The bite away® two is indicated for use to provide temporary relief of the pain and itching resulting from insect stings and bites such as bees, wasps and mosquitoes.
Device Story
Portable, battery-powered, handheld device for OTC use; provides localized heat therapy to insect stings/bites. User activates device via one of two buttons (3-second or 5-second duration); MOSFET powers a 7mm ceramic disc to ~51°C. Device includes LED indicator and audible buzzer for status/error signaling. User manually applies heated disc to affected skin area; device automatically terminates heating after selected duration. User controls treatment duration and site application based on personal heat tolerance; design limits maximum heat delivery to prevent skin injury. Benefits include temporary relief of pain and itching associated with insect bites/stings.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1 series), and electromagnetic compatibility (IEC 60601-1-2). Thermal delivery characteristics were verified via bench testing.
Technological Characteristics
Handheld, battery-powered (2x AAA). Components: hard plastic housing (biocompatible polymers), heated ceramic disc (biocompatible ceramic), PCB, MOSFET, electro-mechanical buttons. Energy source: 3V DC. Dimensions: 6mm heated area. Connectivity: None. Sterilization: Not stated. Software: Embedded microprocessor for timing and error signaling.
Indications for Use
Indicated for temporary relief of pain and itching from insect stings and bites (e.g., bees, wasps, mosquitoes) in patients of all ages.
Regulatory Classification
Identification
A powered heating pad is an electrical device intended for medical purposes that provides dry heat therapy for body surfaces. It is capable of maintaining an elevated temperature during use.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E part 807 of this chapter subject to § 890.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 12, 2025
MibeTec, GmbH
Oliver Grunert
Head of Development, Production, Regulatory and Quality
Münchener Str. 15
Brehna, 06796
Germany
Re: K250047
Trade/Device Name: bite away two
Regulation Number: 21 CFR 890.5740
Regulation Name: Powered Heating Pad
Regulatory Class: Class II
Product Code: IRT
Dated: August 1, 2025
Received: August 12, 2025
Dear Oliver Grunert:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250047 - Oliver Grunert
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250047 - Oliver Grunert
Page 3
Sincerely,
Amber T. Ballard -S
Amber Ballard, PhD
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250047 | |
| Device Name bite away® two | |
| Indications for Use (Describe) The bite away® two is indicated for use to provide temporary relief of the pain and itching resulting from insect stings and bites such as bees, wasps and mosquitoes. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
mibeTec, GmbH bite away® two device
Page 1 of 5
Date Prepared: September 12, 2025
I. SUBMITTER
Manufacturer Name:
mibeTec, GmbH
Munchener Strasse 15
Brehna, Germany
D-06796
Telephone +49 34954 247 489
Mfg. Establishment Registration Number: 3015733772
Official Contact:
Oliver Grunert
Head of Development, Production, Regulatory and Quality
mibetec.regulatory@dermapharm.com
II. DEVICE NAME
Name of Device: bite away® two
Common or Usual Name: Powered Heating Pad
Classification Name: Pad, Heating, Powered (21 CFR 890.5740)
Regulatory Class: 2
Product Code: IRT
510(k) Identification: K250047
III. PREDICATE DEVICE
bite away® neo (K220514)
IV. DEVICE DESCRIPTION
The mibeTec bite away® two device is a light weight, portable, handheld, battery powered, user-operated device that produces mild heat for direct contact with the affected areas of the skin. The heat is initiated by the user through the activation of the unit by depressing a non-locking button. The device is provided with two (2) side-by-side buttons. The user has the choice of a short 3 second heat treatment or a 5 second heat treatment depending on which button is depressed. The activation of the unit is signified with the illumination of an LED light and an audible chirp through the use of an electronic buzzer. An audible error message is activated when there is a user error. The device is not connected to the user as the user is in complete control of the heat treatment and as such, self-delivers the heat treatment to themselves by contacting the device to their own anatomy / treatment site. As in any user-controlled heat therapy, the user determines the tolerance to such delivered heat
{5}
510(k) Summary
mibeTec, GmbH bite away® two device
Page 2 of 5
and applies and removes the heat treatment per their own tolerance; thereby virtually alleviating the risks of inadvertent overheating of the skin. The short duration of treatment time of 3 and 5 seconds and the small surface area of the heated ceramic plate that contacts the patient further alleviates risk of overheating the skin as the device automatically stops heating the element after 3 or 5 seconds; limiting the maximum amount of heat to be delivered to the site.
The mibeTec bite away® two device utilizes two AAA batteries (1.5 volts each) that power a MOSFET that heats a 7mm ceramic disc to approx. 51°C +/- 1% when activated by the user through the use of one of the two electro-mechanical NOC (normal open contact) buttons (3 seconds or 5 seconds of heating). The device consists of a few major components: a hard plastic outer case, a printed circuit board, two activation buttons, two AAA batteries, and a heated ceramic plate that contacts the patient.
## Material Composition
The mibeTec bite away® two device housing is fabricated with biocompatible polymers and the heated disc is fabricated with a biocompatible ceramic material. The electronics are fabricated on a standard electrical circuit board with standard electronics (capacitors, resistor, microprocessors, etc.).
## V. INDICATIONS FOR USE
The bite away® two is indicated for use to provide temporary relief of the pain and itching resulting from insect stings and bites such as bees, wasps and mosquitoes.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The bite away® two device shares indications for use and design principles with the following predicate devices: bite away® neo (K220514), a Class II medical device that was cleared for marketing in the United States under K220514.
## Indications for Use
The bite away® two device and the bite away® neo (K220514) predicate device are substantially equivalent with respect to their indications for use as they are both indicated for the delivery of mild heat to the skin/dermis to provide temporary relief of the pain and itching resulting from insect stings and bites such as bees, wasps and mosquitoes.
Additionally, bite away® two and the bite away® neo device share indication for use principles of being OTC (over the counter).
{6}
510(k) Summary
mibeTec, GmbH bite away® two device
# SUMMARY: TABLE OF SUBSTANTIAL EQUIVALENCE
The bite away® two device is substantially equivalent to the bite away® neo predicate device in the following respects:
| | Subject Device | Predicate Device | Substantially Equivalent |
| --- | --- | --- | --- |
| | mibeTec, GmbH bite away® two | mibeTec, GmbH bite away® neo | |
| Illustration | | | |
| 510(k) Number | K250047 | K220514 | |
| Intended Use | Delivery of mild heat to the skin/dermis | Delivery of mild heat to the skin/dermis | Identical |
| Indications for Use | The bite away® two is indicated for use to provide temporary relief of the pain and itching resulting from insect stings and bites such as bees, wasps and mosquitoes. | The bite away® neo is indicated for use to provide temporary relief of the pain and itching resulting from insect stings and bites such as bees, wasps and mosquitoes. | Identical |
| Design | Handheld device that heats a ceramic disc with a resistor to temperatures of 51°C | Handheld device that heats a ceramic disc with a resistor to temperatures of 51.5°C | Yes |
| Temperature range of Skin surface | 45.8 °C – 46.7 °C | not publicly available | Yes |
| Temperature Range (device surface) | 50 °C – 53 °C | 50 °C – 53 °C | Yes |
| max. Temperature of Skin surface | 47.1 °C | 48.6 °C | Yes |
| Operator Directed/Applied to the Skin | Yes | Yes | Yes |
| Dry Weight | 32 grams | 40 grams | Yes |
| Weight with batteries | 55 grams | 86 grams | Yes |
| Duration of Use | 3 and 5 seconds | 3 and 5 seconds | Yes |
| Dimension of Heated Area on the Device | 6 mm | 7 mm | Yes |
| Power Source | AAA Batteries | AA Batteries | Yes |
| Voltage | 3 volts DC | 3 volts DC | Yes |
| Energy Transfer Source | Heated ceramic disc | Heated ceramic disc | Yes |
| Hard Plastic Outer Case | Yes | Yes | Yes |
| LED Light | Yes | Yes | Yes |
| Audible Signal | Yes | Yes | Yes |
| Microprocessor | Yes | Yes | Yes |
| Placed Directly on Insect Bite for Treatments | Yes | Yes | Yes |
| Classification Name | Powered Heating Pad | Powered Heating Pad | Yes |
| OTC Use | Yes | Yes | Yes |
| Product Class | 2 | 2 | Yes |
{7}
510(k) Summary
mibeTec, GmbH bite away® two device
Page 4 of 5
| Product Code | IRT | IRT | Yes |
| --- | --- | --- | --- |
| CFR Section | 890.5740 | 890.5740 | Yes |
## VII. PERFORMANCE TESTING
### Non-Clinical Testing
#### Biocompatibility Testing
The mibeTec bite away® two device was evaluated against the FDA Guidance document entitled “Use of International Standard ISO 10993-1, ‘Biological Evaluation of Medical Devices – Part 1: Evaluation of Testing within a Risk Management Process.’” The battery of testing included:
- Cytotoxicity – ISO 10993-5
- Sensitization – ISO 10993-10
- Irritation – ISO 10993-23
The device meets the requirements therein.
#### Electrical Safety
The mibeTec bite away® two device demonstrated compliance with the appropriate sections of the following electrical safety standards:
- IEC 60601-1:2005/AMD 1: 2012/AMD2:2020
- IEC 60601-1-6: 60601-1-6:2010/AMD1:2013/AMD2:2020
- IEC 62366: 62366-1:2015/AMD1:2020
- IEC 60601-1-11:2015/AMD1:2020
The device meets the requirements therein.
#### Electromagnetic Compatibility
The mibeTec bite away® two device demonstrated compliance with the appropriate sections of the following electromagnetic compatibility standards:
- IEC 60601-1-2:2014/AMD1:2020
- IEC TR 60601-4-2:2016
The device meets the requirements therein.
#### Software Validation and Verification
The software features of the mibeTec bite away® two device have been tested for functionality. The sequence of events under normal operating conditions has been verified and all features have been demonstrated to work properly. Furthermore, all error messages have been verified to operate properly when each programmed fault is triggered.
The bite away® two device was evaluated for thermal delivery characteristics to demonstrate that it performs as intended.
{8}
510(k) Summary
mibeTec, GmbH bite away® two device
Page 5 of 5
## Clinical Testing
Clinical testing was not required to demonstrate safety and efficacy.
## VIII. CONCLUSION
The non-clinical testing demonstrates that the subject device is as safe and effective and performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.