K243811 · Erchonia Corporation · OLI · Jan 10, 2025 · General, Plastic Surgery
Device Facts
Record ID
K243811
Device Name
Erchonia Zerona® VZ8
Applicant
Erchonia Corporation
Product Code
OLI · General, Plastic Surgery
Decision Date
Jan 10, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Erchonia Zerona® VZ8 is indicated for use as a non-invasive dermatological aesthetic treatment for the reduction of body circumference. Zerona® VZ8 is intended to be distributed for Over-the-Counter (OTC) use.
Device Story
Zerona® VZ8 is a floor-model, low-level laser system for non-invasive body circumference reduction. Device uses eight 405nm laser diodes to emit light, which is collected and processed through a proprietary lens and line refractor to create a scanned line pattern. System operates by emulsifying adipose tissue, causing release of fat/lipids into interstitial space. Intended for OTC use; patient positions target areas 3-4 inches from laser arms. Treatment protocol involves 6 sessions over 2 weeks (3 sessions/week, every other day). Device features LCD touchscreen interface. Benefits include non-invasive fat reduction without surgery, pain, wounds, or recovery garments, allowing patients to maintain daily activities. Safety and efficacy are supported by clinical foundation of predecessor devices (Violet ZERONA® Z6 OTC and Emerald).
Clinical Evidence
No clinical study results submitted. Substantial equivalence supported by bench testing (IEC 60601-1, IEC 60601-1-2, IEC 60825-1) and clinical data from predicate devices.
Technological Characteristics
Low-level laser system; 8 laser diodes (405nm ±10nm); 23mW (±2mW) output per head; proprietary lens and line refractor; floor-model with adjustable arms; LCD touchscreen; mains power converted to DC; non-sterile; complies with IEC 60601-1, IEC 60601-1-2, IEC 60825-1, and 21 CFR 1040.10/11.
Indications for Use
Indicated for non-invasive dermatological aesthetic treatment for the reduction of body circumference in patients seeking non-surgical fat reduction.
Regulatory Classification
Identification
A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 10, 2025
Erchonia Corporation % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K243811
Trade/Device Name: Erchonia Zerona® VZ8 Regulation Number: 21 CFR 878.5400 Regulation Name: Low Level Laser System For Aesthetic Use Regulatory Class: Class II Product Code: OLI Dated: December 11, 2024 Received: December 11, 2024
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510/k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by TANISHA TANISHA L. HITHE -S L. HITHE -> Date: 2025.01.10
12:00:56 -05 12:00:56 -05'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K243811
Device Name Erchonia Zerona® VZ8
Indications for Use (Describe)
The Erchonia Zerona® VZ8 is indicated for use as a non-invasive dermatological aesthetic treatment for the reduction of body circumference.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K243811
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
## Owner Information
#### Name and Address of Sponsor / Manufacturer
Erchonia Corporation 112 Southchase Blvd. Fountain Inn, SC 29644 Telephone: 888-242-0571 Fax: 321-473-1608
#### Establishment Registration Number 2032513
## Name and Address of Official Correspondent
Erchonia Corporation 112 Southchase Blvd. Fountain Inn, SC 29644 Contact: Travis Sammons Telephone: 888-242-0571 Fax: 321-473-1608 Email: tsammons@erchonia.com
#### Date Prepared
11/01/2024
#### Device Information
Trade Name: Erchonia Zerona® VZ8 Model#: VZ8 Common Name: Fat Reducing Low Level Laser Classification Name: Low level laser system for aesthetic use Requlation Description: Low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use. (21 CFR 878.5400) Classification: Class II Panel: General & Plastic Surgery Product Code: OLI
## Predicate Device
Primary Predicate Trade Name: Erchonia® Violet ZERONA® Z6 OTC 510K: K231474 Common Name: Fat Reducing Low Level Laser Classification Name: Low level laser system for aesthetic use. (21 CFR 878.5400) Product Code: OLI
Secondary Predicate
{5}------------------------------------------------
Trade Name: Erchonia® Emerald 510K: K192544 Common Name: Fat Reducing Low Level Laser Classification Name: Low level laser system for aesthetic use. (21 CFR 878.5400) Product Code: OLI
# Device Description
The Zerona® VZ8 laser is designed for clients seeking noninvasive circumference reduction without invasive surgery. Zerona® VZ8 allows the patient to continue their daily activities without interruptions from surgery, pain, wounds or garments. The Zerona® VZ8 works by emulsifying adipose tissue which then releases into the interstitial space. Zerona VZ8 was built on the clinical foundation of its predecessors, Violet ZERONA® Z6 and ZERONA® Z6 OTC, which were proven through clinical studies to be safe and effective in the application of circumference reduction.
The Zerona® VZ8 is identical to the predecessor device, the Erchonia Violet ZERONA® Z6 OTC cleared under K231474, with the difference in the number of laser diodes (eight compared to six). The Installation and Proper Use Reference Guide for use for the Zerona® VZ8 is identical to the predecessor laser, the Erchonia® Violet ZERONA® Z6 OTC: consisting of the similar treatment set-up and same treatment protocol involving 6 treatments occurring over 2 weeks: 3 treatments per week; each treatment every other day.
The Zerona® VZ8 emits a 405-nanometer wavelength with a tolerance of ±10 nanometer, from each of the eight specially created and patented electronic diodes. Laser devices are typically constructed to emit a "spot" of light. The Zerona® VZ8 laser utilizes internal mechanics that collects the light emitted from the laser diode and processes it through a proprietary patented lens, and then redirects the beam with a line refractor. The laser applicator heads, each produce an output power of 23mW (± 2mW) measured. Laser diodes and adjustable laser arms are positioned no greater than 3-4 inches away from the client's target treatment areas.
The software incorporated into the operation of the Zerona® VZ8 complies with FDA and ISO Software Development and Validation regulations.
## Intended Use
Zerona VZ8 is indicated for use as a non-invasive dermatological aesthetic treatment for the reduction of body circumference.
Zerona® VZ8 is intended to be distributed for Over-the-Counter (OTC) use.
# Comparison of Technological Characteristics with the Predicate Device(s)
The Erchonia® Zerona® VZ8 is substantially equivalent to both its primary and secondary predicate devices. The primary predicate device, the Erchonia® Violet ZERONA® Z6 OTC (cleared under K231474), is the predecessor to the newly designed Zerona® VZ8 (subject device). Both devices share the same operational principles, including wavelength, power, energy source, and delivery method. The sole technological distinction between the subject and primary predicate device is the incorporation of two additional diode lasers, which does not compromise substantial equivalence, safety, or effectiveness, nor does it introduce new safety or effectiveness concerns. Additionally, the Erchonia® Zerona® VZ8 is substantially equivalent to the secondary predicate device, the Erchonia® Emerald. Both devices share an identical design, energy source, number of diodes, and total fluence.
{6}------------------------------------------------
| Device | Erchonia Zerona® VZ8 | Erchonia® Violet ZERONA®<br>Z6 OTC | Erchonia® Emerald |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | Unknown | K231474 | K192544 |
| | Subject Device | Primary Predicate | Secondary Predicate |
| Manufacturer | Erchonia | Erchonia | Erchonia |
| Device Design | Image: Erchonia Zerona VZ8 device | Image: Erchonia Violet ZERONA Z6 OTC device | Image: Erchonia Emerald device |
| Indication for<br>Use | The Zerona® VZ8 is indicated for use<br>as a non-invasive dermatological<br>aesthetic treatment for the reduction of<br>body circumference. | The Violet ZERONA Z6 OTC Laser is<br>indicated for use as a non-invasive<br>dermatological aesthetic treatment for the<br>reduction of body circumference. | Erchonia® Emerald (Model# SHL)<br>Laser is indicated for use as a non-<br>invasive dermatological aesthetic<br>treatment for the reduction of body<br>circumference in individuals with a<br>Body Mass Index (BMI) up to 40 kg/m2 |
| Laser Power | $23mW \pm 2mW$ | $23mW \pm 2mW$ | $16mW \pm 1.25mW$ |
| Wavelength | 400nm to 415nm | 400nm to 415nm | 522nm to 542 nm |
| Energy<br>Source | Multi diode collected then line<br>dispersed (coherent) | Multi diode collected then line dispersed<br>(coherent) | Multi diode collected then line<br>dispersed (coherent) |
| Number of<br>Laser Diodes | 8 | 6 | 10 |
| Energy<br>Delivery | Floor model device with probe head | Floor model device with probe head | Floor model device with probe head |
| Total<br>Treatment<br>Time Applied | 20 minutes | 20 minutes | 30 minutes |
| Total Fluency | 221 J | 165 J | 288 J |
| Target Size<br>Per Diode | Line pattern, electronically scanned<br>over area of treatment | Line pattern, electronically scanned over<br>area of treatment | Line pattern, electronically scanned<br>over area of treatment |
| User<br>Interface | LCD Touchscreen | LCD Touchscreen | LCD Touchscreen |
| Principles of<br>Operation | Mains power, converted to DC,<br>powering semi-conductor diodes | Mains power, converted to DC, powering<br>semi-conductor diodes | Mains power, converted to DC,<br>powering semi-conductor diodes |
| Mechanism of<br>Action | Low level light energy used as an<br>adjunct to emulsify adipose tissue | Low level light energy used as an adjunct<br>to emulsify adipose tissue | Low level light energy used as an<br>adjunct to emulsify adipose tissue |
| Product Code | OLI | OLI | OLI |
{7}------------------------------------------------
## Performance Standards
The Erchonia® Zerona® VZ8 complies with FDA's performance standards for light-emitting products (21CFR 1040.10 and 21 CFR 1040.11 by Laser Notice #50).
## Sterilization and Shelf-Life
The device is not provided sterile.
The device is not affected by shelf-life because it is an electro-mechanical device that is not sterile and whose components will not degrade over time while simply sitting in storage prior to initial use.
## Performance Testing-Animal
No animal testing conducted
## Performance Testing-Clinical
No clinical study results are being submitted as part of this submission
#### Risk Assessment
The Erchonia® Zerona® VZ8 is acceptable in accordance with IEC 60601 and IEC 60825, by virtue of Engineering and third-party verification. All identified risks have been mitigated to ensure the lowest acceptable risk possible using the ISO 14971 standard framework.
#### Performance Data
Safety and EMC testing was conducted on the Erchonia® Zerona® VZ8. The device complies with the IEC 60601-1, IEC 60601-1-2 and IEC 60825-1 standards.
#### Compliance with Voluntary Standards
The Erchonia® Zerona® VZ8 complies with the following voluntary standards: IEC 60601-1-2:2014+AMD1:2020 Edition 4.1 IEC 60601-1:2005+AMD1:2012+AMD2:2020 Edition 3.2 IEC 60825-1:2014 Edition 3.0
## Substantially Equivalent Discussion
The Erchonia Zerona® VZ8 is substantially equivalent to its predecessor, the Erchonia® Violet ZERONA® Z6 OTC, cleared under K231474. Both devices share the same laser wavelength, power, energy source, and treatment duration. The sole technological distinction between the Zerona® VZ8 (subject device) and the Violet ZERONA® Z6 OTC (predicate device) is the incorporation of two additional laser diodes [8 vs 6], resulting in improved laser coverage of the treatment area. The safety and efficacy of employing additional laser diodes has been substantiated by the Erchonia Emerald (secondary predicate device), cleared under K192544. Sharing a comparable design with the subject device, the Erchonia Emerald was supported by safety and efficacy data from a full-scale controlled and powered clinical trial utilizing ten laser diodes compared to the subject device's eight. Moreover, the subject device's total fluence [221J] remains within the safety and efficacy parameters established by the secondary predicate [288J].
## Conclusion
Any differences between the subject device and predicate do not render the device not substantially equivalent, do not affect safety or effectiveness, or raise different questions of safety and effectiveness. Therefore, the Erchonia Zerona® VZ8 (subject device) is substantially equivalent to the Erchonia® Violet ZERONA® Z6 OTC (predicate device), indicated for use as a non-invasive dermatological aesthetic treatment for the reduction of body circumference.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.