K161722 · Black & Black Surgical, Inc. · QPB · Jul 26, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161722
Device Name
VITRUVIAN ULTIMATE ASPIRATOR
Applicant
Black & Black Surgical, Inc.
Product Code
QPB · General, Plastic Surgery
Decision Date
Jul 26, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Aesthetic body contouring.
Device Story
The Vitruvian Ultimate Aspirator is a suction lipoplasty system used for aesthetic body contouring. The device consists of a powder-coated steel housing containing a dual 1/2 hp rocking piston vacuum pump. It operates via a pressure control valve with a rotary knob, allowing the user to adjust vacuum levels. A dual-activated air-powered footswitch controls operation. The device includes a digital vacuum meter display with LED numerical and color graphic readouts. It utilizes a 0.3-micron bacterial filter to protect the system. The device is intended for use by clinicians in a surgical or clinical setting. The vacuum output is used to facilitate the removal of tissue during lipoplasty procedures. The device does not perform automated analysis or control the suction process via software; the software is limited to displaying vacuum levels.
Clinical Evidence
Bench testing only. The device was evaluated for compliance with IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), IEC 60601-1-6 (usability), and IEC 62366 (usability). Internal calibration testing confirmed vacuum display accuracy within acceptable criteria.
Technological Characteristics
Powder-coated steel housing; dual 1/2 hp rocking piston motor; 120/230V 50/60Hz power; 0.3-micron bacterial filter; rotary pressure control valve; LED numerical and color graphic display; dual-activated air-powered footswitch. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, and IEC 62366.
Indications for Use
Indicated for aesthetic body contouring.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is in blue and consists of the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". To the left of the FDA logo is the Department of Health & Human Services logo.
Black & Black Surgical, Inc. % Mark Job Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
June 9, 2021
Re: K161722
Trade/Device Name: Vitruvian Ultimate Aspirator Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: QPB
Dear Mark Job:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated July 26, 2016. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Cindy Chowdhury, OHT4: Office of Surgical and Infection Control Devices, 240-402-6647, Cindy.Chowdhury@fda.hhs.gov.
Sincerely,
Cindy Chowdhury -S
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 26, 2016
Black & Black Surgical, Inc. Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313
Re: K161722
Trade/Device Name: Vitruvian Ultimate Aspirator Regulation Number: 21 CFR 878.5040 Regulation Name: Suction Lipoplasty System Regulatory Class: Class II Product Code: MUU Dated: July 18, 2016 Received: July 19, 2016
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161722
Device Name
Black & Black Vitruvian Ultimate Aspirator
Indications for Use (Describe) Aesthetic body contouring.
**Type of Use (Select one or both, as applicable)**
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness is being submitterd in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
#### Submitter Information: 1.
Prepared: June 15, 2016
Est. Reg. #: 3006142527
Norman M. Black. CEO. normanblack@blackandblacksurgical.com Black & Black Surgical. Inc. 5238 Royal Woods Parkway Suite 170 Tucker, GA 30084 770-414-4880 T / 770-414-4879 F
Contact Person: Cynthia Rees, Regulatory Engineer, crees@blackandblacksurgical.com
- 2. Name of Device: Vitruvian Ultimate Aspirator
Proprietary Name: Vitruvian Ultimate Aspirator
Common Name: Aspirator Pump, Liposuction Aspirator, Power Suction Pump
- 3. Classification: Suction Lipoplasty System, Class II - 21 CFR §878.5040 (1998)
- 4. Product Code: MUU
#### 5. Substantial Equivalence:
The Vitruvian Ultimate Aspirator is substantially equivalent to the aspiration device listed below in terms of use, design, operating principles, materials and performance.
Byron Medical: K981215
#### 6. Device Description:
The Vitruvian Aspirator is a pre-assembled device consisting of:
- . Housing: Powder Coated Steel
- . Vacuum Pump: Dual ½ hp dual voltage rocking piston motor with Maximum Suction flow rate of 125PSI, 6.4 cfm, 27.5in HG(711 mm of Hg)
- . Vacuum Meter Display: The device displays the vacuum levels digitally. The software does not control the unit in any way and does not accept any input.
- . Power Operations: 120/230 volt 50/60hz
- . Control: Pressure control valve located on the face of the device with rotary control knob with both LED numerical and Color Graphic LED display
- . Footswitch: Dual activated air powered
- IV Pole .
- . Filter: Bacterial filter having a pore size of 0.3 microns or less.
- 7. Indications for Use: Aesthetic body contouring.
{5}------------------------------------------------
#### 8. Comparison to Predicate
| Item | Predicate Device | Subject Device | Similarity | |
|-----------------------------------------|-----------------------------------------------|------------------------------------------------------|------------|--|
| Device name | PSI-TEC Aspirator | Vitruvian Ultimate Aspirator | N/A | |
| Device<br>Manufacturer | Byron Medical. Inc. | Black & Black Surgical, Inc. | N/A | |
| 510(k) Reference | K981215 | K161722 | N/A | |
| FDA Product Code | MUU | MUU | Same | |
| FDA Classification | Suction Lipoplasty System | Suction Lipoplasty System | Same | |
| FDA Regulation # | 878.5040 | 878.5040 | Same | |
| Indication for Use /<br>Intended Use | Aesthetic body contouring | Aesthetic body contouring | Same | |
| Mechanical Operation | | | | |
| Item | Predicate Device | Subject Device | Similarity | |
| Pump type | Two cylinder piston | Two cylinder piston | Same | |
| AC Power input | 115-240V, 60-50Hz | 120V/230V, 50/60Hz | Similar | |
| Maximum Vacuum | 28 in of Hg | 27.5 in of Hg | Similar | |
| Maximum flow rate | 4.75 CFM | 6.4 CFM | Similar | |
| Sound level | 52 dB | 50 dB | Similar | |
| Pressure control | Knob to control vacuum | Knob to control vacuum | Same | |
| Footswitch | Dual activated air powered | Dual activated air powered | Same | |
| Display | Vacuum gage | Vacuum gage w/LED color graphic read out | Different | |
| Safety<br>Features<br>(over flow valve) | Manual (canister) | Manual (canister) | Same | |
| Housing | Powder Coated Steel | Powder Coated Steel | Same | |
| IV Pole for<br>Solutions: | Yes | Yes | Same | |
| Filter port: | Yes | Yes | Same | |
| Filter | 0.3 Micron | 0.3 Micron | Same | |
| Dimensions | 20.18 in (H) x 22.96 in (D) x 18.32<br>in (W) | 18 in (H) x18 in (W) x38 in (H)(45cm x45cm<br>x97cm) | Similar | |
| Weight | 80 lbs. /36.288 Kilograms | 100 lbs | Similar | |
The Black & Black suction aspirator is substantially equivalent in function and intended use to the Byron PSI-TEC (K981215). Both the Byron PSI-TEC and the Black Vitruvian Aspirator have Dual vacuum motors, a Vacuum gage read-out, metal housing, 4 casters, an IV pole, fitting for suction tubing, and a 0.3 micron filter. Both devices are used for aesthetic body contouring.
The steps of operation of the Vitruvian Aspirator can be found in the IFU.
{6}------------------------------------------------
# 9. Performance Testing
The device complies with the following standards:
| Tested/Evaluated Standard | Results |
|------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| IEC/EN 60601-1-<br>2:2007/AC:2010<br>Clause 4.1 | Complies |
| IEC/EN 60601-1-<br>2:2007/AC:2010<br>Clause 5 | Complies |
| IEC 60601-1:2005 (Third<br>Edition) + CORR. 1:2006 +<br>CORR. 2:2007 + A1:2012 (or<br>IEC 60601-1: 2012 reprint) | All Requirements/Tests<br>either Passed or were<br>N/A |
| IEC 60601-1-6:2010 (Third<br>Edition) | All Requirements/Tests<br>either Passed or were<br>N/A |
| IEC 62366: 2007 (First<br>Edition) | All Principles Evaluated<br>either Passed or were<br>N/A |
Performance Testing Performed at Manufacturing Facility:
- Vitruvian Ultimate Aspirator Vacuum Display Calibration All test results are within acceptable criteria range
### 10. Substantial Equivalence Conclusion:
Based on the information contained in this submission, it is concluded that the Vitruvian Ultimate Aspirator is substantially equivalent in function and intended use to the identified predicate device, the Byron PSI-TEC Aspirator, K981215.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.