It is to be used for taking cytological and histological biopsies of soft tissue.
Device Story
Hand-operated, non-electronic fine biopsy needle; features Chiba tip for tissue sampling. Used by radiologists under imaging guidance (ultrasound, X-ray) for soft tissue biopsies. Device includes transparent Luer Lock connector, sharpened cannula, external echogenic treatment for ultrasound visibility, centimeter markings, and sliding stopper. Single-use, sterile (ETO). Provides tissue samples for pathological examination. Benefits include safe, guided tissue access for diagnostic purposes.
Clinical Evidence
Bench testing only. No clinical data. Testing included appearance, cleanliness, dimensions, needle point, lumen patency, bond strength, stiffness, corrosion/breakage resistance, scale identification, and ISO 80369-7 Luer lock performance. Biocompatibility testing (ISO 10993) confirmed no cytotoxicity, irritation, sensitization, systemic toxicity, or pyrogenicity.
Technological Characteristics
Materials: Stainless Steel (X5CrNi18-10) needle, ABS needle base, TPE depth stop. Features: Chiba tip, echogenic treatment, centimeter markings, sliding stopper. Connectivity: None. Sterilization: ETO. Standards: ISO 6009 (color coding), ISO 9626 (needle tube), ISO 80369-7 (Luer lock), ISO 10993 (biocompatibility).
Indications for Use
Indicated for patients requiring cytological and histological biopsies of soft tissue. Not intended for use in bone.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 9, 2022
Promisemed Hangzhou Meditech Co., Ltd. Zearou Yang Regulatory Affairs Manager No. 1388 Cangxing Street, Cangqian Community, Yuhang District Hangzhou City, Zhejiang 311121 China
Re: K213683
Trade/Device Name: Promisemed Fine Biopsy Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW
Dear Zearou Yang:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated April 22, 2022. Specifically, FDA is updating this SE Letter with corrected Indications for Use as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Long Chen, Ph.D., OHT4: Office of Surgical and Infection Control Devices, 301-796-6389, long.chen@fda.hhs.gov.
Sincerely,
Digitally signed by Long H. Chen
Long H. Chen -S-s
Date: 2022.05.09 14:58:54 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 22, 2022
Promisemed Hangzhou Meditech Co., Ltd. Mr. Zearou Yang Regulatory Affairs Manager No. 1388 Cangxing Street, Cangqian Community, Yuhang District Hangzhou City, Zhejiang 311121 China
Re: K213683
Trade/Device Name: Promisemed Fine Biopsy Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsv instrument Regulatory Class: Class II Product Code: KNW Dated: Februry 11, 2022 Received: February 24, 2022
Dear Mr. Yang:
We have reviewed your Section 510/k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A ) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Long H. Chen Long H. Chen -S -s Date: 2022.04.22 14:15:46 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213683
Device Name Promisemed Fine Biopsy Needle
| Indications for Use (Describe) | It is to be used for taking cytological and histological biopsies of soft tissue. |
|--------------------------------|-----------------------------------------------------------------------------------|
|--------------------------------|-----------------------------------------------------------------------------------|
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size:16px">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">❍</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### 1. Date Prepared
April 25, 2022.
#### 2. Submitter's Information
### Name of Sponsor:
Promisemed Hangzhou Meditech Co., Ltd.
### Address:
No. 1388 Cangxing Street, Cangqian Community, Yuhang District, Hangzhou City, 311121 Zhejiang, China
### Contact Name: Zearou Yang
### Telephone No.:
+86 571 88772985
Fax No.:
+86 571 88772985
#### Email Address:
zearou.yang@promisemed.ca
### 3. Trade Name, Common Name, Classification
Trade/Product Name: Promisemed Fine Biopsy Needle
Common Name: Fine Biopsy Needle
Classification name: Instrument, biopsy
Regulation Number: 876.1075
Device Class: Class II
Product Code: KNW
#### 4. Identification of Predicate Device
K970872: Pan® Aspirating Needle (Chiba)
### 5. Description of the Device
The device is a hand-operated, non-electronic, with Chiba tip for tissue pathological examination/testing.
- -Trasparent Luer Lock connector.
- Sharpened cannula for a safe and painless insertion and penetration. -
{5}------------------------------------------------
- External echogenic treatment for a correct positioning under ultrasound guidance.
- -Centimeter markings and sliding stopper.
- -Sterilized by ETO, shelf life 5 years.
lt's a relatively invasive procedure and is performed by radiologist under guidance of imaging techniques such as ultrasound, X-ray.
This is a single-use device.
Not intended for use in bone.
## 6. Indication
It is to be used for taking cytological and histological biopsies of soft tissue.
# 7. Similarities and Differences of the Proposed Devices to the Predicate Devices
Promisemed Fine Biopsy Needle is substantially equivalent to the predicate device, Pan Aspirating Needle (Chiba), K970872 in that these devices have same intended use and technological characteristics. The basic technological and operating principles are the same for both devices. Both the subject and predicate devices are disposable, sterile, single patient use devices. The differences between the subject device and predicate device do not affect the basic design principle, usage of the subject device.
| Items | Subject Device | Predicate Device<br>(K970872) | Comments |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Trade Name | Promisemed Fine Biopsy<br>Needle | Pan® Aspirating Needle (Chiba) | |
| Manufacturer | Promisemed Hangzhou<br>Meditech Co., Ltd | Gallini International, Inc. | |
| Device Class | Class II | Class II | Same |
| Product Code | KNW | KNW | Same |
| Regulation number | 876.1075 | 876.1075 | Same |
| Regulation Name | Instrument, biopsy | Instrument, biopsy | Same |
| Intended Use/<br>Indications for<br>Use | It is to be used for taking cytological<br>and histological biopsies of soft<br>tissue. | It is to be used for taking cytological<br>and histological biopsies of soft tissue. | Same |
| Operati<br>ng<br>Principl<br>e | It's a relatively invasive<br>procedure and is performed by<br>radiologist under guidance of<br>imaging techniques such as<br>ultrasound, X-ray. | It's a relatively invasive procedure<br>and is performed by radiologist<br>under guidance of imaging<br>techniques such as ultrasound, X-<br>ray. | Same |
| Materials | - Needle: Stainless Steel<br>(X5CrNi18-10);<br>- Needle base: Acrylonitrile<br>Butadiene Styrene (ABS); | - Needle: Stainless Steel<br>(X5CrNi18-10);<br>- Needle base: Acrylonitrile<br>Butadiene Styrene (ABS); | Same |
A detailed comparison to the predicate is provided in Table 1.
{6}------------------------------------------------
| | - Depth stop: Thermoplastic Elastomer (TPE); | - Depth stop: Thermoplastic Elastomer (TPE); | 510(K) Submission, K213083 | | | | |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|------|
| Reuse durability | Single Use | Single Use | Same | | | | |
| Performance: | | | | | | | |
| Echogenic | Yes | Yes | Same | | | | |
| Connector transparency | Transparent/lock | Transparent/lock | Same | | | | |
| Needle Gauge | 16G, 18G, 19G, 20G, 21G, 22G, 23G | 18G, 20G, 21G, 22G, 23G, 25G | Subject device has two gauge sizes (16G, 19G) more than predicate device.<br>The 16G OD of needle tube which providing more samples volume and reduce operation time during clinical procedure, therefore, introduce any new concerns for safety or efficacy. | | | | |
| Needle Length | 90mm, 100mm, 150mm, 200mm | 50mm, 80mm, 90mm, 150mm, 200mm | Different<br>Subject device has less needle length than predicate device. | | | | |
| Inner needle hub color | Per ISO 6009:<br>White/16G, Pink/18G, Cream/19G, Yellow/20G, Light green/21G, Black/22G, Light blue/23G. | Per ISO 6009:<br>Pink/18G, Yellow/20G, Light green/21G, Black/22G, Light blue/23G, Orange/25G | Each color is in accordance with ISO 6009. | | | | |
| Needle tube tip type | Chiba | Chiba | Same | | | | |
| Bond between outer needle tube and hub | according to the nominal OD of the outer needle, the outer needle and hub shall not break when pull the needle with the axial static force specified in the below table for 10s.<br>OD Force /N 23G 34 | | | | | | Same |
| Bond between inner needle tube and hub | The inner needle and hub shall not break when pull the needle with the axial static force specified in the below table for 10s.<br>OD Force /N 23G 10 | | | | | | Same |
| Needle tube performance | Per ISO 9626: | Per ISO 9626: | Same | | | | |
{7}------------------------------------------------
| | | | 510(k) Submission, K213083 |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Fine Biopsy Needle | - Resistance to breakage;<br>- Resistance to corrosion; | - Resistance to breakage;<br>- Resistance to corrosion; | |
| Scale identification | The scale line shall form a ring on the<br>outside of the needle tube and be<br>clearly discernible. The distance<br>from the tip of the inner needle to<br>the front edge of the first scale line<br>shall be 10mm±1.5mm as well as any<br>two adjacent tick marks | The scale line shall form a ring on the<br>outside of the needle tube and be<br>clearly discernible. The distance from<br>the tip of the inner needle to the front<br>edge of the first scale line shall be<br>10mm±1.5mm as well as any two<br>adjacent tick marks | Same |
| Lucer lock<br>Connector<br>performance | Per ISO 80369-7:<br>- Positive pressure liquid leakage;<br>- Sub-atmospheric pressure air<br>leakage;<br>- Stress cracking;<br>- Resistance to separation from<br>axial load;<br>- Resistance to separation from<br>unscrewing;<br>- Resistance to overriding; | Per ISO 80369-7:<br>- Positive pressure liquid leakage;<br>- Sub-atmospheric pressure air<br>leakage;<br>- Stress cracking;<br>- Resistance to separation from axial<br>load;<br>- Resistance to separation from<br>unscrewing;<br>- Resistance to overriding; | Same |
| Sterilizatio<br>n method | EO Sterilization | EO Sterilization | Same |
| Shelf life | 5 years | 5 years | Same |
| Biocompatibility | Complied with ISO10993 series<br>standards, and the following<br>tests are performed<br>- Cytotoxicity: No cytotoxicity;<br>- Skin Irritation: No evidence of<br>skin irritation;<br>- Skin Sensitization: No evidence<br>of sensitization;<br>- Acute Systemic Toxicity: No<br>systemic toxicity; | Complied with ISO10993 series<br>standards, and the following tests<br>are performed<br>- Cytotoxicity: No cytotoxicity;<br>- Skin Irritation: No evidence of<br>skin irritation;<br>- Skin Sensitization: No evidence<br>of sensitization;<br>- Acute Systemic Toxicity: No<br>systemic toxicity; | Same |
| | - Pyrogen: Non pyrogenic: | - Pyrogen: Non pyrogenic: | |
#### Performance Testing Summary 8.
The bench testing performed verifies that the performance of the subject device is substantially equivalent in terms of critical performance characteristics to the predicate device. These tests include:
- Appearance;
- Cleanliness;
- Color code;
- Dimensions;
- Needle point;
- Patency of lumen;
- Tip structure;
{8}------------------------------------------------
- Bond between outer needle tube and hub;
- Bond between inner needle tube and hub;
- Sheath;
- Stiffness;
- Resistance to breakage;
- Resistance to corrosion;
- Scale identification;
- Positive pressure liquid leakage;
- Sub-atmospheric pressure air leakage;
- Stress cracking;
- Resistance to separation from axial load;
- Resistance to separation from unscrewing;
- Resistance to overriding;
The performance test results are indicated that performance of Fine Biopsy Needles meet the specific requirements. We concluded the proposed devices are substantially equivalent to the identified predicate devices in general appearance, dimension, needle tube performance, color, scale and connector.
#### 9. Conclusion
Based on the information provided within this 510(k) submission, proposed subject device is substantially equivalent to the predicate device and is as safe, as effective and performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.