Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate and Xylazine
K244034 · Ever Global (Vietnam) Enterprise Corporation · LZA · Feb 6, 2025 · General Hospital
Device Facts
Record ID
K244034
Device Name
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate and Xylazine
Applicant
Ever Global (Vietnam) Enterprise Corporation
Product Code
LZA · General Hospital
Decision Date
Feb 6, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs, fentanyl citrate, and xylazine in accordance with ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical gloves to Permeation by Chemotherapy Drugs.
Device Story
Disposable, non-sterile, powder-free nitrile examination glove; blue color. Provides single-use barrier protection for examiner's hands/fingers to prevent cross-contamination. Tested for permeation resistance against 57 chemotherapy/hazardous drugs, fentanyl citrate, and xylazine per ASTM D6978-05. Used in clinical settings by healthcare professionals. Output is physical barrier protection; affects clinical decision-making by enabling safe handling of hazardous substances. Benefits patient/clinician by reducing exposure to hazardous drugs and contaminants.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM D6319-19 (dimensions, physical properties, freedom from holes, powder residual) and ASTM D6978-05 (permeation resistance to 57 hazardous drugs). Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
Nitrile compound; non-sterile; powder-free. Dimensions and physical properties comply with ASTM D6319-19. Permeation resistance tested per ASTM D6978-05. Biocompatibility per ISO 10993-5/10. No software or energy source.
Indications for Use
Indicated for use as a disposable patient examination glove worn on the examiner's hands or fingers to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs, fentanyl citrate, and xylazine.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate (K193555)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 6, 2025
Ever Global (Vietnam) Enterprise Corporation Elizabeth Deng U.S. Agent Long Thanh Industrial Zone Taman Village, Dong Nai Province 810000 Vietnam
Re: K244034
Trade/Device Name: Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate and Xylazine Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, ODO, OPJ Dated: December 27, 2024 Received: December 30, 2024
Dear Elizabeth Deng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/3 description: The image shows the text "ALLAN GUAN -S" in a large, bold, sans-serif font. The text is black and the background is white. The letters are evenly spaced and the text is centered.
For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K244034
#### Device Name
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate and Xylazine
#### Indications for Use (Describe)
A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs, fentanyl citrate, and xylazine in accordance with ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical gloves to Permeation by Chemotherapy Drugs.
| Tested Chemotherapy and | Concentration(mg/ml) | Minimum Breakthrough Detection Time(Min) |
|------------------------------|----------------------|------------------------------------------|
| Other Liquid Hazardous Drug | | |
| 01 Arsenic Trioxide | 1.0 | > 240 |
| 02 Azacitidine (Vidaza) | 25.0 | > 240 |
| 03 Bendamustine HCl | 5.0 | > 240 |
| 04 Bleomycin Sulfate | 15.0 | > 240 |
| 05 Bortezomib (Velcade) | 1.0 | > 240 |
| 06 Busulfan | 6.0 | > 240 |
| 07 Carboplatin | 10.0 | > 240 |
| 08 Carfilzomib | 2.0 | > 240 |
| 09 Carmustine (BCNU) | 3.3 | 21.5 |
| 10 Cetuximab (Erbitux) | 2.0 | > 240 |
| 11 Chloroquine | 50.0 | > 240 |
| 12 Cisplatin | 1.0 | > 240 |
| 13 Cladribine | 1.0 | > 240 |
| 14 Cyclophosphamide | 20.0 | > 240 |
| 15 Cyclosporine A | 100.0 | > 240 |
| 16 Cytarabine | 100.0 | > 240 |
| 17 Cytovene (Ganciclovir) | 10.0 | > 240 |
| 18 Dacarbazine | 10.0 | > 240 |
| 19 Daunorubicin | 5.0 | > 240 |
| 20 Decitabine | 5.0 | > 240 |
| 21 Docetaxel | 10.0 | > 240 |
| 22 Doxorubicin Hydrochloride | 2.0 | > 240 |
| 23 Epirubicin (Ellence) | 2.0 | > 240 |
| 24 Etoposide | 20.0 | > 240 |
| 25 Fludarabine | 25.0 | > 240 |
| 26 Fluorouracil | 50.0 | > 240 |
| 27 Fulvestrant | 50.0 | > 240 |
| 28 Gemcitabine | 38.0 | > 240 |
| 29 Idarubicin | 1.0 | > 240 |
| 30 Ifosfamide | 50.0 | > 240 |
| 31 Irinotecan | 20.0 | > 240 |
| 32 Mechlorethamine HCl | 1.0 | > 240 |
| 33 Melphalan | 5.0 | > 240 |
| 34 Methotrexate | 25.0 | > 240 |
| 35 Mesna | 100.0 | > 240 |
| 36 Mitomycin C | 0.5 | > 240 |
| 37 Mitoxantrone | 2.0 | > 240 |
| 38 Oxaliplatin | 5.0 | > 240 |
| 39 Paclitaxel | 6.0 | > 240 |
| 40 Paraplatin | 10.0 | > 240 |
| 41 Pemetrexed | 25.0 | > 240 |
| 42 Pertuzumab | 30.0 | > 240 |
| 43 Raltitrexed | 0.5 | > 240 |
| 44 Retrovir | 10.0 | > 240 |
| 45 Rituximab | 10.0 | > 240 |
| 46 Temsirolimus | 25.0 | > 240 |
| 47 Thiotepa | 10.0 | 13.6 |
| 48 Topotecan HCl | 1.0 | > 240 |
| 49 Trastuzumab | 21.0 | > 240 |
| 50 Triclosan | 2.0 | > 240 |
| 51 Trisenox | 1.0 | > 240 |
| 52 Vinblastine | 1.0 | > 240 |
| 53 Vincristine Sulfate | 1.0 | > 240 |
| 54 Vinorelbine | 10.0 | > 240 |
| 55 Zoledronic Acid | 0.8 | > 240 |
| 56 Fentanyl Citrate | 100.0 mcg/ 2.0 ml | > 240 |
FORM FDA 3881 (8/23)
{4}------------------------------------------------
### Warning: do not use with Carmustine and Thiotepa.
The maximum testing time is 240 minutes. Please note that the following drug has an extremely low permeation time:
| Carmustine (BCNU) | 3.3 mg/ml | 21.5 minutes |
|-------------------|------------|--------------|
| Thiotepa | 10.0 mg/ml | 13.6 minutes |
Xylazine Permeation Resistance Claim - Under the testing conditions of ASTM D6978-05, Xylazine HCl Injection (100.0 mg/ml) was found to have no breakthrough detected for up to 240 minutes.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'EG' in a stylized font, with the 'E' in yellow and the 'G' in yellow. The letters are set against a blue background that resembles a hand. The words 'EVER GLOBAL' are at the top of the logo, and the words 'ENTERPRISE CORP' are at the bottom.
USTRIAL ZONE, TAMAN VILLAGE, DONG (0251)351 4023
# 510(k) SUMMARY
Elizabeth Deng
909 4659188
Xylazine
Class I
K244034
21 CFR 880.6250
LZA, LZC, QDO, OPJ
5748 Eaglewood Place
Rancho Cucamonga, California Rancho Cucamonga, CA 91739
baxianunited48(@yahoo.com
And Fentanyl Citrate And Xylazine
#### 0.0 Summary Preparation Date: February 5, 2025
## 1.0 Submitter:
Submitter's name: Submitter's address:
Phone number: Fax number: Name of contact person:
Ever Global (Vietnam) Enterprise Corporation Long Thanh Industrial Zone, Taman Village, Dong Nai Province, Long Thanh District, VN 810000 84-61-3514022 84-61-3514023 Jerry Lin
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs
Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs And Fentanyl Citrate And
Non-powdered Patient Examination Glove
### 2.0 US Agent:
US Representative Name: Company Address:
Telephone Number: Contact Email Address:
### 3.0 Name of the Device
Proprietary/Trade name:
Common Name:
Classification Name: Device Classification: Regulation Number: Product Code: 510(K) Number:
## 4.0 Predicate device
Device Name:
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate Ever Global (Vietnam) Enterprise Corporation K193555
Company name: 510(K) Number:
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features a blue square with a stylized yellow handprint in the background. The letters "E" and "G" are overlaid in yellow on top of the handprint. The words "EVER GLOBAL" are at the top of the square, and "ENTERPRISE CORP." are at the bottom.
ONG THANH INDUSTRIAL ZONE, TAMAN VILLAGE, DONG AX: (0251)351 4023 02-
SUPERIEURVIETNAM
# 510(k) SUMMARY
## 5.0 Device Description:
"Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate And Xylazine" is a patient examination glove made from nitrile compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of this medical device is to provide single use barrier protection for the wearer and the device meets the specifications for Barrier Protection and tensile properties as defined in ASTM D6319-19, Standard specification for Nitrile Examination Gloves.
### 6.0 Device Indications for use:
A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs, fentanyl citrate, and xylazine in accordance with ASTM D6978-05 standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs.
| No. | Tested Chemotherapy and Other Liquid<br>Hazardous Drug and Concentration | Minimum Breakthrough<br>Detection Time (Min.) |
|-----|--------------------------------------------------------------------------|-----------------------------------------------|
| 1 | Arsenic Trioxide (1.0 mg/ml) | >240 |
| 2 | Azacitidine (Vidaza) (25.0 mg/ml) | >240 |
| 3 | Bendamustine HCl (5.0 mg/ml) | >240 |
| 4 | Bleomycin Sulfate (15.0 mg/ml) | >240 |
| 5 | Bortezomib (Velcade) (1.0 mg/ml) | >240 |
| 6 | Busulfan (6.0 mg/ml) | >240 |
| 7 | Carboplatin (10.0 mg/ml) | >240 |
| 8 | Carfilzomib (2.0 mg/ml) | >240 |
| 9 | Carmustine (BCNU), (3.3 mg/ml) | 21.5 |
| 10 | Cetuximab (Erbitux) (2.0 mg/ml) | >240 |
| 11 | Chloroquine (50.0 mg/ml) | >240 |
| 12 | Cisplatin (1.0 mg/ml) | >240 |
| 13 | Cladribine (1.0 mg/ml) | >240 |
| 14 | Cyclophosphamide (20.0 mg/ml) | >240 |
| 15 | Cyclosporin A (100.0 mg/ml) | >240 |
| 16 | Cytarabine (100.0 mg/ml) | >240 |
| 17 | Cytovene (Ganciclovir) (10.0 mg/ml) | >240 |
| 18 | Dacarbazine (DTIC), (10.0 mg/ml) | >240 |
| 19 | Daunorubicin (5.0 mg/ml) | >240 |
| 20 | Decitabine (5.0 mg/ml) | >240 |
| 21 | Docetaxel (10.0 mg/ml) | >240 |
| 22 | Doxorubicin Hydrochloride (2.0mg/ml) | >240 |
| 23 | Epirubicin (Ellence) (2.0 mg/ml) | >240 |
| 24 | Etoposide, (20.0 mg/ml) | >240 |
| 25 | Fludarabine (25.0 mg/ml) | >240 |
| 26 | Fluorouracil, (50.0 mg/ml) | >240 |
| 27 | Fulvestrant (50.0 mg/ml) | >240 |
| 28 | Gemcitabine (38.0 mg/ml) | >240 |
| 29 | Idaribicin (1.0 mg/ml) | >240 |
| 30 | Ifosfamide (50.0 mg/ml) | >240 |
| 31 | Irinotecan (20.0 mg/ml) | >240 |
| 32 | Mechlorethamine HCl (1.0 mg/ml) | >240 |
| 33 | Melphalan (5.0 mg/ml) | >240 |
| 34 | Methotrexate (25 mg/ml) | >240 |
| 35 | Mesna (100 mg/ml) | >240 |
| 36 | Mitomycin C (0.5 mg/ml) | >240 |
| 37 | Mitoxantrone (2.0 mg/ml) | >240 |
| 38 | Oxaliplatin (5.0 mg/ml) | >240 |
| 39 | Paclitaxel (Taxol), (6.0 mg/ml) | >240 |
| 40 | Paraplatin (10.0 mg/ml) | >240 |
| 41 | Pemetrexed (25.0 mg/ml) | >240 |
| 42 | Pertuzumab (30.0 mg/ml) | >240 |
| 43 | Raltitrexed (0.5 mg/ml) | >240 |
| 44 | Retrovir (10.0 mg/ml) | >240 |
| 45 | Rituximab (10.0 mg/ml) | >240 |
| 46 | Temsirolimus (25.0 mg/ml) | >240 |
| 47 | Thiotepa (10.0 mg/ml) | 13.6 |
| 48 | Topotecan HCl (1.0 mg/ml) | >240 |
| 49 | Trastuzumab (21.0 mg/ml) | >240 |
| 50 | Triclosan (2.0 mg/ml) | >240 |
| 51 | Trisenox (1.0 mg/ml) | >240 |
| 52 | Vinblastine (1.0 mg/ml) | >240 |
| 53 | Vincristine Sulfate (1.0 mg/ml) | >240 |
| 54 | Vinorelbine (10.0 mg/ml) | >240 |
| 55 | Zoledronic Acid (0.8 mg/ml) | >240 |
| 56 | Fentanyl Citrate Injection (100.0mcg/2.0ml) | >240 |
Table 1 – Tested for use with 57 chemotherapy and other liguid hazardous drugs
Premarket Notification 510(k)
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate and Xylazine
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'EG' in a stylized, interconnected design, with the 'E' in a larger, bolder font than the 'G'. The letters are colored in yellow and are set against a blue background that resembles the shape of a hand. The words 'EVER GLOBAL' are written above the letters, and 'ENTERPRISE CORP.' is written below.
Long Thanh Industrial Zone, Taman Village, Dong NAI 810000 VIETNAM TEL: (0251) 896 6676 FAX: (0251)351 4023 024 EMAIL: INFO@EGVNC0.COM
#### FACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM
# 510(k) SUMMARY
Premarket Notification 510(k)
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'E', 'G', and 'C' in a stylized, overlapping design. The letters are in yellow and are set against a blue background. The words 'EVER GLOBAL' are at the top of the logo, and 'ENTERPRISE CORP.' is at the bottom.
RIAL ZONE, TAMAN VILLAGE, DONG (0251)351 402
# 510(k) SUMMARY
Warning: do not use with Carmustine and Thiotepa.
The maximum testing time is 240 minutes. Please note that the following drug has an extremely low permeation time:
| Carmustine (BCNU), 3.3 mg/ml | 21.5 minutes |
|------------------------------|--------------|
| Thiotepa, 10.0 mg/ml | 13.6 minutes |
Xylazine Permeation Resistance Claim - Under the testing conditions of ASTM D6978-05, Xylazine HCl Injection (100.0 mg/ml) was found to have no breakthrough detected for up to 240 minutes.
# 7.0 Comparison of device technological characteristics:
| Device<br>Characteristic | Predicate Device | Subject Device | Comparison |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product Name | Disposable Powder Free Nitrile<br>Examination Glove, Blue Color,<br>Tested For Use With<br>Chemotherapy Drugs and<br>Fentanyl Citrate | Disposable Powder Free Nitrile<br>Examination Glove, Blue Color,<br>Tested For Use With<br>Chemotherapy Drugs and<br>Fentanyl Citrate and Xylazine | similar |
| 510(K) No. | K193555 | K244034 | n/a |
| Product Owner | Ever Global (Vietnam) Enterprise<br>Corporation | Ever Global (Vietnam) Enterprise<br>Corporation | same |
| Product Code | LZA, LZC, QDO | LZA, LZC, QDO, OPJ | same |
| Regulation | 21 CFR 880.6250 | 21 CFR 880.6250 | same |
| Class | I | I | same |
| Indications for<br>Use | The nitrile powder free patient<br>examination glove is a non-sterile<br>disposable device intended for<br>medical purposes that is worn on<br>the examiner's hands or finger to<br>prevent contamination between<br>patient and examiner. In addition. | The nitrile powder free patient<br>examination glove is a non-sterile<br>disposable device intended for<br>medical purposes that is worn on<br>the examiner's hands or finger to<br>prevent contamination between<br>patient and examiner. In addition. | similar |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'FG' in yellow, with the 'F' being larger than the 'G'. The letters are set against a blue background that resembles a hand. The words 'EVER GLOBAL' are at the top of the logo, and 'ENTERPRISE CORP.' is at the bottom.
Long Thanh Industrial Zone, Taman Village, Dong NAI 810000 VIETNAM TEL: (0251) 896 6676 FAX: (0251)351 4023 024 EMAIL: INFO@EGVNC0.COM FACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM
# 510(k) SUMMARY
| | these gloves were tested for use<br>with chemotherapy drugs and<br>fentanyl citrate in accordance<br>with ASTM D6978-05 standard<br>practice for assessment of<br>medical gloves to permeation by<br>chemotherapy drugs. | these gloves were tested for use<br>with chemotherapy drugs and<br>fentanyl citrate and xylazine in<br>accordance with ASTM D6978-<br>05 standard practice for<br>assessment of medical gloves to<br>permeation by chemotherapy<br>drugs. | | | | | |
|----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|---------|----------|-----------------------------------------------------------------------------------|-------------------------------------------------------------|
| Power free | Yes | Yes | same | | | | |
| Size | XS/S/M/L/XL | XS/S/M/L/XL | same | | | | |
| Over-The-<br>Counter Use | Yes (21 CFR 801 Subpart C) | Yes (21 CFR 801 Subpart C) | same | | | | |
| Single Use | Yes | Yes | same | | | | |
| Non-Sterile | Yes | Yes | same | | | | |
| Shelf life | No | No | same | | | | |
| Dimensions-<br>Length | Complies with ASTM D6319-10<br>Short cuff ≥230mm<br>Long cuff ≥300mm | Complies with ASTM D6319-19<br>Short cuff ≥230mm<br>Long cuff ≥300mm | same | | | | |
| Dimensions -<br>Palm Width | Complies with ASTM D6319-10<br>XS: 70 ± 10 mm;<br>S: 80 ± 10 mm;<br>M: 95 ± 10 mm;<br>L: 110 ± 10 mm;<br>XL: 120 ± 10 mm | Complies with ASTM D6319-19<br>XS: 70 ± 10 mm;<br>S: 80 ± 10 mm;<br>M: 95 ± 10 mm;<br>L: 110 ± 10 mm:<br>XL: 120 ± 10 mm | same | | | | |
| Dimensions -<br>Thickness | Complies with ASTM D6319-10<br>Palm ≥0.05 mm;<br>Finger ≥0.05 mm<br>Cuff ≥ 0.05 mm | Complies with ASTM D6319-19<br>Palm ≥0.05 mm;<br>Finger ≥0.05 mm<br>Cuff ≥ 0.05 mm | same | | | | |
| Physical<br>Properties | Complies with ASTM D6319-10<br>Tensile Strength<br>Before aging≥14MPa<br>After aging ≥14MPa<br>Complies with ASTM D6319-10<br>Elongation:<br>Before Aging: ≥500%<br>After Aging: ≥400% | Complies with ASTM D6319-19<br>Tensile Strength<br>Before aging≥14MPa<br>After aging ≥14MPa<br>Complies with ASTM D6319-19<br>Elongation:<br>Before Aging: ≥ 500%<br>After Aging: ≥400% | same | | | | |
| Residual<br>powder | Complies with ASTM D6319-10<br>< 2mg per glove | Complies with ASTM D6319-19<br>< 2mg per glove | same | | | | |
| Freedom from<br>Holes | In accordance with ASTM<br>D6319-10 (G-1 with AQL 2.5) | In accordance with ASTM<br>D6319-19 (G-1 with AQL 2.5) | same | | | | |
| Biocompatibility | AAMI/ANSI/ISO 10993-10 &<br>AAMI/ANSI/ISO 10993-5 Passes<br>Not a skin irritant, Not a skin<br>sensitizer & No cytotoxicity<br>reaction | AAMI/ANSI/ISO 10993-10 &<br>AAMI/ANSI/ISO 10993-5 Passes<br>Not a skin irritant, Not a skin<br>sensitizer & No cytotoxicity<br>reaction | same | | | | |
| Resistance of<br>Gloves to<br>Permeation by<br>Chemotherapy<br>Drugs | no.<br>1 | Test Chemotherapy<br>Drug and<br>Concentration<br>Arsenic Trioxide<br>(1.0 mg/ml) | Minimum<br>Breakthrough<br>Detection<br>Time (Min.)<br>>240 | similar | no.<br>1 | Test Chemotherapy<br>Drug and<br>Concentration<br>Arsenic Trioxide<br>(1.0 mg/ml) | Minimum<br>Breakthrough<br>Detection<br>Time (Min.)<br>>240 |
Premarket Notification 510(k)
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate and Xylazine
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'EG' in a stylized, golden font, superimposed over a blue background that resembles a hand. The words 'EVER GLOBAL' are printed above the letters, and 'ENTERPRISE CORP.' is printed below.
LONG THANH INDUSTRIAL ZONE, TAMAN VILLAGE, DONG NAI 810000 VIETNAM FAX: (0251)351 4023 024 TEL: (0251) 896 6676 FMAIL . INFO@EGVNC0 COM
FACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM
# 510(k) SUMMARY
| 2 | Azacitidine (Vidaza)<br>(25.0 mg/ml) | >240 | 2 | Azacitidine (Vidaza)<br>(25.0 mg/ml) | >240 |
|----|-------------------------------------------------|------|----|-------------------------------------------------|------|
| 3 | Bendamustine HCl<br>(5.0 mg/ml) | >240 | 3 | Bendamustine HCl<br>(5.0 mg/ml) | >240 |
| 4 | Bleomycin Sulfate<br>(15.0 mg/ml) | >240 | 4 | Bleomycin Sulfate<br>(15.0 mg/ml) | >240 |
| 5 | Bortezomib<br>(Velcade) (1.0<br>mg/ml) | >240 | 5 | Bortezomib<br>(Velcade) (1.0<br>mg/ml) | >240 |
| 6 | Busulfan (6.0<br>mg/ml) | >240 | 6 | Busulfan (6.0<br>mg/ml) | >240 |
| 7 | Carboplatin (10.0<br>mg/ml) | >240 | 7 | Carboplatin (10.0<br>mg/ml) | >240 |
| 8 | Carfilzomib (2.0<br>mg/ml) | >240 | 8 | Carfilzomib (2.0<br>mg/ml) | >240 |
| 9 | Carmustine<br>(BCNU), (3.3<br>mg/ml) | 6.2 | 9 | Carmustine<br>(BCNU), (3.3<br>mg/ml) | 21.5 |
| 10 | Cetuximab (Erbitux)<br>(2.0 mg/ml) | >240 | 10 | Cetuximab (Erbitux)<br>(2.0 mg/ml) | >240 |
| 11 | Chloroquine (50.0<br>mg/ml) | >240 | 11 | Chloroquine (50.0<br>mg/ml) | >240 |
| 12 | Cisplatin (1.0<br>mg/ml) | >240 | 12 | Cisplatin (1.0<br>mg/ml) | >240 |
| 13 | Cladribine (1.0<br>mg/ml) | >240 | 13 | Cladribine (1.0<br>mg/ml) | >240 |
| 14 | Cyclophosphamide<br>(20.0 mg/ml) | >240 | 14 | Cyclophosphamide<br>(20.0 mg/ml) | >240 |
| 15 | Cyclosporin A<br>(100.0 mg/ml) | >240 | 15 | Cyclosporin A<br>(100.0 mg/ml) | >240 |
| 16 | Cytarabine (100.0<br>mg/ml) | >240 | 16 | Cytarabine (100.0<br>mg/ml) | >240 |
| 17 | Cytovene<br>(Ganciclovir) (10.0<br>mg/ml) | >240 | 17 | Cytovene<br>(Ganciclovir) (10.0<br>mg/ml) | >240 |
| 18 | Dacarbazine<br>(DTIC), (10.0<br>mg/ml) | >240 | 18 | Dacarbazine<br>(DTIC), (10.0<br>mg/ml) | >240 |
| 19 | Daunorubicin (5.0<br>mg/ml) | >240 | 19 | Daunorubicin (5.0<br>mg/ml) | >240 |
| 20 | Decitabine (5.0<br>mg/ml) | >240 | 20 | Decitabine (5.0<br>mg/ml) | >240 |
| 21 | Docetaxel (10.0<br>mg/ml) | >240 | 21 | Docetaxel (10.0<br>mg/ml) | >240 |
| 22 | Doxorubicin<br>Hydrochloride<br>(2.0mg/ml) | >240 | 22 | Doxorubicin<br>Hydrochloride<br>(2.0mg/ml) | >240 |
| 23 | Epirubicin (Ellence)<br>(2.0 mg/ml) | >240 | 23 | Epirubicin (Ellence)<br>(2.0 mg/ml) | >240 |
| 24 | Etoposide, (20.0<br>mg/ml) | >240 | 24 | Etoposide, (20.0<br>mg/ml) | >240 |
| 25 | Fludarabine (25.0<br>mg/ml) | >240 | 25 | Fludarabine (25.0<br>mg/ml) | >240 |
| 26 | Fluorouracil, (50.0<br>mg/ml) | >240 | 26 | Fluorouracil, (50.0<br>mg/ml) | >240 |
| 27 | Fulvestrant (50.0<br>mg/ml) | >240 | 27 | Fulvestrant (50.0<br>mg/ml) | >240 |
| 28 | Gemcitabine (38.0<br>mg/ml) | >240 | 28 | Gemcitabine (38.0<br>mg/ml) | >240 |
| 29 | Idaribicin (1.0<br>mg/ml) | >240 | 29 | Idaribicin (1.0<br>mg/ml) | >240 |
| 30 | Ifosfamide (50.0<br>mg/ml) | >240 | 30 | Ifosfamide (50.0<br>mg/ml) | >240 |
| 31 | Irinotecan (20.0<br>mg/ml) | >240 | 31 | Irinotecan (20.0<br>mg/ml) | >240 |
| 32 | Mechlorethamine<br>HCl (1.0 mg/ml) | >240 | 32 | Mechlorethamine<br>HCl (1.0 mg/ml) | >240 |
| 33 | Melphalan (5.0<br>mg/ml) | >240 | 33 | Melphalan (5.0<br>mg/ml) | >240 |
| 34 | Methotrexate (25<br>mg/ml) | >240 | 34 | Methotrexate (25<br>mg/ml) | >240 |
| 35 | Mesna (100 mg/ml) | >240 | 35 | Mesna (100 mg/ml) | >240 |
| 36 | Mitomycin C (0.5<br>mg/ml) | >240 | 36 | Mitomycin C (0.5<br>mg/ml) | >240 |
| 37 | Mitoxantrone (2.0<br>mg/ml) | >240 | 37 | Mitoxantrone (2.0<br>mg/ml) | >240 |
| 38 | Oxaliplatin (5.0<br>mg/ml) | >240 | 38 | Oxaliplatin (5.0<br>mg/ml) | >240 |
| 39 | Paclitaxel (Taxol),<br>(6.0 mg/ml) | >240 | 39 | Paclitaxel (Taxol),<br>(6.0 mg/ml) | >240 |
| 40 | Paraplatin (10.0<br>mg/ml) | >240 | 40 | Paraplatin (10.0<br>mg/ml) | >240 |
| 41 | Pemetrexed (25.0<br>mg/ml) | >240 | 41 | Pemetrexed (25.0<br>mg/ml) | >240 |
| 42 | Pertuzumab (30.0<br>mg/ml) | >240 | 42 | Pertuzumab (30.0<br>mg/ml) | >240 |
| 43 | Raltitrexed (0.5<br>mg/ml) | >240 | 43 | Raltitrexed (0.5<br>mg/ml) | >240 |
| 44 | Retrovir (10.0<br>mg/ml) | >240 | 44 | Retrovir (10.0<br>mg/ml) | >240 |
| 45 | Rituximab (10.0<br>mg/ml) | >240 | 45 | Rituximab (10.0<br>mg/ml) | >240 |
| 46 | Temsirolimus (25.0<br>mg/ml) | >240 | 46 | Temsirolimus (25.0<br>mg/ml) | >240 |
| 47 | Thiotepa (10.0<br>mg/ml) | 13.6 | 47 | Thiotepa (10.0<br>mg/ml) | 13.6 |
| 48 | Topotecan HCl (1.0<br>mg/ml) | >240 | 48 | Topotecan HCl (1.0<br>mg/ml) | >240 |
| 49 | Trastuzumab (21.0<br>mg/ml) | >240 | 49 | Trastuzumab (21.0<br>mg/ml) | >240 |
| 50 | Triclosan (2.0<br>mg/ml) | >240 | 50 | Triclosan (2.0<br>mg/ml) | >240 |
| 51 | Trisenox (1.0<br>mg/ml) | >240 | 51 | Trisenox (1.0<br>mg/ml) | >240 |
| 52 | Vinblastine (1.0<br>mg/ml) | >240 | 52 | Vinblastine (1.0<br>mg/ml) | >240 |
| 53 | Vincristine Sulfate<br>(1.0 mg/ml) | >240 | 53 | Vincristine Sulfate<br>(1.0 mg/ml) | >240 |
| 54 | Vinorelbine (10.0<br>mg/ml) | >240 | 54 | Vinorelbine (10.0<br>mg/ml) | >240 |
| 55 | Zoledronic Acid (0.8<br>mg/ml) | >240 | 55 | Zoledronic Acid (0.8<br>mg/ml) | >240 |
| 56 | Fentanyl Citrate<br>Injection, (100<br>mcg/2ml) | >240 | 56 | Fentanyl Citrate<br>Injection, (100<br>mcg/2ml) | >240 |
| | | | 57 | Xylazine HCl (100<br>mg/ml) | >240 |
Premarket Notification 510(k) Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate and Xylazine
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features a large, stylized "EG" in yellow, set against a blue background that resembles a hand. The words "EVER GLOBAL" are at the top of the logo, and "ENTERPRISE CORP." is at the bottom, both in a smaller font size.
LONG THANH INDUSTRIAL ZONE, TAMAN VILLAGE, DONG NAI 810000 VIETNAM FAX: (0251)351 4023 024 TEL: (0251) 896 6676 MAIL . INFO@EGVNCO COM
FACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM
# 510(k) SUMMARY
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'EG' in a stylized, interconnected design, with the 'E' being larger and more prominent than the 'G'. The letters are in a bright yellow color, and they are set against a blue background that resembles a hand. The words 'EVER GLOBAL' are written in black at the top of the logo, and 'ENTERPRISE CORP.' is written in black at the bottom.
LONG THANH INDUSTRIAL ZONE, TAMAN VILLAGE, DONG FAX: (0251)351 4023 024
SUPERIEURVIETNAM
# 510(k) SUMMARY
### 8.0 Assessment of Non-Clinical Performance Data:
The following bench testing was conducted for design elements and performance characteristics deemed appropriate to demonstrate equivalence to the predicate device. Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate And Xylazine met the predetermined acceptance criteria. No new safety or performance issues were raised during testing:
| Test | Test Method | Purpose | Acceptance Criteria | Results |
|----------------------------------------------------------------------|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Dimension | ASTM D6319-19 | Determine the<br>geometrical<br>dimension of gloves | Length:<br>Short cuff: ≥230mm<br>Long cuff: ≥300mm<br>Thickness:<br>Palm: ≥ 0.05 mm<br>Finger: ≥ 0.05 mm<br>Cuff: ≥ 0.05 mm<br>Palm Width:<br>XS: 70 ± 10 mm<br>S: 80 ± 10 mm<br>M: 95 ± 10 mm<br>L: 110 ± 10 mm<br>XL: 120 ± 10 mm | Pass |
| Freedom from<br>holes (Water leak) | 21 CFR 800.20.<br>&<br>ASTM D5151-19 | Detect the holes on<br>the gloves. | G-I/ AQL 2.5 | Pass |
| Tensile strength<br>(Before aging/<br>After aging) | ASTM D6319-19 | Evaluate the tensile<br>(tenson) properties of<br>the gloves. In<br>addition, it also<br>determines the<br>influence of elevated | Before Aging: ≥14MPa<br>After Aging: ≥14MPa | Pass |
| Elongation (Before<br>aging/ After aging) | ASTM D6319-19 | temperature on the<br>physical properties of<br>gloves. | Before Aging: ≥500%<br>After Aging: ≥400% | Pass |
| Powder Residual | ASTM D6319-19 | Determine the<br>average powder mass<br>found on the gloves | < 2mg per glove | Pass |
| Biocompatibility-<br>cytotoxicity | AAMI/ANSI/ISO<br>10993-5 | Determine the<br>cytotoxicity potential<br>of glove | No in vitro cytotoxic as<br>described in<br>ISO 10993-5 | Pass |
| Biocompatibility-<br>Skin Sensitization | AAMI/ANSI/ISO | Determine the<br>potential of glove to<br>promote skin<br>sensitization after<br>repeated applications | No dermal reactions<br>indicative of delayed<br>contact hypersensitivity | Pass |
| and Irritation | 10993-10 | Determine the<br>potential of gloves to<br>promote skin<br>irritation after<br>repeated applications. | No skin irritation,<br>cumulative irritation<br>index to be 0. | Pass |
| Resistance of<br>Gloves to<br>Permeation by<br>Chemotherapy<br>Drugs | ASTM D6978-05 | Assessment of<br>medical gloves to<br>permeation by<br>chemotherapy drugs. | The resistance of our<br>device to permeation<br>were challenged against<br>57 liquid hazardous<br>drugs | Pass |
Premarket Notification 510(k) Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate and Xylazine
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the text "EVER GLOBAL (VIETNAM) ENT. CORP" in bold, black font. The text is all uppercase. There is a line underneath the text.
Image /page/13/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'EG' in a stylized, interconnected design, with the 'E' in the upper left and the 'G' in the lower right. The letters are in yellow and are set against a blue background that appears to be a stylized hand. The words 'EVER GLOBAL' are at the top of the logo, and 'ENTERPRISE CORP' is at the bottom, both in a smaller font size.
ONG THANH INDUSTRIAL ZONE, TAMAN VILLAGE, DONG X: (0251)351 4023 02-
# 510(k) SUMMARY
## 9.0 Assessment of Clinical Performance Data:
Clinical data is not needed for this type of device.
## 10.0 Conclusion:
The conclusions drawn from the nonclinical tests demonstrate that the subject device Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate And Xylazine is as safe, as effective, and performs as well as or better than the legally marketed device, K193555.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.