Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate
K193555 · Ever Global (Vietnam) Enterprise Corp · LZC · Jun 4, 2020 · General Hospital
Device Facts
Record ID
K193555
Device Name
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate
Applicant
Ever Global (Vietnam) Enterprise Corp
Product Code
LZC · General Hospital
Decision Date
Jun 4, 2020
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs and fentanyl citrate in accordance with ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical gloves to Permeation by Chemotherapy Drugs.
Device Story
Disposable non-sterile nitrile examination glove; provides barrier protection for examiner's hands/fingers; prevents cross-contamination between patient and examiner. Tested for permeation resistance against 55 chemotherapy drugs and fentanyl citrate per ASTM D6978-05. Used in clinical settings by healthcare professionals. Output is physical barrier protection; informs clinical decision-making regarding safe handling of hazardous drugs. Benefits include protection of healthcare personnel from exposure to chemotherapy agents and fentanyl.
Clinical Evidence
Bench testing only. Testing included dimensions (ASTM D6319-10), physical properties (tensile strength/elongation), water leak/pinhole testing (ASTM D5151-06), powder residual (ASTM D6124-06), biocompatibility (ISO 10993-5, ISO 10993-10), and chemotherapy/fentanyl permeation resistance (ASTM D6978-05).
Technological Characteristics
Nitrile compound; non-sterile; powder-free. Dimensions: 230mm (short) or 300mm (long) length; 0.05mm min thickness. Physical properties: 14 MPa min tensile strength; 500% min elongation (before aging). Biocompatibility: passes ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).
Indications for Use
Indicated for use as a disposable patient examination glove worn on the examiner's hands or fingers to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs and fentanyl citrate.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs (K190403)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
June 4, 2020
Ever Global (Vietnam) Enterprise Corp % Elizabeth Deng U.S. Representative Elizabeth Deng 5748 Eaglewood Place Ranch Cucamonga, California 91730
Re: K193555
Trade/Device Name: Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZC, LZA, QDO Dated: April 29, 2020 Received: May 5, 2020
Dear Elizabeth Deng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Ever Global (Vietnam) Enterprise Corporation
Long Thanh Industrial Zone, Taman Village, Long Thanh District, Dong Nai Province, Vietnam
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | | | | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. |
|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|--|---------------------------------------------------------------------------------------------|
| Indications for Use | | | | |
| 510(k) Number (if known) | K193555 | | | |
| Device Name | Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate | | | |
| Indications for Use (Describe) | A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or<br>fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with<br>chemotherapy drugs and fentanyl citrate in accordance with ASTM D6978-05 Standard Practice for Assessment of<br>Resistance of Medical gloves to Permeation by Chemotherapy Drugs. | | | |
| Test Chemotherapy Drug | Concentration (mg/ml) | Minimum Breakthrough Detection Time (Min.) | | |
| 01 Arsenic Trioxide | 1.0 | > 240 | | |
| 02 Azacitidine (Vidaza) | 25.0 | > 240 | | |
| 03 Bendamustine HCl | 5.0 | > 240 | | |
| 04 Bleomycin Sulfate | 15.0 | > 240 | | |
| 05 Bortezomib (Velcade) | 1.0 | > 240 | | |
| 06 Busulfan | 6.0 | > 240 | | |
| 07 Carboplatin | 10.0 | > 240 | | |
| 08 Carfilzomib | 2.0 | > 240 | | |
| 09 Carmustine (BCNU) | 3.3 | 6.2 | | |
| 10 Cetuximab (Erbitux) | 2.0 | > 240 | | |
| 11 Chloroquine | 50.0 | > 240 | | |
| 12 Cisplatin | 1.0 | > 240 | | |
| 13 Cladribine | 1.0 | > 240 | | |
| 14 Cyclophosphamide | 20.0 | > 240 | | |
| 15 Cyclosporine A | 100.0 | > 240 | | |
| 16 Cytarabine mg/ml) | 100.0 | > 240 | | |
| 17 Cytovene (Ganciclovir) | 10.0 | > 240 | | |
| 18 Dacarbazine | 10.0 | > 240 | | |
| 19 Daunorubicin | 5.0 | > 240 | | |
| 20 Decitabine | 5.0 | > 240 | | |
| 21 Docetaxel | 10.0 | > 240 | | |
| 22 Doxorubicin Hydrochloride | 2.0 | > 240 | | |
| 23 Epirubicin (Ellence) | 2.0 | > 240 | | |
| 24 Etoposide | 20.0 | > 240 | | |
| 25 Fludarabine | 25.0 | > 240 | | |
| 26 Fluorouracil | 50.0 | > 240 | | |
| 27 Fulvestrant | 50.0 | > 240 | | |
| 28 Gemcitabine | 38.0 | > 240 | | |
| 29 Idarubicin | 1.0 | > 240 | | |
| 30 Ifosfamide | 50.0 | > 240 | | |
| 31 Irinotecan | 20.0 | > 240 | | |
| 32 Mechlorethamine HCl | 1.0 | > 240 | | |
| 33 Melphalan | 5.0 | > 240 | | |
| 34 Methotrexate | 25.0 | > 240 | | |
| 35 Mesna | 100.0 | > 240 | | |
| 36 Mitomycin C | 0.5 | > 240 | | |
| 37 Mitoxantrone | 2.0 | > 240 | | |
| FORM FDA 3881 (7/17) | Page 1 of 2 | PSC Publishing Services (301) 443-6740 | | |
{3}------------------------------------------------
# Ever Global (Vietnam) Enterprise Corporation
Long Thanh Industrial Zone, Taman Village, Long Thanh District, Dong Nai Province, Vietnam
| 38 Oxaliplatin | 5.0 | > 240 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|--------------|
| 39 Paclitaxel | 6.0 | > 240 |
| 40 Paraplatin | 10.0 | > 240 |
| 41 Pemetrexed | 25.0 | > 240 |
| 42 Pertuzumab | 30.0 | > 240 |
| 43 Raltitrexed | 0.5 | > 240 |
| 44 Retrovir | 10.0 | > 240 |
| 45 Rituximab | 10.0 | > 240 |
| 46 Temsirolimus | 25.0 | > 240 |
| 47 Thiotepa | 10.0 | 13.6 |
| 48 Topotecan HCl | 1.0 | > 240 |
| 49 Trastuzumab | 21.0 | > 240 |
| 50 Triclosan | 2.0 | > 240 |
| 51 Trisenox | 1.0 | > 240 |
| 52 Vinblastine | 1.0 | > 240 |
| 53 Vincristine Sulfate | 1.0 | > 240 |
| 54 Vinorelbine | 10.0 | > 240 |
| 55 Zoledronic Acid | 0.8 | > 240 |
| Warning: do not use with Carmustine and Thiotepa.<br>The maximum testing time is 240 minutes. Please note that the following drug has an extremely low permeation time: | | |
| Carmustine (BCNU) | 3.3 mg/ml | 6.2 minutes |
| Thiotepa | 10.0 mg/ml | 13.6 minutes |
| Fentanyl Permeation Resistance Claim - Under the testing conditions of ASTM D6978-05, Fentanyl Citrate<br>Injection (100mcg/2mL) was found to have no breakthrough detected up to 240 minutes. | | |
| Type of Use (Select one or both, as applicable) | | |
| Image: Checkbox | Over-The-Counter Use (21 CFR 801 Subpart C) | |
| Image: Unchecked checkbox | Prescription Use (Part 21 CFR 801 Subpart D) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995.<br>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | |
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
## K193555
#### 1.0 Submitter:
| Submitter's name : | Ever Global (Vietnam) Enterprise Corp. | |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------|--|
| Submitter's address : | Long Thanh Industrial Zone<br>Taman Village Dong Nai Province, VN 810000 | |
| Phone number : | 84-61-3514022 | |
| Fax number : | 84 -61-3514023 | |
| Name of contact person: | Jerry Lin | |
| Summary Preparation Date: | May 27th,2020 | |
| 2.0 US Agent: | | |
| US representative name: | Elizabeth Deng | |
| Company address: | 5748 Eaglewood Place<br>Rancho Cuamonga, California<br>Rancho Cucamonga, CA 91739 | |
| Telephone number: | 909 4659188 | |
| Contact email: | Baxianunited48@Yahoo.Com | |
| 3.0 Name of the Device | | |
| Proprietary/Trade name: | Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For<br>Use With Chemotherapy Drugs and Fentanyl Citrate | |
| Common Name: | Nitrile Examination Gloves | |
| Classification Name: | Patient Examination Glove | |
| Device Classification: | Class I | |
| Regulation Number: | 21 CFR 880.6250 | |
| Product Code: | LZA, LZC, QDO | |
| 4.0 Predicate device | | |
| Device Name: | Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For<br>Use With Chemotherapy Drugs | |
| Company name: | Ever Global (Vietnam) Enterprise Corp. | |
| 510(K) Number: | K190403 | |
#### 5.0 Device Description:
"Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate" is a patient examination glove made from nitrile compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of this medical device is to provide single use barrier protection for the wearer and the device meets the specifications for Barrier Protection and tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.
#### 6.0 Indication for use:
# Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate
A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs.
{5}------------------------------------------------
| No. | Test Chemotherapy Drug and<br>Concentration | Minimum Breakthrough<br>Detection Time (Min.) |
|-----|---------------------------------------------|-----------------------------------------------|
| 1 | Arsenic Trioxide (1.0 mg/ml) | > 240 |
| 2 | Azacitidine (Vidaza) (25.0 mg/ml) | > 240 |
| 3 | Bendamustine HCl (5.0 mg/ml) | > 240 |
| 4 | Bleomycin Sulfate (15.0 mg/ml) | > 240 |
| 5 | Bortezomib (Velcade) (1.0 mg/ml) | > 240 |
| 6 | Busulfan (6.0 mg/ml) | > 240 |
| 7 | Carboplatin (10.0 mg/ml) | > 240 |
| 8 | Carfilzomib (2.0 mg/ml) | > 240 |
| 9 | Carmustine (BCNU), (3.3 mg/ml) | 6.2 |
| 10 | Cetuximab (Erbitux) (2.0 mg/ml) | > 240 |
| 11 | Chloroquine (50.0 mg/ml) | > 240 |
| 12 | Cisplatin (1.0 mg/ml) | > 240 |
| 13 | Cladribine (1.0 mg/ml) | > 240 |
| 14 | Cyclophosphamide (20.0 mg/ml) | > 240 |
| 15 | Cyclosporine A (100.0 mg/ml) | > 240 |
| 16 | Cytarabine (100.0 mg/ml) | > 240 |
| 17 | Cytovene (Ganciclovir) (10.0 mg/ml) | > 240 |
| 18 | Dacarbazine (10.0 mg/ml) | > 240 |
| 19 | Daunorubicin (5.0 mg/ml) | > 240 |
| 20 | Decitabine (5.0 mg/ml) | > 240 |
| 21 | Docetaxel (10.0 mg/ml) | > 240 |
| 22 | Doxorubicin Hydrochloride (2.0 mg/ml) | > 240 |
| 23 | Epirubicin (Ellence) (2.0 mg/ml) | > 240 |
| 24 | Etoposide (20.0 mg/ml) | > 240 |
| 25 | Fludarabine (25.0 mg/ml) | > 240 |
| 26 | Fluorouracil (50.0 mg/ml) | > 240 |
| 27 | Fulvestrant (50.0 mg/ml) | > 240 |
| 28 | Gemcitabine (38.0 mg/ml) | > 240 |
| 29 | Idarubicin (1.0 mg/ml) | > 240 |
| 30 | Ifosfamide (50.0 mg/ml) | > 240 |
| 31 | Irinotecan (20.0 mg/ml) | > 240 |
| 32 | Mechlorethamine HCl (1.0 mg/ml) | > 240 |
| 33 | Melphalan (5.0 mg/ml) | > 240 |
| 34 | Methotrexate (25 mg/ml) | > 240 |
| 35 | Mesna (100 mg/ml) | > 240 |
| 36 | Mitomycin C (0.5 mg/ml) | > 240 |
| 37 | Mitoxantrone (2.0 mg/ml) | > 240 |
| 38 | Oxaliplatin (5.0 mg/ml) | > 240 |
| 39 | Paclitaxel (6.0mg/ml) | > 240 |
| 40 | Paraplatin (10.0 mg/ml) | > 240 |
| 41 | Pemetrexed (25.0 mg/ml) | > 240 |
| 42 | Pertuzumab (30.0 mg/ml) | > 240 |
| 43 | Raltitrexed (0.5 mg/ml) | > 240 |
| 44 | Retrovir (10.0 mg/ml) | > 240 |
| 45 | Rituximab (10.0 mg/ml) | > 240 |
| 46 | Temsirolimus (25.0 mg/ml) | > 240 |
| 47 | Thiotepa (10.0 mg/ml) | 13.6 |
| 48 | Topotecan HCl (1.0 mg/ml) | > 240 |
| 49 | Trastuzumab (21.0 mg/ml) | > 240 |
| 50 | Triclosan (2.0 mg/ml) | > 240 |
| 51 | Trisenox (1.0 mg/ml) | > 240 |
| 52 | Vinblastine (1.0 mg/ml) | > 240 |
| 53 | Vincristine Sulfate (1.0 mg/ml) | > 240 |
| 54 | Vinorelbine (10.0 mg/ml) | > 240 |
| 55 | Zoledronic Acid (0.8 mg/ml) | > 240 |
Table 1 Tested for use with 55 chemotherapy drugs.
{6}------------------------------------------------
Warning: do not use with Carmustine and Thiotepa.
The maximum testing time is 240 minutes. Please note that the following drug has an extremely low permeation time:
| Carmustine (BCNU), 3.3 mg/ml | 6.2 minutes |
|------------------------------|--------------|
| Thiotepa, 10.0 mg/ml | 13.6 minutes |
Fentanyl Permeation Resistance Claim - Under the testing conditions of ASTM D6978-05, Fentanyl Citrate Injection (100mcg/2mL) was found to have no breakthrough detected for up to 240 minutes.
{7}------------------------------------------------
7.0 Comparison of Technological characteristics between the predicate and subject devices:
| Device Characteristic | Predicate Device | Subject Device | Comparison |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|
| Product name | Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs | Disposable Powder Free Nitrile Examination Glove, Blue Color, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate | similar |
| 510(K) No. | K190403 | K193555 | N/A |
| Product Owner | Ever Global (Vietnam) Enterprise Corporation | Ever Global (Vietnam) Enterprise Corporation | same |
| Product Code | LZA, LZC | LZA, LZC, QDO | similar |
| Regulation | 21 CFR 880.6250 | 21 CFR 880.6250 | same |
| Class | I | I | same |
| Intended Use | The Nitrile Powder Free patient examination glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs. | The Nitrile Powder Free patient examination glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs And Fentanyl Citrate in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs. | similar |
| Powder free | Yes | Yes | same |
| Size | Small/ Medium/Large/X Large | X Small/ Small/Medium/Large/X Large | X Small is additional. |
| Single Use | YES | YES | same |
| Non-Sterile | No | No | same |
| Dimensions-Length | Complies with ASTM D6319-10 230 mm min. | Short cuff ≥230mm<br>Long cuff ≥300mm | Long cuff is additional. |
| Dimensions - Palm Width | Complies with ASTM D6319-10 | Complies with ASTM D6319-10 | Similar.<br>X Small is additional. |
| | Small<br>Medium<br>Large | X Small 70±10<br>Small 80±10<br>Medium 95 ±10<br>Large 110±10 | |
Table 2 Comparison of Technological Characteristics
{8}------------------------------------------------
| Dimensions - Thickness | | Complies with ASTM D6319-10<br>Palm 0.05mm min.<br>Finger 0.05 mm min.<br>Cuff 0.05 mm min | Complies with ASTM D6319-10<br>Palm 0.05mm min.<br>Finger 0.05 mm min.<br>Cuff 0.05 mm min | same | |
|--------------------------------------|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------|
| Physical<br>Properties | | Tensile Strength:<br>Before Aging 14 MPa, min.<br>After Aging 14 MPa, min.<br>Elongation:<br>Before Aging 500% min.<br>After Aging 400% min. | Tensile Strength:<br>Before Aging 14 MPa, min.<br>After Aging 14 MPa, min.<br>Elongation:<br>Before Aging 500% min.<br>After Aging 400% min. | same | |
| Residual<br>powder | | Complies with ASTM D6319-10<br>< 2mg per glove | Complies with ASTM D6319-10<br>< 2mg per glove | same | |
| Freedom from<br>Holes | | In accordance with ASTM D6319-<br>10 and ASTM D5151- 06<br>(reapproved 2015), G-1, AQL 2.5 | In accordance with ASTM D6319-<br>10 and ASTM D5151-06<br>(reapproved 2015), G-1, AQL 2.5 | same | |
| Biocompatibility | | AAMI/ANSI/ISO 10993-10<br>Passes<br>Not a skin irritant & Not a skin<br>sensitizer | AAMI/ANSI/ISO 10993-10 &<br>AAMI/ANSI/ISO 10993-5<br>Passes<br>Not a skin irritant, Not a skin<br>sensitizer & No cytotoxicity<br>reaction | same | |
| Chemotherapy drugs tested | | Breakthrough Detection Time in Minutes | | Comparison | |
| | | Predicate<br>device | Subject Device | | |
| | Device Characteristic | K190403 | Short Cuff | Long Cuff | |
| Arsenic Trioxide (1.0 mg/ml) | | - | > 240 | | Different |
| Azacitidine (Vidaza) (25.0 mg/ml) | | - | > 240 | | Different |
| Bendamustine HCI (5.0 mg/ml) | | - | > 240 | | Different |
| | Bleomycin Sulfate (15.0 mg/ml) | - | > 240 | | Different |
| | Bortezomib (Velcade) (1.0 mg/ml) | - | > 240 | | Different |
| Busulfan (6.0 mg/ml) | | - | > 240 | | Different |
| Carboplatin (10.0 mg/ml) | | - | > 240 | | Different |
| Carfilzomib (2.0 mg/ml) | | - | > 240 | | Different |
| Carmustine (BCNU), (3.3 mg/ml) | | 6.2 | 21.5 | 37.5 | Similar |
| Cetuximab (Erbitux) (2.0 mg/ml) | | - | > 240 | | Different |
| Chloroquine (50.0 mg/ml) | | - | > 240 | | Different |
| | Cisplatin (1.0 mg/ml) | > 240 | > 240 | | Same |
| | Cladribine (1.0 mg/ml) | - | > 240 | | Different |
| | Cyclophosphamide (20.0 mg/ml) | > 240 | > 240 | | Same |
| | Cyclosporin A (100.0 mg/ml) | - | > 240 | | Different |
| | Cytarabine (100.0 mg/ml) | - | > 240 | | Different |
| Dacarbazine (DTIC), (10.0 mg/ml) | > 240 | > 240 | Same | | |
| Daunorubicin (5.0 mg/ml) | - | > 240 | Different | | |
| Decitabine (5.0 mg/ml) | - | > 240 | Different | | |
| Docetaxel (10.0 mg/ml) | - | > 240 | Different | | |
| Doxorubicin Hydrochloride (2.0mg/ml) | > 240 | > 240 | Same | | |
| Epirubicin (Ellence) (2.0 mg/ml) | - | > 240…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.