K080782 · Taewoong Medical Co., Ltd. · ESW · Jan 30, 2009 · General, Plastic Surgery
Device Facts
Record ID
K080782
Device Name
NITI-S ESOPHAGEAL STENT
Applicant
Taewoong Medical Co., Ltd.
Product Code
ESW · General, Plastic Surgery
Decision Date
Jan 30, 2009
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.3610
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Niti-S Esophageal Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors.
Device Story
Niti-S Esophageal Stent is a self-expanding, metallic tubular prosthesis designed to restore esophageal structure and function. The device consists of a Nitinol (Nickel-Titanium alloy) wire stent and a flexible introducer system. During clinical use, a physician deploys the stent into the esophagus; the stent exerts outward radial force on the luminal surface to establish patency in the presence of malignant strictures. The device is provided in various diameters (16-20 mm body; 24-28 mm head) and lengths (30-120 mm) to accommodate individual patient anatomy. The introducer system utilizes a co-axial tube design where the stent is constrained by an outer sheath; pulling the sheath releases the stent for deployment. The device is intended for prescription use.
Clinical Evidence
Bench testing only. Biocompatibility assessment performed on patient-contact materials. No clinical data provided.
Technological Characteristics
Self-expanding stent constructed of Nickel-Titanium alloy (Nitinol). Available in body diameters of 16, 18, or 20 mm and lengths of 30, 50, 70, 90, or 120 mm. Head diameters are 24, 26, or 28 mm with 15 mm length. Deployment via co-axial introducer system with outer sheath. Non-powered, mechanical device.
Indications for Use
Indicated for patients with esophageal strictures resulting from intrinsic or extrinsic malignant tumors.
Regulatory Classification
Identification
An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
Predicate Devices
Microvasive Ultraflex Esophageal Stent System (K940838)
Niti-S Stent & Introducer, Model Esophageal (K041648)
Cook Esophageal Z Stent with DUA Anti-Reflux (K011591)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: March 12, 2008
#### 1. Submitter:
## JAN 30 2009
0807 82
| Name: | Taewoong Medical Co., Ltd.<br>1-5 Gomak-ri, Wolgot-myeon,<br>Gimpo-si, Gyeonggi-do,<br>Republic of Korea 415871 |
|----------|------------------------------------------------------------------------------------------------------------------|
| Contact: | J.H. Nam /Director<br>Phone +82 31 996-0641 |
#### 2. Device:
| Proprietary Name: | Niti-S Esophageal Stent |
|-----------------------|-------------------------|
| Common Name: | Prosthesis, Esophageal |
| Classification Name: | Esophageal Prosthesis |
| Classification: | 21 CFR 878.3610 |
| Product Code: | ESW |
| Third Party Reviewed: | NO |
#### 3. Predicate Device:
Microvasive Ultraflex Esophageal Stent System Boston Scientific Corporation K940838
Niti-S Stent & Introducer, Model Esophageal, Taewoong Medical Co., Ltd K041648
Cook Esophageal Z Stent with DUA Anti-Reflux Wilson-Cook Medical K011591
#### 4. Description:
The proposed Niti-S Esophageal Stent consists of an implantable metallic stent and a flexible introducer system. The stent is a rigid, flexible, and expandable tubular device
Taewoong Medical Co., Ltd.
11
{1}------------------------------------------------
produ made of a Nitinol wire that is intended to be implanted to restore the structure and/or function of the esophagus. This device also includes the introducer. Upon deployment, the stent imparts an outward radial force on the luminal surface of the duct to establish patency.
#### 5. Indications for use:
Niti-S Esophageal Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors.
#### Technological Characteristics: 6.
The proposed Niti-S Esophageal Stent is a self-expanding Nickel Titanium alloy (Nitinol) stent mounted on an introducer. The stent is available in multiple sizes. The proposed stent's body diameter is 16, 18 or 20 mm with a 30, 50, 70, 90, or 120 mm length and each head's diameter is 24, 26 or 28 mm and 15 mm in length. Physician preference and individual patient condition and/or anatomy will determine the appropriate size.
Niti-S Esophageal Stent is substantially equivalent to Microvasive Ultraflex Esophageal Stent System, Boston scientific Corporation K940838, Cook Esophageal Z Stent with DUA Anti-Reflux, Wilson-Cook K011591 and Niti-S Stent & Introducer, Model Esophageal, Taewoong Medical Co., Ltd K041648. The devices have the same intended use and self-expanding stents constructed of Nitinol. The stents are mounted on an introducer system; however the designs are different. The predicate Microvasive Ultraflex Esophageal Stent System is constrained onto the introducer system with knitted loop by pulling string. The Niti-S Stent & Introducer, Model Esophageal and the proposed Niti-S Esophageal Stent utilized a co-axial tube and the stents are constrained onto the introducer system shaft by the outer sheath by pulling outer sheath.
#### 7. Performance Data:
Laboratory testing regarding characteristics was performed on the Niti-S Esophageal Stent to verify its safety and performance. A biocompatibility assessment was performed on the patient contact materials of the Niti-S Esophageal Stent.
#### 8. Conclusions:
Taewoong Medical Co., Ltd. concludes that the Niti-S Esophageal Stent is safe and effective. It is substantially equivalent to the predicate devices, Microvasive Ultraflex Esophageal Stent System, Cook Esophageal Z Stent with DUA Anti-Reflux and Niti-S Stent & Introducer, Model Esophageal.
END
K080982
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Taewoong Medical Co., Ltd. c/o Cathryn N. Cambria, RAC Consultant Arkin Consulting Group, LLC 5536 Trowbridge Drive DUNWOODY GEORGIA 30338
## JAN 3 0 2009
Re: K080782
> Trade/Device Name: Taewoong Niti-S Esophageal Stent System Regulation Number: 21 CFR §878.3610 Regulation Name: Esophageal prosthesis Regulatory Class: II Product Code: FGE Dated: December 23, 2008 Received: December 30, 2008
Dear Ms. Cambria:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppor/index.html.
Sincerely yours.
Laura B. Merrill
anine M. M. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K080782
10
510(k) Submission - Niti-S Esophageal Stent
510(k) Number K
080782
Device Name: Niti-S Esophageal Stent
Indication for use: Niti-S Esophageal Stent is intended for use in esophageal w strictures caused by intrinsic and/or extrinsic malignant tumors.
Prescription Use (Per 21CFR801 Subpart D) Over-The-Counter Use (Per 21CFR807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Arvito Whay
Division Sign-Off Division of Reproductive, Abdominal, and Radiological Devices 8079 510(k) Number
Taewoong Medical Co., Ltd.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.