Alma FemiLift Pixel CO2 Laser System, Delivery Devices and Accessories
K250071 · Alma Lasers, Inc. · GEX · Sep 24, 2025 · General, Plastic Surgery
Device Facts
Record ID
K250071
Device Name
Alma FemiLift Pixel CO2 Laser System, Delivery Devices and Accessories
Applicant
Alma Lasers, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 24, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Alma FemiLift Pixel CO2 Laser System and Delivery Device Accessories are intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery), podiatry, gynecology, neurosurgery, orthopedics (soft tissue), arthroscopy (knee).
Device Story
The Alma FemiLift Pixel CO2 Laser System delivers light energy for soft tissue surgery. The system includes a console with an optical bench, microcontroller, power supply, cooling system, and air-purge system; a touchscreen control panel; an articulated arm; and various handpieces. The new FemiLift handpiece features a 90-degree mirror to deflect the laser beam, allowing the operator to hold the device parallel to the target tissue, unlike standard perpendicular handpieces. It utilizes a single-use sterile sleeve to provide a barrier between the handpiece and the patient. The system is operated by clinicians in clinical settings. By enabling precise ablation, vaporization, and coagulation, the device assists in procedures ranging from dermatological resurfacing to complex neurosurgical and gynecological resections, potentially improving patient outcomes through controlled tissue interaction.
Clinical Evidence
Bench testing only. Performance data included viral barrier testing of the FemiLift single-use sleeve, comparative power output testing against the predicate CO2 Focus 50mm handpiece, biocompatibility (ISO 10993-1), sterilization validation (ISO 11135), packaging validation (ASTM D4332, D4196), and electrical/EMC safety testing (IEC 60601 series, IEC 60825-1).
Technological Characteristics
CO2 laser system; articulated arm delivery. FemiLift handpiece includes 90-degree mirror and single-use sterile sleeve. Materials biocompatible per ISO 10993-1. Sterilization via ISO 11135. Electrical safety per IEC 60601-1, 60601-1-2, 60601-2-22; laser safety per IEC 60825-1. Software validation per IEC 62304.
Indications for Use
Indicated for surgical ablation, vaporization, excision, incision, and coagulation of soft tissue across dermatology, plastic surgery, podiatry, otolaryngology, gynecology, neurosurgery, orthopedics, general/thoracic surgery, dental/oral surgery, and genitourinary specialties. Specific applications include skin resurfacing, lesion removal, tumor resection, and various endoscopic/laparoscopic procedures. Contraindications not explicitly listed.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Alma Lasers Pixel CO2 Laser System, Delivery Device and Accessories (K103501)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
September 24, 2025
Alma Lasers, Inc
Jessica Rivera-Montejo
Director - Regulatory Affairs and Quality Assurance
485 Half Day Rd
Suite100
Buffalo Grove, Illinois 60089
Re: K250071
Trade/Device Name: Alma FemiLift Pixel CO2 Laser System, Delivery Devices and Accessories
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: August 27, 2025
Received: August 27, 2025
Dear Jessica Rivera-Montejo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}
K250071 - Jessica Rivera-Montejo
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K250071 - Jessica Rivera-Montejo
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.09.24
18:15:04 -04'00"
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K250071
Device Name
Alma FemiLift Pixel CO2 Laser System, Delivery Devices and Accessories
Indications for Use (Describe)
The Alma FemiLift Pixel CO2 Laser System and Delivery Device Accessories are intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery), podiatry, gynecology, neurosurgery, orthopedics (soft tissue), arthroscopy (knee). The Alma FemiLift Pixel CO2 Laser System is cleared for use for the particular indications as follows:
Dermatology & Plastic Surgery
The ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:
- Laser skin resurfacing
- Laser dermabrasion
- Laser burn debridement.
Laser skin resurfacing (ablation and/or vaporization) for the treatment of:
- Wrinkles, rhytids, and furrows (including fines lines and texture irregularities)
Clinical literature demonstrates that skin resurfacing of wrinkles, rhytids, and furrows with CO2 laser increases the amount of sub-epidermal collagen.
Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of:
- Keratoses, including actinic and seborrheic keratosis, seborrheic vulgaris, seborrheic wart, and verruca seborrhoica;
- Vermilionectomy of the lip;
- Cutaneous horns;
- Solar/actinic elastosis;
- Cheilitis, including actinic cheilitis;
- Lentigines, including lentigo maligna or Hutchinson’s malignant freckle;
- Uneven pigmentation/ dyschromia;
- Acne scars;
- Surgical scars;
- Keloids including acne keloidalis nuchae;
- Hemangiomas (including Buccal, port wine and pyogenic granulomas/granuloma pyogenicum/granuloma telagiectaticum);
- Tattoos;
- Telangiectasia;
- Removal of small skin tumors, including periungual (Koenen) and subungual fibromas;
- Superficial pigmented lesions;
- Aden sebaceous hypertrophy or sebaceous hyperplasia;
- Rhinophyma reduction;
- Cutaneous papilloma (skin tags);
- Milia;
- Debridement of eczematous or infected skin;
- Basal and squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis (BP) lesions;
FORM FDA 3881 (6/20)
Page 1 of 5
PSC Publishing Services (301) 443-6740
{4}
- Nevi, including spider, epidermal and protruding;
- Neurofibromas;
- Laser de-epithelialization;
- Trichoepitheliomas;
- Xanthelasma Palpebrarum;
- Syringoma
Laser ablation, vaporization and /or excision for complete or partial nail matrixectomy.
Vaporization/coagulation of:
- Benign/malignant vascular/avascular skin lesions;
- Moh's Surgery;
- Lipectomy;
- Verrucae and seborrheic vulgaris, including paronychial, perifungal, and subungual warts.
Laser incision and /or excision of soft tissue for the performance of upper and lower eyelid blepharoplasty.
Laser incision and /or excision of soft tissue for the creation of recipient sites for hair transplantation.
Laser incision and /or excision of soft tissue for the performance of upper and lower eyelid blepharoplasty.
Laser incision and /or excision of soft tissue for the creation of recipient sites for hair transplantation.
## Podiatry
Laser ablation, vaporization and /or excision of soft tissue for the reduction, removal, and/or treatment of:
- Verrucae vulgaris/plantar (warts), including paronychial, perifungal, and subungual warts;
- Fungal nail treatment;
- Porokeratoma ablation;
- Ingrown nail treatment;
- Neuromas/fibromas, including Morton's neuroma;
- Debridement of ulcers;
- Other soft tissue lesions.
Laser ablation, vaporization and /or excision in podiatry for complete or partial matrixectomy.
## Otolaryngology (ENT)
Laser incision, excision, ablation and/or vaporization of soft tissue in otolaryngology the treatment of:
- Choanal atresia,
- Leukoplakia, including oral, larynx, uvula, palatal, upper lateral pharyngeal tissue;
- Nasal obstruction;
- Adult and juvenile papillomatosis polyps;
- Polypectomy of nose and nasal passages;
- Lymphangioma removal;
- Removal of vocal cord/fold nodules, polyps and cysts;
- Removal of recurrent papillomas in the oral cavity, nasal cavity, larynx, pharynx and trachea, including the uvula, palatal, upper lateral pharyngeal tissue, tongue and vocal cords;
- Laser/tumor surgery in the larynx, pharynx, nasal, ear and oral structure and tissue
- Zenker's Diverticulum/pharyngoesophageal diverticulum [endoscopic laser-assisted esophageal diverticulectomy (ELAED)];
- Stenosis, including subglottic stenosis; tonsillectomy (including tonsillar cryptolysis, neoplasma) and tonsil ablation/tonsillotomy;
- Pulmonary bronchial and tracheal lesion removal;
- Benign and malignant nodules, tumors and fibromas (larynx, pharynx, trachea, tracheobronchial/endobronchial);
- Benign and malignant lesions and fibromas (nose and nasal passages);
- Benign and malignant tumors and fibromas; (oral);
- Stapedotomy/stapedectomy;
- Acoustic neuroma in the ear;
- Superficial lesions of the ear, including chondrodermatitis nondularis chronica helices/Winkler's disease;
FORM FDA 3881 (6/20)
{5}
- Telangiectasia/hemangioma of larynx, pharynx and trachea (includes uvula, palatal or upper lateral pharyngeal tissue);
- Cordectomy, cordotomy (for the treatment of vocal fold paralysis/vocal fold motion impairment), and cordal lesions of larynx, pharynx and trachea;
- Myringotomy/tympanostomy (tympanic membrane fenestration);
- Uvulopalatoplasty (LAUP, laser UPPP);
- Turbinectomy and turbinate reduction/ablation);
- Septal spur ablation/reduction and septoplasty;
- Partial glossectomy;
- Tumor resection on oral, subfascial and neck tissues;
- Rhinophyma;
- Verrucae vulgaris (warts);
- Gingivoplasty/gingivectomy.
## Gynecology (GYN)
Laser incision, excision, ablation and/or vaporization and of soft tissue in gynecology (GYN) for the treatment of:
- Conization of the cervix, including cervical intraepithelial neoplasia (CIN), vulvar and vaginal intraepithelial neoplasia (VIN, VAIN);
- Condyloma acuminate, including cervical, genital, vulvar, perineal, and Bowen’s disease, (Erythroplasia of Queyrat) and Bowenoid papulosis (BP) lesions;
- Leukoplakia (vulvar dystrophies);
- Incision and drainage (I&D) of Bartholin’s and Nabothian cysts;
- Herpes vaporization;
- Urethral caruncle vaporization;
- Cervical dysplasia;
- Benign and malignant tumors;
- Hemangiomas.
## GYN Laparoscopy
Vaporization, incision, excision, ablation, or photocoagulation of soft tissue in endoscopic and laparoscopic surgery, including GYN laparoscopy, for treatment of:
- Endometrial lesions, including ablation of endometriosis;
- Excision/lysis of adhesions;
- Salpingostomy
- Oophorectomy/ovariectomy;
- Fimbrioplasty;
- Metroplasty;
- Microsurgery (tubal);
- Uterine myomas and fibroids;
- Ovarian fibromas and follicle cysts;
- Uterosacral ligament ablation;
- Hysterectomy.
## Neurosurgery
Laser incision, excision, ablation and/or vaporization and of soft tissue in neurosurgery for the treatment of:
## Cranial
- Posterior fossa tumors;
- Peripheral neurectomy;
- Benign and malignant tumors and cysts (e.g. gliomas, meningiomas (including basal tumors), acoustic neuromas, lipomas and large tumors);
- Arteriovenous malformation;
- Pituitary gland tumors (transsphenoidal approach).
## Spinal Cord
- Incision/excision and vaporization of benign and malignant tumors and cysts;
- Intra- and extradural lesions;
FORM FDA 3881 (6/20)
{6}
- Laminectomy/ laminotomy/ microdiscectomy.
## Orthopedics
Incision/excision and vaporization and of soft tissue in orthopedic surgery. Applications include:
### Arthroscopy
- Meniscectomy;
- Chondromalacia;
- Chondroplasty;
- Ligament release (lateral and other);
- Excision of plica;
- Partial synovectomy.
### General
- Debridement of traumatic wounds;
- Debridement of decubitus and diabetic ulcers;
- Microsurgery;
- Artificial joint revision;
- PMMA removal.
## General/Thoracic Surgery
Incision, excision and vaporization and of soft tissue in general and thoracic surgery including endoscopic and open procedures. Applications include:
- Debridement of decubitus ulcers, stasis, diabetic and other ulcers;
- Mastectomy;
- Debridement of burns;
- Rectal and anal hemorrhoidectomy;
- Breast biopsy;
- Reduction mammoplasty;
- Cytoreduction for metastatic disease;
- Laparotomy and laparoscopic applications;
- Mediastinal and thoracic lesions and abnormalities;
- Skin tag vaporization;
- Atheroma;
- Cysts, including sebaceous cysts, pilar cysts, and mucous cysts of the lips;
- Pilonidal cyst removal and repair;
- Abscesses;
- Other soft tissue applications.
## Dental/Oral Surgery
Incision/excision and vaporization of soft tissue in dentistry and oral surgery. Applications include:
- Gingivectomy- removal of hyperplasia;
- Gingivoplasty;
- Incisional and excisional biopsy;
- Treatment of ulcerous lesions, including aphthous ulcers;
- Incision of infection when used with antibiotic therapy;
- Frenectomy (frenum release);
- Excision and ablation of benign and malignant lesions;
- Homeostasis;
- Operculectomy;
- Crown lengthening;
- Removal of soft tissue, cysts and tumors;
- Oral cavity tumors and hemangiomas;
- Abscesses;
- Extraction site hemostasis;
- Salivary gland pathologies;
FORM FDA 3881 (6/20)
{7}
- Pre-prosthetic gum preparation;
- Leukoplakia;
- Partial glossectomy;
- Periodontal gum resection.
## Genitourinary
Incision/excision and vaporization and of soft tissue in genitourinary procedures. Applications include:
- Benign and malignant lesions of external genitalia;
- Condyloma;
- Phimosis
- Erythroplasia
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 5 of 5
{8}
510(k) SUMMARY
Alma Lasers Inc.'s Alma FemiLift Pixel CO₂ System
Submitter's Name, Address, Telephone Number, and Contact Person
Alma Lasers Inc.
485 Half Day Rd Ste 100
Buffalo Grove, IL 60089
Contact Person: Jessica Rivera-Montejo, Director Regulatory and Quality
Email: jrivera-montejo@almalasers.com
Date Prepared: September 16, 2025
Name of Device
Trade Name: Alma FemiLift Pixel CO₂ System, Delivery Devices and Accessories
510(k) #: K250071
Classification Name: Laser Instrument, Surgical, Powered Classification
Regulation: 21 CFR 878.4810
Product Code: GEX
Predicate Device
Alma Lasers Pixel CO₂ Laser System, Delivery Device and Accessories (K103501)
Device Description
The Alma FemiLift Pixel CO₂ System is a laser system designed to deliver light energy for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissues in various medical specialties. The system consists of the following major components:
- Laser system console (containing the optical bench assembly and laser, the microcontroller control electronics and system software, the high voltage power supply, the laser cooling system, the compressed air-purge system, and the service panel)
- Control panel with touch-screen technology
- Articulated arm
- Footswitch
- Delivery device handpieces, including:
- Focusing handpieces (FemiLift-Subject Device, CO₂ Focus 50mm, and CO₂ Focus 100mm)
- Pixel handpieces (CO₂ Pixel 7x7 and CO₂ Pixel 9x9),
- Scanning handpiece
{9}
The FemiLift Handpiece is a 50mm focusing handpiece that has been added to the existing suite of handpieces previously cleared under K103501. The FemiLift Handpiece is comprised of a reusable handpiece body and a sterile sleeve ("FemiLift Single-Use Sleeve"). The handpiece body connects to the laser system and directs the laser beam for treatment, using a mirror to deflect the laser beam by 90°. The FemiLift Single-Use Sleeve provides a sterile barrier between the handpiece body and the patient.
The Alma FemiLift Pixel CO₂ System can be used with the CO₂ Focus 50mm and 100mm Handpieces, CO₂ Pixel 7x7 and 9x9 Handpieces, and CO₂ Scanner Handpiece for all of the specific uses indicated for the system, and with the FemiLift Handpiece for the gynecology and aesthetic surgery (dermatology and plastic surgery) indications.
## Intended Use / Indications for Use
The Alma FemiLift Pixel CO₂ System is intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery), podiatry, gynecology, neurosurgery, orthopedics (soft tissue), arthroscopy (knee).
There is no change in indications for use compared to the predicate device, Alma Lasers Pixel CO₂ Laser System (K103501). Therefore, the subject device has the same intended use as the predicate device.
The complete Indications for Use can be found in Appendix A to this 510(k) Summary.
## Comparison of Technological Characteristics
The Alma FemiLift Pixel CO₂ System has the same basic design, materials, operating principle as the predicate device (Alma Lasers Pixel CO₂ Laser System). The primary difference between the devices is the addition of the FemiLift Handpiece, which is a modified version of the CO₂ Focus 50mm handpiece originally cleared with the predicate device. The FemiLift Handpiece includes all of the same components as the CO₂ Focus 50mm Handpiece, except the FemiLift model includes a mirror and single-use sterile sleeve and does not include a standoff or plume extractor nozzle, which are unnecessary for use of the FemiLift Handpiece.
The principles of operation for the FemiLift Handpiece and the CO₂ Focus 50mm Handpiece are nearly identical, with the only difference being the angle at which the operator positions the handpiece relative to the target tissue. Due to the deflection of the laser beam by 90° from the FemiLift model's mirror, the operator holds the FemiLift Handpiece parallel to the target tissue, whereas the CO₂ Handpiece (which does not include a mirror) is held perpendicular to the target tissue. However, these different positions do not change the angle at which the laser beam treatment reaches the patient, which is always perpendicular to the target tissue.
## Performance Data
Performance testing was performed for the new FemiLift handpiece, which was the primary difference between the subject device and the predicate device. The company assessed the performance of the FemiLift Handpiece through viral barrier testing of the FemiLift Single-Use Sleeve and through comparative testing between the handpiece power output produced by
{10}
the FemiLift Handpiece versus the CO₂ Focus 50mm Handpiece. The viral barrier testing demonstrated that the FemiLift Single-Use Sleeve is impermeable to penetration by microorganisms. The comparative testing demonstrated that the handpiece power output is substantially equivalent between the FemiLift Handpiece and the CO₂ Focus 50mm Handpiece.
The following additional testing was performed to support substantial equivalence:
- Biocompatibility testing per ISO 10993-1;
- Sterilization validation testing per ISO 11135;
- Packaging validation testing per ASTM D4332 and ASTM D4196;
- Electrical safety and EMC testing per IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-22; IEC 60825-1; IEC 62304
- Software validation testing.
In all instances, the FemiLift Handpiece functioned as intended and the results met the predefined acceptance criteria.
## Conclusion
The Alma FemiLift Pixel CO₂ System has the same intended use and indications for use as the predicate device (Alma Lasers Pixel CO₂ Laser System, K103501).
The primary technological difference is the addition of the FemiLift Handpiece, which has similar technological characteristics and principles of operation as the predicate device's CO₂ Focus 50mm Handpiece. Although the FemiLift Handpiece includes a mirror and sterile sleeve whereas the predicate CO₂ Focus 50mm Handpiece does not, these differences do not raise different questions of safety or effectiveness because they do not change the overall intended therapeutic use of the laser system, or affect the fundamental scientific technology used to achieve the device's intended use. Furthermore, performance testing of the technological differences showed that the new FemiLift Handpiece is as safe and effective as the predicate device's CO₂ Focus 50mm Handpiece.
Therefore, the Alma FemiLift Pixel CO₂ System is substantially equivalent to the predicate device (K103501).
{11}
# Appendix A: Complete Indications for Use Statement
The Alma FemiLift Pixel CO₂ System is intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: gynecology and aesthetic surgery (dermatology and plastic surgery), podiatry, gynecology, neurosurgery, orthopedics (soft tissue), arthroscopy (knee).
The Alma FemiLift Pixel CO₂ System is cleared for use for the particular indications as follows:
## Dermatology & Plastic Surgery
The ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:
- Laser skin resurfacing
- Laser dermabrasion
- Laser burn debridement
Laser skin resurfacing (ablation and/or vaporization) for the treatment of:
- Wrinkles, rhytids, and furrows (including fines lines and texture irregularities)
Clinical literature demonstrates that skin resurfacing of wrinkles, rhytids, and furrows with CO₂ laser increases the amount of sub-epidermal collagen.
Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of:
- Keratoses, including actinic and seborrheic keratosis, seborrheic vulgaris, seborrheic wart, and verruca seborrhoica
- Vermilionectomy of the lip
- Cutaneous horns
- Solar/actinic elastosis
- Cheilitis, including actinic cheilitis
- Lentigines, including lentigo maligna or Hutchinson's malignant freckle
- Uneven pigmentation/ dyschromia
- Acne scars
- Surgical scars
- Keloids including acne keloidalis nuchae
- Hemangiomas (including Buccal, port wine and pyogenic granulomas/granuloma pyogenicum/granuloma telagiectaticum)
- Tattoos
- Telangiectasia
- Removal of small skin tumors, including periungual (Koenen) and subungual fibromas
- Superficial pigmented lesions
- Aden sebaceous hypertrophy or sebaceous hyperplasia
- Rhinophyma reduction
- Cutaneous papilloma (skin tags)
- Milia
- Debridement of eczematous or infected skin
- Basal and squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis (BP) lesions
- Nevi, including spider, epidermal and protruding
{12}
- Neurofibromas
- Laser de-epithelialization
- Trichoepitheliomas
- Xanthelasma Palpebrarum
- Syringoma
Laser ablation, vaporization and /or excision for complete or partial nail matrixectomy.
Vaporization/coagulation of:
- Benign/malignant vascular/avascular skin lesions
- Moh's Surgery
- Lipectomy
- Verrucae and seborrheic vulgaris, including paronychial, perifungal, and subungual warts
Laser incision and /or excision of soft tissue for the performance of upper and lower eyelid blepharoplasty.
Laser incision and /or excision of soft tissue for the creation of recipient sites for hair transplantation.
## Podiatry
Laser ablation, vaporization and /or excision of soft tissue for the reduction, removal, and/or treatment of:
- Verrucae vulgaris/plantar (warts), including paronychial, perifungal, and subungual warts
- Fungal nail treatment
- Porokeratoma ablation
- Ingrown nail treatment
- Neuromas/fibromas, including Morton's neuroma
- Debridement of ulcers
- Other soft tissue lesions
Laser ablation, vaporization and /or excision in podiatry for complete or partial matrixectomy.
## Otolaryngology (ENT)
Laser incision, excision, ablation and/or vaporization of soft tissue in otolaryngology the treatment of:
- Choanal atresia
- Leukoplakia, including oral, larynx, uvula, palatal, upper lateral pharyngeal tissue
- Nasal obstruction
- Adult and juvenile papillomatosis polyps
- Polypectomy of nose and nasal passages
- Lymphangioma removal
- Removal of vocal cord/fold nodules, polyps and cysts
- Removal of recurrent papillomas in the oral cavity, nasal cavity, larynx, pharynx and trachea including the uvula, palatal, upper lateral pharyngeal tissue, tongue and vocal cords
- Laser/tumor surgery in the larynx, pharynx, nasal, ear and oral structures and tissue
- Zenker's Diverticulum/pharyngoesophageal diverticulum [endoscopic laser-assisted esophageal diverticulectomy (ELAED)]
- Stenosis, including subglottic stenosis
{13}
- Tonsillectomy (including tonsillar cryptolysis, neoplasma) and tonsil ablation/tonsillotomy
- Pulmonary bronchial and tracheal lesion removal
- Benign and malignant nodules, tumors and fibromas (larynx, pharynx, trachea, tracheobronchial/endobronchial)
- Benign and malignant lesions and fibromas (nose and nasal passages)
- Benign and malignant tumors and fibromas (oral)
- Stapedotomy/stapedectomy
- Acoustic neuroma in the ear
- Superficial lesions of the ear, including chondrodermatitis nondularis chronica helices/Winkler's disease
- Telangiectasia/hemangioma of larynx, pharynx and trachea (includes uvula, palatal or upper lateral pharyngeal tissue)
- Cordectomy, cordotomy (for the treatment of vocal fold paralysis/vocal fold motion impairment) and cordal lesions of larynx, pharynx and trachea
- Myringotomy/tympanostomy (tympanic membrane fenestration)
- Uvulopalatoplasty (LAUP, laser UPPP)
- Turbinectomy and turbinate reduction/ablation
- Septal spur ablation/reduction and septoplasty
- Partial glossectomy
- Tumor resection on oral, subfascial and neck tissues
- Rhinophyma
- Verrucae vulgaris (warts)
- Gingivoplasty/gingivectomy
## Gynecology
Laser incision, excision, ablation and/or vaporization and of soft tissue in gynecology (GYN) for the treatment of:
- Conization of the cervix, including cervical intraepithelial neoplasia (CIN), vulvar and vaginal intraepithelial neoplasia (VIN, VAIN)
- Condyloma acuminate, including cervical, genital, vulvar, perineal, and Bowen's disease, (Erythroplasia of Queyrat) and Bowenoid papulosis (BP) lesions
- Leukoplakia (vulvar dystrophies)
- Incision and drainage (I&D) of Bartholin's and Nabothian cysts
- Herpes vaporization
- Urethral caruncle vaporization
- Cervical dysplasia
- Benign and malignant tumors
- Hemangiomas
## GYN Laparoscopy
Vaporization, incision, excision, ablation, or photocoagulation of soft tissue in endoscopic and laparoscopic surgery, including GYN laparoscopy, for treatment of:
- Endometrial lesions, including ablation of endometriosis
- Excision/lysis of adhesions
- Salpingostomy
- Oophorectomy/ovariectomy
- Fimbrioplasty
{14}
- Metroplasty
- Microsurgery (tubal)
- Uterine myomas and fibroids
- Ovarian fibromas and follicle cysts
- Uterosacral ligament ablation
- Hysterectomy.
## Neurosurgery
Laser incision, excision, ablation and/or vaporization and of soft tissue in neurosurgery for the treatment of:
## Cranial
- Posterior fossa tumors
- Peripheral neurectomy
- Benign and malignant tumors and cysts (e.g. gliomas, meningiomas (including basal tumors), acoustic neuromas, lipomas and large tumors)
- Arteriovenous malformation
- Pituitary gland tumors (transsphenoidal approach)
## Spinal Cord
- Incision/excision and vaporization of benign and malignant tumors and cysts
- Intra- and extradural lesions
- Laminectomy/ laminotomy/ microdiscectomy
## Orthopedics
Incision/excision and vaporization of soft tissue in orthopedic surgery. Applications include:
### Arthroscopy
- Meniscectomy
- Chondromalacia
- Chondroplasty
- Ligament release (lateral and other)
- Excision of plica
- Partial synovectomy
### General
- Debridement of traumatic wounds
- Debridement of decubitus and diabetic ulcers
- Microsurgery
- Artificial joint revision
- PMMA removal
## General/Thoracic Surgery
Incision, excision and vaporization and of soft tissue in general and thoracic surgery including endoscopic and open procedures. Applications include:
- Debridement of decubitus ulcers, stasis, diabetic and other ulcers
- Mastectomy
{15}
- Debridement of burns
- Rectal and anal hemorrhoidectomy
- Breast biopsy
- Reduction mammoplasty
- Cytoreduction for metastatic disease
- Laparotomy and laparoscopic applications
- Mediastinal and thoracic lesions and abnormalities
- Skin tag vaporization
- Atheroma
- Cysts, including sebaceous cysts, pilar cysts, and mucous cysts of the lips
- Pilonidal cyst removal and repair
- Abscesses
- Other soft tissue applications
## Dental/Oral Surgery
Incision/excision and vaporization of soft tissue in dentistry and oral surgery. Applications include:
- Gingivectomy – removal of hyperplasia
- Gingivoplasty
- Incisional and excisional biopsy
- Treatment of ulcerous lesions, including aphthous ulcers
- Incision of infection when used with antibiotic therapy
- Frenectomy (Frenum release)
- Excision and ablation of benign and malignant lesions
- Homeostasis
- Operculectomy
- Crown lengthening
- Removal of soft tissue, cysts, and tumors
- Oral cavity tumors and hemangiomas
- Abscesses
- Extractions site hemostasis
- Salivary gland pathologies
- Pre-prosthetic gum preparation
- Leukoplakia
- Partial glossectomy
- Periodontal gum resection
## Genitourinary
Incision/excision and vaporization of soft tissue in genitourinary procedures. Applications include:
- Benign and malignant lesions of external genitalia
- Condyloma
- Phimosis
- Erythroplasia
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.