar, differences not affecting safety or performance |
| Aspiration modes | Flow mode or vacuum mode | Flow mode or vacuum mode | Same |
| Cassette/ Cartridge and Tubing Sets | The cartridge consists of a collection bag to collect aspiration fluids, tubing for irrigation and aspiration and the tubing for the administration set.
Fluid circulation is achieved by the interaction of the membrane and the pump pistons. | The cassette/cartridge consists of a collection bag to collect aspiration fluids, tubing for irrigation and aspiration.
Fluid circulation is achieved by the interaction of the peristaltic or venturi pump. | Similar, differences not affecting safety or performance |
| Console screen | Glass Color display Touchscreen (19”) with programmable Graphical User Interface | Glass Color display Touchscreen (19” & 15”) with programmable Graphical User Interface | Similar, differences not affecting safety or performance . |
| Multi-function footswitch | Multi-function footswitch with a dual-linear pedal and programmable function buttons to control all functions including laser.
Wired and Wireless connection to the console.
Battery-operated with wall charger on Device and back-up power/connection cable. | Multi-function footswitch with a dual-linear pedal and programmable function buttons to control all functions.
Wired and Wireless connection to the console.
Battery-operated with back-up power/connection cable. | Same |
| Phaco Principle of operation | The Phaco Module can be used for Phaco emulsification and fragmentation. The module provides ultrasound power with continuous autotuning. The module drives a connected Phaco handpiece. The handpiece contains piezo-electric elements which transforms the electrical drive energy into an ultrasonic output at the distal end of the handpiece needle. | The Phaco Module can be used for Phaco emulsification and fragmentation. The module provides ultrasound power with continuous autotuning. The module drives a connected Phaco handpiece. The handpiece contains piezo-electric elements which transforms the electrical drive energy into an ultrasonic output at the distal end of the handpiece needle. | Same |
p4 of 7
{8}
This AG
Traditional 510(k) Premarket Submission K250501
System Sophi
This
Think In System
| Attribute | Predicate Device EVA NEXUS Ophthalmic Surgical System | Subject Device System Sophi | Device Comparison |
| --- | --- | --- | --- |
| Phaco frequency | 40kHz Auto-tuning | 40kHz Auto-tuning | Same |
| Vitrectomy | TDC Two Dimensional Cutter | Bi-Blade Vitrectomy Cutter by MID Labs , USA | The Bi-Blade Vitrectomy Cutter is a 3^{rd}-Party market-cleared accessory (K153168) |
| Vitrectomy Power source for cutter activation | Pneumatic pressure pulse | Pneumatic pressure pulse | Same |
| Diathermy | Various forceps and pencils Accessories to the predicate device | Various forceps and pencils by Kirwan Surgical Products, USA | The Diathermy items are 3^{rd}-Party market-cleared accessories |
| Diathermy Principle of operation | Provides coagulation for anterior and posterior eye-segment surgery. | The Diathermy Module provides coagulation for anterior eye-segment surgery. | Same |
| Diathermy Type | Bipolar Coagulation | Bipolar Coagulation | Same |
| Diathermy Driving Frequency | 1 MHz (±10%) | 0.5 MHz (±10%) | Similar, differences not affecting safety or performance. |
| Electrical Safety Testing | passed | passed | Same |
| EMC/EMD Testing | passed | passed | Same |
| Complies with ISO 10993-1 | passed | passed | Same |
# 11 Non-Clinical Performance Data
To demonstrate safety and effectiveness of System Sophi and to show substantial equivalence to the predicate device, This AG completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met. The System Sophi passed the following testing in accordance with internal requirements, FDA guidance, and international standards as described below, supporting its safety and effectiveness, and its substantial equivalence to the predicate device:
- Electrical Safety Testing was performed to demonstrate the Sophi System meets the following standard:
- IEC 60601-1:2020, "Medical electrical equipment – Part 1: General requirements for safety 1: collateral standard: Safety Requirements for Medical Electrical Systems"
- Electromagnetic Compatibility (EMC) Testing was conducted to demonstrate that the System Sophi meets the following standards:
- IEC 60601-1-2:2020, "Medical electrical equipment – Part 1-2: General requirements for safety and essential performance – Collateral Standard: Electromagnetic compatibility – Requirements and tests."
p5 of 7
{9}
This AG
Traditional 510(k) Premarket Submission K250501
System Sophi
This
Think In System
- Additional EMC testing for common electromagnetic emitters was conducted in accordance with FDA guidance, "Electromagnetic Compatibility (EMC) of Medical Devices"
- Wireless coexistence testing was conducted to demonstrate compliance with IEEE ANSI USEMCSC C63.27-2021, "American National Standard for Evaluation of Wireless Coexistence"
- Testing was performed to demonstrate the Sophi System meets the requirements for lens removal devices, vitrectomy devices, and high frequency surgical equipment per the following standards:
- IEC 80601-2-58, "Medical electrical equipment - Part 2-58: Particular requirements for the basic safety and essential performance of lens removal devices and vitrectomy devices for ophthalmic surgery"
- Basic safety and essential performance of high frequency surgical equipment was conducted per IEC 60601-2-2:2017, "Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories"
- Biocompatibility: The biocompatibility profile of the direct and indirect tissue contacting components of the device were assessed for cytotoxicity, sensitization, Irritation, acute systemic toxicity, and material-mediated pyrogenicity as recommended in FDA’s 2020 Biocompatibility Guidance, "Biocompatibility Testing of Medical Devices - Standards Specific Information for the Accreditation Scheme for Conformity Assessment (ASCA) Pilot Program: Guidance for Industry, Accreditation Bodies, Testing Laboratories, and Food and Drug Administration Staff" and relevant parts of ISO 10993 Biological Evaluation of Medical Devices standard series.
- Software verification and validation testing: The System Sophi software was developed and verified in accordance with FDA’s 2023 guidance, "Content of Premarket Submissions for Device Software Functions." The software is classified as an Enhanced Document level per FDA’s guidance document.
- Cybersecurity testing was performed in accordance with FDA guidance, "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions."
- Sterilization validation was performed per ISO 11135:2014 to demonstrate that the sterilization process for the sterile accessories reliably sterilizes the product with a sterility assurance level (SAL) of 10⁻⁶. EO/ECH residuals testing was performed per ISO 10993-7:2008 to demonstrate the sterile accessories maintain acceptable residuals levels post-sterilization. Endotoxin testing (LAL) was performed per ANSI/AAMI ST72:2019 to demonstrate that the sterile accessories maintain acceptable endotoxin levels.
- Packaging validation/shelf life studies were performed to validate the packaging and claimed shelf life for the sterile accessories. Transit conditioning was conducted per ISO 11607-1, ASTM D7386 and ASTM D4169. Aging studies were conducted per ASTM F1980. Whole packaging integrity, seal integrity, visual inspection, and label adhesion/legibility tests were performed.
- Sterilization validation was performed per ISO 17665-1:2006 to validate the sterilization instructions of the reusable handpiece.
p6 of 7
{10}
This AG
Traditional 510(k) Premarket Submission K250501
System Sophi
This
Think In System
## 12 Clinical Performance Data
No clinical data was necessary to determine the substantial equivalence of this device.
## 13 Statement of Substantial Equivalence
The System Sophi has the same intended use as the predicate device and similar technological characteristics. The differences in technological characteristics do not raise new or different questions of safety and effectiveness. Based on the results of the risk assessment and all applicable testing, the System Sophi is determined to perform as intended during normal anticipated usage and is at least as safe, as effective, and performs as well as or better than the legally marketed predicate device and therefore can be determined to be substantially equivalent to the predicate device.
p7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.