K243505 · Fresenius Medical Care Renal Therapies Group, LLC · KDI · May 30, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K243505
Device Name
5008X Hemodialysis System
Applicant
Fresenius Medical Care Renal Therapies Group, LLC
Product Code
KDI · Gastroenterology, Urology
Decision Date
May 30, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 5008X™ system is indicated for intermittent hemodialysis treatment for patients with acute and chronic renal failure in a healthcare facility. Therapy options for patients weighing more than 40 kg include: Hemodiafiltration (HDF), Hemodialysis (HD), Hemofiltration (HF) and Isolated Ultrafiltration. The 5008X Machine is equipped with a bibag system, but it may also be used with liquid bicarbonate using the connector included with the machine. The bibag System is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. The bibag System is intended to be used as one component in the preparation of dialysate according to a physician’s prescription in a 3-stream proportioning hemodialysis machine equipped with the bibag module. The 5008X Machine is available equipped with a Crit-Line Clip (CLiC) Monitor System. The optional CLiC system is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.
Device Story
Electromechanical hemodialysis system; controls extracorporeal blood circuit and dialysate circuit; pumps blood through dialyzer for toxin removal/ultrafiltration. Components: 5008X machine, DIASAFE plusUS filters, bloodlines, CitraSure disinfectant. Optional CLiC monitor uses optical sensors in blood chamber to measure hematocrit, oxygen saturation, and blood volume change in real-time. Clinician/nurse uses CLiC output to adjust fluid removal rates to prevent dialysis symptoms (nausea, cramping, vomiting). Machine manages fluid mixing, heating, degassing, and monitoring (venous/arterial pressure, air/blood detection). Used in hospitals/clinics by trained staff. Benefits: optimized fluid removal, ultrapure dialysate production, and integrated monitoring.
Clinical Evidence
No clinical studies were performed. Evidence consists of bench testing, including structural integrity, leak testing, pump performance, biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1-2), and software verification/validation.
Indicated for intermittent hemodialysis (HD, HDF, HF, isolated ultrafiltration) in healthcare facilities for patients >40 kg with acute or chronic renal failure.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 30, 2025
Fresenius Medical Care Renal Therapies Group, LLC
Laura Reed
Director, Submissions Management & RA In-Center
920 Winter Street
Waltham, Massachusetts 02451
Re: K243505
Trade/Device Name: 5008X Hemodialysis System
Regulation Number: 21 CFR 876.5860
Regulation Name: High Permeability Hemodialysis System
Regulatory Class: Class II
Product Code: KDI, KOC, FJK, FIP, NII
Dated: November 12, 2024
Received: May 2, 2025
Dear Laura Reed:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243505 - Laura Reed
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243505 - Laura Reed
Page 3
Sincerely,
Maura Rooney -S
Maura Rooney
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K243505
Device Name
5008X Hemodialysis System
Indications for Use (Describe)
Machine:
The 5008X™ system is indicated for intermittent hemodialysis treatment for patients with acute and chronic renal failure in a healthcare facility. Therapy options for patients weighing more than 40 kg include: Hemodiafiltration (HDF), Hemodialysis (HD), Hemofiltration (HF) and Isolated Ultrafiltration.
The 5008X Machine is equipped with a bibag system, but it may also be used with liquid bicarbonate using the connector included with the machine.
The bibag System is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. The bibag System is intended to be used as one component in the preparation of dialysate according to a physician’s prescription in a 3-stream proportioning hemodialysis machine equipped with the bibag module.
The 5008X Machine is available equipped with a Crit-Line Clip (CLiC) Monitor System.
The optional CLiC system is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.
DIASAFEPlusUS Filter:
The DIASAFE® plusUS is intended for the preparation of ultrapure dialysate and sterile, non- pyrogenic substitution fluid from pretreated water and is not intended to provide primary purification. Attention must still be given to the chemical and microbiological quality of water and concentrates and the maintenance of supply systems (e.g. RO system, central delivery system).
The microbiological quality (microbial count [CFU/mL] and endotoxin measurement [EU/mL]) of the incoming water should be < 200 CFU/mL and < 2 EU/mL, respectively.
The DIASAFE® plusUS can only be used with Fresenius Medical Care dialysis machines fitted for use with DIASAFE® plusUS.
5008X Tubing Set:
The blood tubing sets are intended to be used only with the Fresenius Medical Care 5008X Hemodialysis Device, which is indicated for intermittent hemodialysis treatment for patients with acute kidney injury or chronic kidney disease in a healthcare facility. Therapy options include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.
The HDF blood tubing sets are indicated for use with a prescribed dialyzer. For use with a compatible dialyzer, as per the labeling.
Additional Indications for Use for the blood tubing set equipped with Twister AFRC
The Fresenius Twister Access Flow Reversing Connector (AFRC) is indicated for use during hemodialysis to reverse the blood flow to and from the arterial and venous vascular access devices in order to obtain an access flow measurement.
Additional Indications for Use for the blood tubing set equipped with CLiC
The blood tubing set is intended to be used with the Crit-Line Clip Monitor System (CLiC) on the 5008X Hemodialysis Device.
The CLiC is used to non-invasively measure hematocrit, oxygen saturation, and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse,
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740
{4}
under physician direction, intervenes (i.e., increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.
CitraSure Disinfectant:
CitraSure is intended for thermal disinfection of the Fresenius Medical Care 5008X Hemodialysis System when delivered and diluted per the 5008X heat disinfection program.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
PSC Publishing Services (301) 443-6740
{5}
K243505
Page 1 of 15
FRESENIUS MEDICAL CARE
510(k) Summary
5008X Hemodialysis System Traditional 510(k)
# 1. 510(K) SUMMARY
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR § 807.92.
## 1.1. Submitter's Information
Name: Fresenius Medical Care Renal Therapies Group, LLC
Address: 920 Winter Street
Waltham, MA 02451-1457
Contact Person: Laura Reed, Regulatory Affairs Director
Phone: (781) 491-7581
Alternate Contact: Denise Oppermann, Vice President Regulatory Affairs – North America
Phone: (781) 996-9103
Preparation Date: 12 November 2024
## 1.2. Device Name
Trade Name: 5008X Hemodialysis System
Common Name: Hemodialysis Delivery Device
Regulation Name: High Permeability Hemodialysis System
Regulatory Class: Class II per 21 CFR §876.5860
Product Code: KDI, KOC, FJK, FIP, NII
Product Code Name: Dialyzer, high permeability with or without sealed dialysate system
Accessories, blood circuit, hemodialysis
Set, tubing, blood, with and without anti-regurgitation valve
Subsystem, water purification
Disinfectant, dialysate delivery system
FDA Review Panel: Gastroenterology/Urology
## 1.3. Legally Marketed Predicate Devices
None of the predicate devices have been subject to any design-related recalls.
### 1.3.1. 5008X Machine Predicate Device
The 5008X Machine (K231534) is the legally marketed predicate device. The 2008T BlueStar Hemodialysis Machine (K231125) is the legally marketed secondary predicate device.
### 1.3.2. 5008X Bloodlines Predicate Device
The 5008X Bloodlines (K231534) are the legally marketed predicate devices. The Combiset SMARTECH Hemodialysis Bloodlines (K201207) are the legally marketed secondary predicate devices.
{6}
FRESENIUS MEDICAL CARE
510(k) Summary
5008X Hemodialysis System Traditional 510(k)
K243505
Page 2 of 15
# 1.4. Device Description
## 1.4.1. Device Identification
The 5008X Hemodialysis System consists of the following:
- The 5008X Hemodialysis Machine
- Two (2) DIASAFE®plusUS Filters (hereinafter referred to as the “Diasafe filter(s)”)
- The 5008X Bloodlines are available in five (5) configurations:
- 5008X HD/HDF – Standard Blood Tubing Set (hereinafter referred to as the “HD/HDF Standard Bloodline”)
- 5008X HD/HDF – With Twister Blood Tubing Set (hereinafter referred to as the “HD/HDF Twister Bloodline”)
- 5008X HD/HDF – With CLiC Blood Tubing Set (hereinafter referred to as the “HD/HDF CLiC Bloodline”)
- 5008X HD Pre-Flush – Standard Blood Tubing Set (hereinafter referred to as the “HD Standard Bloodline”)
- 5008X HD Pre-Flush – With Twister Blood Tubing Set (hereinafter referred to as the “HD Twister Bloodline”)
- CitraSure™ Disinfectant
## 1.4.2. Device Characteristics
The 5008X Hemodialysis Machine is an electromechanical device. Software controls the machine during hemodialysis treatment, including fluid flow, mixing, heating, and alarms.
The Diasafe filter is a non-sterile dialysis fluid filter that produces ultrapure dialysate and sterile, non-pyrogenic substitution fluid as defined in ISO 23500-5. The filter reduces microbial contaminants including endotoxins in the dialysate during hemodialysis treatment.
The 5008X Bloodlines are single-use, ethylene oxide (EO) sterilized bloodlines.
CitraSure™ disinfectant is a liquid chemical disinfectant for chemical heat disinfection of the 5008X Hemodialysis Machine.
## 1.4.3. Environment of Use
The 5008X Hemodialysis System including its components (Diasafe filters, bloodlines, disinfectant, and accessories) is intended for use in healthcare facilities, such as hospitals and dialysis clinics where intermittent dialysis treatment is performed.
## 1.4.4. Brief Written Description of the Device
The 5008X Hemodialysis Machine provides hemodialysis treatment by controlling and monitoring both the dialysate circuit and the extracorporeal blood circuit. The machine pumps blood from the patient’s body through an extracorporeal circuit, one component of which is the
{7}
FRESENIUS MEDICAL CARE
510(k) Summary 5008X Hemodialysis System Traditional 510(k)
K243505
Page 3 of 15
dialyzer. The dialyzer contains a semi-permeable membrane that uses diffusion to transfer toxins and ultrafiltration to transport excess water from the blood into the dialysate circuit. In this separate dialysate circuit, the dialysate concentrates are mixed with purified water, heated, degassed, filtered (Diasafe filters), and delivered to the dialyzer. Balancing chambers control the dialysate during treatment. Substitution fluid can also be delivered to the patient using the Bloodline if hemofiltration (HF) or hemodiafiltration (HDF) is being performed. During treatment, the extracorporeal blood circuit is monitored for venous and arterial blood pressures as well as for the presence of air and blood. The 5008X Hemodialysis Machine accommodates the following accessory devices and options:
- Accessory Devices
- Diasafe Filter (K231534)
- Liquid bicarbonate or the bibag (K162716) concentrate system (central delivery of bicarbonate is not available on the 5008X)
- Acid concentrates compliant with ISO 13958:2014 or ISO 23500-4:2019
- Dialyzers compliant with ISO 8637-1 may be used for HD. Fresenius Medical Care dialyzers indicated for HDF and HF must be used if HDF or HF is being performed.
- 5008X Bloodlines
- 5008X Fluid Sampling Accessory
- 5008X Venous Pressure Replacement with Transducer Protector
- 5008X HD/HDF Medication Line
- Options
- CLiC (Crit-Line in a Clip Monitor): K121599 (Stand-alone CLiC)
- BTM (Blood Temperature Monitor)
- CDX (Clinical Data Exchange) – K231125 (CDX cleared with 2008T HD Machine)
- Access Flow Monitor
## 1.4.5. Materials of Use
### 1.4.5.1. 5008X Machine and Diasafe Filters
The 5008X Machine and Diasafe filters are classified as externally communicating, blood path, indirect, long-term contact (>30 days) duration, (Category C) devices in accordance with FDA guidance Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (08 September 2023). A list of the hydraulic materials for the machine is provided in Table 1.
{8}
FRESENIUS MEDICAL CARE
510(k) Summary
5008X Hemodialysis System Traditional 510(k)
K243505
Page 4 of 15
Table 1: Machine Hydraulic Materials
| Component Material | Material Type |
| --- | --- |
| Plastic/Rubber | PPE+PS (Polyphenylene ether + polystyrene)
EPDM (Ethylene Propylene Diene Monomer Rubber)
PP (Polypropylene)
PVDF (Polyvinylidene fluoride)
PVC (Polyvinyl chloride)
PPSU (Polyphenylsulfone)
PESU (Polyethersulfone)
PPS (Polyphenylene Sulfide)
PTFE (Polytetrafluoroethylene)
PEEK (Polyetheretherketone)
PAEK (Polyaryletherketone)
FKM (Fluorinated, carbon-based synthetic rubber)
Platinum Cross-linked Silicone |
| Metals | Titanium
Stainless Steel |
| Other | Borosilicate Glass
Graphite
Ceramic |
A list of Diasafe filter materials is provided in Table 2.
Table 2: Diasafe Filter Materials
| Component | Material Type |
| --- | --- |
| Filter fiber | Polysulfone and Polyvinylpyrrolidone |
| Housing
Cap/Flange | Polypropylene |
| Potting resin | Polyurethane |
| O-Ring
Plastic Tabs
Sealing Disc Ring | Silicone |
## 1.4.5.2. 5008X Bloodlines
The 5008X Bloodlines are classified as externally communicating, circulating blood, long-term contact (> 30 days) duration, (Category C) devices in accordance with FDA guidance Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1:
{9}
FRESENIUS MEDICAL CARE
K243505
Page 5 of 15
510(k) Summary
5008X Hemodialysis System Traditional 510(k)
Evaluation and testing within a risk management process (08 September 2023). The materials for the bloodline components are listed in Table 3.
Table 3: 5008X Bloodline Materials
| Component | Material Type |
| --- | --- |
| Tubing and Components | Polyvinylchloride (PVC)
Polycarbonate (PC)
Polypropylene (PP)
Polyethylene (PE)
Acrylonitrile Butadiene Styrene (ABS)
Polyisoprene
Silicone
TPE
Styrene/Butadiene Copolymer
Polystyrene |
| Transducer Protector | Polyester Polybutylene Terephthalate (PBT)
Polytetrafluoroethylene (PTFE)
PC |
| Twister | Acrylic and XT
Polyisoprene
Silicone |
| Integrated Crit-Line Blood Chamber | PVC
PE
PC |
| Bonding Solvents | TetraMEK (95% Tetrahydrofuran/5% MEK)
Cyclohexanone
Loctite 3311 |
## 1.4.5.3. CitraSure™ Disinfectant
CitraSure™ has no direct or indirect contact with the patient and is rinsed from the 5008X Machine after the disinfection cycle is complete. CitraSure™ is comprised of citric acid, lactic acid, malic acid, potassium sorbate, sodium benzoate, and dialysis-quality water.
## 1.4.6. Key Performance Specifications/Characteristics
The key performance specifications and characteristics for the 5008X Machine, Diasafe Filter, 5008X Bloodline, and CitraSure disinfectant are outlined in Table 4.
Table 4: Key Performance Characteristics
| Feature | Specification/Characteristic |
| --- | --- |
| 5008X Machine | |
{10}
FRESENIUS MEDICAL CARE
K243505
Page 6 of 15
510(k) Summary
5008X Hemodialysis System Traditional 510(k)
Table 4: Key Performance Characteristics
| Feature | Specification/Characteristic |
| --- | --- |
| Maximum Blood Flow Rate | 600 mL/min |
| Maximum Dialysate Flow Rate | 1000 mL/min |
| Maximum Substitution Flow Rate | With AutoSub plus (automatic substitution): 400 mL/min
With manual substitution: 600 mL/min |
| Net Fluid Removal | 0–4000 mL/hr
Accuracy: ± (1%UF + 0.15% of balanced fluid volume)
Balanced fluid volume = substitution fluid flow + dialysate flow |
| Dialysis Time | Intermittent, typical time 4 hours |
| Dialysis Fluid Composition | Volumetric, selectable: (Maximum ranges)
Acid adjustment range: 125–151 mEq/L Na+
Bicarbonate adjustment range: 25–40 mEq/L Bicarbonate (postreaction, after mixing with the acid and purified water).
Monitoring conductivity average accuracy: ± 1.5% |
| Dialysis Fluid Temperature | Range 34°C–39°C. Fixed alarm window 33°C (or 32°C during BTM recirculation measurement and preparation) or above 40°C. |
| Heparin Administration Rate | 0.1 to 9.9 mL/hr
Accuracy: ± 0.5mL or ± 6% |
| Diasafe Filter | |
| Bacterial and Endotoxin Filtration | Produces sterile, non-pyrogenic substitution fluid from dialysis fluid with a maximum incoming water quality of 200 CFU/mL (bacteria) and 2 EU/mL (endotoxin) |
| Maximum Number Disinfection Cycles | 100 CitraSure cycles
13 Pure Bright Bleach cycles |
| Use Life | 90 days maximum or if maximum disinfection cycles have been reached |
| 5008X Bloodline | |
| Maximum Blood Flow Rate | 600 mL/min |
| Minimum Labeled Arterial Pressure | –300 mmHg |
| Maximum Labeled Venous Pressure | +500 mmHg |
| Arterial Pump Segment [Inner/Outer Diameter (ID/OD)] | 8.0 mm/12.0 mm |
{11}
FRESENIUS MEDICAL CARE
510(k) Summary
5008X Hemodialysis System Traditional 510(k)
K243505
Page 7 of 15
Table 4: Key Performance Characteristics
| Feature | Specification/Characteristic |
| --- | --- |
| Twister Component | The Twister component facilitates measurement of a patient’s access flow by reversing blood flow to and from the arterial and venous vascular access sites while maintaining a closed extracorporeal circuit |
| CLiC Chamber (Optional) – Hematocrit (HCT) | The measured HCT using the CLiC Bloodline has a standard deviation X2 of ≤ 3%, and an average bias of ≤ 1% (of the average HCT) as measured on control blood chambers |
| CLiC Chamber (Optional) – Oxygen Saturation (O2 Sat) | The measured O2 Sat using the CLiC Bloodline has a standard deviation X2 of ≤ 3%, and an average bias of ≤ 2% (of the average O2 Sat) as measured on control blood chambers |
| CitraSure™ Disinfectant | |
| Minimum Disinfection Temperature | 70°C |
| Minimum Contact Time | 10 minutes |
| Use Life | 90 Days |
| Shelf Life | 2 years |
## 1.5. Intended Use
The 5008X Hemodialysis System is intended for use in acute and chronic hemodialysis therapy. Therapy options include hemodialysis (HD), hemodiafiltration (HDF), hemofiltration (HF), and isolated ultrafiltration (ISO).
## 1.6. Indications for Use
### 1.6.1. 5008X Machine
The 5008X™ system is indicated for intermittent hemodialysis treatment for patients with acute and chronic renal failure in a healthcare facility. Therapy options for patients weighing more than 40 kg include: Hemodiafiltration (HDF), Hemodialysis (HD), Hemofiltration (HF) and Isolated Ultrafiltration.
The 5008X™ Machine is equipped with a bibag system, but it may also be used with liquid bicarbonate using the connector included with the machine. The bibag system is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. The bibag system is intended to be used as one component in the preparation of dialysate according to a physician’s prescription in a 3-stream proportioning hemodialysis machine equipped with the bibag module.
The 5008X™ Machine is available equipped with a Crit-Line Clip (CLiC) Monitor System.
The optional CLiC system is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient
{12}
FRESENIUS MEDICAL CARE
510(k) Summary 5008X Hemodialysis System Traditional 510(k)
K243505
Page 8 of 15
and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.
## 1.6.2. Diasafe Filter
The DIASAFE®plusUS is intended for the preparation of ultrapure dialysate and sterile, nonpyrogenic substitution fluid from pretreated water and is not intended to provide primary purification. Attention must still be given to the chemical and microbiological quality of water and concentrates and the maintenance of supply systems (e.g. RO system, central delivery system).
The microbiological quality (microbial count [CFU/mL] and endotoxin measurement [EU/mL]) of the incoming water should be < 200 CFU/mL and < 2 EU/mL, respectively.
The DIASAFE®plusUS can only be used with Fresenius Medical Care dialysis machines fitted for use with DIASAFE®plusUS.
## 1.6.3. 5008X Tubing Set
The blood tubing sets are intended to be used only with the Fresenius Medical Care 5008X Hemodialysis Device, which is indicated for intermittent hemodialysis treatment for patients with acute kidney injury or chronic kidney disease in a healthcare facility. Therapy options include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.
The HDF blood tubing sets are indicated for use with a prescribed dialyzer. For use with a compatible dialyzer, as per the labeling.
**Additional Indications for Use for the blood tubing set equipped with Twister AFRC**
The Fresenius Twister Access Flow Reversing Connector (AFRC) is indicated for use during hemodialysis to reverse the blood flow to and from the arterial and venous vascular access devices in order to obtain an access flow measurement.
**Additional Indications for Use for the blood tubing set equipped with CLiC**
The blood tubing set is intended to be used with the Crit-Line Clip Monitor System (CLiC) on the 5008X Hemodialysis Device.
The CLiC is used to non-invasively measure hematocrit, oxygen saturation, and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e., increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.
{13}
FRESENIUS MEDICAL CARE
510(k) Summary 5008X Hemodialysis System Traditional 510(k)
K243505
Page 9 of 15
## 1.6.4. CitraSure™ Disinfectant
CitraSure™ is intended for thermal disinfection of the Fresenius Medical Care 5008X Hemodialysis System when delivered and diluted per the 5008X heat disinfection program.
## 1.7. Comparison of Technological Characteristics with the Predicate Device
The following technological characteristics of the 5008X Hemodialysis System are substantially equivalent to those of the predicate 5008X Hemodialysis System (K231534):
- Intended Use
- Indications for Use
- Design Specifications
- Technological Characteristics
- Principle of Operation
- Performance Requirements
The following technological characteristics of the 5008X Hemodialysis Machine are substantially equivalent to the 2008T BlueStar Hemodialysis Machine (K231125):
- Clinical Data Exchange (CDX)
- Heparin Administration
- Access Flow Monitor (AFM)
The following technological characteristics of the 5008X Bloodlines are substantially equivalent to the Combiset SMARTECH Hemodialysis Bloodlines (K201207):
- Heparin Line
- Saline Line
## 1.8. Sterilization Testing
### 1.8.1. 5008X Bloodline
The 5008X Bloodlines are sterilized by exposure to 100% ethylene oxide (EO). The sterility assurance level (SAL) is 10⁻⁶. Sterility and non-pyrogenicity are claimed for the fluid pathway of the bloodline.
### 1.8.2. 5008X Bloodline EO Residual Testing
Residual testing for EO and ethylene chlorohydrin (ECh) was performed in accordance with AAMI/ANSI/ISO 10993-7:2008/(R)2012 Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals. Acceptable results (i.e., < 4.6 mg/device for EO and ECh) were obtained for the subject bloodlines.
{14}
FRESENIUS MEDICAL CARE
510(k) Summary
5008X Hemodialysis System Traditional 510(k)
K243505
Page 10 of 15
## 1.8.3. 5008X Bloodline Bacterial Endotoxin (Pyrogenicity) Testing
The subject bloodlines were tested for bacterial endotoxin (pyrogenicity) with Limulus Amebocyte Lysate (LAL) and determined to be non-pyrogenic (< 20 EU/device) in accordance with ANSI/AAMI/ST72:2019 Bacterial Endotoxins – Test Methodologies, Routine Monitoring, and Alternatives to Batch Testing.
## 1.8.4. 5008X Bloodline Sterile Barrier Testing
Sterility of the blood pathway is maintained by the sterile fluid path of the device itself which includes the following components:
- Vented Caps
- Transducer Protectors (TPs)
- Tubing and all other components that make up the structural integrity of the bloodline’s fluid path
The tubing and other components were qualified as part of the sterile barrier through the structural integrity test adapted from ISO 8637-2:2018. Testing was performed on samples after aging and distribution simulation.
## 1.9. Performance Data
Performance testing was conducted for the 5008X Hemodialysis System. Results of performance testing support substantial equivalence, safety, and efficacy of the 5008X Hemodialysis System.
## 1.9.1. 5008X Hemodialysis Machine
The following performance tests were conducted on the 5008X Hemodialysis Machine to support the determination of substantial equivalence:
- System Level Performance Testing
- Functional Design Verification
- Electrical Safety and Electromagnetic Compatibility (EMC)
- Simulated Shipping and Distribution
- Software Verification and Validation
## 1.9.2. 5008X Bloodline
Testing conducted to support the determination of substantial equivalence for the 5008X Bloodlines and accessories is summarized in Table 5.
Table 5: 5008X Bloodline Performance Testing Summary
| Test Performed | Test Objective |
| --- | --- |
| Structural Integrity | Demonstrate that the bloodlines and accessories can withstand 1.5X the labeled maximum positive and negative pressures. |
| Positive and Negative | Demonstrate that the portions of the bloodline and accessories can withstand |
{15}
K243505
Page 11 of 15
510(k) Summary
5008X Hemodialysis System Traditional 510(k)
| Test Performed | Test Objective |
| --- | --- |
| Leak Test | pressures resulting from faulty conditions of the dialysis machine. |
| Pump Segment Performance | Evaluate performance characteristics of the bloodline pump segment over the range of inlet pressures (normally 0 mmHg to -250 mmHg), flow rates, and treatment time. |
| Blood Pump Endurance | Demonstrate that the blood pump segment does not break or detach at maximum blood flow rates and pressures for 12 hours. |
| Simulated Use | Demonstrate that bloodlines perform with no tubing failures (kinking, collapsing, or disconnection) under simulated use conditions for not less than 4 hours, and those bloodline components which interface with hemodialysis machine are removable after treatment is complete without resulting in external leaks. |
| Needle Access Port Test | Demonstrate that the needle access ports, including the medication line accessory with access port, can withstand 1.5X the manufacturer’s recommended maximum positive and negative pressures after multiple accesses. |
| Male and Female Luer Connectors Tests (Performance) | Demonstrate that the components of the bloodlines and accessories meet the performance requirements of ISO 80369-7 2021, Section 6. |
| Male and Female Luer Connectors Tests (Dimensions) | Demonstrate that all the Luer components of the bloodlines and accessories, meet the dimensional requirements of ISO 80369-7: 2021, Section 5. |
| Male and Female Luer Connectors (Raw Material) | Demonstrate that all the Luer components of the bloodlines and accessories, including new or changed, are manufactured with a raw material with a nominal modulus of elasticity either in flexure or in tension greater than 700 MPa. |
| Visual Inspection After Simulation of Transportation | Demonstrate that shipping case, packaging configuration, and palletization pattern maintain the product’s structural integrity during manual handling and motorized freight. |
| Tensile Testing | Demonstrate that all bonded and glued engagements in the bloodline between components, and between components and tubing can withstand a specific tensile force based on the tubing type and position in the bloodline. |
| Labeling Content | Demonstrate that the Instructions for Use, unit labels, case labels, and shipping carton artwork for the bloodlines and accessories meet the requirements of ISO 8637-2:2018 and Guidance for Industry and FDA Staff: Hemodialysis Blood Tubing Sets – Premarket Notification [510(k)] Submissions. |
| Labeling Content – Color Coding | Verify that the color-coded components meet the requirements of ISO 8637-2:2018 and Guidance for Industry and FDA Staff: Hemodialysis Blood Tubing Sets – Premarket Notification [510(k)] Submissions |
| Readability of Barcode with Human Readable Identification Codes | Demonstrate that the barcode information on the outer container labels and unit labels for the bloodlines is capable of being scanned. |
{16}
FRESENIUS MEDICAL CARE
K243505
Page 12 of 15
510(k) Summary
5008X Hemodialysis System Traditional 510(k)
| Test Performed | Test Objective |
| --- | --- |
| Transparency of Transducer Protector | Demonstrate that the machine side of the transducer protector (TP) is clear to allow for visual inspection of blood contamination during use. |
| Transducer Protector Leak Test | The TP will be capable of maintaining a secure and leak-free connection to the hemodialysis machine when subjected to 2X the maximum labeled pressure. |
| Arterial Pressure Dome Leak Test | The pressure dome will be capable of maintaining a secure and leak-free connection to the hemodialysis machine when subjected to 2X the maximum labeled pressure. |
| Measure Gap between Lenses | Demonstrate that the distance between the lenses from the CLiC Blood Chamber is 0.078 ± 0.005 in. to verify the optical property. |
| Functional CLiC Chamber Test (Hematocrit) | Demonstrate that the CLiC integrated blood chamber optical properties enable the measurement of Hematocrit in the range of 10–60 HCT. |
| Functional CLiC Chamber Test (O2 Sat) | Demonstrate that the CLiC integrated blood chamber optical properties enable the measurement of Saturated Oxygen in the range of 30%–100% where the HCT ≥ 10. |
| Tubing Compliance Test | Demonstrate that 5008X bloodline tubing is capable of being occlusively clamped by the venous line clamp of the 5008X hemodialysis machine. |
| Clamp Occlusion Test | Demonstrate that the performance of the manual clamp to guarantee air and liquid tightness occlusion at the desired pressure and time by pressurizing a tube with the clamp under test closed on it. |
| Twister Rotation Test | Demonstrate that the Twister component can twist 180° ± 1° to reverse the blood flow. |
| Twister Torque Test | Demonstrate that the torque needed to rotate the Twister component from 0° to 180° and back is ≤ 7 inch-pound. |
| Check Valve Cracking Pressure | Demonstrate that the opening pressure (cracking pressure) of the Antisiphon valve is lower than 400 mbar. |
| Check Valve Pressure Drop | Demonstrate that the antisiphon valve presents a pressure drop in a specific range when substitution fluid is pumped through the valve at a certain flow rate. |
| Tube Transparency | Evaluate the transparency of the tube by observing the interface of air and liquid during the passage of air bubbles. |
| Tube Resistance to Kinking | Demonstrate that the tubing does not kink at the worst-case bend radius when packaged. The kinking of a tube at the maximum expected bend radius is evaluated measuring the percentage of flow reduction against the tube curved. |
| Tube Clamping | Demonstrate that the performance of the tube to resist clamp occlusions by subjecting the sample to a desired pressure and alternatively occluding with a clamp for a defined number of cycles. |
| Tube Patency | Demonstrate that the performance of the tube to maintain its patency after being clamped and then reopened by measuring the flow reduction of the re- |
{17}
K243505
Page 13 of 15
FRESENIUS MEDICAL CARE
510(k) Summary
5008X Hemodialysis System Traditional 510(k)
| Test Performed | Test Objective |
| --- | --- |
| | opened tube compared to tube not clamped. |
| Air Bubble Trapping | Demonstrate that the Venous Chamber is able to trap incoming air bubbles larger than 20 μL. |
| Recirculation Connector Packaging Burst Strength | Demonstrate that the minimum burst strength of the perimeter seal around the recirculation connector flexible package as it is internally pressurized and enclosed within restraining plates. |
| Recirculation Connector Packaging Peel Strength | Demonstrate that the seal strength of the seal joint around recirculation connector flexible package. |
| Recirculation Connector Packaging Dye Penetration Test | Demonstrate that the seal of the recirculation connector flexible package does not contain micro channels. Micro channel is defined as one that forms a 50 μm (0.002”) wide channel in the edge seals of a porous or nonporous packaging and presents a leak. |
| Recirculation Connector Packaging Microbial Resistance | Demonstrate that the ability of the recirculation connector flexible package material to resist passage of the challenge organism. |
| Viral Retentiveness Test for Transducer Protector | Demonstrate that the membrane inside the TP can prevent passage of bacteriophage (PhiX174) from the patient side to the machine side up to a pressure of 600 mmHg for 1 hr. |
## 1.9.3. Biocompatibility Testing
Biocompatibility testing for the 5008X Machine and 5008X Bloodlines patient contacting components was conducted in accordance with ISO 10993-1:2018 and FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (08 September 2023). The following endpoints were evaluated to support the biological safety of the 5008X Hemodialysis Machine:
- 5008X Machine
- Cytotoxicity
- Sensitization
- Intracutaneous Irritation
- Pyrogenicity (Material Mediated)
- Hemocompatibility, ASTM Hemolysis (Direct)
- Subchronic Toxicity
- Genotoxicity
- Chemical Characterization
- Toxicological Risk Assessment
- 5008X Bloodlines
{18}
K243505
Page 14 of 15
FRESENIUS MEDICAL CARE
510(k) Summary
5008X Hemodialysis System Traditional 510(k)
- Cytotoxicity
- Sensitization
- Intracutaneous Irritation
- Systemic Toxicity
- Pyrogenicity (Material Mediated)
- Hemocompatibility, ASTM Hemolysis (Direct and Indirect)
- Hemocompatibility, Dynamic (Mechanical) Hemolysis
- Hemocompatibility, Complement Activation
- Hemocompatibility, Platelet and Leukocyte Count
- Hemocompatibility, Partial Thromboplastin Time
- Chemical Characterization
- Toxicological Risk Assessment
## 1.9.4. Human Factors Validation Testing
The 5008X Hemodialysis System was validated for safe and effective use in accordance with FDA guidance Applying Human Factors and Usability Engineering to Medical Devices (03 February 2016).
## 1.9.5. Electrical Safety and Electromagnetic Compatibility (EMC)
Electromagnetic compatibility (EMC) testing was conducted on the 5008X Hemodialysis Machine in accordance with IEC 60601-1-2 Edition 4.1 2020-09, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic disturbances - Requirements and tests.
## 1.9.6. Software Verification and Validation Testing
Unit, integration, and system level software verification testing were performed to demonstrate the efficacy of the software and to confirm operation of the machine. The following testing was performed:
- Functional and Performance Verification
- Regression Testing
- Code Reviews
Software verification information within this submission is provided in accordance with the following FDA guidance documents:
- Content of Premarket Submissions for Device Software Functions (14 June 2023)
- Off-The-Shelf Software Use in Medical Devices (11 August 2023)
{19}
FRESENIUS MEDICAL CARE
510(k) Summary 5008X Hemodialysis System Traditional 510(k)
K243505
Page 15 of 15
- Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (27 September 2023)
- Select Updates for the Premarket Cybersecurity Guidance: Section 524B of the FD&C Act (13 March 2024)
- Applying Human Factors and Usability Engineering to Medical Devices (3 February 2016)
## 1.9.7. Animal Studies
No animal studies were performed for the 5008X Hemodialysis System.
## 1.9.8. Clinical Studies
No clinical studies were performed for the 5008X Hemodialysis System.
## 1.10. Conclusion
The indications for use, design, principle of operation, and technological characteristics of the 5008X Hemodialysis System and its components are substantially equivalent to those of the predicate and secondary predicate devices. Test results demonstrate that the differences between the 5008X Hemodialysis System, the predicate, and secondary predicate devices do not raise new concerns with regard to safety or efficacy. FMCRTG concludes that within the meaning of the Medical Device Amendments Act of 1976, the 5008X Hemodialysis System is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.