GCI Sterilization Wrappers

K250321 · George Courey, Inc. · FRG · Jun 6, 2025 · General Hospital

Device Facts

Record IDK250321
Device NameGCI Sterilization Wrappers
ApplicantGeorge Courey, Inc.
Product CodeFRG · General Hospital
Decision DateJun 6, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6850
Device ClassClass 2

Indications for Use

GCI Sterilization Wrappers are intended to be used to enclose reusable textile-based medical devices, such as surgical gowns, towels, and drapes, for sterilization by a health care provider. They are designed to allow effective steam sterilization of the enclosed textile items and to maintain sterility until use, for up to 28 days.

Device Story

GCI Sterilization Wrappers function as a physical barrier to maintain sterility of textile-based medical devices (gowns, towels, drapes) following steam sterilization. Wrappers are non-sterile, reusable textile sheets (99% polyester, 1% carbon) available in various sizes (18"x18" to 60"x60"). Healthcare staff wrap devices using established folding procedures; closure is secured with autoclave tape or elastomer closures. The device is validated for a prevacuum steam sterilization cycle (132°C for 4 minutes, 30-minute dry time) and supports up to 38 wash/sterilization cycles. Color-coded borders assist in size identification. By maintaining a sterile barrier, the device ensures the safety of surgical instruments for clinical use.

Clinical Evidence

No clinical data. Bench testing only. Performance validated per ASTM D5587 (tearing strength), ASTM D5034 (breaking strength), 16 CFR 1610 (flammability), AATCC 127 (hydrostatic pressure), AATCC 42 (impact penetration), and ISO 10993 series (biocompatibility). Sterility maintenance validated for 28 days; reprocessing effectiveness confirmed for 38 cycles; sterilization SAL of 10^-6 demonstrated.

Technological Characteristics

Woven fabric (99% polyester, 1% carbon). Physical barrier principle. Reusable (38 cycles). Sizes: 18"x18" to 60"x60". Color-coded borders. Sterilization: Prevacuum steam (132°C, 4 min exposure, 30 min dry). Biocompatibility per ISO 10993-1, -5, -10. Standards: ASTM D5587, ASTM D5034, AATCC 127, AATCC 42, ISO 9073, AAMI TIR 12, AAMI TIR 30.

Indications for Use

Indicated for use by healthcare providers to enclose reusable textile-based medical devices (gowns, towels, drapes) for steam sterilization and maintenance of sterility for up to 28 days. Validated for use with specific load weights ranging from 1 lb to 25 lbs depending on wrapper size.

Regulatory Classification

Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION June 6, 2025 George Courey Inc. % Sarah Fitzgerald Senior Consultant Emergo by UL 2500 Bee Cave Road Building 1, Suite 300 Austin, Texas 78746 Re: K250321 Trade/Device Name: GCI Sterilization Wrappers Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: May 6, 2025 Received: May 6, 2025 Dear Sarah Fitzgerald: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250321 - Sarah Fitzgerald Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250321 - Sarah Fitzgerald Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Christopher K. Dugard -S Christopher K. Dugard, M.S. Division Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K250321 Device Name GCI Sterilization Wrappers Indications for Use (Describe) GCI Sterilization Wrappers are intended to be used to enclose reusable textile-based medical devices, such as surgical gowns, towels, and drapes, for sterilization by a health care provider. They are designed to allow effective steam sterilization of the enclosed textile items and to maintain sterility until use, for up to 28 days. GCI Sterilization Wrapper sizes and validated load limitations are as follows: | SKU / Model | Size | Border Color | Maximum Validated Load Weight | | --- | --- | --- | --- | | P81818 KBRO | 18"x18" | White | 1 lbs | | P82424 KBRO | 24"x24" | Pink | 1.82 lbs | | P83030 KBRO | 30"x30" | Yellow | 3.45 lbs | | P83636 KBRO | 36"x36" | Blue | 5.8 lbs | | P84545 KBRO | 45"x45" | Mint | 10.98 lbs | | P85454 KBRO | 54"x54" | Red | 18.5 lbs | | P86060 KBRO | 60"x60" | Green | 25 lbs | GCI Sterilization Wrappers are reusable through 38 wash, dry, and sterilization cycles. Pack assembly should follow established folding, assembly, and wrapping procedures as defined by the clinical staff for aseptic presentation of the pack and its contents. Closure can include autoclave tape and elastomer closures. Sterilization: A prevacuum steam sterilization cycle (4 minutes exposure at $132^{\circ}\mathrm{C}$, Dry Time: 30 minutes) has been validated. Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 {4} K250321 510(k) Summary GCI Sterilization Wrappers 1. Submission Sponsor George Courey, Inc. 6620, Ernest-Cormier Laval, Quebec, H7C 2T5 Canada Contact: Cort Naab Title: Director of Surgical Solutions Telephone: 450-661-6620 E-mail: cnaab@georgecourey.com 2. Submission Correspondent Emergo by UL 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327-9997 Email: lst.us.EmergoFDASubmissions@ul.com Contact: Sarah Marie Fitzgerald Title: Senior Consultant, Quality and Regulatory Affairs 3. Date Prepared June 6, 2025 4. Device Identification Trade/Proprietary Name: GCI Sterilization Wrappers Common/Usual Name: Sterilization Wraps Classification Name: Sterilization Wrap Regulation Number: 880.6850 Product Code: FRG Class: 2 5. Legally Marketed Predicate Device Predicate Device name: Standard Supreme Sterilization Wrapper 510(k) number: K172207 Manufacturer: Standard Textile Co., Inc. Common/Usual Name: Sterilization Wraps Classification Name: Sterilization Wrap Regulation Number: 880.6850 Product Code: FRG Class: 2 6. Indication for Use Statement K250321 {5} GCI Sterilization Wrappers are intended to be used to enclose reusable textile-based medical devices, such as surgical gowns, towels, and drapes, for sterilization by a health care provider. They are designed to allow effective steam sterilization of the enclosed textile items and to maintain sterility until use, for up to 28 days. GCI Sterilization Wrapper sizes and load limitation are as follows: | SKU / Model | Size | Border Color | Maximum Validated Load Weight | | --- | --- | --- | --- | | P81818 KBRO | 18”x18” | White | 1 lbs | | P82424 KBRO | 24”x24” | Pink | 1.82 lbs | | P83030 KBRO | 30”x30” | Yellow | 3.45 lbs | | P83636 KBRO | 36”x36” | Blue | 5.8 lbs | | P84545 KBRO | 45”x45” | Mint | 10.98 lbs | | P85454 KBRO | 54”x54” | Red | 18.5 lbs | | P86060 KBRO | 60”x60” | Green | 25 lbs | GCI Sterilization Wrappers are reusable through 38 wash, dry, and sterilization cycles. Pack assembly should follow established folding, assembly, and wrapping procedures as defined by the clinical staff for aseptic presentation of the pack and its contents. Closure can include autoclave tape and elastomer closures. Sterilization: A prevacuum steam sterilization cycle (4 minutes exposure at 132°C, Dry Time: 30 minutes) has been validated. ## 7. Device Description GCI Sterilization Wrappers are intended to enclose a textile-based medical device for sterilization and maintain sterility of the enclosed device until used. The proposed devices are medical devices that are provided non-sterile. They may be reused up to 38 times. They are available in multiple sizes to accommodate textile-based medical devices, ranging from 18”x18” to 60”x60”. The size is indicated by color coding of the wrap borders for easy identification. Load sizes should be large enough to prevent too much overlapping of the wrapper material (such that all but one side of the wrapper should be a single layer when properly folded) to allow appropriate sterilant penetration. Load sizes should be small enough to overlap by at least 2 inches on the remaining side to ensure there are no gaps. ## 8. Technological Characteristics Comparison Table 1 compares the subject wraps to the predicate device with respect to intended use, principles of operation, technological characteristics, materials, and performance. Table 2 compares the specific indications for use. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device. Table 1. Technological Characteristics Comparison Table | Attribute | SUBJECT: K250321 GCI Surgical Wrappers | PREDICATE: K172207 Sterilization Wrappers | Comparison | | --- | --- | --- | --- | | Class | II | II | Same | | Product Code | FRG | FRG | Same | K250321 {6} | Attribute | SUBJECT: K250321 GCI Surgical Wrappers | PREDICATE: K172207 Sterilization Wrappers | Comparison | | --- | --- | --- | --- | | Regulation (21 CFR) | 880.6850 | 880.6850 | Same | | Intended Use | Enclose another medical device for sterilization and maintain sterility of the enclosed device until used. | Enclose another medical device for sterilization and maintain sterility of the enclosed device until used. | Same | | Method of Action | Physical barrier | Physical barrier | Same | | Durability | Reusable | Reusable | Same | | Use Life | 38 uses | 75 uses | Similar* | | Colors | Green with additional colors on edging (white, pink, yellow, blue, mint, red, green) | Red, Yellow, Blue, or Green | Similar* Colors provide easy reference. | | Available Sizes | - 18"x18" 24"x24" - 30"x30" 36"x36" 45"x45" 54"x54" - 60"x60" - | 12"x12" 18"x18" 23"x23" 27"x27" 30"x30" 35"x35" 45"x45" 54"x54" 54"x72" - 70"x70" | Within Same Range (Similar*) | | Maximum Wrapped Package Content Weights | Appropriate wrapping described in IFU | 12.9 lbs | Similar* AAMI ST77 recommends maximum of 25 lbs, and both follow this. | | Device Materials | 99% polyester, 1% carbon | Polyester, cotton | Similar* | | Fabric | Woven | Woven | Same | | Cut-n-Sew | Yes | Yes | Same | | Performance Testing | Conducted | Conducted | Same See Table 3. | | Sterilization | Steam sterilization cycle – 4 minutes exposure at 132°C (~270°F), Dry time 30 minutes | Steam sterilization cycle – 4 minutes exposure at 270°F (~132°C), Dry time unknown | Similar* | | Sterility | Provided non-sterile | Provided non-sterile | Same | | Biocompatibility | Assessed per ISO 10993-1 and tested per the ISO 10993 Series | Assessed per ISO 10993-1 and tested per the ISO 10993 Series | Same | * Discussion on Similarity: In all cases of differences, one or more of the following are true: - The subject device was assessed in alignment with relevant standards and/or guidance documents and met expectations. K250321 Page 3 of 6 {7} - The subject device met the pre-determined acceptance criteria based on the expected use of the device. - The characteristic is not relevant to the performance, safety, or effectiveness of the device, but rather is provided for convenience (size, border colors). - The information for the predicate device is unknown based on publicly available information, but as the subject device meets expectations, no further comparison is needed. Therefore, all pre-determined acceptance criteria have been met, the device performs as intended, and any differences are not relevant to the safety, effectiveness, or performance of the subject device as compared to the predicate device. Table 2 provides a detailed comparison of the indications for use. The predicate has a significant amount of additional descriptive data. Table 2. Indications for Use Comparison | Proposed Indications (K250321) | | | | Predicate K172207 Cleared Indications | | --- | --- | --- | --- | --- | | GCI Sterilization Wrappers are intended to be used to enclose reusable textile-based medical devices, such as surgical gowns, towels, and drapes, for sterilization by a health care provider. They are designed to allow effective steam sterilization of the enclosed textile items and to maintain sterility until use, for up to 28 days. | | | | Standard Supreme Sterilization Wrappers are intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. Standard Supreme Sterilization Wrappers will function as a sterilization wrap when processed according to instructions. Standard Supreme Sterilization Wrappers are reusable through 75 wash, dry, and sterilization cycles. They are manufactured and distributed as non-sterile sterilization wraps that are intended to be sterilized and processed by health care facilities and/or contract sterilization / laundry companies. Pack Assembly: Pack assembly should follow established folding, assembly and wrapping procedures as defined by the clinical staff for aseptic presentation of the pack and its contents. The oblong and envelope folds have been validated for use on textile packs. Closure can include autoclave tape and elastomer closures. Sterilization: A prevacuum sterilization cycle (4-minutes exposure at 132°C, Dry Time: 30 minutes) has been validated. | | GCI Sterilization Wrappers sizes and load limitation are as follows: | | | | | | SKU / Model | Size | Border Color | Maximum Validated Load Weight | | | P81818 KBRO | 18"x18" | White | 1 lbs | | | P82424 KBRO | 24"x24" | Pink | 1.82 lbs | | | P83030 KBRO | 30"x30" | Yellow | 3.45 lbs | | | P83636 KBRO | 36"x36" | Blue | 5.8 lbs | | | P84545 KBRO | 45"x45" | Mint | 10.98 lbs | | | P85454 KBRO | 54"x54" | Red | 18.5 lbs | | | P86060 KBRO | 60"x60" | Green | 25 lbs | | | GCI Sterilization Wrappers are reusable through 38 wash, dry, and sterilization cycles. Pack assembly should follow established folding, assembly, and wrapping procedures as defined by the clinical staff for aseptic presentation of the pack and its contents. Closure can include autoclave tape and elastomer closures. Sterilization: A prevacuum steam sterilization cycle (4 minutes exposure at 132°C, Dry Time: 30 minutes) has been validated. | | | | | K250321 {8} K250321 Page 5 of 6 | Proposed Indications (K250321) | Predicate K172207 Cleared Indications | | | | | --- | --- | --- | --- | --- | | | configuration including 85.6 pounds of packs in a 9.1 cubic foot chamber – load density of 10.4 lbs./cu.ft. Table 1: Standard Supreme Sterilization Wrappers Color and Model #’s | | | | | | Sizes | Misty | Jade | Ceil | | | 12”x12” | 21520502 | - | - | | | 18”x18” | 21521502 | 21531502 | 21541502 | | | 23”x23” | 21522002 | 21532002 | 21542002 | | | 27”x27” | - | 21532302 | - | | | 30”x30” | 21522802 | 21532802 | 21542802 | | | 35”x35” | 21523002 | 21533002 | 21543002 | | | 45”x45” | 21525002 | 21535002 | 21545002 | | | 54”x54” | 21526002 | 21536002 | 21546002 | | | 54”x72” | 21528202 | 21538202 | 21548202 | | | 70”x70” | 21528802 | 21538802 | 21548802 | # 9. Non-Clinical Performance Data Performance testing was conducted in alignment with FDA guidance "Premarket Notification [510(k)] Submissions for Sterilizers Intended for Use in Health Care Facilities" and "Addendum to: Guidance on Premarket Notification [510(k)] Submissions for Sterilizers Intended for Use in Health Care Facilities." Testing was conducted to relevant standards including ASTM D5587, 16 CFR 1610, AATCC 127, AATCC 42, ISO 10993-5, ISO 10993-10, ISO 9073, ASTM D5034, AAMI TIR 12, and AAMI TIR 30. All testing results passed pre-determined acceptance criteria. In alignment with requirements, supporting the performance of the subject device. Table 3. Summary of Non-Clinical Testing (Averages) | Attribute | Acceptance Criteria | SUBJECT: K250321 GCI Surgical Wrappers | PREDICATE: K172207 Sterilization Wrappers | Comparison | | --- | --- | --- | --- | --- | | Tearing Strength (ASTM D5587) | >20 PSI Standard fabric direction | >21 PSI Standard fabric direction | Unknown | Similar* | | Breaking Strength (ASTM D5034) | >20 PSI | >163 PSI | >20 PSI | Similar* | | Flammability (16 CFR 1610) | Class 1 | Class 1 | Unknown | Similar* | | Water Resistance: Hydrostatic Pressure (AATCC 127) | Water Resistant, >20 cm H2O | Water Resistant, >27 cm H2O | Unknown | Similar* | | Water Resistance: Impact Penetration (AATCC 42) | < 4.5 g | < 0.3 g | Unknown | Similar* | | Linting | < 5 log IPM | < 5 log IPM | < 5 log IPM | Same | {9} | Attribute | Acceptance Criteria | SUBJECT: K250321 GCI Surgical Wrappers | PREDICATE: K172207 Sterilization Wrappers | Comparison | | --- | --- | --- | --- | --- | | Sterility Maintenance (ANSI/AAMI/ISO 17665-1) | Sterility maintained for labeled time | 28 Days | Unknown | Similar* | | Cytotoxicity (ISO 10993-5) | Non-Cytotoxic | Non-Cytotoxic | Non-Cytotoxic | Same | | Irritation (ISO 10993-10) | Non-Irritating | Non-Irritating | Unknown | Similar* | | Sensitization (ISO 10993-10) | Non-Sensitizing | Non-Sensitizing | Unknown | Similar* | | Use Life and Reprocessing Effectiveness (AAMI TIR12 and TIR30) | Reprocessing effectiveness confirmed for labeled use life | Use life 38 cycles Reprocessing effectiveness confirmed | Use life 75 cycles Reprocessing effectiveness unknown | Similar* | | Sterilization Validation (ANSI/AAMI/ISO 17665-1) | SAL of 10-6 shall be demonstrated. | pass | pass | Same | Note: PSI is equivalent to $\mathrm{lbf / in}^2$ ## 10. Clinical Performance Data N/A ## 11. Conclusion Statement The conclusion drawn from the testing demonstrates that the subject device is at least as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K172207, Class II (21 CFR 880.6850, Product Code FRG). K250321 Page 6 of 6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...