K172207 · Standard Textile Co., Inc. · FRG · Aug 25, 2017 · General Hospital
Device Facts
Record ID
K172207
Device Name
Standard Supreme Sterilization Wrapper
Applicant
Standard Textile Co., Inc.
Product Code
FRG · General Hospital
Decision Date
Aug 25, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Standard Supreme Sterilization Wrappers are intended to enclose another medical device that is to be sterilized by a health care provider. They are intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. Standard Supreme Sterilization Wrappers will function as a sterlization wrap when processed according to instructions. Standard Supreme Sterilization Wrappers are reusable through 75 wash, dry, and sterilization cycles. They are manufactured and distributed as non-sterily sterilization wraps that are intended to be sterlized and processed by heath care facilities and/or contract sterilization/laundry companies.
Device Story
Standard Supreme Sterilization Wrappers are woven fabric sheets used to wrap medical devices for steam sterilization. Healthcare staff wrap devices using established folding procedures (oblong or envelope folds) and secure them with autoclave tape or elastomer closures. The wrapped packs are processed in a prevacuum steam sterilizer (4 minutes at 270°F). The wrappers act as a sterile barrier, maintaining the sterility of the enclosed instruments until they are opened for clinical use. The device is reusable for up to 75 cycles of washing, drying, and sterilization. The primary benefit is providing a durable, reusable sterile barrier for surgical instruments in clinical settings.
Woven 55/45 cotton/polyester blend fabric. Vat dyed. Dimensions range from 12"x12" to 70"x70". Sterilization via prevacuum steam (270°F, 4 min). Complies with ASTM D-5034-2013 for strength and AATCC 61-1993 for colorfastness. Non-toxic per MEM Elution.
Indications for Use
Indicated for use by healthcare providers to enclose medical devices for steam sterilization and to maintain sterility until use. Reusable for up to 75 wash, dry, and sterilization cycles.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 25, 2017
Standard Textile Co., Inc. % Jonathan Kahan Partner Hogan Lovells US LLP 555 Thirteenth Street, NW Washington, District of Columbia 20004-1109
Re: K172207
Trade/Device Name: Standard Supreme Sterilization Wrapper Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: July 21, 2017 Received: July 21, 2017
Dear Jonathan Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Tara A. Ryan -S
Lori Wiggins, MPT, CLT Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K172207
#### Device Name
Standard Supreme Sterilization Wrappers
#### Indications for Use (Describe)
Standard Supreme Sterilization Wrappers are intended to enclose another medical device that is to be sterilized by a health care provider. They are intended to allow sterilization of the enclosed medical device and also to maintain steriity of the enclosed device until used.
Standard Supreme Sterilization Wrappers will function as a sterlization wrap when processed according to instructions. Standard Supreme Sterilization Wrappers are reusable through 75 wash, dry, and sterilization cycles. They are manufactured and distributed as non-sterily sterilization wraps that are intended to be sterlized and processed by heath care facilities and/or contract sterilization/laundry companies.
#### Pack Assembly:
Pack assembly should follow established folding, assembly and wrapping procedures as defined by the clinical staff for aseptic presentation of the pack and its contents. The oblorg and envelope folds have been validated for use on textile packs. Closure can include autoclave tape and elastomer closures.
#### Sterilization:
A prevacuum steam sterilization cycle (4 minute exposure at 270°F) has been validated for use on textile packs. The maximum pack size validated for use was 0.58 cubic feet with a pack weight of 12.9 pounds – pack density of 22.4 lbs./cu.ft. The maximum chamber loading configuration included 85.6 pounds of packs in a 9.1 cubic foot chamber – load density of 10.4 lbs./cu.ft.
| Sizes | Misty | Jade | Ceil |
|---------|----------|----------|----------|
| 12"x12" | 21520502 | -- | -- |
| 18"x18" | 21521502 | 21531502 | 21541502 |
| 23"x23" | 21522002 | 21532002 | 21542002 |
| 27"x27" | -- | 21532302 | -- |
| 30"x30" | 21522802 | 21532802 | 21542802 |
| 35"x35" | 21523002 | 21533002 | 21543002 |
| 45"x45" | 21525002 | 21535002 | 21545002 |
| 54"x54" | 21526002 | 21536002 | 21546002 |
| 54"x72" | 21528202 | 21538202 | 21548202 |
| 70"x70" | 21528802 | 21538802 | 21548802 |
Type of Use (Select one or both, as applicable)
□ Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
## 510K SUMMARY STANDARD SUPREME STERILIZATION WRAPPERS (K172207)
## Contact Information
| Standard Textile Co., Inc. | Contact Person: Brad Bushman |
|----------------------------|---------------------------------------------------|
| One Knollcrest Drive | Phone (513) 761-9255 Ext 2455; Fax (513) 679-8389 |
| Cincinnati, Ohio 45237 | Summary Prepared on 8/24/2017 |
## Device Information
| Device Name: | Standard Supreme Sterilization Wrappers |
|----------------------|----------------------------------------------|
| Class: | Class II |
| Common/Usual Name: | Non-sterile Sterilization Wrappers |
| Classification Name: | Sterilization Wrap |
| | Regulation 21 CFR 880.6850, Product Code FRG |
## Device Description
Standard Supreme Sterilization Wrappers are made from Standard Supreme fabric. This fabric uses intimately blended cotton and polyester yarns, is woven into fabric and then dyed. The fabric comes in three colors - misty, ceil and jade. The dyes used to achieve the three shades include Teresil Navy, Novasol Blue, VatGreen1, Vat Yellow1, and Vat Red31.
| Standard Supreme Wrapper Product Line Table | | |
|---------------------------------------------|---------------------------------|-------------|
| Item | Description | Size (inch) |
| 21520502 | Standard Supreme Wrapper, Misty | 12x12 |
| 21521502 | Standard Supreme Wrapper, Misty | 18x18 |
| 21522002 | Standard Supreme Wrapper, Misty | 23x23 |
| 21522802 | Standard Supreme Wrapper, Misty | 30x30 |
| 21523002 | Standard Supreme Wrapper, Misty | 35x35 |
| 21525002 | Standard Supreme Wrapper, Misty | 45x45 |
| 21526002 | Standard Supreme Wrapper, Misty | 54x54 |
| 21528202 | Standard Supreme Wrapper, Misty | 54x72 |
| 21528802 | Standard Supreme Wrapper, Misty | 70x70 |
| 21531502 | Standard Supreme Wrapper, Jade | 18x18 |
| 21532002 | Standard Supreme Wrapper, Jade | 23x23 |
| 21532302 | Standard Supreme Wrapper, Jade | 27x27 |
| 21532802 | Standard Supreme Wrapper, Jade | 30x30 |
| 21533002 | Standard Supreme Wrapper, Jade | 35x35 |
| 21535002 | Standard Supreme Wrapper, Jade | 45x45 |
| 21536002 | Standard Supreme Wrapper, Jade | 54x54 |
| 21538202 | Standard Supreme Wrapper, Jade | 54x72 |
| 21538802 | Standard Supreme Wrapper, Jade | 70x70 |
| 21541502 | Standard Supreme Wrapper, Ceil | 18x18 |
| 21542002 | Standard Supreme Wrapper, Ceil | 23x23 |
| 21542802 | Standard Supreme Wrapper, Ceil | 30x30 |
| 21543002 | Standard Supreme Wrapper, Ceil | 35x35 |
| 21545002 | Standard Supreme Wrapper, Ceil | 45x45 |
| 21546002 | Standard Supreme Wrapper, Ceil | 54x54 |
| 21548202 | Standard Supreme Wrapper, Ceil | 54x72 |
| 21548802 | Standard Supreme Wrapper, Ceil | 70x70 |
{4}------------------------------------------------
# Intended Use/Indications for Use
Standard Supreme Sterilization Wrappers are intended to enclose another medical device that is to be sterilized by a health care provider. They are intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. Standard Supreme Sterilization Wrappers will function wrap when processed according to instructions. Standard Supreme Sterilization Wrappers are reusable through 75 wash, dry, and sterilization cycles. They are manufactured and distributed as non-sterile sterilization wraps that are intended to be sterilized and processed by health care facilities and/or contract sterilization/laundry companies.
# Pack Assembly:
Pack assembly should follow established folding, assembly and wrapping procedures as defined by the clinical staff for aseptic presentation of the pack and its contents. The oblong and envelope folds have been validated for use on textile packs. Closure can include autoclave tape and elastomer closures.
# Sterilization:
A prevacuum steam sterilization cycle (4 minute exposure at 270°F) has been validated for use on textile packs. The maximum pack size validated for use was 0.58 cubic feet with a pack weight of 12.9 pounds – pack density of 22.4 lbs./cu.ft. The maximum chamber loading configuration included 85,6 pounds of packs in a 9,1 cubic foot chamber - load density of 10,4 lbs./cu.ft.
## Performance
Standard Supreme Sterilization Wrappers have successfully completed the non-clinical performance tests listed below.
- a. Whole Package Challenge and 30 Day Shelf Life Study Standard Supreme Sterilization Wrappers demonstrated their ability in these tests to be an effective sterile barrier after 75 cycles of soiling, washing, drying and sterilization when used as a sequential double wrap for textile packs.
- b. Strength through 75 wash, dry and sterilization cycles ASTM #D-5034-2013 >50 psi
- c. Linting < 5.0 log IPM. Resists the rigors of linting associated with flexing and abrasion.
- d. Cytotoxicity MEM Elution (MG023) Non-Toxic
- e. Sterilization Sterilization efficiency/penetration for the Standard Supreme Sterilization Wrappers was demonstrated with a half-cycle exposure condition showing no growth of the indicator organism in worst case challenge load configurations. A pre-vacuum steam sterilization cycle (4 minute exposure at 270°F) has been validated for use on textile packs. The maximum pack size validated for use was 0.58 cubic feet with a pack weight of 12.9 pounds - pack density of 22.4 Ibs./cu.ft. The maximum chamber loading configuration included 85.6 pounds of packs in a 9.1 cubic foot chamber - load density of 10.4 lbs./cu.ft.
- f. Use Life - 75 processing (wash, dry and sterilization).
- g. Colorfastness to Commercial Laundering AATCC #61-1993(4A) >3.5.
{5}------------------------------------------------
# Substantial Equivalence Comparison
Standard Supreme Sterilization Wrappers have been shown to be substantially equivalent to WrapPel "T" sterilization wraps K923408. All fabrics used are woven, dyed and treated with a fluoropolymer finish. Fabrics are spread, cut and sewn into the similar if not the exact wrapper size. All products can use color coded thread around the edges to provide easy size identification. Performance through 75 processing demonstrated that Standard Supreme Sterilization Wrappers and WrapPel "T" Sterilization Wrappers have the strength for reuse. Both subject and predicate wrappers can be steam sterilized.
In practice Standard Supreme Sterilization Wrappers differ from WrapPel "T" Sterilization Wrappers in that it is a 55/45 blended cotton/polyester fabric versus WrapPel "T" Sterilization Wrappers which are 100% polyester.
| Substantial Equivalence Table<br>Standard Supreme Sterilization Wrappers (K172207)<br>vs.<br>WrapPel "T" Sterilization Wrappers (K923408) | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|----------------------------------------|
| | Standard Supreme Sterilization<br>Wrappers | WrapPel "T" Sterilization<br>Wrappers |
| <b>Manufacturing</b> | | |
| Quality System | 21 CFR 820 | 21 CFR 820 |
| Fabric | Woven (1:1) | Woven (1:1) |
| Finishing | Vat Dyed | Disperse Dyed &<br>Flurocarbon Treated |
| Cut-n-Sew | Yes | Yes |
| Finished Dimensions | Similar/Same | Similar/Same |
| <b>Performance</b> – Non-clinical tests were used to evaluate the physical and performance characteristics<br>of Standard Supreme Sterilization Wrappers. Performance through 75 processing demonstrated that<br>Standard Supreme Sterilization Wrappers and WrapPel "T" Sterilization Wrappers have the strength<br>for reuse. Both wrappers can be steam sterilized, are colorfast, low lint and non-toxic. | | |
| Strength | >20 psi | >20 psi |
| Colorfastness | Yes | Yes |
| Linting | Completed | Completed |
| Toxicity | Non-Toxic | Non-Toxic |
| Use Life | 75 wash, dry, & sterilizations | 75 wash, dry, & sterilizations |
| <b>Sterilization - Pre-vacuum Steam Cycles</b> | | |
| Cycle - 4 minute<br>exposure at 270°F. | Yes | Yes |
| <b>Intended Use, i.e., traditional wrapping practices using oblong and envelope folds</b> | | |
| Fabric Packs | Yes | Yes |
{6}------------------------------------------------
# Substantial Equivalence Conclusion Statement
Based on intended use, bench testing and technological characteristics, Standard Supreme Sterilization Wrappers are substantially equivalent to the predicate device. The non-clinical tests demonstrate that Standard Supreme Sterilization Wrappers are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.