Disposable Biopsy Needle is intended for use in obtaining an aspirate for cytology.
Device Story
Disposable Biopsy Needle is a single-use, sterile, EO-sterilized manual instrument for obtaining soft tissue aspirates for cytology. It consists of a cutting cannula, inner stylet, operating handle, depth stopper, and protective sheath. The device is operated by a clinician under imaging guidance (ultrasound or X-ray). The clinician inserts the needle to the biopsy site, using centimeter markings and an adjustable depth stopper for positioning. The inner stylet is removed, and the cutting cannula is manipulated to collect the sample. The device features echogenic tips for visualization. The output is a tissue aspirate used for diagnostic evaluation. The device benefits patients by enabling minimally invasive tissue sampling for pathology, aiding in the diagnosis of various soft tissue conditions.
Clinical Evidence
No clinical data provided. Bench testing included packaging/shelf-life (ASTM F1980), biocompatibility (ISO 10993 series), sterilization (ISO 11135), and performance testing (dimensions, connection firmness, stiffness, corrosion resistance, leakage, echogenicity, puncture force). A comparative bench study across various tissue types (liver, kidney, breast, thyroid, prostate, spleen, pancreas, lymph nodes, lung) demonstrated no significant difference in sampling weight or cell integrity compared to predicate devices.
Technological Characteristics
Materials: Stainless steel (SUS304) cannula/stylet, Polycarbonate (PC) handle, Thermoplastic Elastomer (TPE) depth stopper. Features: Luer hub, centimeter markings, echogenic tip, adjustable depth stopper. Tip types: Chiba or Greene. Sterilization: EO gas. Single-use. No software or electronic components.
Indications for Use
Indicated for soft tissue biopsies including liver, kidney, breast, thyroid, prostate, spleen, pancreas, lymph nodes, lung, and soft tissue tumors. Not intended for use in bone.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 22, 2025
Zhejiang CuraWay Medical Technology Co., Ltd.
Tu Mengjing
Regulatory Affairs Specialist
Room 106, Building 1, No. 600, 21st Avenue,
Baiyang Sub-district, Qiantang New District
Hangzhou, Zhejiang 310018
China
Re: K244018
Trade/Device Name: Disposable Biopsy Needle
Regulation Number: 21 CFR 876.1075
Regulation Name: Gastroenterology-Urology Biopsy Instrument
Regulatory Class: Class II
Product Code: KNW
Dated: December 27, 2024
Received: December 27, 2024
Dear Tu Mengjing:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K244018 - Tu Mengjing
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K244018 - Tu Mengjing
Page 3
Sincerely,
Jessica Carr -S
Jessica Carr, PhD
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K244018 | |
| Device Name Disposable Biopsy Needle | |
| Indications for Use (Describe) Disposable Biopsy Needle is intended for use in soft tissue biopsies such as liver, kidney, breast, thyroid, prostate, spleen, pancreas, lymph nodes, lung and various soft tissue tumors. It is not intended for use in bone. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
1 / 9
# 510(K) Summary
# K244018
| Company Name/Owner | Zhejiang CuraWay Medical Technology Co., Ltd. |
| --- | --- |
| Contact person/Author | Mengjing Tu
Telephone number: +86-0571-87016876 |
| Date prepared | April 10, 2025 |
| Contact details Address | Room 106, Building 1, No. 600, 21^{st} Avenue, Baiyang Sub-district, Qiantang New District, 310018 Hangzhou City, Zhejiang Province, China |
| Trade name | Disposable Biopsy Needle |
| Common name | Biopsy Needle |
| Classification name | Instrument, Biopsy |
| Review panel | Gastroenterology/Urology |
| Regulation number | 21 CFR 876.1075 |
| Product code | KNW |
| Predicate device | Pan® Aspirating Needle (Chiba) (K970872)
Promisemed Fine Biopsy Needle (K213683) |
{5}
2 / 9
# 1. Device Description
Disposable Biopsy Needle is intended for use in obtaining an aspirate for cytology. Disposable Biopsy Needle has five models: BN-MAR-1, BN-MAR-2, BN-MAR-3, BN-MAR-4, BN-MAR-8.
The Disposable Biopsy Needle consists of operating handle of Inner stylet, operating handle of cutting cannula, cutting cannula, inner stylet, depth stopper and protective sheath.
The models differ as follows:
- **Cutting cannula tip type**: The cannula tip type of BN-MAR-1, BN-MAR-2, BN-MAR-3 is chiba. The cannula tip type of BN-MAR-4, BN-MAR-8 is greene.
- **Operating handle designs**:
**BN-MAR-1, BN-MAR-3, BN-MAR-4 Models**:
a. BN-MAR-1, BN-MAR-3, and BN-MAR-4 have the same operating handle, including the operating handle of inner stylet and operating handle of cutting cannula.
b. The Operating handle of Inner stylet connector connects to the Operating handle of cutting cannula connector via a straight snap-fit design, allowing for easy disconnection with one hand to remove the Inner stylet.
**BN-MAR-2, BN-MAR-8 Models**:
a. The Operating handle of Inner stylet of BN-MAR-2 connects to the Operating handle of cutting cannula via a straight snap-fit design connector, allowing easy disconnection with one hand to remove the Inner stylet.
b. The Operating handle of inner stylet of BN-MAR-8 connects to the Operating handle of cutting cannula via a female Luer taper and male Luer thread taper, which can meet the sealing performances in clinical applications.
- **Bevel Angle of the Cutting cannula tip**: There are three kinds of bevel angle of the Cutting cannula tip, which are $\alpha$, $\beta$, $\delta$. The bevel angle of the cutting cannula tip of BN-MAR-1 and BN-MAR-2 is $\alpha$. The bevel angle of the needle tip of BN-MAR-3 is $\beta$. The bevel angle of the Cutting cannula tip of BN-MAR-4 and BN-MAR-8 is $\delta$.
Each model has various specifications depending on different sizes such as needle lengths, needle diameters, etc.
The cutting cannula has numerically ordered centimeter markings on it to provide
{6}
reference for depth placement. In the distal area of the cutting cannula, an ultrasound enhancement is available to promote accurate placement under ultrasound or X-ray guidance.
An adjustable depth stopper allows the user to restrict forward movement, localizing the needle tip to the biopsy site.
The Operating handle of Inner stylet is color-coded, which indicates gauge size of the needle. It is available in several needle gauge sizes and lengths.
The Disposable Biopsy Needle is a single-use device, supplied in a sterile state sterilized by EO gas. Re-sterilization by users is forbidden. The shelf life is defined for 3 years, which is verified.
## 2. Intended use / Indications
### Intended use:
Disposable Biopsy Needle is intended for use in obtaining an aspirate for cytology.
### Indications:
Disposable Biopsy Needle is intended for use in soft tissue biopsies such as liver, kidney, breast, thyroid, prostate, spleen, pancreas, lymph nodes, lung and various soft tissue tumors. It is not intended for use in bone.
## 3. Substantial equivalence comparison with predicate device
Detailed substantial comparison was made between subject device and predicate devices.
Pan® Aspirating Needle (Chiba) (K970872) and Promisemed Fine Biopsy Needle (K213683) were selected as predicate devices to the Disposable Biopsy Needle (Subject device).
The subject device and predicate devices are based on the same technological elements and intended use. There are several design differences, but do not cause a risk to the safety and effectiveness of the product. Detailed information please refer to comparison table below. Please refer to the following Table 1 for details.
3 / 9
{7}
Table 1: Substantial equivalence comparison of Disposable Biopsy Needle
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Comment |
| --- | --- | --- | --- | --- |
| Proprietary Name | Disposable Biopsy Needle | Pan® Aspirating Needle (Chiba) | Promisemed Fine Biopsy Needle | - |
| Classification Regulation | 876.1075 | 876.1075 | 876.1075 | Same |
| Product Code | KNW | KNW | KNW | Same |
| 510(k) Number | K244018 | K970872 | K213683 | - |
| Intended Use | Disposable Biopsy Needle is intended for use in obtaining an aspirate for cytology. | It is to be used for taking cytological and histological biopsies of soft tissue. | It is to be used for taking cytological and histological biopsies of soft tissue. | Same |
| Visualization technique | The introduction of the needle into the body should be carried out under imaging guidance (ultrasound, X-Ray, CT). | It's a relatively invasive procedure and is performed by radiologist under guidance of imaging techniques such as ultrasound, X-ray. | It's a relatively invasive procedure and is performed by radiologist under guidance of imaging techniques such as ultrasound, X-ray. | Difference 1 |
| Materials | -Cutting cannula and Inner stylet: Stainless Steel (SUS304)-Operating handle of cutting cannula, Operating handle of inner stylet: Polycarbonate (PC)- Depth stopper: Thermoplastic Elastomer (TPE) | - Needle: Stainless Steel (X5CrNi18-10);- Needle base: Acrylonitrile Butadiene Styrene (ABS)- Depth stop: Thermoplastic Elastomer (TPE) | - Needle: Stainless Steel (X5CrNi18-10);- Needle base: Acrylonitrile Butadiene Styrene (ABS)- Depth stop: Thermoplastic Elastomer (TPE) | Difference 2 |
{8}
| Comparison Elements | Subject Device | | Predicate Device 1 | | Predicate Device 2 | | Comment |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Single use | YES | | YES | | YES | | Same |
| Needle Hub | Luer hub | | Luer hub | | Luer hub | | Same |
| Performance | | | | | | | |
| Echogenic | YES | | YES | | YES | | Same |
| Connector transparency | Transparent/lock | | Transparent/lock | | Transparent/lock | | Same |
| Needle Gauge | 16G, 17G, 18G, 19G, 20G, 21G, 22, 23G, 24G, 25G | | 18G, 20G, 21G, 22G, 23G, 25G | | 16G, 18G, 19G, 20G, 21G, 22G, 23G | | Difference 3 |
| Needle Length | 40, 50, 60, 70, 80, 90, 100, 110, 120, 130, 140, 150, 160, 170, 180, 190, 200, 210, 220, 230, 240, 250 mm | | 50, 80, 90, 150, 200 mm | | 90mm, 100mm, 150mm, 200mm | | Difference 4 |
| Needle tube tip type | Chiba, Greene | | Chiba | | Chiba | | Difference 5 |
| Bond between outer needle tube and hub | according to the nominal OD of the outer needle, the outer needle and hub shall not break when pull the needle with the axial static force specified in the below table for 10s. | | according to the nominal OD of the outer needle, the outer needle and hub shall not break when pull the needle with the axial static force specified in the below table for 10s. | | according to the nominal OD of the outer needle, the outer needle and hub shall not break when pull the needle with the axial static force specified in the below table for 10s. | | Same |
| | OD | Force/N | OD | Force/N | OD | Force/N | |
| | 23G | 34 | 23G | 34 | 23G | 34 | |
| Biocompatibility | ISO 10993 series standards | | ISO 10993 series standards | | ISO 10993 series standards | | Same |
| Anatomical | Soft tissue biopsies such as liver, | | Soft tissue | | Soft tissue | | Difference 6 |
| | soft tissue biopsies such as liver, | | soft tissue biopsies | | soft tissue biopsies | | |
{9}
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Comment |
| --- | --- | --- | --- | --- |
| sites | kidney, breast, thyroid, prostate, spleen, pancreas, lymph nodes , lung and various soft tissue tumors. It is not intended for use in bone. | | | |
| Sterilization method | EO sterilization | EO sterilization | EO sterilization | Same |
## Difference 1: Visualization technique
The difference between the Subject device and the predicate device is the subject device not imaging under the CT. The usability of the subject device under CT has been verified.
## Difference 2: Materials
The difference between the Subject device and the predicate device is the PC material. PC do not come into contact with patients, so it will not bring safety risks
## Difference 3: Needle Gauge
Subject device has two gauge sizes (17G, 24G) more than predicate devices. The 17G OD and 24G OD of Cutting cannula of subject device are within the predicate devices'. It will not cause new risks to device safety and effectiveness.
## Difference 4: Needle Length
It can be seen from the table above that subject device provides more selections on the Length of Cutting Cannula, namely 40mm, 60mm, 70mm,
{10}
110mm, 120mm, 130mm, 140mm, 160mm, 170mm, 180mm, 190mm, 210mm, 220mm, 230mm, 240mm, 250 mm. The 60mm, 70mm, 110mm, 120mm, 130mm, 140mm, 160mm, 170mm, 180mm, 190mm are within the length range (50mm to 200mm) of predicate devices. Besides, considering the different shape of patients and the distance between needle insertion point and lesion, 40mm, 210mm, 220mm, 230mm, 240mm and 250mm are provided by subject device. The puncture length is judged by the clinician, and we offer a wider selection of puncture lengths that do not raise new issues affecting the safety and effectiveness of the devices. The additional puncture lengths are similar to other marketed products with the same intended use.
## Difference 5: Cutting Cannula tip type
The puncture process: The Chiba type has a blade on the tip of Cutting cannula for puncture, while the Greene type has a blade on the tip of Inner stylet, but both use a combination of the Cutting cannula and Inner stylet to achieve puncture, and the process is the same.
The sampling process: When sampling with the Chiba type, the Inner stylet is withdrawn, and only the Cutting cannula is operated to take a sample at the lesion site. The tip of Cutting cannula can move back and forth with the operator's technique, but high demands are placed on the operator's technique to avoid damaging the tissue around the needle. When sampling with the Greene type, the Inner stylet is withdrawn, and only the Cutting cannula is operated to take a sample at the lesion site. The tip of Cutting cannula is flat and dull, and it is not easy to move back and forth with the operator's technique. Despite this difference in design, performance testing in the indicated tissue types demonstrated substantially equivalent safety and effectiveness.
## Difference 6: Anatomical sites
Though the indications for use descriptions are not identical, the specific indications designating anatomical sites are within the same intended use as the predicate and do not raise different questions of safety and effectiveness.
Based on the risk analysis of the above differences, the differences do not affect the safety and effectiveness of the equipment.
7 / 9
{11}
4. Non-Clinical Tests
The following testing was conducted to demonstrate the safe and effective use of subject devices and the standards that subject devices complied with:
- Packaging and shelf-life testing per ASTM F1980.
- Biocompatibility evaluation per ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11 and ISO 10993-23.
- Sterilization validation per ISO 11135: 2014.
- EO Residuals testing per ISO 10993-7: 2008.
- Performance testing including Appearance, Basic dimensions, Sample collection space and accessibility, Luer connection, Connection firmness, Stiffness, Toughness, Corrosion resistance, Leakage, Ultrasound detectability, X-ray detectability, Protective sheath, Puncture force, Chemical Characteristics, EO and ECH Residual, Sterility test, Bacterial endotoxin.
- A comparative study was conducted between CuraWay’s Disposable Biopsy Needle and the predicate devices across various tissues including liver, kidney, breast, thyroid, prostate, spleen, pancreas, lymph nodes, and lungs. The results showed no significant difference in sampling weight or cell integrity between the subject and predicate devices.
5. Clinical data
No clinical data was provided.
8 / 9
{12}
6. Conclusion
The indications for use of subject device are equivalent to the predicate device. The standards testing and nonclinical performance testing demonstrate that the subject device technological characteristics are equivalent to the predicate device for biopsy of soft tissue. In conclusion, the subject device is substantially equivalent to the predicate devices.
9 / 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.