K244050 · Apyx Medical Corporation · QPB · May 12, 2025 · General, Plastic Surgery
Device Facts
Record ID
K244050
Device Name
AYON Body Contouring System (AYON SYSTEM)
Applicant
Apyx Medical Corporation
Product Code
QPB · General, Plastic Surgery
Decision Date
May 12, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AYON Body Contouring System is intended for the fragmentation, emulsification, and aspiration of subcutaneous fatty tissue for the purpose of aesthetic body contouring. • The AYON Ultrasound Sub-System is used for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring. • The AYON Infiltration and Aspiration Sub-System(s) are used for the infiltration and suction / aspiration of fluids and tissue during surgical procedures. Infiltration and Suction is designed to operate as a stand-alone function.
Device Story
AYON Body Contouring System comprises three subsystems: Ultrasound, Infiltration, and Aspiration. Ultrasound subsystem uses piezoelectric crystal handpiece and titanium alloy probes (37 kHz) to fragment/emulsify subcutaneous fat. Infiltration subsystem uses peristaltic pump to deliver fluids via cannulas. Aspiration subsystem uses dual suction units (operable separately or in series) to remove fluids/tissue. Used in operating environments by trained physicians. System features touchscreen GUI for centralized control. Infiltration and aspiration functions can operate independently of ultrasound. Benefits include aesthetic body contouring via controlled tissue fragmentation and removal. Device accommodates Apyx Medical electrosurgical generators.
Clinical Evidence
No clinical studies were required. Substantial equivalence is supported by mechanical, physical, electrical, and biocompatibility bench testing. Aspiration performance tested per ISO 10079-1/4; electrical safety per IEC 60601-1-1; EMC per IEC 60601-1-2/4-2; reprocessing per ISO 17665; biocompatibility per ISO 10993-1.
Technological Characteristics
System includes ultrasound amplifier, handpiece (piezoelectric), and titanium alloy probes (2.2-3.7mm diameter). Infiltration uses peristaltic pump. Aspiration uses dual suction units. Operates at 37 kHz nominal frequency. Power: 100-120 VAC. Wireless footswitches for control. Connectivity: touchscreen GUI. Sterilization: per ISO 17665. Electrical safety: IEC Class I, Type BF.
Indications for Use
Indicated for fragmentation, emulsification, and aspiration of subcutaneous fatty tissue for aesthetic body contouring in patients undergoing surgical procedures.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
May 12, 2025
Apyx Medical Corporation
Angela Huber
Global Director of Regulatory Affairs
5115 Ulmerton Road
Clearwater, Florida 33760
Re: K244050
Trade/Device Name: AYON Body Contouring System (AYON SYSTEM)
Regulation Number: 21 CFR 878.5040
Regulation Name: Suction Lipoplasty System
Regulatory Class: Class II
Product Code: QPB
Dated: December 31, 2024
Received: April 11, 2025
Dear Angela Huber:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K244050 - Angela Huber
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K244050 - Angela Huber
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alicia Hemphill - Digitally signed by Alicia Hemphill -S
Date: 2025.05.12 15:21:00 -05'00'
Alicia L. Hemphill (Johnson), M.S.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K244050 | |
| Device Name AYON(TM) Body Contouring System (AYON SYSTEM) | |
| Indications for Use (Describe) The AYON Body Contouring System is intended for the fragmentation, emulsification, and aspiration of subcutaneous fatty tissue for the purpose of aesthetic body contouring. • The AYON Ultrasound Sub-System is used for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring. • The AYON Infiltration and Aspiration Sub-System(s) are used for the infiltration and suction / aspiration of fluids and tissue during surgical procedures. Infiltration and Suction is designed to operate as a stand-alone function. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Page 1 of 7 K244050
## 510(k) SUMMARY
### 1. General Information
**Submitted by:**
Apyx Medical Corporation
5115 Ulmerton Road
Clearwater, Florida 33760-4004
United States of America
**Contact Person:**
Angela Huber, PhD, RAC,
Global Director of Regulatory Affairs
Phone: 218-343-4881
Email: angela.huber@apyxmedical.com
**Date Prepared:**
May 5, 2025
**Trade Name (Model Numbers):**
AYON™ Body Contouring System
(AYON SYSTEM)
**Common Name:**
Liposuction System
**Classification:**
Class II per 21CFR 878.5040 - Suction
Lipoplasty System
**Product Code**
QPB
**Predicate Device:**
VASER 2.1 Lipo System (K110306)
### 2. Proposed Indications for Use
The AYON Body Contouring System is intended for the fragmentation, emulsification, and aspiration of subcutaneous fatty tissue for the purpose of aesthetic body contouring.
- The AYON Ultrasound Sub-System is used for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring.
- The AYON Infiltration and Aspiration Sub-System(s) are used for the infiltration and suction / aspiration of fluids and tissue during surgical procedures. Infiltration and Suction is designed to operate as a stand-alone function.
### 3. Device Description
The AYON Body Contouring System includes an Ultrasound Sub-system, an Infiltration Sub-system, and an Aspiration Sub-system. The Ultrasound Sub-system includes an ultrasound amplifier, an ultrasound handpiece, and ultrasound probes which are to be used with compatible skin ports and plugs. The Infiltration Sub-system includes a peristaltic pump and weighing scale and utilizes sterile infiltration tubing, handles and cannulas to infiltrate tissue
{5}
Page 2 of 7 K244050
with fluids. The Aspiration Sub-system includes two suction units that have the capability to function separately or together in series, such that it can be used as two separate suction units by two operators, or as a single suction unit (in series) by a single operator. The Aspiration Sub-system utilizes sterile suction tubing, handles and cannulas to aspirate fluids and soft tissue. The Infiltration and Aspiration Sub-systems are designed to operate independent of the Ultrasound Sub-system such that the system may be offered with only Infiltration and Aspiration to support standard suction-assisted liposuction (SAL) or with Ultrasound enabled to support ultrasound-assisted liposuction (UAL).
Apyx Medical devices/accessories for the system include the following:
- Infiltration and aspiration handles
- Ultrasound handpiece
- Ultrasound handpiece cables
- Ultrasound probes
- Footswitches: Two (2) wireless footswitches to control ultrasound and infiltration
- Skin ports and skin port tool
- AYON Base Tower mounting poles and canister mounts
- Power cords
Additional commercially available devices/accessories include the following:
- Infiltration and aspiration cannulas
- Aspiration filters
- Sterile, single-use infiltration and aspiration tubing
- Wrench
- Sterilization tray and cleaning brushes
- Collection canisters and liners
- Infiltration fluid warming unit(s)
The AYON Body Contouring System may also accommodate other equipment, including commercially available electrosurgical generators owned by Apyx Medical, such as the Apyx One Console.
## 4. Performance Data
The relevant FDA guidance documents were followed for all aspects of performance testing.
### a. Mechanical and Physical Bench Testing
Each subsystem was tested separately and as a combined system to demonstrate system compatibility of the subsystems and devices/accessories. The subsystem testing included:
- The ultrasound subsystem was assessed for overall performance, tip vibrational amplitude, thermal effects on tissue, voltage and current outputs.
- The infiltration subsystem was assessed for overall performance, and compatibility with commercially available cannulas and tubing.
- The aspiration subsystem testing was performed in accordance with ISO 10079-1 and ISO 10079-4.
{6}
Page 3 of 7 K244050
b. Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety testing was performed in accordance with IEC 60601-1-1. Electromagnetic Compatibility testing was performed in accordance with IEC 60601-1-2 and 60601-4-2. Wireless emissions testing was conducted for the wireless footswitch.
c. Software and Cybersecurity
Software and Cybersecurity Verification and Validation testing were performed in accordance with ISO 62304.
d. Biocompatibility Testing
Biological evaluation per ISO 10993-1 was conducted on the patient contacting materials (direct and indirect contacting).
e. Reprocessing
The reprocessing of sterile products was tested in accordance with ISO 17665.
5. Clinical Performance
No clinical studies were required to support the substantial equivalence of the device. These types of devices, including the predicate device, have been on the market for many years with proven safety and efficacy. The non-clinical testing detailed in this submission support the substantial equivalence of the device.
6. Substantial Equivalence
The Subject Device for this submission is the AYON Body Contouring System. The identified Predicate Device is the Sound Surgical VASER 2.1 Lipo System, manufactured by Solta Medical under K110306. A comparison between the Subject Device and the Predicate Device is provided in Table 1.
Table 1. Substantial Equivalence Table
| | AYON Body Contouring System
Subject Device
(This Submission) | VASER 2.1 Lipo System
Predicate Device
K110306 | Comments |
| --- | --- | --- | --- |
| Overall System | AYON Body Contouring System | VASER 2.1 Lipo System | |
| Components of the System | Includes an Ultrasound Sub-System | Includes the Sound Surgical VASER 2.1 ultrasound system | Identical |
| | Includes an Aspiration Sub-System | Includes an aspiration system as part of the Sound Surgical VentX 2.1 system | Identical |
{7}
Page 4 of 7
K244050
| | Includes an Infiltration Sub-System | Includes an infiltration system as part of the Sound Surgical VentX 2.1 system | Identical |
| --- | --- | --- | --- |
| Intended Use | **Overall System:**
The AYON Body Contouring System is intended for the fragmentation, emulsification, and aspiration of subcutaneous fatty tissue for the purpose of aesthetic body contouring. | **Overall System:**
The Sound Surgical VASER 2.1 Lipo System is intended for the fragmentation, emulsification and aspiration of subcutaneous fatty tissue for the purpose of aesthetic body contouring. | Identical |
| | **Overall System:** | **Overall System:**
The Sound Surgical VASER 2.1 Lipo System is also indicated for use in the following surgical specialties for the fragmentation, emulsification and aspiration of soft tissue:
• Neurosurgery;
• Gastrointestinal and affiliated organ surgery;
• Urological surgery;
• Plastic and reconstructive surgery;
• General surgery;
• Gynecological surgery;
• Thoracic surgery; and
• Laparoscopic surgery. | The Subject Device is not pursuing the expanded indications for additional surgical specialties included in the intended use for the Predicate Device. |
| | **Ultrasound Sub-System:**
The AYON Ultrasound Sub-System is used for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring. | **Ultrasound Sub-System:**
The Sound Surgical VASER 2.1 is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring. | Same intent and purpose of use with minor changes. |
| | **Infiltration and Aspiration Sub-Systems:**
The AYON Infiltration and Aspiration Sub-System(s) are used for the infiltration and suction / aspiration of fluids and tissue during surgical procedures. Infiltration and Suction is designed to operate as a stand-alone function. | **Infiltration and Aspiration Sub-Systems:**
The Sound Surgical VentX 2.1 is intended for the suction or aspiration of fluids and tissue during surgical procedures. The VentX 2.1 is designed to operate with the Sound Surgical VASER 2.1 or as a stand-alone system. | Same intent and purpose of use with minor changes for clarity. |
4
{8}
Page 5 of 7
K244050
| Product Code /Regulation | QPB, 21 CFR 878.5040 | QPB, 21 CFR 878.5040 | Identical |
| --- | --- | --- | --- |
| User | Used by trained Physicians in an operating environment. | Used by trained Physicians in an operating environment. | Identical |
| Mechanism of Action | The ultrasound system fragments and emulsifies subcutaneous fatty tissue. The infiltration and aspiration systems utilize sterile infiltration tubing, a handle and cannula to infiltrate the tissue with fluids, and sterile suction tubing, a handle and cannula to aspirate fluids and soft tissue. | The ultrasound system fragments and emulsifies subcutaneous fatty tissue. The infiltration and aspiration systems utilize sterile infiltration tubing, a handle and cannula to infiltrate the tissue with fluids, and sterile suction tubing, a handle and cannula to aspirate fluids and soft tissue. | Identical |
| User Interface/Controls | Single touchscreen utilizing a graphical user interface, on/off switches | Knobs and dials in various locations, on/off switches. | The single touchscreen of the Subject Device enables ease of use through control of the entire system in one centralized location. |
| **Ultrasound Sub-System** | | | |
| Voltage and Frequency | AC Powered
Voltage: 100-120 VAC
Frequency: 60/50 Hz | AC Powered
Voltage: 115C/230VAC
Frequency: 60/50 Hz | Similar |
| Electrical Safety | IEC Class I, Type BF | IEC Class I, Type BF | Identical |
| Method of Fragmentation and Emulsification | Ultrasonic | Ultrasonic | Identical |
| Operating Frequency | 37 kHz nominal | 36 kHz nominal | Similar |
| Ultrasound Handpiece Transducer Material | Piezoelectric crystal | Piezoelectric crystal | Identical |
| Ultrasound Probes | Types:
Diameters – 2.2mm to 3.7mm
Lengths – 18cm to 33cm | Types:
Diameters – 1.0mm to 6.0mm
Lengths – 7cm to 40cm | Similar |
| | Material:
Titanium alloy | Material:
Titanium alloy | Identical |
| Ultrasound Vibratory Power / Amplitude | Three different power or amplitude settings:
Low power = 8 watts
Medium power = 20 watts | Power or amplitude settings ranging from 10% to 100% | Similar |
5
{9}
Page 6 of 7
K244050
| | High power = 28 watts | | |
| --- | --- | --- | --- |
| Modes of Operation | **Standard Mode** - Delivers the ultrasonic energy in pulses of two different amplitudes. The pulses alternate between full amplitude for 50 milliseconds and an amplitude of 50% of the full amplitude for 50 milliseconds. | **VASER (Pulse) Mode** - Delivers the ultrasonic energy in pulses of two different amplitudes. The pulses alternate between full amplitude for 50 milliseconds and an amplitude of 50% of the full amplitude for 50 milliseconds. | Identical |
| | **Continuous Mode** – Delivers the ultrasonic energy continuously at full amplitude. | **Continuous Mode** – Delivers the ultrasonic energy continuously at full amplitude. | Identical |
| Aspiration Sub-System | | | |
| Power | AC Powered
Voltage: 120 VAC
Frequency: 50-60 Hz
Power: 950 VA | AC Powered
Voltage 120 VAC
Frequency 60 Hz
Power: 1000 VA | Similar |
| Electrical Safety | IEC Class I, Type BF | IEC Class I, Type BF | Identical |
| Suction Vacuum | 27 in Hg at sea level | 20 in Hg at 5000 ft. (~25 in Hg at sea level) | Similar^{1} |
| Infiltration Sub-System | | | |
| Power | AC Powered
Voltage: 120 VAC
Frequency: 50-60 Hz
Power: 950 VA | AC Powered
Voltage 120 VAC
Frequency 60 Hz
Power: 1000 VA | Similar |
| Electrical Safety | IEC Class I, Type BF | IEC Class I, Type BF | Identical |
| Infiltration Pump | Peristaltic | Peristaltic | Identical |
| Infiltration Rate | Adjustable flow rate from 40-1000mL/min | Adjustable flow rate from 50-550mL/min | Similar^{2} |
## 7. Substantial Equivalence Determination
The Subject and Predicate Devices have the same intended use, the minor clarifications do not change safety of the device or use of the device. The devices have identical mechanisms of action to produce the fragmentation and emulsification of subcutaneous fatty tissues. The technological features of the two devices are either identical or very similar in nature. Minor differences in the Ultrasound Sub-systems were found not to have an impact on the substantial equivalence of the two devices through comparative bench testing. Based on the information contained in this submission, it is concluded that the AYON Body Contouring System is substantially equivalent in function and intended use to the identified predicate device, the Sound Surgical VASER 2.1 Lipo System, manufactured by Solta Medical under
1 Suction vacuum for the Subject Device falls within the vacuum range of 0 to 29.9 in Hg included in the Product Code description and falls below the maximum vacuum range for other FDA cleared devices. Vitruvian Ultimate Aspirator (K161722), PSI-TEC Aspirator (K981215).
2 The infiltration flow rates for the Subject Device are identical or fall below the maximum flow rates for other FDA cleared infiltration and infusion pumps. Reference devices; Klein Surgical Infiltration pump (K031432) and the Wells Johnson Infusion pump (K991437).
{10}
Page 7 of 7 K244050
K110306.
## 8. Conclusion
The performance data presented in this submission, supports substantially equivalent performance between the subject device and its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.