K220514 · mibeTec GmbH · IRT · Jun 23, 2022 · Physical Medicine
Device Facts
Record ID
K220514
Device Name
bite away neo
Applicant
mibeTec GmbH
Product Code
IRT · Physical Medicine
Decision Date
Jun 23, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5740
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Delivery of mild heat to the skin/dermis
Device Story
Portable, hand-held, battery-powered device; delivers mild heat to skin/dermis via ceramic disc. User-operated; initiates treatment by depressing non-locking button. Two treatment durations: 3 or 5 seconds. Device includes microcontroller, MOSFET transistor, thermistor, and buzzer. Activation signaled by LED and audible chirp; audible alarm for user error. Device automatically stops heating after selected duration to prevent overheating. Used in home/OTC setting; patient self-administers by contacting device to affected site. Heat therapy intended to alleviate pain/itching from insect bites. Benefits include portable, controlled, user-directed relief.
Clinical Evidence
No clinical studies were performed. Safety and effectiveness supported by bench testing, including biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and usability (IEC 60601-1-6).
Indicated for temporary relief of pain and itching from insect stings and bites (e.g., bees, wasps, mosquitoes). OTC use.
Regulatory Classification
Identification
A powered heating pad is an electrical device intended for medical purposes that provides dry heat therapy for body surfaces. It is capable of maintaining an elevated temperature during use.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E part 807 of this chapter subject to § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.
June 23, 2022
MibeTec. GmbH % Kenneth Kleinhenz Official Correspondent QSR Consulting 10807 Dakota Ranch Road Santee, California 92071
Re: K220514
Trade/Device Name: bite away® neo Regulation Number: 21 CFR 890.5740 Regulation Name: Powered heating pad Regulatory Class: Class II Product Code: IRT Dated: March 30, 2022 Received: April 1, 2022
Dear Kenneth Kleinhenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220514
Device Name bite away® neo
Indications for Use (Describe)
The bite away® neo is indicated for use to provide temporary relief of the pain and itching from insect stings and bites such as bees, wasps and mosquitoes.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Date Prepared: 21 June 2022
#### I. SUBMITTER
Manufacturer Name:
mibeTec, GmbH Munchener Strasse 15 Brehna, Germany D-17493 +49 34954 247 489 telephone
Mfg. Establishment Registration Number: 3015733772
Official Contact:
Kenneth K. Kleinhenz Regulatory Affairs Telephone (619) 244-9573 Kleinhenz64@gmail.com
### II. DEVICE
| Name of Device: | bite away® neo |
|------------------------|--------------------------------------|
| Common or Usual Name: | Heating pad |
| Classification Name: | Powered Heating Pad (21 CFR 890.5740 |
| Regulatory Class Name: | II |
| Product Code: | IRT |
| 510(K) Identification: | K220514 |
### PREDICATE DEVICE III.
Riemser Bite Away, K160943
### DEVICE DESCRIPTION IV.
## Design Characteristics
The bite away® neo device is a light weight, portable, hand-held, battery powered, user- operated device that produces mild heat for direct contact with the affected areas of the skin. The heat is initiated by the user through the activation of the unit by depressing a non-locking button. The device is provided with two (2) side-by-side buttons. The user has the choice of a short 3 second heat treatment or a 5 second heat treatment depending on which button is depressed. The activation of the unit is signified with the illumination of an LED light and an audible chirp through the use of an electronic buzzer. An audible alarm is activated when there is a user error. The device is not connected to the user as the user is in complete control of the heat treatment and as such, self-delivers the heat treatment to themselves by contacting the device to their own anatomy / treatment site. The short duration of 3 and 5 seconds of heating the device's heated ceramic plate that contacts the patient further alleviates risk of overheating the skin as the device automatically stops heating the element after 3 or 5 seconds; limiting the maximum amount of heat to be delivered to the site.
{4}------------------------------------------------
The bite away® neo device utilizes two AA batteries (1.5 volts each) that power a resistor that heats a 7mm ceramic disc to approximately 50°C when activated by the user through the use of one of the two electro-mechanical NOC (normal open contact) buttons (3 seconds of heating). The device consists of a few major components: a hard plastic outer case, a printed circuit board, an activation button. AA batteries, and a heated ceramic plate that contacts the printed circuit board consists of the following components: electro-mechanical NOC buttons, microcontroller with programmed firmware, MOSFET transistor heating element, ceramic pad, resistors, coils, capacitor, thermistor temperature sensor and buzzer. The bite away® neo device is approximately 7 inches long and weighs approximately 40 grams.
## Material Composition
The bite away® neo device housing is fabricated with biocompatible polymers and the heated disc is fabricated with a biocompatible ceramic material. The electronics are fabricated on a standard electrical circuit board with standard electronics (capacitors, resistor, microprocessors, etc.).
### V. INDICATIONS FOR USE
The bite away® neo is indicated for use to provide temporary relief of the pain and itching resulting from insect stings and bites such as bees, wasps and mosquitoes.
## INTENDED USE
Delivery of mild heat to the skin / dermis
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The bite away neo Device shares indications for use and design principles with the following predicate device: bite away (Riemser, K160943); a Class II medical device that was cleared for marketing in the United States under K160943.
## Indications For Use
The bite away® neo Device and the bite away (Riemser, K160943) predicate device are substantially equivalent with respect to their indications for use as they are both indicated for the same intended use of temporary relief from the pain and itching resulting from insect stings and bites. Additionally, the bite away® neo and the bite away predicate device shares indications for use principles of being OTC (over the counter).
{5}------------------------------------------------
## Design and Materials
The design principles of the bite away® neo and the bite away predicate device (K160943) are substantially equivalent as they all share common design principals of being a useroperated, hand-held, light weight, battery powered, reusable devices that deliver mild heat to the skin/dermis as a means to transfer heat from the device to the user- directed anatomical location. All devices share the common design feature of user-applied heat therapy that is directed and controlled solely by the user. The bite away® neo and the bite away predicate device (K160943) share a common material and design principal of utilizing a ceramic disc to transfer the heat from the device to the skin. The bite away® neo and the bite away predicate device also share common design and material features of utilizing a light-weight polymer outer case to hold and contain the electronics and AA batteries that power the devices also share the common design feature of being form-fit packaged in a paperboard box. The bite away® neo and the bite away predicate device (K160943) share the design feature of being over-the-counter (OTC) devices under 21 CFR 890.5740. The delivery of heat to the insect bite is delivered and controlled by the user. Device weight does not play a role in the delivery of heat. A weight difference of 15 grams is virtually indistinguishable by the user. Therefore, differences in device weight do not affect safety or efficacy. Differences in contact time by 1 second (5 sec. versus 6 sec) do not affect safety or efficacy and only plays a role in greater safety as it creates a larger margin of safety against burns to the skin. The reduced heat contact time of 1 second does not affect efficacy given the fact that the user can reapply the heat with an additional heat treatment if necessary. The subject device and the predicate device have the same indications for use with respect to insect bites and only differ in the claimed mechanism of action (blood flow), which has no affect on safety or efficacy because the mechanisms of action does not affect the subjects use of the device.
| | Subject Device | Predicate Device<br>(K160943) | Rationale for no Effect on<br>Safety and Efficacy |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | bite away® neo | bite away | |
| Manufacturer | mibeTec, GmbH | Riemser Pharma | Validated Process |
| Device Name | bite away® neo | bite away | Device name does not imply any claims.<br>No additional risks |
| Intended Use | Delivery of mild heat to the<br>skin/dermis | Delivery of mild heat to the<br>skin/dermis | Identical. No additional risks |
| Indications for<br>Use | The bite away® neo is indicated<br>for use to provide temporary<br>relief of the pain and itching<br>resulting from insect stings and<br>bites such as bees, wasps and<br>mosquitoes. | The bite away is indicated for use to<br>provide temporary relief of the pain<br>and itching resulting from insect<br>stings and bites such as bees, wasps<br>and mosquitoes by temporarily<br>increasing localized blood<br>flow. | Indicated for same insect bites. Absence<br>of mechanism of action claims does not<br>affect safety or efficacy. No additional<br>risks |
| Design | Hand held device that heats a<br>ceramic disc with a resistor to<br>temperatures of 51.5°C | Hand held device that heats a<br>ceramic disc with a resistor to<br>temperatures of 51.5°C | Identical Maximum No additional<br>risks Temperature |
| | Subject Device | Predicate Device<br>(K160943) | Rationale for no Effect on<br>Safety and Efficacy |
| | bite away® neo | bite away | |
| Operator<br>Directed/Appli<br>ed to the Skin | Yes | Yes | Identical. No<br>additional risks |
| Dry Weight | 40 grams | 25 grams | Weight does not play a role in the<br>delivery of heat. Insignificant<br>difference; essentially undetectable by<br>user |
| Weight with<br>batteries | 86 grams | Not publicly available | Weight does not play a role in the delivery<br>of heat. Insignificant difference;<br>essentially undetectable by user |
| Duration of<br>Use | 3 and 5 seconds | 3 and 6 seconds | Greater safety with less heating time |
| Dimension of<br>Heated Area on<br>the Device | 7 mm | 9 mm | Identical temperature range; no<br>additional risk. . Small disc size<br>delivers heat to less surface area;<br>reducing risk. User can apply to<br>multiple areas for more surface area<br>treatments. |
| Temperature<br>Range (device<br>surface) | 50 - 53°C | 50 - 53°C | Identical temperature range. No<br>additional risks |
| Temperature<br>max at skin<br>surface | 48.6°C | Not publicly available | Ceramic pad touching skin is identical.<br>Min. and max. temperature output from<br>device deliver same heat to the skin. |
| Power Source | AA Batteries | AA Batteries | Identical. No additional risks |
| Voltage | 3 volts DC | 3 volts DC | Identical. No additional risks |
| Energy Transfer<br>Source | Heated ceramic disc | Heated ceramic disc | Identical. No additional risks |
| Hard Plastic<br>Outer<br>Case | Yes | Yes | Identical. No additional risks |
| LED Light | Yes | Yes | Identical. No additional risks |
| Audible Signal | Yes | Yes | Identical. No additional risks |
| Microprocessor | Yes | Yes | Identical. No additional risks |
| Placed Directly<br>on Insect Bite<br>for Treatments | Yes | Yes | Identical. No additional risks |
| Classification<br>Name | Powered Heating Pad | Powered Heating Pad | Identical. No additional risks |
| OTC Use | Yes | Yes | Identical. No additional risks |
| Product Class | 2 | 2 | Identical. No additional risks |
| Product Code | IRT | IRT | Identical. No additional risks |
| CFR Section | 890.5740 | 890.5740 | Identical. No additional risks |
| Intended Use | Delivery of mild heat to the<br>skin/dermis | Delivery of mild heat to the<br>skin/dermis | Identical. No additional risks |
The bite away® neo device is substantially equivalent to the bite away predicate device (K160943) in the following respects:
{6}------------------------------------------------
{7}------------------------------------------------
## VII. PERFORMANCE DATA
## Biocompatibility Testing
The patient contact polymers and ceramics were evaluated against the international standard ISO 10993-1 (Biological Evaluation of Medical Devices) and the FDA Guidance Document entitled, "Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process." The battery of testing included.
Cytotoxicity Sensitization Irritation
## Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the bite away neo® device, consistent with the appropriate sections of the following electrical standards: IEC 60601-1 (MOD), IEC 60601-1-2. and IEC 60601-1-11. The bite away neo® device complies with IEC 60601-1 (MOD) standard for safety, IEC 60601-1-2 standard for EMC compliance, and IEC 60601-1-11 standard for medical electrical equipment in a home healthcare environment.
## Software Verification and Validation
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software would not directly result in serious injury or death to the patient or operator. The bite away neo device complies with the applicable sections of IEC 62366.
## Usability Studies
Usability testing was conducted on the bite away neo® device consistent with the appropriate sections of the following usability standards: IEC 60601-1-6. The bite away neo® device complies with the applicable sections of IEC 60601-1-6.
## Clinical Studies
No clinical studies were performed to support safety or effectiveness of the subject device.
## VIII. CONCLUSIONS
The nonclinical testing demonstrates that the subject device is as safe and effective, and performs as well as or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.