K231474 · Erchonia Corporation · OLI · Jul 21, 2023 · General, Plastic Surgery
Device Facts
Record ID
K231474
Device Name
Erchonia Violet ZERONA® Z6 OTC
Applicant
Erchonia Corporation
Product Code
OLI · General, Plastic Surgery
Decision Date
Jul 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Violet ZERONA® Z6 OTC is indicated for use as a non-invasive dermatological aesthetic treatment for the reduction of body circumference. Violet ZERONA® Z6 OTC is intended to be distributed for Over-the-Counter (OTC) use.
Device Story
The Erchonia Violet ZERONA® Z6 OTC is a floor-model, low-level laser system for non-invasive body circumference reduction. It utilizes six electronic diodes emitting 405nm violet light to emulsify adipose tissue, releasing lipids into the interstitial space. The device features adjustable arms and a proprietary lens/line refractor system to scan light over a 516 cm² treatment area. It is intended for over-the-counter use, with patients undergoing a protocol of six 20-minute treatments over two weeks. The device is operated via an LCD touchscreen. By reducing treatment time from 40 to 20 minutes compared to its predecessor, it offers improved patient convenience while maintaining equivalent clinical efficacy in circumference reduction.
Clinical Evidence
Open-label, single-arm clinical study (n=25, 84% female, mean age 49.96, mean BMI 26.43). Subjects received six 20-minute treatments over two weeks. Primary endpoint: change in combined waist-hips-bilateral thighs circumference. Results showed a mean reduction of 3.91 inches (95% CI: 3.26 - 4.56), meeting the success criteria of -3.72±5% inches. No adverse events reported. Data confirms non-inferiority to the red-laser predicate (K162578) with reduced treatment time.
Technological Characteristics
Floor-model laser system; 6 semiconductor diodes; 405nm (±10nm) wavelength; 23mW (±2mW) output power per head; proprietary lens and line refractor; LCD touchscreen interface; mains power converted to DC; complies with IEC 60601-1, IEC 60601-2, IEC 60825-1, and 21 CFR 1040.10/11.
Indications for Use
Indicated for non-invasive dermatological aesthetic treatment for the reduction of body circumference in patients seeking non-surgical fat reduction.
Regulatory Classification
Identification
A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
July 21, 2023
Erchonia Corporation Travis Sammons Clinical Affairs Manager 650 Atlantis Rd. Melbourne, Florida 32904
Re: K231474
Trade/Device Name: Erchonia Violet ZERONA® Z6 OTC Regulation Number: 21 CFR 878.5400 Regulation Name: Low Level Laser System For Aesthetic Use Regulatory Class: Class II Product Code: OLI Dated: May 18, 2023 Received: May 22, 2023
Dear Travis Sammons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jianting Wang -S
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231474
Device Name Violet ZERONA® Z6 OTC
Indications for Use (Describe)
The Violet ZERONA® Z6 OTC is indicated for use as a non-invasive dermatological aesthetic treatment for the reduction of body circumference.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary K231474
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
# Owner Information Name and Address of Sponsor / Manufacturer
Erchonia Corporation 650 Atlantis Road Melbourne, FL 32904 Telephone: 321-473-1251 Fax: 321-473-1608
# Establishment Registration Number
2032513
# Name and Address of Official Correspondent
Erchonia Corporation 650 Atlantis Road Melbourne, FL 32904 Contact: Travis Sammons Telephone: 321-473-1251 Fax: 321-473-1608 Email: tsammons@erchonia.com
## Date Prepared
7/21/2023
## Device Information
Trade Name: Erchonia Violet ZERONA® Z6 OTC Model#: VZ6 Common Name: Fat Reducing Low Level Laser Classification Name: Low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use. (21 CFR 878.5400) Classification: Class II Panel: General & Plastic Surgery Product Code: OLI
## Predicate Device
Erchonia ZERONA® Z6 OTC previously cleared under K162578.
## Reference Device
Erchonia Zerona® Z6 OTC cleared under K143007.
The procedure administration protocol performed with the subject device, Erchonia Violet ZERONA® Z6 OTC, is identical to that evaluated in the comparative study which supported 510(k) clearance for K143007, with the difference in this current study being the use of a violet rather than red laser and a reduction in per-treatment administration time from 40 minutes to 20 minutes. Both the subject and reference devices administered the same treatment protocol of 6 treatments occurring over 2 weeks.
{4}------------------------------------------------
#### Device Description
The Violet ZERONA® Z6 OTC laser is designed for client's seeking noninvasive circumference reduction without invasive surgery. Violet ZERONA® Z6 OTC allows the patient to continue their daily activities without interruptions from surgery, pain, wounds or garments. The Violet ZERONA® Z6 OTC works by emulsifying adipose tissue which then releases into the interstitial space. The Violet ZERONA® Z6 was built on the clinical foundation of its predecessors, Zerona® and Zerona®-AD, and ZERONA® Z6 OTC, which were proven through clinical studies to be safe and effective in the application of circumference reduction.
The Erchonia Violet ZERONA® Z6 OTC is identical to the predecessor device, the Erchonia ZERONA® Z6 OTC cleared under K162578, with the differences being light output (visible violet compared to visible red) and duration of treatment administrations (20 minutes compared to 40 minutes, which is automatically programmed into the device). The Installation and Proper Use Reference Guide for use for the Erchonia® Violet ZERONA® Z6 OTC are identical to those for use of the predecessor laser, the Erchonia® ZERONA® Z6 OTC; consisting of the same device set-up and same treatment protocol involving 6 treatments occurring over 2 weeks: 3 treatments per week; each treatment every other day.
The Violet ZERONA® Z6 OTC emits a 405-nanometer wavelength with a tolerance of ±10 nanometer, from each of the six specially created and patented electronic diodes. Laser devices are typically constructed to emit a "spot" of light. The Violet ZERONA® Z6 OTC laser utilizes internal mechanics that collects the light emitted from the laser diode and processes it through a proprietary patented lens, and then redirects the beam with a line refractor. The laser applicator heads, each produce an output power of 23mW (± 2mW) measured. Laser diodes and adjustable laser arms are positioned no greater than 3-4 inches away from the client's target treatment areas.
The software incorporated into the operation of the Violet ZERONA® Z6 OTC complies with FDA and ISO Software Development and Validation regulations.
#### Intended Use
The Violet ZERONA® Z6 OTC is indicated for use as a non-invasive dermatological aesthetic treatment for the reduction of body circumference.
Violet ZERONA® Z6 OTC is intended to be distributed for Over-the-Counter (OTC) use.
{5}------------------------------------------------
# Comparison of Technological Characteristics with the Predicate Device
The following table compares the subject device (Erchonia Violet ZERONA® Z6 OTC) to the predicate device (Erchonia ZERONA® Z6 OTC). The comparison provides detailed information regarding the basis for the determination of substantial equivalence.
| Device | Erchonia Violet ZERONA® Z6 OTC | Erchonia ZERONA® Z6 OTC | |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| 510(k) # | N/A | K162578 | Differences |
| | Subject Device | Predicate Device | |
| Device Design | Image: Erchonia Violet ZERONA® Z6 OTC | Image: Erchonia ZERONA® Z6 OTC | None |
| Indication for<br>Use | The Violet ZERONA Z6 OTC Laser is indicated<br>for use as a non-invasive dermatological aesthetic<br>treatment for the reduction of body<br>circumference. | The ZERONA Z6 OTC Laser is indicated for use<br>as a non-invasive dermatological aesthetic<br>treatment for the reduction of body<br>circumference. | None |
| Laser Power | $23mW \pm 2mW$ | $17.25mW \pm 1.25mW$ | Clinical study completed<br>(refer to Performance Testing<br>-Clinical) |
| Wavelength | Violet: 400nm- 415nm | Red: 630nm – 640nm | Clinical study completed<br>(refer to Performance Testing<br>-Clinical) |
| Energy Source | Multi diode collected then line dispersed<br>(coherent) | Multi diode collected then line dispersed<br>(coherent) | None |
| Number of<br>Laser Diodes<br>Applied Per<br>Treatment Area | 6 | 6 | None |
| Energy Delivery | Floor model device with probe head | Floor model device with probe head | None |
| Total<br>Treatment<br>Time Applied | 20 minutes | 40 minutes | Clinical study completed<br>(refer to Performance Testing<br>-Clinical) |
| Treatment<br>Frequency | 3 x week, 2 weeks | 3 x week, 2 weeks | None |
| Total Fluency | 165 J | 248 J | Clinical study completed<br>(refer to Performance Testing<br>-Clinical) |
| Target Size Per<br>Diode | Line pattern, electronically scanned over area of<br>treatment (516 cm²) | Line pattern, electronically scanned over area of<br>treatment (516 cm²) | None |
| User Interface | LCD Touchscreen | LCD Touchscreen | None |
| Principles of<br>Operation | Mains power, converted to DC, powering semi-<br>conductor diodes | Mains power, converted to DC, powering semi-<br>conductor diodes | None |
| Mechanism of<br>Action | Low level light energy used as an adjunct to<br>emulsify adipose tissue | Low level light energy used as an adjunct to<br>emulsify adipose tissue | None |
| Product Code | OLI | OLI | None |
{6}------------------------------------------------
## Performance Standards
The Erchonia Violet ZERONA® Z6 OTC complies with FDA's performance standards for light-emitting products (21CFR 1040.10 and 21 CFR 1040.11 by Laser Notice #50).
#### Risk Assessment
The Erchonia Violet ZERONA® Z6 OTC is acceptable in accordance with IEC 60601 edition 3.1. by virtue of Engineering and third-party verification. All identified risks have been mitigated to ensure the lowest acceptable risk possible using the ISO 14971 standard framework.
#### Electromagnetic Compatibility and Electrical Safety
The Erchonia Violet ZERONA® Z6 OTC utilizes the same components as the predicate device, Erchonia ZERONA® Z6 OTC, with the exception of the laser diode. The Erchonia Violet ZERONA® Z6 OTC emits violet light and the Erchonia ZERONA® Z6 OTC emits red light, the change does not alter the safety and EMC testing previously conducted on the Erchonia ZERONA® Z6 OTC. Therefore, the safety and EMC testing which was conducted on the Erchonia ZERONA® Z6 OTC device applies to the Erchonia Violet ZERONA® Z6 OTC. The device testing complies with the current IEC 60601-1, IEC 60601-2 and IEC 60825-1 standards.
#### Compliance with Voluntary Standards
The Erchonia Violet ZERONA® Z6 OTC complies with the following voluntary standards: IEC 60601-1-2:2014 Edition 4.0 IEC 60601-1:2005 Edition 3.1 IEC 60825-1:2014 Edition 3.0
#### Performance Testing-Clinical
BACK GROUND: The purpose of this clinical study was to determine the effectiveness of the Erchonia Violet ZERONA® Z6 OTC laser device, in providing noninvasive body circumference reduction.
STUDY DESIGN: This clinical study was an open-label single-arm design to evaluate the efficacy of application of the Erchonia Violet ZERONA® Z6 OTC violet laser application to that of application of the Erchonia® ZERONA Z6 OTC red laser for the reduction of body circumference.
SUBJECTS: Twenty-five (25) subjects, 84% of whom were female completed the study. Mean subject age was 49.96 years, mean baseline body mass index (BMI) was 26.43kg/m², and mean baseline combined waist-hips-bilateral thighs circumference was 123.16.
STUDY PROCEDURES: Each subject received six (6) treatments with the Erchonia Violet ZERONA® Z6 OTC across a two-week period. The study procedure was identical to that evaluated in the comparative study, with the difference being evaluated in this study being use of a violet rather than red laser and a reduction in per-treatment administration time from 40 minutes to 20 minutes.
STUDY MEASURES: The study primary outcome measure of combined waist-hips-bilateral thighs circumference (inches) was measured at baseline and at completion of the two-week treatment phase (study endpoint).
STUDY RESULTS: Success for the current study group was pre-determined as mean change in combined circumference measurements of -3.72±5% inches (-3.53 to -3.91 inches). The mean change in total body circumference measurement at study endpoint relative to baseline was -3.91 inches, falling within the prespecified maximally clinically acceptable difference range (-3.53 to -3.91 inches), and exceeding the preestablished lower boundary of -3.53 inches; thereby establishing study primary success.
{7}------------------------------------------------
SAFETY: No adverse events were reported or observed for any subject across study duration.
CONCLUSION: The study results demonstrate that application of the Erchonia Violet ZERONA® Z6 OTC violet laser is as effective in providing clinically meaningful circumference reduction as the Erchonia ZERONA® Z6 OTC red laser, with half the required per-treatment administration time (from 40 to 20 minutes), which is a significant time savings to the patient.
# Substantially Equivalent Discussion
The Erchonia Violet ZERONA® Z6 OTC is substantially equivalent to the predicate device, the Erchonia ZERONA® Z6 OTC previously submitted under K162578. Both the subject device and the predicate device have identical device design, number of diodes, energy delivery, treatment frequency, and mechanism of action. The differences in laser diodes (wavelength and laser power) and treatment time applied per area, is justified in the outcome of the clinical study that demonstrates the difference does not render the device not substantially equivalent, does not affect the safety or effectiveness, or raise different questions of safety and effectiveness.
The clinical data provided in this submission demonstrates that the application of the Erchonia Violet ZERONA® Z6 OTC (violet laser) maintains the same safety profile and is as effective in providing clinically meaningful circumference reduction as the Erchonia ZERONA® Z6 OTC (red laser), with half the required per-treatment administration time, which is a significant time savings to the patient. (See Table 1)
| | Erchonia Violet<br>ZERONA® Z6 OTC | Erchonia ZERONA® Z6<br>OTC |
|----------------------------------------------|-----------------------------------|----------------------------|
| Laser Diode | Violet | Red |
| n | 25 | 22 |
| Total Treatment Time Applied | 20 minutes | 40 minutes |
| Combined Circumference<br>Reduction (inches) | 3.91 (95% CI: 3.26 - 4.56) | 3.72 (95% CI: 2.78 – 4.66) |
| Side Effects | None | None |
Table 1: Combined Circumference Reduction Across Study Duration by Device
## Conclusion
Any differences between the subject device and predicate do not render the device not substantially equivalent, do not affect safety or effectiveness, or raise different questions of safety and effectiveness. Therefore, the Erchonia Violet ZERONA® Z6 OTC (subject device) is substantially equivalent to the Erchonia ZERONA® Z6 OTC (predicate device), indicated for use as a non-invasive dermatological aesthetic treatment for the reduction of body circumference.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.