The devices are intended to be used for monitoring and recording of, and to generate alarms for multiple physiological parameters of adults, pediatrics, and neonates.
Device Story
IntelliVue MX400/MX450/MX500/MX550 are multiparameter patient monitors acquiring physiological signals (ECG, SpO2, BP, CO2, EEG, etc.) for display, alarm generation, and data management. Used in surgical, cardiac, medical, and neonatal care environments; operated by trained healthcare professionals at bedside or during hospital transport. Monitors feature touchscreen interfaces and support external modules (e.g., Masimo SedLine) via plug-in interfaces. Output includes real-time waveforms, numerics, and trends. New software (P.01) adds animated 'Visual Patient' avatar, configurable alarm management, and updated alarm sounds. Device aids clinical decision-making by providing continuous physiological status updates. Benefits include real-time patient monitoring and support for anesthetic/sedation management.
Clinical Evidence
No clinical trials were required. Substantial equivalence is supported by non-clinical bench testing, electrical safety, electromagnetic compatibility, software verification/validation, human factors, and usability testing. Clinical studies previously performed for the integrated Masimo SedLine parameter remain valid as the measurement technology is unchanged.
Technological Characteristics
Multiparameter monitor with touchscreen; powered by AC mains or battery. Connectivity via LAN/wireless. Plug-in module 867186 provides isolated digital communication. Software P.01 includes animated avatar, alarm management, and SSO. Complies with IEC 62304, ANSI AAMI ES60601-1, IEC 60601-1-2, 60601-1-8, and 60601-1-6.
Indications for Use
Indicated for use by healthcare professionals for monitoring physiological parameters of adult, pediatric, and neonatal patients. Specific measurements include ECG (diagnostic rhythm/morphology), ST segment (adults only), BIS (adult/pediatric brain state/anesthesia aid), SSC Sepsis Protocol (adults), IPI (adult/pediatric 1-12 years), PPV (sedated adult patients on mechanical ventilation), NMT (adult/pediatric neuromuscular transmission), Masimo rainbow SET (SpO2, pulse rate, SpCO, SpMet, SpHb, RRac), Masimo O3 (cerebral rSO2, adults >=40kg, pediatrics 5-40kg), and SedLine (adult brain state/anesthesia monitoring). Intended for hospital use, including transport.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 22, 2023
Philips Medizin Systeme Boeblingen GmbH Siegfried Breitling Regulatory Affairs Specialist Hewlett-Packard-Strasse 2 Boeblingen, BW 71034 Germany
Re: K223574
Trade/Device Name: IntelliVue Patient Monitor MX400 (866060), Intelli Vue Patient Monitor MX450 (866062), IntelliVue Patient Monitor MX500 (866064), IntelliVue Patient Monitor MX550 (866066) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, MLD, DSJ, DSK, DXN, DRT, DPS, DQA, DSF, MSX, CCK, CBQ, NHO, NHP, NHQ, CBS, CBR, CCL, BZC, BZQ, LKD, KLK, KOI, GWQ, GWS, FLL Dated: November 29, 2022 Received: November 30, 2022
Dear Siegfried Breitling:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jennifer W. Shih -S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223574
### Device Name
IntelliVue Patient Monitor MX400 (866060); IntelliVue Patient Monitor MX450 (866062); IntelliVue Patient Monitor MX500 (866064); IntelliVue Patient Monitor MX550 (866066)
### Indications for Use (Describe)
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are only for use on one patient at a time.
The monitors are not therapeutic devices.
The monitors are for prescription use only.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC11).
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.
The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only. The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
The Intelli Vue NMT Module is intended to be used as an objective neuromuscular transmission monitor, using accelerometry for measuring the muscle contraction following an electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients.
The Masimo rainbow SET measurement is indicated for the nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.
The noninvasive Masimo O3 Regional Oximeter System and accessories are intended for use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors. The Masimo O3 Regional Oximeter System and accessories are indicated for use on adults ≥40 kg and on pediatrics ≥5 kg and <40 kg, in healthcare environments.
The SedLine Sedation Monitor is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The agents include: Alfentani, Desflurane, Nitrous Oxide, Propofol, Remifentanil, and Sevoflurane. The Sedation Monitor is intended for use with adult patients (18 years of age and older) in the operating room (OR), intensive care unit (ICU), and clinical research laboratory.
The monitors are intended for use by trained healthcare professionals in a hospital environment. They are not intended for home use.
The MX400/MX450/MX500/MX550 monitors are additionally intended for use in transport situations within hospital environments.
{3}------------------------------------------------
### Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
## 510(k) Summary
| 510(k) Summary | |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1.1 Submitter | |
| Date Prepared | November 29, 2022 |
| Submitter/Owner | Philips Medizin Systeme Böblingen GmbH<br>FDA Establishment Number 9610816<br>Hewlett-Packard-Str. 2<br>71034 Böblingen<br>Germany |
| Key Contact | Siegfried Breitling<br>Regulatory Affairs Specialist<br>siegfried.breitling@philips.com<br>Phone: +49 151 200 443 77 |
| 510(k) Submission Type | This is a traditional 510(k). |
| 1.2 Device | |
| Trade Name | IntelliVue Patient Monitor MX400<br>IntelliVue Patient Monitor MX450<br>IntelliVue Patient Monitor MX500<br>IntelliVue Patient Monitor MX550 |
| Common Name | Multiparameter Patient Monitor |
| Classification Name | Panel & Name: Cardiovascular Devices<br>Subpart & Division: 21 CFR §870.1025<br>Regulatory Class: II<br>Classification Product Code: MHX<br>Subsequent Product Codes: MLD, DSJ, DSK, DXN, DRT, DPS, DQA, DSF,<br>MSX, CCK, CBQ, NHO, NHP, NHQ, CBS, CBR, CCL, BZC, BZQ, LKD, KLK, KOI,<br>GWQ, GWS, FLL |
| 1.3 Predicate Device | |
| | 510(k) No. | Company | Device Name | Product Code |
|------------------|------------|---------|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| Predicate Device | K182979 | Philips | IntelliVue Patient Monitor MX400<br>IntelliVue Patient Monitor MX450<br>IntelliVue Patient Monitor MX500<br>IntelliVue Patient Monitor MX550 | MHX<br>DRT, DSI, DSJ, DSK, DXN, MLD, MSX |
Image /page/4/Picture/6 description: The image shows the Philips logo, which is a blue shield shape. Inside the shield, the word "PHILIPS" is written in white at the top. Below the text, there are two wavy lines and four stars, also in white. The overall design is simple and recognizable, representing the brand's identity.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is plain white, which makes the blue letters stand out.
| | K172890 | Masimo | SedLine Sedation Monitor | OLW |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|--------|--------------------------|--------------------|
| | | | | GXY, OLT, OMC, ORT |
| The subject devices are substantially equivalent to the legally marketed predicate devices. | | | | |
| 1.4 | Device Description | | | |
| IntelliVue Patient Monitors MX400, MX450, MX500 and MX550 - description of the device per 21<br>CFR 807.92(a) (4) | | | | |
| The IntelliVue Patient Monitors MX400, MX450, MX500 and MX550 acquire multiple physiological<br>patient signals, display measurement values, waves and trends, generate physiological and<br>technical alarms, provide data recording and support patient data management. | | | | |
| The monitors support multiple non-invasive and invasive measurements such as ECG, arrhythmia,<br>ST, QT, SpO2, respiration rate, pulse rate, heart rate, invasive and non-invasive blood pressure,<br>temperature, CO2, tcpO2/ tcpCO2, C.O., CCO, intravascular SO2, SvO2, ScvO2, EEG, BIS, NMT, and<br>gas analysis. | | | | |
| The monitors offer a monitoring solution optimized for the surgical, cardiac, medical and neonatal<br>care environments. They are located at the patient bedside vicinity and can also be used during<br>patient transport inside hospitals. The monitors have a color display with touchscreen as a primary<br>input device. They also support a specialized remote control, keyboard and pointing devices such as<br>a mouse. External displays can be connected to a built-in video port to provide an adaptive<br>duplicate image of the primary display. The monitors can interact with several compatible external<br>measuring and auxiliary devices locally at the bedside or in transport situations and with the Central<br>Station via LAN or wireless link. | | | | |
### Intended Use and Indication for Use 1.5
## Intended Use as required per 21 CFR 807.92(a)(5)
Intended Use:
The devices are intended to be used for monitoring and recording of, and to generate alarms for multiple physiological parameters of adults, pediatrics, and neonates.
## Indications for Use:
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are only for use on one patient at a time.
Image /page/5/Picture/11 description: The image shows the Philips logo, which is a blue shield shape. Inside the shield, there are two wavy lines that resemble water or sound waves. Above and below the waves are four-pointed star shapes. Above the shield, the word "PHILIPS" is written in a sans-serif font, also in blue.
{6}------------------------------------------------
The monitors are not therapeutic devices. The monitors are for prescription use only.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC11).
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.
The SSC Sepsis Protocol, in the Protocol Watch clinical decision support tool, is intended for use with adult patients only.
The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
The IntelliVue NMT Module is intended to be used as an objective neuromuscular transmission monitor, using accelerometry for measuring the muscle contraction following an electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients.
The Masimo rainbow SET measurement is indicated for the nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion conditions, and for patients who are well or poorly perfused.
The noninvasive Masimo O3 Regional Oximeter System and accessories are intended for use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors. The Masimo O3 Regional Oximeter System and accessories are indicated for use on adults ≥40 kg and on pediatrics ≥5 kg and <40 kg, in healthcare environments.
Image /page/6/Picture/13 description: The image shows the Philips logo, which is a blue shield shape. Inside the shield, there are two wavy lines that resemble water or sound waves. Above and below the wavy lines, there are four-pointed stars. The word "PHILIPS" is written in white letters at the top of the shield.
{7}------------------------------------------------
The SedLine Sedation Monitor is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The agents include: Alfentanil, Desflurane, Fentanyl, Isoflurane, Nitrous Oxide, Propofol, Remifentanil, and Sevoflurane. The SedLine Sedation Monitor is intended for use with adult patients (18 years of age and older) in the operating room (OR), intensive care unit (ICU), and clinical research laboratory.
The monitors are intended for use by trained healthcare professionals in a hospital environment. They are not intended for home use. The MX400/MX450/MX550 monitors are additionally intended for use in transport situations within hospital environments.
### Comparison of Intended Uses for Subject Device and Predicate 1.6
The modified IntelliVue MX400, MX450, MX500 and MX550 have the same intended use and indications as the predicate devices. Additionally, the MX500 and MX550 devices inherited the Masimo SedLine Sedation Monitor indication:
- The SedLine Sedation Monitor is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The agents include: Alfentanil, Desflurane, Fentanyl, Isoflurane, Nitrous Oxide, Propofol, Remifentanil, and Sevoflurane. The SedLine Sedation Monitor is intended for use with adult patients (18 years of age and older) in the operating room (OR), intensive care unit (ICU), and clinical research laboratory.
### 1.7 Comparison of Technological Characteristics with Predicate Device
| Similarities | |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item of<br>Comparison | Description/Rationale |
| Device Design | device design of subject devices is the same as for the predicate devices unchanged hardware and architecture monitor hardware entirely unchanged device design of added external measurements Masimo SedLine, is identical to their predicate devices |
| Materials | same materials used as those of the predicate devices materials used in the new plug-in module SedLine (867186), are the same as those of other legally marketed plug-in modules biocompatibility aspects do not apply because the devices do not have patient contact. materials used in the added external measurement Masimo SedLine, are identical to their predicate devices. biocompatibility aspects of accessories are not affected, because all accessories of the predicate devices remain unchanged |
| Energy Source | powered by AC mains or from built-in battery, same as predicate devices devices do not deliver energy to the patients for their function, same as predicate devices the added external measurement Masimo SedLine, does not have its own power supply; It is supplied by its host patient monitor. This is the same as for its predicate device |
| Software/Hardware<br>Features | proposed modification does not introduce any new technological hardware features predicate devices already support plug-in modules, with these modules offering built-in measurements as well as interfaces to external measurements hardware and software of the added external measurement Masimo SedLine is unchanged from its predicate device measurements of physiological parameters are exactly the same as in the predicate devices |
| Physiological<br>Parameters | existing physiological parameters of the predicate devices remain unchanged the added external measurement Masimo SedLine is the same as those of its respective predicate device |
| Performance<br>specifications | specifications of all measurement characteristics, including measurement principles, methods, algorithms, and all detailed performance specifications remain unchanged performance specifications of the added external measurement Masimo SedLine remains unchanged |
| Operating Principle<br>and Mechanism of<br>Action | unchanged from the predicate devices |
| Human Interface | human interface of the IntelliVue MX400, MX450, MX500 and MX550 remain exactly the same the newly added external measurement Masimo SedLine does not have a human interface of its own |
| Measurement<br>Accessories | all accessories of the predicate devices IntelliVue Patient Monitor MX400 IntelliVue Patient Monitor MX450 IntelliVue Patient Monitor MX500 IntelliVue Patient Monitor MX550 as well as those of the added external measurement Masimo SedLine are re-used without any change |
| Differences | |
| Hardware | one new plug-in module SedLine (product number 867186)is created, which acts as a digital communication interface between the IntelliVue Patient Monitors MX400/MX450/MX500/MX550 and the external Masimo measurement devices the new plug-in module does not alter or modify the contents of the transferred data the new plug-in module also provides electrical isolation for both power and communication between patient monitor and the external SedLine measurement device; the patient isolation circuit is the same as for other legally marketed plug-in modules for the IntelliVue patient monitor family the predicate external measurement device Masimo SedLine also connects to a host monitor / backboard device using an isolated digital interface, where the isolation must be provided by that host monitor / backboard device; the new plug-in module 867186 provides exactly this isolated digital interface as optional plug-in module, specialized for the IntelliVue Patient Monitors MX400, MX450, MX500 and MX550 |
Image /page/7/Picture/10 description: The image shows the Philips logo, which is a blue shield shape. Inside the shield, there are two wavy lines that resemble sound waves or water waves. Above and below the waves are four-pointed star shapes. The word "PHILIPS" is written in white at the top of the shield.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the text "K223574" in bold font at the top. Below that is the text "510(k) Summary" in a larger, bold font. The text is aligned to the left and appears to be the title or heading of a document.
Image /page/8/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are all capitalized and evenly spaced. The font is sans-serif and the overall impression is clean and modern.
Image /page/8/Picture/4 description: The image shows the Philips logo, which is a blue shield shape. Inside the shield, there are two wavy lines and four stars. The word "PHILIPS" is written in white at the top of the shield.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white.
Image /page/9/Picture/4 description: The image shows the Philips logo. The logo is a blue shield shape with the word "PHILIPS" in white at the top. Inside the shield, there are two wavy lines and two stars. The wavy lines are white and positioned horizontally across the center of the shield. The stars are four-pointed and located on either side of the wavy lines.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and appear to be a sans-serif font. The background is plain white, which makes the blue letters stand out.
| Software | The software of the subject devices IntelliVue MX400, MX450, MX500 and MX550 has been extended so that the new plug-in module will be accepted as compatible and the physiological data provided by the Masimo SedLine measurement device will be displayed on the patient monitor screen. Additionally, a new software version designated as IntelliVue Software P.01 introduced the following new features: New alarm sounds designated as "Philips 2021" sounds. New Visual Patient feature, which displays patients's vital signs in an animated patient avatar alongside the conventional vital sign waveforms and numerics. Configurable alarm management, limiting changes in alarm settings to pre-defined authorized personnel. Additionally, the new Software revision will further enhance the following existing features: Early Warning Score Validation followed by user identification via Single Sign On Additional SPO2 fallback configuration Additional Lead Diagram presentation for ECG Enhancement of presentation of the global "alarm off" state |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Physiological<br>Parameters | For marketing reasons, the proposed modification does not provide the parameters and graphical presentations DSA (SedLine) which the predicate device is providing. |
Image /page/10/Picture/4 description: The image shows the Philips logo, which is a blue shield shape. Inside the shield, there are two wavy lines and four stars. The word "PHILIPS" is written in blue at the top of the shield.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
### Substantial Equivalence Summary
Operational and technological characteristics form the basis for the determination of substantial equivalence of the subject devices with the legally marketed predicate devices. The subject devices are substantially equivalent to the predicate devices.
### 1.8 Performance Data
### Non-Clinical Tests – Harmonized Standards
The subject devices have passed all safety tests for demonstrated compliance with the recognized standards below.
| Standard | FDA<br>Recognition # | Title # |
|------------------------------------------------------------------------------------------------|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 62304 Edition 1.1<br>2015-06 consolidated<br>version | 13-79 | Medical device software - Software life cycle processes |
| ANSI AAMI ES60601-<br>1:2005/(R)2012 and<br>A1:2012,<br>C1:2009/(R)2012 and<br>A2:2010/(R)2012 | 19-4 | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential<br>performance |
| IEC 60601-1-2 Edition<br>4.0 2014-02 | 19-8 | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential<br>performance - Collateral Standard: Electromagnetic<br>disturbances - Requirements and tests |
| 60601-1-8 Edition 2.2<br>2020-07<br>CONSOLIDATED<br>VERSION | 5-131 | Medical electrical equipment - Part 1-8: General<br>requirements for basic safety and essential<br>performance - Collateral Standard: General<br>requirements tests and guidance for alarm systems in<br>medical electrical equipment and medical electrical<br>systems |
| 60601-1-6 Edition 3.1<br>2013-10 | 5-89 | Medical electrical equipment - Part 1-6: General<br>requirements for basic safety and essential<br>performance - Collateral standard: Usability |
Image /page/11/Picture/9 description: The image shows the Philips logo, which is a blue shield shape. Inside the shield, there are two wavy lines and two four-pointed stars. The word "PHILIPS" is written in white at the top of the shield.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is the only element in the image and it is centered.
## Non-clinical Bench Tests
No new issues of substantial equivalence are introduced because of using this device
### Clinical Studies
The subject devices, like the primary predicate devices, did not require clinical trials. Any clinical studies performed for the Masimo SedLine parameter are still valid as the measurements are not modified; they are only being connected to an additional host patient monitor.
Compliance to the FDA Quality System Regulations, FDA recognized standards, FDA guidance documents, harmonized standards, verification and validation, software validation, usability validation, and risk management activities have taken place for the subject devices.
Based upon the design, intended use, indications for use, classification, usability and safety testing, the subject devices are substantially equivalent to the listed predicate devices.
No new issues of substantial equivalence are introduced as a result of using this device.
### 1.9 CONCLUSIONS
The results of the substantial equivalence assessment, taken together with non-clinical bench testing, electrical safety and electromagnetic compatibility, software verification and validation, human factors, usability and interoperability testing, demonstrate that the modified devices do not raise different questions of safety and effectiveness when compared to the predicate, perform as intended, and have performance characteristics that are substantially equivalent to the predicate devices.
Image /page/12/Picture/12 description: The image shows the Philips logo, which is a blue shield shape. Inside the shield, there are two wavy lines and two four-pointed stars. Above the shield, the word "PHILIPS" is written in white letters.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.