Niti-S Esophageal Stent; Esophageal TTS Stent

K243619 · Taewoong Medical Co., Ltd. · ESW · Oct 30, 2025 · General, Plastic Surgery

Device Facts

Record IDK243619
Device NameNiti-S Esophageal Stent; Esophageal TTS Stent
ApplicantTaewoong Medical Co., Ltd.
Product CodeESW · General, Plastic Surgery
Decision DateOct 30, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3610
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Niti-S Esophageal Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas. The Esophageal TTS Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas.

Device Story

Device consists of implantable Nitinol wire stent and disposable flexible introducer system. Stent restores esophageal structure/function by exerting outward radial force on luminal surface. Suture attached to one end allows for repositioning during placement. Used in clinical setting by physicians for management of malignant esophageal strictures and fistulas. Introducer system facilitates delivery and deployment. Stent remains in situ to maintain patency. Benefits include relief of obstruction and closure of fistulas. Subject device is identical to predicates except for change in suture material from Nylon to Polyester.

Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, intracutaneous reactivity, acute/subchronic systemic toxicity, pyrogenicity, genotoxicity, and subcutaneous implantation. Performance testing included shelf-life and tensile testing.

Technological Characteristics

Implantable Nitinol wire stent in diamond shape; lengths 60-150mm; midsection diameters 16-22mm; dumbbell diameters 24-28mm. Coverage: Silicone full covered or bare. Radiopaque markers: 8 platinum-iridium, 2 316L stainless steel. Suture: Polyester. Delivery system: Co-axial tube type (OTW or TTS). Sterilization: Ethylene Oxide (EO). No electrical components or software.

Indications for Use

Indicated for patients with esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas.

Regulatory Classification

Identification

An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.

Special Controls

The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION October 30, 2025 Taewoong Medical Co., Ltd. % Matthew Krueger Principal Consultant Biologics Consulting Group 100 Daingerfield Rd, Suite 400 Alexandria, Virginia 22314 Re: K243619 Trade/Device Name: Niti-S Esophageal Stent; Esophageal TTS Stent Regulation Number: 21 CFR 878.3610 Regulation Name: Esophageal Prosthesis Regulatory Class: Class II Product Code: ESW Dated: September 29, 2025 Received: September 29, 2025 Dear Matthew Krueger: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K243619 - Matthew Krueger Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K243619 - Matthew Krueger Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, SIVAKAMI VENKATACHALAM -S for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} Niti-S Esophageal Stent Page 9 of 46 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K243619 | ? | | Please provide the device trade name(s). | | ? | | Niti-S Esophageal Stent; Esophageal TTS Stent | | | | Please provide your Indications for Use below. | | ? | | The Niti-S Esophageal Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas. The Esophageal TTS Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} K243619 Page 1 of 7 Traditional 510(k) Summary Taewoong Medical Niti-S Esophageal Stent Esophageal TTS Stent 1. SUBMITTER INFORMATION Applicant: Taewoong Medical Contact: Yongjin Jeff Kim Phone: +82 70 4649 1543 Email: jinjeff@stent.net Address: 14, Gojeong-ro, Wolgot-myeon, Gimpo-si, Gyeonggi-do, 10022, Republic of Korea 2. CORRESPONDENT INFORMATION Contact: Matthew Krueger Title: Principal Consultant Firm: Biologics Consulting Group 3. DATE PREPARED: SEPTEMBER 25, 2024 4. DEVICE INFORMATION Device Name: Esophageal TTS Stent Niti-S Esophageal Stent Common Name: Prosthesis, Esophageal Regulation Number: 21 C.F.R. 878.3610 Regulation Name: Esophageal prosthesis Product Code: ESW Regulatory Class: Class II 5. PREDICATE DEVICE INFORMATION Device Name: Esophageal TTS Stent, Niti-S Esophageal Stent 510(k) Number: K240522, K080782 Manufacturer: Taewoong Medical The predicate devices have not been subject to a design related recall. {5} K243619 Page 2 of 7 Traditional 510(k) Summary Taewoong Medical Niti-S Esophageal Stent Esophageal TTS Stent # 6. DEVICE DESCRIPTION The Esophageal TTS Stents and Niti-S Esophageal Stents that are the subject of this 510(k) are identical to the devices cleared in K240522 and K080782, respectively, with the exception of the change in suture material for both stents. The indications for use for the Niti-S Esophageal Stent have also been updated to align with those of the Esophageal TTS Stents. The Esophageal TTS Stents and Niti-S Esophageal Stents consist of an implantable metallic stent and a disposable, flexible introducer system for placement of the stent. The stent is a flexible and expandable tubular device made of Nitinol wire that is intended to be implanted to restore the structure and/or function of the esophagus. The introducer is a disposable system for delivery and deployment of the stent at the target position. Upon deployment, the stent imparts an outward radial force on the luminal surface of the esophagus to establish patency. A length of suture is attached to one end of the stent and can be used for stent repositioning during initial placement, in accordance with the instructions for use. # 7. INDICATIONS FOR USE The Esophageal TTS Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas. The Niti-S Esophageal Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas. # 8. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The table below compares the intended use and the technological characteristics of the subject device and predicate device. Table 1: Comparator Table for Subject and Predicate Devices | | Subject Device | Predicate Device K080782, K240522 | Comments | | --- | --- | --- | --- | | Device Name | Niti-S Esophageal Stent, Esophageal TTS Stent | Niti-S Esophageal Stent (K080782), Esophageal TTS Stent (K240522) | Identical | | Common Name | Esophageal Stent | Esophageal Stent | Identical | | Manufacturer | Taewoong Medical Co., Ltd | Taewoong Medical Co., Ltd | Identical | | Product Code | ESW | ESW | Identical | | Regulation | 21 CFR 878.3610 Esophageal prosthesis | 21 CFR 878.3610 Esophageal prosthesis | Identical | {6} K243619 Page 3 of 7 Traditional 510(k) Summary Taewoong Medical Niti-S Esophageal Stent Esophageal TTS Stent | | Subject Device | Predicate Device K080782, K240522 | Comments | | --- | --- | --- | --- | | Indications for use | Niti-S Esophageal Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas. Esophageal TTS Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas. | (K080782) Niti-S Esophageal Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors. (K240522) Esophageal TTS Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas. | **Substantially Equivalent:** • Original use of K080782: esophageal strictures from malignant tumors • Now includes: “occlusion of concurrent esophageal fistulas”. This indication was previously cleared for the identical stent implant as cleared in K240522. • Stent Justification: o Same materials o Same design o Same purposes o Same target site o Same manufacturing process • Therefore, the Niti-S Esophageal Stent with the modified Indications for Use is Substantially Equivalent to the previously cleared Esophageal TTS Stent (K240522). | Page 3 of 7 {7} K243619 Page 4 of 7 Traditional 510(k) Summary Taewoong Medical Niti-S Esophageal Stent Esophageal TTS Stent | | Subject Device | Predicate Device K080782, K240522 | Comments | | --- | --- | --- | --- | | Design (Stents) | | | **Substantially Equivalent:** - Dimensionally the same - General design is the same - Only the suture material has changed from Nylon to Polyester - Performance testing, biocompatibility testing, and shelf life testing demonstrate that the change in suture material does not affect safety or effectiveness of the device. Thus, the stent with the modified suture material is substantially equivalent to the predicate device. | | | - Nitinol wire in diamond shape - Lengths: 60 mm, 80 mm, 100 mm, 120 mm, 140 mm, 150 mm - Midsection Diameters: 16 mm, 18 mm, 20 mm, 22 mm - Dumbbell Diameters: 24 mm, 26 mm, 28 mm - Coverage: Silicone Full Covered, Silicone Both Bare - Radiopaque Markers: 8 platinum-iridium markers and 2 316L stainless steel markers - Suture material: Polyester | - Nitinol wire in diamond shape - Lengths: 60 mm, 80 mm, 100 mm, 120 mm, 140 mm, 150 mm - Midsection Diameters: 16 mm, 18 mm, 20 mm, 22 mm - Dumbbell Diameters: 24 mm, 26 mm, 28 mm - Coverage: Silicone Full Covered, Silicone Both Bare - Radiopaque Markers: 8 platinum-iridium markers and 2 316L stainless steel markers - Suture material: Nylon | | Page 4 of 7 {8} K243619 Page 5 of 7 Traditional 510(k) Summary Taewoong Medical Niti-S Esophageal Stent Esophageal TTS Stent | | Subject Device | Predicate Device K080782, K240522 | Comments | | --- | --- | --- | --- | | Design (Introducer) | **Niti-S Esophageal Stent** OTW (Over-The-Wire) type Stent Delivery System Co-axial tube type Usable Length: 70cm Diameter: 16 Fr (5.3mm), 20 Fr (6.7mm) **Esophageal TTS Stent** TTS (Through-The-Scope) type Stent Delivery System Co-axial tube type Usable Length: 180 and 220cm Diameter: 10.5 Fr (3.5mm) | **Niti-S Esophageal Stent** (K080782) OTW (Over-The-Wire) type Stent Delivery System Co-axial tube type Usable Length: 70cm Diameter: 16 Fr (5.3mm), 20 Fr (6.7mm) **Esophageal TTS Stent** (K240522) TTS (Through-The-Scope) type Stent Delivery System | Identical | | Single Use | Yes | Yes | Identical | | Sterile | EO Sterilization | EO Sterilization | Identical | | Method of Placement | OTW (Over-The-Wire), TTS (Through-The-Scope) | OTW (Over-The-Wire), TTS (Through-The-Scope) | Identical | | Method of Deployment | Release by pulling outer sheath | Release by pulling outer sheath | Identical | | Packaging | PET tray with Tyvek lid placed in a Tyvek pouch and then placed inside of a Manila paper box | PET tray with Tyvek lid placed in a Tyvek pouch and then placed inside of a Manila paper box | Identical | ## 9. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING ### Biocompatibility Testing Biocompatibility testing per ISO 10993-1 and FDA Guidance was performed to demonstrate the biological safety of the subject stents with the new polyester suture material. Page 5 of 7 {9} K243619 Page 6 of 7 Traditional 510(k) Summary Taewoong Medical Niti-S Esophageal Stent Esophageal TTS Stent Stent materials are considered implants in contact with tissue for &gt; 30 days. Table 2 outlines the testing conducted. Table 2: Biocompatibility Testing Conducted | Test | Standard | | --- | --- | | Cytotoxicity MEM Extract | ISO 10993-5:2009 | | Sensitization Guinea pig maximization test | ISO 10993-10:2021 | | Intracutaneous Reactivity New Zealand rabbit | ISO 10993-23:2021 | | Acute Systemic Toxicity ICR mice | ISO 10993-11:2017 | | Pyrogen Test New Zealand rabbit | ISO 10993-11:2017 USP <151> | | 90-Day Subchronic Systemic Toxicity Rats | ISO 10993-11: 2017 | | In Vitro Mammalian Chromosomal Aberration (Genotoxicity) Chinese hamster lung cell line (CHL) | ISO 10993-3: 2015 ISO/TR 10993-33: 2015 | | Bacterial Reverse Mutation (Genotoxicity) | ISO 10993-3: 2015 ISO/TR 10993-33: 2015 | | 4-Week Subcutaneous Implantation Testing Rats | ISO 10993-6: 2018 | | 13-Week Subcutaneous Implantation Testing Rats | ISO 10993-6: 2018 | | Chemical Characterization Exhaustive extraction | ISO 10993-18:2020 | ## Electrical Safety Not applicable. The subject device contains no electric components, generates no electrical emissions, and uses no electrical energy of any type. ## Electromagnetic Compatibility (EMC) Not applicable. The subject device contains no electric components, generates no electrical emissions, and uses no electrical energy of any type. ## Software Not applicable. The subject device contains no software. Page 6 of 7 {10} K243619 Page 7 of 7 Traditional 510(k) Summary Taewoong Medical Niti-S Esophageal Stent Esophageal TTS Stent ## Performance Testing Other performance testing was conducted to show that the device meets its design requirements and performs as intended. The performance tests include: - Shelf-life testing - Tensile testing ## 10. CONCLUSION Based on the non-clinical performance and biocompatibility testing, the Esophageal TTS Stents and Niti-S Esophageal Stents are as safe and effective as the predicate devices and support a determination of substantial equivalence. Page 7 of 7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...