ALMA LASERS PIXEL CO2 LASER SYSTEM, DELIVERY DEVICE AND ACCESSORIES
K103501 · Alma Lasers, Inc. · GEX · Jan 14, 2011 · General, Plastic Surgery
Device Facts
Record ID
K103501
Device Name
ALMA LASERS PIXEL CO2 LASER SYSTEM, DELIVERY DEVICE AND ACCESSORIES
Applicant
Alma Lasers, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 14, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K103501 · Jan 14, 2011
ALMA LASERS PIXEL CO2 LASER SYSTEM, DELIVERY DEVICE AND ACCESSORIES
Alma Lasers, Inc.
Published clinical literature
Clinical literature is cited to support the physiological mechanism (increase in sub-epidermal collagen) and clinical utility of CO2 laser skin resurfacing for the treatment of wrinkles, rhytids, and furrows.
CO2 laser; skin resurfacing; collagen; clinical literature
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Clinical literature on CO2 laser skin resurfacing; Literature review / Clinical observation
Patients undergoing CO2 laser skin resurfacing
Not applicable for this study
Increase in sub-epidermal collagen
Indications for Use
The Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories are intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery), podiatry, gynecology, neurosurgery, orthopedics (soft tissue), arthroscopy (knee).
Device Story
CO2 laser system delivers light energy for soft tissue surgery; includes console with optical bench, microcontroller, cooling system, LCD touch-screen, 7-joint articulated arm, footswitch, and various handpieces (focusing and Pixel). Operated by physicians in clinical settings (OR, clinic). Laser energy ablates, vaporizes, excises, incises, or coagulates soft tissue. Output controlled by clinician via interface; provides precise tissue interaction for diverse surgical procedures. Benefits include controlled tissue removal and hemostasis across multiple medical specialties.
Clinical Evidence
No clinical data presented. Substantial equivalence is based on technical characteristics and intended use comparisons to the predicate device.
Technological Characteristics
CO2 laser system; 7-joint articulated arm delivery; LCD touch-screen interface; microcontroller-based control electronics; laser cooling system; compressed air-purge system. Components include optical bench, high voltage power supply, and various handpieces (focusing and Pixel).
Indications for Use
Indicated for soft tissue ablation, vaporization, excision, incision, and coagulation across multiple specialties: dermatology/plastic surgery (resurfacing, lesion removal, blepharoplasty, hair transplant sites), podiatry (warts, nail matrixectomy, ulcers), ENT (lesions, tonsillectomy, airway surgery), gynecology (cervical/vulvar lesions, laparoscopy, hysterectomy), neurosurgery (cranial/spinal tumors), orthopedics (arthroscopy, wound debridement), general/thoracic surgery (mastectomy, ulcers, cysts), dental/oral surgery (gingivectomy, biopsies, lesions), and genitourinary procedures. Prescription use only.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories (K080463)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like shapes extending upwards. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird-like figure. The text is in all capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Alma Lasers, Inc. % Ms. Tatiana Epstein Vice President, QA and RA 485 Half Day Road, Suite 100 Buffalo Grove, Illinois 60089
Re: K103501
Trade/Device Name: Alma Lasers Pixel CO2" Laser System, Delivery Devices and Accessories Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic laser Regulatory Class: Class II Product Code: HQF Dated: January 04, 2011 Received: January 07, 2011
Dear Ms. Epstein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
JAN 1 4 2011
{1}------------------------------------------------
Page 2 - Ms. Tatiana Epstein
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Aty B. 12 h
fr.
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
2
Enclosure
{2}------------------------------------------------
K10 3 50 | 510(k) Number (if known):
Alma Lasers Pixel CO2TM Laser System, Delivery Devices and Accessories Device Name:
Indications for Use:
The Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories are intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery), podiatry, gynecology, neurosurgery, orthopedics (soft tissue), arthroscopy (knee).
The Alma Lasers Pixel CO2™ Laser System is cleared for the particular indications as follows:
## Dermatology & Plastic Surgery
The ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:
- . laser skin resurfacing
- laser derm-abrasion .
- . laser burn debridement.
Laser skin resurfacing (ablation and/or vaporization) for the treatment of:
- wrinkles, rhytids, and furrows (including fines lines and texture irregularities) .
Clinical literature demonstrates that skin resurfacing of wrinkles, rhytids, and furrows with CO2 laser increases the amount of sub-epidermal collagen.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ilapdsall for min
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103501
3
{3}------------------------------------------------
510(k) Number (if known): K10 350 |
Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories Device Name:
Indications for Use - Continued from the previous page:
#### Dermatology & Plastic Surgery, continued
Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of:
- keratoses, including actinic and seborrheic keratosis, seborrhoecae vulgares, seborrheic . wart, and verruca seborrheica;
- . vermillionectomy of the lip;
- cutaneous horns; .
- . solar/actinic elastosis;
- chelitis, including actinic chelitis; .
- lentigines, including lentigo maligna or Hutchinson's malignant freckle; .
- uneven pigmentation/ dyschromia;
- acne scars; .
- surgical scars; .
- keloids including acne keloidalis nuchae; .
- hemangiomas (including Buccal, port wine and pyogenic granulomas/granuloma . pyogenicum/granuloma telagiectaticum);
- tattoos: .
- . telangiectasia:
- . removal of small skin tumors, including periungual (Koenen) and subungual fibromas:
- . superficial pigmented lesions;
- . adenosebaceous hypertrophy or sebaceous hyperplasia;
- . rhinophyma reduction: .
- . cutaneous papilloma (skin tags);
- milia; .
- debridement of eczematous or infected skin: .
- basal and squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis (BP) lesions;
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nil RP. Sylvan fix
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number: K103501
Page 2 of 10
\$\gamma\$
{4}------------------------------------------------
510(k) Number (if known): K10 350 /
Device Name: Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories
Indications for Use - Continued from the previous page:
#### Dermatology & Plastic Surgery, continued
- nevi, including spider, epidermal and protruding; .
- . neurofibromas;
- laser de-epithelialization; .
- tricoepitheliomas; .
- xanthelasma palpebrarum; .
- syringoma .
Laser ablation, vaporization and /or excision for complete or partial nail matrixectorny.
Vaporization/coagulation of:
- benign/malignant vascular/avascular skin lesions;
- . Moh's Surgery;
- . Lipectomy;
- Verrucae and seborrhoecae vulgares, including paronychial, periungal, and subungaual ● warts.
Laser incision and /or excision of soft tissue for the performance of upper and lower eyelid blepharoplasty.
Laser incision and /or excision of soft tissue for the creation of recipient sites for hair transplantation.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nat Al. Dail
Division Sign 06
fix. mxa
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K10.35
Page 3 of 10
ડ
{5}------------------------------------------------
510(k) Number (if known): K10 3501
Alma Lasers Pixel CO2TM Laser System, Delivery Devices and Accessories Device Name:
Indications for Use - Continued from the previous page:
### Podiatry
Laser ablation, vaporization and for excision of soft tissue for the reduction, removal, and/or treatment of:
- verrucae vulgares/plantar (warts), including paronychial, periungal, and subungual warts;
- fungal nail treatment; .
- porokeratoma ablation: �
- ingrown nail treatment; .
- . neuromas/fibromas, including Morton's neuroma;
- debridement of ulcers; .
- . other soft tissue lesions.
Laser ablation, vaporization and /or excision in podiatry for complete or partial matrixectomy.
# Otolaryngology (ENT)
Laser incision, excision, ablation and/or vaporization of soft tissue in otolaryngology the treatment of:
- choanal atresia,
- leukoplakia, including oral, larynx, uvula, palatal, upper lateral pharyngeal tissue;-.
- . nasal obstruction;
- . adult and juvenile papillomatosis polyps;
- . polypectomy of nose and nasal passages;
- . lymphangioma removal:
- removal of vocal cord/fold nodules, polyps and cysts: .
- . removal of recurrent papillomas in the oral cavity, nasal cavity, larynx, pharynx and trachea, including the uvula, palatal, upper lateral pharyngeal tissue, tongue and vocal cords:
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
N.L.B. de A.L. S.
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number
Page 4 of 10
6
{6}------------------------------------------------
510(k) Number (if known): K10 350)
Device Name: Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories
Indications for Use - Continued from the previous page:
### Otolaryngology (ENT), continued
- lascr/tumor surgery in the larynx, pharynx, nasal, ear and oral structures and tissue; .
- Zenker's Diverticulum/pharyngoesophageal diverticulum [endoscopic laser-assisted . esophagodivertuculostomy (ELAED)];
- stenosis, including subglottic stenosis;
- tonsillectomy (including tonsillar cryptolysis, neoplasma) and tonsil ablation/tonsillotomy;
- . pulmonary bronchial and tracheal lesion removal;
- benign and malignant nodules, tumors and fibromas (larynx, pharynx, trachea, . tracheobronchial/endobronchial);
- . benign and malignant lesions and fibromas (nose and nasal passages);
- . benign and malignant tumors and fibromas; (oral);
- . stapedotomy/stapedectomy;
- acoustic neuroma in the ear; .
- . superficial lesions of the ear, including chondrodermatitis nondularis chronica helices/Winkler's disease;
- . telangiectasia/hemangioma of larynx, pharynx and trachea (includes uvula, palatal or upper lateral pharyngeal tissue);
- . cordectomy, cordotomy (for the treatment of vocal fold paralysis/vocal fold motion impairment), and cordal lesions of larynx, pharynx and trachea;
- . myringotomy/tympanostomy (tympanic membrane fenestration);
- . uvulopalatoplasty (LAUP, laser UPPP);
- . turbinectomy and turbinate reduction/ablation);
- . septal spur ablation/reduction and septoplasty:
- partial glossectomy: .
- tumor resection on oral, subfacial and neck tissues; .
- . rhinophyma;
- verrucae vulgares (warts); .
- gingivoplasty/gingivoctomy.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
currence of CDRH, Office of Device Evaluation (ODE)
Neil K. Regle for m
Division of Surgical, Orthopedic. and Restorative Devices
510(k) Number K103501
5 Page of 10
{7}------------------------------------------------
K10 3501 510(k) Number (if known):
Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories Device Name:
Indications for Use - Continued from the previous page:
## Gynecology (GYN)
Laser incision, excision, ablation and/or vaporization and of soft tissue in gynecology (GYN) for the treatment of:
- conization of the cervix, including cervical intraepithelial neoplasia (CIN), vulvar and . vaginal intraepithelial neoplasia (VIN, VAIN);
- condyloma acuminate, including cervical, genital, vulvar, perineal, and Bowen's disease, . (Erythroplasia of Queyrat) and Bowenoid papulosa (BP) lesions;
- . leukoplakia (vulvar dystrophies);
- incision and drainage (I&D) of Bartholin's and nubuthian cysts; .
- . herpes vaporization;
- urethral caruncle vaporization;
- cervical dysplasia; ●
- . benign and malignant tumors;
- hemangiomas. ◆
# GYN Laparoscopy
Vaporization, incision, excision, ablation, or photocoagulation of soft tissue in endoscopic and laparoscopic surgery, including GYN laparoscopy, for treatment of:
- . endometrial lesions, including ablation of endometriosis;
- . excision/lysis of adhesions;
- . salpingostomy;
- . oophorectomy/ovariectomy;
- . fimbroplasty;
- . metroplasty;
- . microsurgery (tubal);
- uterine myomas and fibroids; .
V Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nxt R.P. Dal fir mkn
(Division Sigh-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K10-3501
Page 6 of 10
-157
{8}------------------------------------------------
510(k) Number (if known): K10 3501
Device Name: Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories
Indications for Use - Continued from the previous page:
### GYN Laparoscopy, continued
- . ovarian fibromas and follicle cysts;
- . uterosacral ligament ablation;
- . hysterectomy.
#### Neurosurgery
Laser incision, excision, ablation and/or vaporization and of soft tissue in neurosurgery for the treatment of:
### Cranial
- . posterior fossa tumors;
- peripheral neurectomy;
- benign and malignant tumors and cysts (e.g. gliomas, menigiomas (including basal e tumors), acoustic neuromas, lipomas and large tumors);
- . arteriovenous malformation;
- . pituitary gland tumors (transphenoidal approach).
#### Spinal Cord
- incision/excision and vaporization of benign and malignant tumors and cysts:
- intra- and extradural lesions;
- laminectomy/ laminotomy/ microdiscectomy. ●
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
N.S.P.R.B.L. for asks
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103501
Page 7 of 10
ਹਵਿਲ
<sup>9</sup>
{9}------------------------------------------------
510(k) Number (if known): K10 350 1
Device Name: Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories
Indications for Use - Continued from the previous page:
### Orthopedics
Incision/excision and vaporization and of soft tissue in orthopedic surgery. Applications include:
#### Arthroscopy
- . menisectomy;
- . chodromalacia;
- � chondroplasty;
- ligament release (lateral and other); .
- . excision of plica;
- partial synovectomy.
#### General
- deridement of traumatic wounds;
- debridement of decubitus and diabetic ulcers; ●
- microsurgery;
- artificial joint revision; ●
- PMMA removal.
### General/Thoracic Surgery
Incision, excision and vaporization and of soft tissue in general and thoracic surgery including endoscopic and open procedures. Applications include:
- debridement of decubitus ulcers, stasis, diabetic and other ulcers;
- mastectomy;
- debridement of burns;
- rectal and anal hemorrhoidectomy;
- breast biopsy; ●
V Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) |
|---------------------|
|---------------------|
Division of Surgical, Orthopedic,
and Restorative Devices
| 510(k) Number | K10-350 |
|---------------|---------|
|---------------|---------|
Page 8 of 10ਜ ਦਿਰ
{10}------------------------------------------------
K10 350 | 20 510(k) Number (if known):
Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories Device Name:
Indications for Use - Continued from the previous page:
## General/Thoracic Surgery, continued
- reduction mammoplasty; .
- cytoreduction for metastatic disease; .
- laparotomy and laparoscopic applications; .
- mediastinal and thoracic lesions and abnormalities; .
- skin tag vaporization; �
- atheroma; ♥
- cysts, including sebaceous cysts, pilar cysts, and mucous cysts of the lips; .
- pilonidal cyst removal and repair;
- abscesses;
- other soft tissue applications.
#### Dental/Oral Surgery
Incision/excision and vaporization of soft tissue in dentistry and oral surgery. Applications include:
- gingivectomy- removal of hyperplasias; .
- . gingivoplasty;
- incisional and excisional biopsy; ●
- treatment of ulcerous lesions, including aphthous ulcers; ●
- incision of infection when used with antibiotic therapy; ●
- frenectomy (frenum release); .
- excision and ablation of benign and malignant lesions;
- homeostasis;
- operculectomy;
- crown lengthening;
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
| 510(k) Number | K103501 |
|---------------|--------------|
| | Page 9 of 10 |
| | 160 |
{11}------------------------------------------------
510(k) Number (if known): K10 3501
Device Name: Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories
Indications for Use - Continued from the previous page:
## Dental/Oral Surgery, continued
- o removal of soft tissue, cysts and tumors;
- oral cavity tumors and hemangiomas; .
- . abscesses;
- . extraction site hemostasis;
- salivary gland pathologies;
- . preprosthetic gum preparation;
- . leukoplakia;
- . partial glossectomy;
- periodontal gum resection. .
### Genitourinary
Incision/excision and vaporization and of soft tissue in genitourinary procedures. Applications include:
- . benign and malignant lesions of external genitalia;
- . condyloma;
- . phimosis
- erythroplasia. .
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Net. R.P. Side for mka
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number. K103501.
Page 10 of 10
484
{12}------------------------------------------------
# Appendix 7-510(k) Summary for the Alma Lasers Pixel CO2TM Laser System, Delivery Devices and Accessories
Summary Preparation Date: November 24, 2010
## · General Information
| Sponsor/<br>510(k) Owner<br>Address and<br>Establishment<br>Registration # | Sponsor | | |
|----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|------------------------------------|-------------------------------------------------------------|
| | Alma Lasers, Inc.<br>485 Half Day Rd. Suite No. 100<br>Buffalo Grove, IL 60089, USA<br>FDA Registration #: 3004167969 | | |
| | Tatiana Epstein<br>VP QA&RA<br>Alma Lasers, Inc. | Telephone:<br>Facsimile:<br>Email: | (224) 377-2011<br>(224) 377-2050<br>tatianae@almalasers.com |
| Contact Person: | Main Contact:<br>Tatiana Epstein<br>VP QA&RA<br>Alma Lasers, Inc. | Telephone:<br>Facsimile:<br>Email: | (224) 377-2011<br>(224) 377-2050<br>tatianae@almalasers.com |
| | Secondary Contact:<br>Avi Farbstein<br>EVP and GM North America | Telephone:<br>Facsimile: | (224) 377-2011<br>(224) 377-2050 |
## · Names
Alma Lasers Pixel CO2TM Laser System, Delivery Device Names: Devices and Accessories
Primary Classification Names: Laser Instrument, Surgical, Powered; GEX
### · Predicate Devices
Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories (K080463), cleared 07/23/2008
#### • Product Description
The Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories are used to deliver light energy and are intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in the medical specialties, including: aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynecology, neurosurgery, orthopedics (soft tissue), arthroscopy (knee) , general and thoracic surgery (including open and endoscopic), oral surgery and genitourinary surgery.
The Alma Lasers Pixel CO2™ CO2 Laser System consists of the following major components (refer to Error! Reference source not found.):
- 6. Laser system console (containing the optical bench assembly and laser, the microcontroller control electronics and system software, the high voltage power supply, the laser cooling system, the compressed air-purge system, and the service panel)
{13}------------------------------------------------
- 7. LCD control panel with touch-screen technology
- 8. 7-joint articulated arm
- Footswitch 9.
- 10. Delivery Device Handpieces:
- . Focusing handpieces
- Pixel handpieces .
## · Indications for Use
The Alma Lasers Pixel CO2TM Laser System, Delivery Device and Accessories are used to deliver light energy and are intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in the medical specialties, including: aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynecology, neurosurgery, orthopedics (soft tissue), arthroscopy (knee) , general and thoracic surgery (including open and endoscopic), oral surgery and genitourinary surgery.
The intended use is identical to that previously cleared for the Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories (K080463).
The Indications for Use statement can be found in Appendix 4.
## · Rationale for Substantial Equivalence
The Alma Lasers Pixel CO2TM Laser System, Delivery Device and Accessories has the following similarities to the previously cleared Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories (K080463):
- Has the same intended use and indications for use, .
- . Uses the same operating principle (technology),
- . Incorporates the same basic design,
- Incorporates the same materials, .
- Is packaged using the same materials and processes. .
In summary, the Alma Lasers Pixel CO2™ Laser System, Delivery Device and Accessories is substantially equivalent to the predicate devices.
#### · Safety and Effectiveness Information
The review of the indications for use and technical characteristics provided demonstrates that the Alma Lasers Pixel CO2M Laser System, Delivery Device and Accessories is substantially equivalent to the predicate device.
#### · Conclusion
Alma Lasers Pixel CO2™ Laser System, Delivery Device and Accessories was found to be substantially equivalent to the predicate Alma Lasers ThermoXEL™ CO2 Laser System and Delivery Device Accessories (K080463).
{14}------------------------------------------------
510(k) Number (if known): K10
Device Name: Alma Lasers Pixel CO2TM Laser System, Delivery Devices and Accessories
Indications for Use - Continued from the previous page:
### Dermatology & Plastic Surgery, continued
Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of:
- . keratoses, including actinic and seborrheic keratosis, seborrhoecae vulgares, seborrheic wart, and verruca seborrheica;
- . vermillionectomy of the lip;
- cutaneous horns; .
- solar/actinic elastosis; .
- . chelitis, including actinic chelitis;
- lentigines, including lentigo maligna or Hutchinson's malignant freckle;
- . uneven pigmentation/ dyschromia;
- . acne scars:
- . surgical scars;
- . keloids including acne keloidalis nuchae;
- . hemangiomas (including Buccal, port wine and pyogenic granulomas/granuloma pyogenicum/granuloma telagiectaticum);
- tattoos;
- . telangiectasia;
- removal of small skin tumors, including periungual (Koenen) and subungual fibromas;
- . superficial pigmented lesions;
- adenosebaceous hypertrophy or sebaceous hyperplasia; .
- . rhinophyma reduction;
- cutaneous papilloma (skin tags); .
- . milia;
- debridement of eczematous or infected skin;
- basal and squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis (BP) lesions;
Over-The-Counter Use Prescription Use AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 2 of 10
ਜਵੰਤੇ
{15}------------------------------------------------
510(k) Number (if known): K10
Device Name: Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories
Indications for Use - Continued from the previous page:
#### Dermatology & Plastic Surgery, continued
- nevi, including spider, epidermal and protruding; .
- . neurofibromas;
- laser de-epithelialization; .
- . tricoepitheliomas;
- xanthelasma palpebrarum; .
- syringoma .
Laser ablation, vaporization and for excision for complete or partial nail matrixectomy.
Vaporization/coagulation of:
- � benign/malignant vascular/avascular skin lesions;
- . Moh's Surgery;
- . Lipectomy;
- Verrucae and seborrhoecae vulgares, including paronychial, periungal, and subungaual warts.
Laser incision and /or excision of soft tissue for the performance of upper and lower eyelid blepharoplasty.
Laser incision and /or excision of soft tissue for the creation of recipient sites for hair transplantation.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 3 of 10
ਜੋਵਿੰਧ
క్రిశో
{16}------------------------------------------------
510(k) Number (if known): K10
Device Name: Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories
Indications for Use - Continued from the previous page:
# Podiatry
Laser ablation, vaporization and /or excision of soft tissue for the reduction, removal, and/or treatment of:
- . verrucae vulgares/plantar (warts), including paronychial, periungal, and subungual warts;
- fungal nail treatment; .
- . porokeratoma ablation:
- . ingrown nail treatment;
- . neuromas/fibromas, including Morton's neuroma;
- debridement of ulcers:
- . other soft tissue lesions.
Laser ablation, vaporization and /or excision in podiatry for complete or partial matrixectomy.
## Otolaryngology (ENT)
Laser incision, excision, ablation and/or vaporization of soft tissue in otolaryngology the treatment of:
- ◆ choanal atresia,
- leukoplakia, including oral, larynx, uvula, palatal, upper lateral pharyngeal tissue; .
- nasal obstruction: ●
- . adult and juvenile papillomatosis polyps;
- polypectomy of nose and nasal passages; .
- lymphangioma removal; .
- . removal of vocal cord/fold nodules, polyps and cysts;
- removal of recurrent papillomas in the oral cavity, nasal cavity, larynx, pharynx and . trachea, including the uvula, palatal, upper lateral pharyngeal tissue, tongue and vocal cords:
Prescription Use > (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 4 of 10
155 ---
{17}------------------------------------------------
510(k) Number (if known): K10
Device Name: Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories
Indications for Use - Continued from the previous page:
### Otolaryngology (ENT), continued
- laser/tumor surgery in the larvnx, pharvnx, nasal, ear and oral structures and tissue; . ●
- . Zenker's Diverticulum/pharvngoesophageal diverticulum fendoscopic laser-assisted esophagodivertuculostomy (ELAED)];
- . stenosis, including subglottic stenosis;
- . tonsillectomy (including tonsillar cryptolysis, neoplasma) and tonsil ablation/tonsillotomy;
- pulmonary bronchial and tracheal lesion removal;
- . benign and malignant nodules, tumors and fibromas (larynx, pharynx, trachea, tracheobronchial/endobronchial);
- . benign and malignant lesions and fibromas (nose and nasal passages);
- benign and malignant tumors and fibromas; (oral); .
- stapedotomy/stapedectomy; .
- . acoustic neuroma in the ear;
- . superficial lesions of the ear, including chondrodermatitis nondularis chronica helices/Winkler's disease;
- telangiectasia/hemangioma of larynx, pharynx and trachea (includes uvula, palatal or . upper lateral pharyngeal tissue);
- cordectomy, cordotomy (for the treatment of vocal fold paralysis/vocal fold motion . impairment), and cordal lesions of larynx, pharynx and trachea;
- . myringotomy/tympanostomy (tympanic membrane fenestration);
- . uvulopalatoplasty (LAUP, laser UPP);
- . turbinectomy and turbinate reduction/ablation);
- . septal spur ablation/reduction and septoplasty;
- . partial glossectomy:
- tumor resection on oral, subfacial and neck tissues; ●
- rhinophyma;
- verrucae vulgares (warts); .
- gingivoplasty/gingivoctomy. . .
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{18}------------------------------------------------
510(k) Number (if known): K10
Device Name: Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories
Indications for Use - Continued from the previous page:
# Gynecology (GYN)
Laser incision, excision, ablation and/or vaporization and of soft tissue in gynecology (GYN) for the treatment of:
- conization of the cervix, including cervical intraepithelial neoplasia (CIN), vulvar and ● vaginal intraepithelial neoplasia (VIN, VAIN);
- condyloma acuminate, including cervical, genital, vulvar, perineal, and Bowen's disease, ● (Erythroplasia of Queyrat) and Bowenoid papulosa (BP) lesions;
- leukoplakia (vulvar dystrophies); .
- incision and drainage (I&D) of Bartholin's and nubuthian cysts; .
- herpes vaporization; .
- . urethral caruncle vaporization;
- . cervical dysplasia:
- benign and malignant tumors;
- hemangiomas. .
#### GYN Laparoscopy
Vaporization, incision, excision, ablation, or photocoagulation of soft tissue in endoscopic and laparoscopic surgery, including GYN laparoscopy, for treatment of:
- endometrial lesions, including ablation of endometriosis; .
- excision/lysis of adhesions;
- . salpingostomy;
- . oophorectomy/ovariectomy;
- fimbroplasty;
- metroplasty;
Prescription Use
- microsurgery (tubal);
(Part 21 CFR 801 Subpart D)
- uterine myomas and fibroids; .
>
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 6 of 10
{19}------------------------------------------------
510(k) Number (if known): K10
Device Name: Alma Lasers Pixel CO2TM Laser System, Delivery Devices and Accessories
Indications for Use - Continued from the previous page:
### GYN Laparoscopy, continued
- ovarian fibromas and follicle cysts; .
- uterosacral ligament ablation; ●
- hysterectomy. .
# Neurosurgery
Laser incision, excision, ablation and/or vaporization and of soft tissue in neurosurgery for the treatment of:
### Cranial
- posterior fossa tumors;
- . peripheral neurectomy;
- benign and malignant tumors and cysts (e.g. gliomas, menigiomas (including basal . tumors), acoustic neuromas, lipomas and large tumors);
- . arteriovenous malformation:
- pituitary gland tumors (transphenoidal approach). .
#### Spinal Cord
- incision/excision and vaporization of benign and malignant tumors and cysts; ●
- . intra- and extradural lesions;
- laminectomy/ laminotomy/ microdiscectomy. .
イ Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
{20}------------------------------------------------
510(k) Number (if known): K10
Device Name: Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories
Indications for Use - Continued from the previous page:
# Orthopedics
Incision/excision and vaporization and of soft tissue in orthopedic surgery. Applications include:
#### Arthroscopy
- . menisectomy;
- chodromalacia; ●
- chondroplasty; ●
- ligament release (lateral and other); .
- excision of plica; .
- . partial synovectomy.
#### General
- deridement of traumatic wounds; .
- debridement of decubitus and diabetic ulcers; ◆
- . microsurgery;
- . artificial joint revision;
- . PMMA removal.
### General/Thoracic Surgery
Incision, excision and vaporization and of soft tissue in general and thoracic surgery including endoscopic and open procedures. Applications include:
- debridement of decubitus ulcers, stasis, diabetic and other ulcers;
- . mastectomy;
- debridement of burns; .
- rectal and anal hemorrhoidectomy; .
- . breast biopsy;
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{21}------------------------------------------------
510(k) Number (if known): K10
Device Name: Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories
Indications for Use - Continued from the previous page:
# General/Thoracic Surgery, continued
- reduction mammoplasty; .
- . cytoreduction for metastatic disease;
- laparotomy and laparoscopic applications; .
- mediastinal and thoracic lesions and abnormalities; .
- skin tag vaporization;
- . atheroma:
- cysts, including sebaceous cysts, pilar cysts, and mucous cysts of the lips; .
- pilonidal cyst removal and repair; .
- abscesses; ●
- . other soft tissue applications.
# Dental/Oral Surgery
Incision/excision and vaporization of soft tissue in dentistry and oral surgery. Applications include:
- gingivectomy- removal of hyperplasias;
- . gingivoplasty;
- incisional and excisional biopsy; .
- treatment of ulcerous lesions, including aphthous ulcers; .
- incision of infection when used with antibiotic therapy;
- frenectomy (frenum release); ●
- excision and ablation of benign and malignant lesions; .
- homeostasis; .
- operculectomy:
- crown lengthening; ●
- Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 9 of 10
391
{22}------------------------------------------------
510(k) Number (if known): K10
Device Name: Alma Lasers Pixel CO2™ Laser System, Delivery Devices and Accessories
Indications for Use - Continued from the previous page:
# Dental/Oral Surgery, continued
- removal of soft tissue, cysts and tumors; .
- . oral cavity tumors and hemangiomas;
- abscesses; .
- . extraction site hemostasis;
- . salivary gland pathologies;
- . preprosthetic gum preparation;
- leukoplakia;
- . partial glossectomy;
- periodontal gum resection.
# Genitourinary
Incision/excision and vaporization and of soft tissue in genitourinary procedures. Applications include:
- . benign and malignant lesions of external genitalia;
V
- . condyloma;
- . phimosis
- . erythroplasia.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.