ECHELON LINEAR™ Stapler 30mm Stapler (GTX30); ECHELON LINEAR™ Stapler 60mm Stapler (GTX60); ECHELON LINEAR™ Stapler 90mm Stapler (GTX90); ECHELON LINEAR™ Stapler 3D 30mm Blue Reload (GTXR30B); ECHELON LINEAR™ Stapler 3D 30mm Green Reload (GTXR30G); ECHELON LINEAR™ Stapler 3D 60mm Blue Reload (GTXR60B); ECHELON LINEAR™ Stapler 3D 60mm Green Reload (GTXR60G); ECHELON LINEAR™ Stapler 3D 90mm Blue Reload (GTXR90B); ECHELON LINEAR™ Stapler 3D 90mm Green Reload (GTXR90G)
K243276 · Ethicon Endo-Surgery, LLC · GDW · Feb 25, 2025 · General, Plastic Surgery
Device Facts
Record ID
K243276
Device Name
ECHELON LINEAR™ Stapler 30mm Stapler (GTX30); ECHELON LINEAR™ Stapler 60mm Stapler (GTX60); ECHELON LINEAR™ Stapler 90mm Stapler (GTX90); ECHELON LINEAR™ Stapler 3D 30mm Blue Reload (GTXR30B); ECHELON LINEAR™ Stapler 3D 30mm Green Reload (GTXR30G); ECHELON LINEAR™ Stapler 3D 60mm Blue Reload (GTXR60B); ECHELON LINEAR™ Stapler 3D 60mm Green Reload (GTXR60G); ECHELON LINEAR™ Stapler 3D 90mm Blue Reload (GTXR90B); ECHELON LINEAR™ Stapler 3D 90mm Green Reload (GTXR90G)
Applicant
Ethicon Endo-Surgery, LLC
Product Code
GDW · General, Plastic Surgery
Decision Date
Feb 25, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ETHICON ™ Linear Stapler has application throughout the alimentary tract and in thoracic surgery for transection and resection of internal tissues.
Device Story
Manual surgical stapler; delivers two staggered rows of titanium staples to approximate internal tissues; used in open surgical procedures. Device consists of a handle and reloadable cartridge; requires manual loading of reload; supports up to eight firings per instrument. Incorporates Gripping Surface Technology (GST) and 3D staple geometry for tissue management. Operated by surgeons in surgical facilities. Provides mechanical tissue transection and resection; benefits include reliable wound closure and hemostasis. 90mm variant supports wider tissue transection (e.g., side-to-side anastomosis, sleeve gastrectomy).
Clinical Evidence
No clinical data provided. Evidence consists of bench performance testing (staple line integrity, force to close/fire, lockout functionality) and in-vivo animal studies (acute hemostasis and tissue healing survival studies). Biocompatibility testing performed per ISO 10993-1.
Technological Characteristics
Manual, non-powered surgical stapler. Materials: Titanium Alloy (Ti3Al-2.5V). Features: 2 staggered staple rows, GST cartridge deck, 3D staple geometry. Dimensions: 30mm, 60mm, 90mm staple line lengths. Sterilization: Gamma radiation (SAL 10^-6). No software or electrical components.
Indications for Use
Indicated for patients requiring transection and resection of internal tissues throughout the alimentary tract and in thoracic surgery. No specific age or gender contraindications provided.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
PROXIMATE® Reloadable Linear Stapler and reloads (K020779)
ECHELON 60 Endopath Stapler, Endoscopic Linear Cutter Reload (K070887)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 25, 2025
Ethicon Endo- Surgery. LLC Aysenur Huseyinoglu Erucman Senior Regulatory Affairs Specialist 475 Calle C Guaynabo, PR 00969
#### Re: K243276
Trade/Device Name: ECHELON LINEAR™ Stapler (GTX30); ECHELON LINEAR™ Stapler 60mm Stapler (GTX60); ECHELON LINEAR™ Stapler 90mm Stapler (GTX90); ECHELON LINEAR™ Stapler 3D 30mm Blue Reload (GTXR30B); ECHELON LINEAR™ Stapler 3D 30mm Green Reload (GTXR30G): ECHELON LINEAR™ Stapler 3D 60mm Blue Reload (GTXR60B); ECHELON LINEARTM Stapler 3D 60mm Green Reload (GTXR60G); ECHELON LINEAR™ Stapler 3D 90mm Blue Reload (GTXR90B); ECHELON LINEAR™ Stapler 3D 90mm Green Reload (GTXR90G) Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW, GAG Dated: October 16, 2024 Received: January 28, 2025
Dear Aysenur Huseyinoglu Erucman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Tek N. Tek N. Lamichhane -S Lamichhane -S Date: 2025.02.25 21:04:48 -05'00'
Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
| Submission Number (if known) | K243276 |
|------------------------------|---------|
|------------------------------|---------|
| Device Name | ETHICON™ Linear Stapler 30mm Stapler (GTX30); |
|-------------|---------------------------------------------------------|
| | ETHICON™ Linear Stapler 60mm Stapler (GTX60); |
| | ETHICON™ Linear Stapler 90mm Stapler (GTX90); |
| | ETHICON™ Linear Stapler 3D Reload 30 mm Blue (GTXR30B); |
| | ETHICON™ Linear Stapler 3D Reload 30mm Green (GTXR30G); |
| | ETHICON™ Linear Stapler 3D Reload 60mm Blue (GTXR60B); |
| | ETHICON™ Linear Stapler 3D Reload 60mm Green (GTXR60G); |
| | ETHICON™ Linear Stapler 3D Reload 90mm Blue (GTXR90B); |
| | ETHICON™ Linear Stapler 3D Reload 90mm Green (GTXR90G) |
| Indications for Use (Describe) | The ETHICON ™ Linear Stapler has application throughout the alimentary tract and in thoracic surgery for transection and resection of internal tissues. |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of Use (Select one or both, as applicable) | <div style="display:flex; flex-direction:column;"><span><label><input checked="true" type="checkbox"/>Prescription Use (Part 21 CFR 801 Subpart D)</label></span><span><label><input type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C)</label></span></div> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) Summary
#### I. SUBMITTER
Company: Ethicon Endo-Surgery, LLC 475 Calle C Guaynabo, PR 00969
- Contact: Aysenur Huseyinoglu Erucman Senior Regulatory Affairs Specialist Ethicon Endo-Surgery, Inc. Phone: 201-725-4434 Email: ahuseyi1@its.jnj.com
Date Prepared: February 25, 2025
#### II. Subject DEVICES
#### Trade Names:
- ETHICON™ Linear Stapler 30mm Stapler (GTX30) ●
- ETHICON™ Linear Stapler 60mm Stapler (GTX60) ●
- ETHICON™ Linear Stapler 90mm Stapler (GTX90)
- ETHICON™ Linear Stapler 3D Reload 30 mm Blue (GTXR30B) ●
- ETHICON™ Linear Stapler 3D Reload 30mm Green (GTXR30G) .
- ETHICON™ Linear Stapler 3D Reload 60mm Blue (GTXR60B)
- ETHICON™ Linear Stapler 3D Reload 60mm Green (GTXR60G) ●
- . ETHICON™ Linear Stapler 3D Reload 90mm Blue (GTXR90B)
- ETHICON™ Linear Stapler 3D Reload 90mm Green (GTXR90G) ●
| Classification Name: | Surgical Stapler |
|-----------------------------|------------------------------|
| | Staple, Implantable |
| Classification Regulation : | 21 CFR878.4750 |
| | 21 CFR 878.4740 |
| Regulatory Class: | Class 2 -Staple, Implantable |
| Product Code: | GDW,GAG |
#### III. Predicate DEVICES
| Predicate Device<br>510(k) Number | Predicate Device Name | Predicate Device Model Numbers |
|-----------------------------------|-----------------------------------------------------|----------------------------------------------------------------------------|
| K020779 | PROXIMATE® Reloadable Linear<br>Stapler and reloads | Devices: TX30B, TX30G, TX60B, TX60G<br>Reloads: XR30B, XR30G, XR60B, XR60G |
These predicates have not been subjected to a recall related to these design modifications.
{5}------------------------------------------------
## Reference Device(s):
The below reference devices were used to show equivalency to the 3D technology:
## Reference Device Models
| Reference<br>Device 510(k)<br>Number | Reference Device Name | Reference Device Model Numbers |
|--------------------------------------|-----------------------------------------------------------------|------------------------------------------------------------|
| K223760 | ECHELON LINEAR™ Cutter Reloads | GLCR60B,GLCR60G,GLCR80B, GL<br>CR80G,<br>GLCR100B,GLCR100G |
| K200420 | ECHELON CONTOUR™ Curved<br>Cutter Reloads | GCR40B, GCR40G |
| K163523 | ECHELON CIRCULAR™ Powered<br>Stapler | CDH23P, CDH25P, CDH29P,<br>CDH31P |
| K070887 | ECHELON 60 Endopath Stapler,<br>Endoscopic Linear Cutter Reload | ECR60G |
## ETHICON™ Linear Staplers Product Configuration
| Subject<br>Device<br>Model<br>Numbers | Product Code<br>Description | Staple<br>Line<br>Length | Subject<br>Device<br>Reloads | Predicate Device<br>Model Numbers | Predicate K<br>Number |
|---------------------------------------|-------------------------------------|--------------------------|------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-----------------------|
| GTX30 | ETHICON™<br>Linear 30mm<br>Stapler | 30mm | GTXR30B<br>GTXR30G | TX30B; TX30G<br>Compatible<br>Reloads:<br>XR30B (Blue)<br>XR30G (Green) | K020779 |
| GTX60 | ETHICON™<br>Linear 60mm<br>Stapler | 60mm | GLCR60B<br>GLCR60G | TX60B; TX60G<br>Compatible<br>Reloads:<br>XR60B (Blue)<br>XR60G (Green) | K020779 |
| GTX90 | ETHICONTM<br>Linear 90mm<br>Stapler | 90mm | GLCR90B<br>GLCR90G | There is no 90mm<br>Predicate device,<br>therefore, 30mm<br>and 60mm<br>devices used to<br>show substantial<br>equivalence. | K020779 |
{6}------------------------------------------------
| 30 mm Reloads Codes | 60 mm Reload Codes | 90 mm Reloads Codes | Corresponding Reload Color |
|---------------------|--------------------|---------------------|----------------------------|
| GTXR30B | GTXR60B | GTXR90B | Blue |
| GTXR30G | GTXR60G | GTXR90G | Green |
#### Reload codes and corresponding Reload color
### IV. DEVICE DESCRIPTION
The ETHICON™ Linear Stapler delivers two staggered rows of titanium staples in order to approximate internal tissues. The Subject device is sterile, single patient, disposable device used in conjunction with reloadable ETHICON™ Linear Stapler 3D Reloads to staple tissue in one firing stroke. The Subject device is a next generation Linear Stapler to be used in open procedures requiring a surgical stapler. The Subject device incorporates proven Gripping Surface Technology (GST) Reload Technology, leveraged from Ethicon Endocutters, along with a 3-D staples developed for other Ethicon Endocutters and open mechanical devices.
The ETHICON™ Linear Stapler 30 mm, 60 mm and 90 mm Staplers are sterile, single- patient-use instruments that staple tissue. There are two staggered rows of staples, on either side of the staple line. This device may be used on the general population for routine wound closure via stapling. The 30 mm reload creates a 30 mm staple line. The 60 mm reload creates a 60 mm staple line. The 90 mm reload creates a 90 mm staple line.
The 90mm device and compatible reloads have also been developed. The 90mm device can be used by surgeons who prefer to use a 90mm device when transecting wider tissue such as completing the closure of a side-to-side anastomosis or when performing a sleeve gastrectomy. The Predicate device does not currently offer a 90mm size but the principles of operation are the 30mm and 60mm devices. Design verification testing demonstrates that the 90mm device and compatible reloads do not raise new types of safety or effectiveness questions.
The instruments are shipped without a reload and must be loaded prior to use. A staple retaining cap on the reload protects the staple leg points during shipping and transportation. The instrument may be loaded eight times for a maximum of eight firings per instrument during a single procedure. There are blue and green reload options for each device size.
The instruments' lock-out feature is designed to prevent a used or improperly installed reload from being refired or an instrument from being fired without a reload.
#### INDICATIONS FOR USE V.
The ETHICON™ Linear Stapler has application throughout the alimentary tract and in thoracic surgery for transection and resection of internal tissues.
| Subject Device | Predicate Device | Difference |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| The ECHELON LINEAR TM Stapler<br>has application throughout the<br>alimentary tract and in thoracic<br>surgery for transection and resection<br>of internal tissues. | The Proximate ® Reloadable Linear<br>Stapler has application throughout<br>the alimentary tract and in thoracic<br>surgery for transection and resection<br>of internal tissues. | The indications are the same. |
### Indication for Use Comparison Table
{7}------------------------------------------------
# VI. TECHNOLOGICAL COMPARISON
| Characteristics | Subject Device | Predicate Device |
|----------------------------------------------------|----------------------------------------------------------------------------------------------|-------------------------------------------|
| <b>General Information</b> | | |
| Stapler Code | GTX30, GTX60, GTX90 | TX30B, TX30G<br>TX60B, TX60G |
| Compatible Reloads | GTXR30B, GTXR60B,<br>GTXR90B<br>Green: GTXR30G,<br>GTXR60G, GTXR90G | XR30B,<br>XR60B<br>Green:<br>XR30G, XR60G |
| How it is supplied | Provided sterile, single<br>patient use | Same |
| Packaged | PETG Tray | Same |
| Sterilization Method | Gamma | Same |
| Sterility Assurance<br>Level | SAL 10-6 | Same |
| <b>Design Features</b> | | |
| Operation of stapling<br>and transection of tissue | The device operates by<br>driving staples into a<br>staple forming anvil to<br>staple tissue | Same |
| Cartridge Reload<br>surface | GST Reload Technology<br>for cartridge deck surface | Flat cartridge deck surface |
| Number of staple rows | 2 Staggered Rows | Same |
| Staple material | Titanium Alloy -Ti3Al-<br>2.5V Wire | Titanium Alloy (Ti-6Al-4V) |
| Formed staple geometry | 3D final staple form | 2 D shaped staple form |
| <b>Principle of Operation</b> | | |
| Environment of Use | Surgical facility | Same |
| Intended population | General | Same |
| Energy Type | Manual | Same |
| Maximum number of<br>Firings per device | 8 Firings | Same |
### Subiect and Predicate Device and Reload Comparison Table
### Regulatory History
The Regulatory History of the predicate device ETHICON™ Linear Stapler is provided in the Table below to help clarify the submissions associated with the predicate.
## Regulatory History of the Predicate Device
| Submission Number | Description | Clearance Date |
|-------------------|--------------------------------------------|----------------|
| K020779 | PROXIMATE® TX Reloadable<br>Linear Stapler | 05-JUN-2002 |
{8}------------------------------------------------
#### VII. PERFORMANCE DATA:
The following performance data demonstrate that the Subject device is substantially equivalent to the Predicate device and the differences between the devices were found not to affect safety or performance.
## Bench Performance Testing:
- Device Stapling Performance and Comparison to Predicate ●
- Formed Staple Height (FSH) O
- Staple Form Quality (SFQ) O
- Staple Line Integrity (SLI) o
- Device Functional Requirement ●
- Force to Close O
- Force to Fire o
- Jaw Aperture O
- No Spent Reload/Lockout O
- Human Factor Report/Usability Testing O
- Product Characterization ●
- Staple line Strength Test O
Animal Testing: In-vivo testing evaluations included.
- Acute Hemostasis evaluation study ●
- Tissue Healing response, Survival Study ●
All animal studies passed the criteria for success.
Clinical Studies: The premarket submission did not rely on the assessment of clinical performance data to demonstrate device performance and equivalence.
Bio-compatibility: studies was performed based on ISO 10993-1 and confirmed that the
proposed device is biocompatible for the intended patient contact profile. All biocompatibility studies passed the criteria for success.
Electrical Safety and Electromagnetic Compatibility: This device is not powered and does not include software therefore this does not apply.
# VII. CONCLUSIONS
The conclusions of the testing criteria demonstrate that the Subject device; ETHICON™ Linear Stapler performs substantially equivalent to the Predicate device and does not raise new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.