The devices are intended to be used for monitoring and recording of, and to generate alarms for multiple physiological parameters of adults, pediatrics, and neonates.
Device Story
IntelliVue 6100, 6300, 6500 are multiparameter patient monitors; display units with integrated CPUs and touchscreen interfaces. Inputs include ECG/Resp, SpO2, NIBP, IBP, Temperature; supports external modules via FlexLink connector (e.g., Medtronic Microstream CO2 NanoPod, Predictive Temperature Module). Used in hospitals for stationary or transport monitoring; operated by healthcare professionals. Communicates with Philips Patient Information Center iX and XDS software; provides analog ECG-Out for external devices (defibrillators, CT, IABP). Monitors display real-time physiological data and generate alarms; assists clinicians in patient status assessment and decision-making. Benefits include continuous, non-invasive monitoring of vital signs and pulmonary status to facilitate timely clinical intervention.
Clinical Evidence
No clinical trials were required. Evidence includes non-clinical bench testing, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), usability (IEC 60601-1-6), and alarm system testing (IEC 60601-1-8). Performance specifications for integrated measurements were verified against recognized standards (e.g., ECG, NIBP, IBP, CO2, temperature, pulse oximetry). Clinical data was provided specifically to demonstrate the accuracy performance of the Philips FAST PicoSAT X SpO2 technology.
Technological Characteristics
Stationary/transport monitors with TFT LCD (10" or 15"). Powered by AC mains or internal battery. Materials: chemically strengthened glass, disinfectant-resistant plastics. Connectivity: FlexLink, ECG-Out, network integration with Philips PIC iX. Software: IntelliVue Software R.00.01. Standards: IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-8, 60601-2-25, 60601-2-27, 80601-2-30, 60601-2-34, ISO 80601-2-55, 80601-2-56, 80601-2-61, IEC 62304, IEC 62133-2, UL 2054.
Indications for Use
Indicated for use by health care professionals for monitoring physiological parameters of adult, pediatric, and neonatal patients. Not for therapeutic use. Prescription use only. ECG for diagnostic rhythm/morphology. ST segment monitoring for adults only. SSC Sepsis Protocol for adults only. PPV for sedated, mechanically ventilated adults free from arrhythmia. Masimo rainbow SET for SpO2, pulse rate, SpCO, SpMet, SpHb, and RRac in motion/no-motion and perfusion conditions. Predictive Temperature for oral/axillary/rectal prediction or continuous monitoring in adults/pediatrics. NanoPod capnography for CO2 and respiration rate in neonates/pediatrics/adults. IPI for ventilatory status in adults and pediatric patients (1-12 years).
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 2, 2025
Philips Medizin Systeme Böblingen GmbH
Monica Da Silva
Principal Regulatory Affairs Specialist
Hewlett-Packard-Str. 2
Boeblingen, 71034
Germany
Re: K250453
Trade/Device Name: IntelliVue Patient Monitor 6100 (6100); IntelliVue Patient Monitor 6300 (6300); IntelliVue Patient Monitor 6500 (6500)
Regulation Number: 21 CFR 870.1025
Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm)
Regulatory Class: Class II
Product Code: MHX, MLD, DSJ, DSK, DXN, DRT, DPS, DQA, DSF, MSX, CCK, BZQ, FLL
Dated: August 25, 2025
Received: August 25, 2025
Dear Monica da Silva:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250453 - Monica da Silva
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250453 - Monica da Silva
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JENNIFER W. SHIH -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
# K250453
## Device Name
IntelliVue Patient Monitor 6100 (6100);
IntelliVue Patient Monitor 6300 (6300);
IntelliVue Patient Monitor 6500 (6500)
## Indications for Use (Describe)
**Intended Use:**
The devices are intended to be used for monitoring and recording of, and to generate alarms for multiple physiological parameters of adults, pediatrics, and neonates.
**Indications for Use:**
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are only for use on one patient at a time.
The monitors are not therapeutic devices.
**Caution:** The monitors are for prescription use only.
The ECG measurement is indicated to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11).
ST segment monitoring is indicated for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is indicated for use with adult patients only.
The derived measurement Pulse Pressure Variation (PPV) is indicated for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.
The Predictive Temperature module is indicated to provide an accurate prediction of patient temperature using the oral, axillary, or rectal body sites, or to provide an actual temperature reading in the continuous monitor mode to adult and pediatric patients.
The NanoPod capnography module is indicated to provide professionally trained health care providers with continuous, non-invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate. It is indicated for use with neonatal, pediatric, and adult patients in hospitals, hospital-type facilities, intra-hospital transport.
The NanoPod capnography module provides the clinician with an integrated pulmonary index (IPI). The IPI is based on end tidal carbon dioxide, respiration rate, oxygen saturation and heart rate. The IPI is a single index of an adult or pediatric patient's ventilatory status displayed on a scale of 1 - 10, where 10 indicates optimal pulmonary status. IPI monitoring displays a single value that represents the patient's pulmonary parameters and alerts clinicians to changes in the patient's pulmonary status. The Integrated Pulmonary Index (IPI) is indicated for use with adult and pediatric (1 to 12 years) patients only.
{4}
Warning: The IPI is an adjunct to and not intended to replace vital sign monitoring.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}
PHILIPS
IntelliVue Patient Monitors
6100, 6300, 6500
510(k) Summary
K250453
| 510(k) Summary | | | | |
| --- | --- | --- | --- | --- |
| 1.1 Submitter | | | | |
| Date Prepared | February 14, 2025 | | | |
| Submitter/Owner | Philips Medizin Systeme Böblingen GmbH
FDA Establishment Number 9610816
Hewlett-Packard-Str. 2
71034 Böblingen
Germany | | | |
| Key Contact | Monica da Silva
Principal Regulatory Affairs Specialist
Monica.dasilva@philips.com
Phone: +49 175 3293 046 | | | |
| 510(k) Submission Type | Traditional 510(k) | | | |
| 1.2 Device | | | | |
| Trade Name | IntelliVue Patient Monitor 6100
IntelliVue Patient Monitor 6300
IntelliVue Patient Monitor 6500 | | | |
| Common Name | Multiparameter Patient Monitor | | | |
| Classification Name | Panel & Name: Cardiovascular Devices
Subpart & Division: 21 CFR §870.1025
Regulatory Class: II
Product Code: MHX | | | |
| 1.3 Predicate Device | | | | |
| Predicate Device | 510(k) No. | Company | Device Name | Product Code |
| | K223574 | Philips | IntelliVue Patient Monitor MX550 | MHX |
| The subject devices are substantially equivalent to the legally marketed predicate devices. | | | | |
| 1.4 Device Description | | | | |
| IntelliVue Patient Monitors 6100, 6300, 6500 – description of the device per 21 CFR 807.92(a) (4) | | | | |
| The new IntelliVue Patient Monitors 6100 (867311), 6300 (867313) and 6500 (867315) are display units with built-in CPUs. They integrate a TFT LCD flat panel display with dimensions of 10’ for model | | | | |
PHILIPS
Philips Hospital Patient Monitoring
Page 1 of 9
{6}
PHILIPS
IntelliVue Patient Monitors
6100, 6300, 6500
510(k) Summary
K250453
6100 and 6300 and 15' for the 6500 model). All Monitors are outfitted with a touchscreen interface as a primary input for the device. They also support data input devices such as a dedicated remote control, keyboard and pointing devices such as a mouse.
The new patient monitors have integrated measurements such as ECG/Resp, SpO2, NIBP, IBP and Temperature. In addition to their integrated measurements, these devices can also connect to other external measurement modules through their FlexLink connector. The FlexLink connector serves as an interface between the 6100, 6300 and 6500 Patient Monitors and Medtronic's Microstream™ CO2 NanoPod as well as the Predictive Temperature Module.
The new Patient Monitors run on mains power or on the internal battery. They serve as stationary patient monitors and can also be used as portable patient monitors in hospital transport only.
The new Patient Monitors can communicate with the Philips Central Station (aka Patient Information Center iX, last cleared with K211900) as well as with Philips XDS Software, which is intended for use as an independent display replicating the Patient Monitors screen. They are outfitted with an ECG-Out connector, which may be used to synchronize (via an analog ECG signal output) a connected external device such as a defibrillator, CT scan or Intra-aortic Balloon Pump.
## 1.5 Intended Use and Indication for Use
### Intended Use as required per 21 CFR 807.92(a)(5)
#### Intended Use:
The devices are intended to be used for monitoring and recording of, and to generate alarms for multiple physiological parameters of adults, pediatrics, and neonates.
#### Indications for Use:
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are only for use on one patient at a time.
The monitors are not therapeutic devices.
Caution: The monitors are for prescription use only.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC11).
PHILIPS
Philips Hospital Patient Monitoring
Page 2 of 9
{7}
PHILIPS
IntelliVue Patient Monitors
6100, 6300, 6500
510(k) Summary
K250453
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
The SSC Sepsis Protocol, in the Protocol Watch clinical decision support tool, is indicated for use with adult patients only.
The derived measurement Pulse Pressure Variation (PPV) is indicated for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.
The Predictive Temperature Module is indicated to provide an accurate prediction of patient temperature using the oral, axillary, or rectal body sites, or to provide an actual temperature reading in the continuous monitor mode to adult and pediatric patients.
The NanoPod capnography module is indicated to provide professionally trained health care providers with continuous, non-invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate. It is indicated for use with neonatal, pediatric, and adult patients in hospitals, hospital-type facilities, intra-hospital transport.
The NanoPod capnography module provides the clinician with an integrated pulmonary index (IPI). The IPI is based on end tidal carbon dioxide, respiration rate, oxygen saturation and heart rate. The IPI is a single index of an adult or pediatric patient's ventilatory status displayed on a scale of 1 - 10, where 10 indicates optimal pulmonary status. IPI monitoring displays a single value that represents the patient's pulmonary parameters and alerts clinicians to changes in the patient's pulmonary status.
The Integrated Pulmonary Index (IPI) is indicated for use with adult and pediatric (1 to 12 years) patients only.
Warning: The IPI is an adjunct to and not intended to replace vital sign monitoring.
PHILIPS
Philips Hospital Patient Monitoring
Page 3 of 9
{8}
PHILIPS
IntelliVue Patient Monitors
6100, 6300, 6500
510(k) Summary
K250453
# 1.6 Comparison of Intended Uses for Subject Device and Predicate
The new IntelliVue Patient Monitors 6100, 6300, 6500 have the same intended use as the predicate device IntelliVue Patient Monitor MX550. The subject and predicate device are intended for monitoring, recording and alarming of physiological parameters for adults, pediatrics and neonates. Subject and predicate device are intended for use by trained healthcare professionals in a hospital environment.
For the most part the subject device has the same indications for use as the predicate device. Due to the subject device's compatibility with the Predictive Temperature Module and Medtronic's Microstream™ CO2 NanoPod additional indications for use were added to the subject device's indication for use:
- The Predictive Temperature Module is indicated to provide an accurate prediction of patient temperature using the oral, axillary, or rectal body sites, or to provide an actual temperature reading in the continuous monitor mode to adult and pediatric patients.
- The NanoPod capnography module is indicated to provide professionally trained health care providers with continuous, non-invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate. It is indicated for use with neonatal, pediatric, and adult patients in hospitals, hospital-type facilities, intra-hospital transport.
The NanoPod capnography module provides the clinician with an integrated pulmonary index (IPI). The IPI is based on end tidal carbon dioxide, respiration rate, oxygen saturation and heart rate. The IPI is a single index of an adult or pediatric patient's ventilatory status displayed on a scale of 1 - 10, where 10 indicates optimal pulmonary status. IPI monitoring displays a single value that represents the patient's pulmonary parameters and alerts clinicians to changes in the patient's pulmonary status.
PHILIPS
Philips Hospital Patient Monitoring
Page 4 of 9
{9}
PHILIPS
IntelliVue Patient Monitors
6100, 6300, 6500
510(k) Summary
K250453
| 1.7 Comparison of Technological Characteristics with Predicate Device | |
| --- | --- |
| Item of Comparison | Description/Rationale |
| Similarities | |
| Device Design | - Subject device is designed for stationary and in-hospital transport use as the predicate device.
- Subject device runs on IntelliVue Software, which is the same software platform available in the predicate device.
- Subject device is designed with integrated measurements ECG/Resp, SpO2, NIBP, IBP and Temperature, which can also be displayed in the predicate device when connected to external measurements modules. |
| Energy Source | - Subject device is powered by AC mains or from built-in battery, as the predicate device.
- Subject and predicate device do not deliver energy to patients. |
| Software/Hardware Features | - Subject device does not introduce any new technological hardware features.
- Measurements offered by the subject device are also offered by the predicate device via its compatibility with external measurement modules.
- Software features introduced with IntelliVue Software version R, rely on well-established software platform and architecture which was iteratively developed from the version available in the predicate device. |
| Physiological Parameters | - Measurements embedded in subject device are existing physiological parameters which are also displayed on the predicate device through its connectivity with external measurements. |
| Performance specifications | - Specifications of all measurement characteristics, including measurement principles, methods, algorithms, and all detailed performance specifications are unchanged from existing IntelliVue Patient Monitors i.e. reference device (IntelliVue X3)
- Performance specifications of the external measurement that are compatible with the subject device remain unchanged from their original clearance i.e. Medtronic’s Microstream™ CO2 NanoPod (K213911) and SureTemp® Plus (K030580) for the Predictive Temperature Module. |
| Operating Principle and Mechanism of Action | - Operating principle and mechanism of action from subject device is unchanged from the predicate device. |
PHILIPS
Philips Hospital Patient Monitoring
Page 5 of 9
{10}
PHILIPS
IntelliVue Patient Monitors
6100, 6300, 6500
K250453
Traditional 510(k)
510(k) Summary
| 1.7 Comparison of Technological Characteristics with Predicate Device | |
| --- | --- |
| Item of Comparison | Description/Rationale |
| Human Interface | • human interface of the new IntelliVue Patient Monitors 6100, 6300 and 6500 leverages from the well established IntelliVue Software, which also runs on the predicate device. |
| Measurement Accessories | • Accessories applied to the predicate device are compatible with the subject devices. |
| Differences | |
| Device Design | • Subject device benefits of updated electronic components when compared to predicate device.
• Subject device integrates a new Philips proprietary FlexLink connector |
| Materials | • Subject device is outfitted with chemically strengthened glass when compared to predicate device
• Subject device is outfitted with modified plastic to better withstand disinfectant agents. |
| Physiological Parameters | • Subject device is compatible with external measurements with previously cleared technologies i.e. Medtronic’s Microstream™ CO2 NanoPod (K213911) and SureTemp® Plus (K030580) for the Predictive Temperature Module. |
| Measurement Accessories | • Subject device introduces a new Predictive Temperature FlexLink Cable (P/N 453564951751). |
| Software | • The IntelliVue Software version R.00.01 that is embedded in the subject device introduces:
o the integration of new hardware components and functional elements for the subject device such as display resolution and compatibility with other medical devices i.e. Medtronic’s Microstream™ CO2 NanoPod (P/N 989803209661) and the Predictive Temperature Module (P/N 867394)
o optimizations to user interface and cybersecurity (i.e “Issue Helper Window”, “My Apps Window”, “Extended Alarm Logging” etc..) and
o new applications such as the “Clinical Monitoring Analytics” |
PHILIPS
Philips Hospital Patient Monitoring
Page 6 of 9
{11}
PHILIPS
IntelliVue Patient Monitors
6100, 6300, 6500
510(k) Summary
K250453
| 1.8 Substantial Equivalence Summary | | |
| --- | --- | --- |
| Operational and technological characteristics form the basis for the determination of substantial equivalence of the subject devices with the legally marketed predicate devices. The subject devices are substantially equivalent to the predicate devices. | | |
| 1.9 Performance Data | | |
| Clinical Tests – Harmonized Standards | | |
| The subject devices have passed all safety tests for demonstrated compliance with the recognized standards below. | | |
| Standard | FDA Recognition # | Title # |
| IEC 60601-1:2005 incl. AMD1:2012 and AMD2:2020 | 19-49 | Medical Electrical Equipment – Part 1: General requirements for basic safety and essential performance. |
| IEC 60601-1-2:2014 incl. AMD1:2020 | 19-36 | Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests. |
| IEC 60601-1-6:2010 incl. AMD1:2013 and AMD2:2020 | 5-132 | Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability. |
| IEC 60601-1-8:2006 incl. AMD1:2012 and AMD2:2020 | 5-131 | Medical electrical equipment – Part 1-8: General requirements for basic safety and essential performance – Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems. |
| IEC 60601-2-25:2011 | 3-105 | Medical electrical equipment – Part 2: Particular requirements for the basic safety and essential performance of electrocardiographs. |
| IEC 60601-2-27:2011 | 3-126 | Medical electrical equipment – Part 2: Particular Requirements for the Safety, Including Essential Performance of Electrocardiographic Monitoring Equipment. |
PHILIPS
Philips Hospital Patient Monitoring
Page 7 of 9
{12}
PHILIPS
IntelliVue Patient Monitors
6100, 6300, 6500
510(k) Summary
K250453
| IEC 80601-2-30:2018 | 3-123 | Medical electrical equipment – Part 2: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers. |
| --- | --- | --- |
| IEC 60601-2-34:2011 | 3-115 | Medical electrical equipment – Part 2-34: Particular requirements for the basic safety and essential performance of invasive blood pressure monitoring equipment. |
| ISO 80601-2-55:2018 | 1-140 | Medical electrical equipment – Part 2: Particular requirements for the basic safety and essential performance of respiratory gas monitors. |
| ISO 80601-2-56:2017
incl. AMD1:2018 | 6-421 | Medical electrical equipment – Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. |
| ISO 80601-2-61:2017
incl. COR1:2018 | 1-139 | Medical electrical equipment – Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment |
| IEC 62304:2006
incl. AMD1: 2015 | 13-79 | Medical device software Software life-cycle processes |
| IEC 62133-2:2017
incl. AMD1: 2021 | 19-33 | Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems |
| UL 2054 | 19-11 | Household and Commercial Batteries |
PHILIPS
Philips Hospital Patient Monitoring
Page 9 of 9
{13}
PHILIPS
IntelliVue Patient Monitors
6100, 6300, 6500
510(k) Summary
K250453
| Non-clinical Bench Tests |
| --- |
| No new issues of safety or effectiveness as compared to the predicate are introduced with the new IntelliVue Patient Monitors 6100, 6300, 6500. |
| Clinical Studies |
| The subject devices, like the predicate device, did not require any clinical trials.
The 510(k) included clinical data from Philips initiated laboratory tests with the intent to demonstrate the accuracy performance of the Philips FAST PicoSAT X SpO2 technology.
The activities involved in the development and release of the subject device are compliant with FDA Quality System Regulations, FDA recognized standards, FDA guidance documents and harmonized standards. The subject device was subjected to several verification and validation activities, usability validation and risk management activities.
Based upon the design, intended use, indications for use, classification, usability and safety testing, the subject devices are substantially equivalent to the predicate device. |
# 1.10 CONCLUSION
Substantial equivalence assessment, based on non-clinical bench testing, electrical safety, electromagnetic compatibility, software verification and validation, human factors, interoperability testing and clinical tests (for Philips FAST PicoSAT X SpO2 technology), demonstrate that the new IntelliVue Patient Monitors 6100, 6300 and 6500 do not raise different questions of safety and effectiveness when compared to their predicate device IntelliVue Patient Monitor MX550. The new IntelliVue Patient Monitors 6100, 6300 and 6500 perform as intended and have performance characteristics that are substantially equivalent to IntelliVue Patient Monitor MX550.
PHILIPS
Philips Hospital Patient Monitoring
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.