ECHELON LINEAR™ 60 mm Cutter (GLC60);ECHELON LINEAR™ 80mm Cutter (GLC80);ECHELON LINEAR™ 100 mm Cutter (GLC100);ECHELON LINEAR™ Cutters 60mm Blue Reload (GLCR60B);ECHELON LINEAR™ Cutters 60mm Green Reload (GLCR60G);ECHELON LINEAR™ Cutters 80mm Blue Reload (GLCR80B);ECHELON LINEAR™ Cutters 80mm Green Reload (GLCR80G);ECHELON LINEAR™ Cutters 100mm Blue Reload (GLCR100B);ECHELON LINEAR™ Cutters 100mm Green Reload (GLCR100G)
K223760 · Ethicon Endo-Surgery · GDW · Mar 15, 2023 · General, Plastic Surgery
Device Facts
Record ID
K223760
Device Name
ECHELON LINEAR™ 60 mm Cutter (GLC60);ECHELON LINEAR™ 80mm Cutter (GLC80);ECHELON LINEAR™ 100 mm Cutter (GLC100);ECHELON LINEAR™ Cutters 60mm Blue Reload (GLCR60B);ECHELON LINEAR™ Cutters 60mm Green Reload (GLCR60G);ECHELON LINEAR™ Cutters 80mm Blue Reload (GLCR80B);ECHELON LINEAR™ Cutters 80mm Green Reload (GLCR80G);ECHELON LINEAR™ Cutters 100mm Blue Reload (GLCR100B);ECHELON LINEAR™ Cutters 100mm Green Reload (GLCR100G)
Applicant
Ethicon Endo-Surgery
Product Code
GDW · General, Plastic Surgery
Decision Date
Mar 15, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The ECHELON LINEARTM Cutters have application in gastrointestinal, thoracic, and pediatric surgery for transection, resection, and the creation of anastomoses.
Device Story
Manual surgical stapler; delivers two double-staggered rows of staples while simultaneously dividing tissue. Used in open gastrointestinal, thoracic, and pediatric procedures. Operated by surgeons. Device requires manual loading of reloads; features safety lockouts preventing firing without a reload or with a used reload, and preventing jaw closure or knife exposure when not in home position. Incorporates GST (Gripping Surface Technology) and 3D-shaped anvil pockets for improved staple formation. Provides mechanical transection and hemostasis; benefits include consistent staple line integrity and tissue management during resection/anastomosis.
Clinical Evidence
No clinical data. Evidence provided via bench testing (stapling performance, force requirements, staple line strength) and animal studies (acute hemostasis, tissue healing/survival in GI and pulmonary models). Biocompatibility confirmed.
Technological Characteristics
Manual, non-powered surgical stapler. Materials are biocompatible. Features GST reload technology and 3D-shaped anvil pockets. Includes safety lockouts for firing, clamping, and knife position. Dual-sided firing knobs and modular handle coupling. Sterile, single-use.
Indications for Use
Indicated for use in gastrointestinal, thoracic, and pediatric surgery for transection, resection, and the creation of anastomoses in the general patient population.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
PROXIMATE® Linear Cutters with Safety Lockout and Reloads (K020779)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 15, 2023
Ethicon Endo-Surgery Alicia Butler Senior Regulatory Affairs Specialist 475 Calle C Guaynabo, PR 00969
### Re: K223760
Trade/Device Name: ECHELON LINEAR™ 60 mm Cutter (GLC60):ECHELON LINEAR™ 80mm Cutter (GLC80);ECHELON LINEAR™ 100 mm Cutter (GLC100);ECHELON LINEAR™ Cutters 60mm Blue Reload (GLCR60B);ECHELON LINEAR™ Cutters 60mm Green Reload (GLCR60G):ECHELON LINEARTM Cutters 80mm Blue Reload (GLCR80B):ECHELON LINEAR™ Cutters 80mm Green Reload (GLCR80G);ECHELON LINEAR™ Cutters 100mm Blue Reload (GLCR100B);ECHELON LINEAR™ Cutters 100mm Green Reload (GLCR100G) Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: December 15, 2022 Received: December 15, 2022
Dear Alicia Butler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
| | Digitally signed by |
|------|-----------------------------------------------|
| Mark | Mark Trumbore -S |
| | Trumbore -S Date: 2023.03.15 08:05:38 -04'00' |
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
Device Name
| ECHELON LINEARTM 60 mm Cutter (GLC60); |
|--------------------------------------------------------------------------------------------|
| ECHELON LINEARTM 80mm Cutter (GLC80); |
| ECHELON LINEARTM 100 mm Cutter (GLC100); |
| ECHELON LINEARTM Cutters 60mm Blue Reload (GLCR60B); |
| ECHELON LINEARTM Cutters 60mm Green Reload (GLCR60G); |
| ECHELON LINEARTM Cutters 80mm Blue Reload (GLCR80B); |
| ECHELON LINEARTM Cutters 80mm Green Reload (GLCR80G); |
| ECHELON LINEARTM Cutters 100mm Blue Reload (GLCR100B); |
| ECHELON LINEARTM Cutters 100mm Green Reload (GLCR100G) |
| Indications for Use (Describe) |
| The ECHELON LINEARTM Cutters have application in gastrointestinal, thoracic, and pediatric |
| surgery for transection, resection, and the creation of anastomoses. |
| Type of Use (Select one or both, as applicable) |
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
I. SUBMITTER
Company: Ethicon Endo-Surgery, LLC 475 Calle C Guaynabo, PR 00969
Contact: Alicia Butler Senior Regulatory Affairs Specialist Ethicon Endo-Surgery, Inc. Phone: 513-337-1633 Email: abutler9@its.jnj.com
Date Prepared: December 15, 2022
#### II. SUBJECT DEVICES
# Trade Names:
- ECHELON LINEAR™ 60mm Cutter and Reloads ●
- ECHELON LINEAR™ 80mm Cutter and Reloads ●
- ECHELON LINEAR™ 100mm Cutter and Reloads .
| Common or Usual Name: | Surgical Stapler with Implantable Staples | Classification Name: | Implantable staple (21 CFR 878.4750) |
|-----------------------|-------------------------------------------|----------------------|--------------------------------------|
| Regulatory Class: | Class 2 -Staple, Implantable | Product Code: | GDW |
| Common or Usual Name: | Surgical Stapler with Implantable Staples | Classification Name: | Surgical Stapler (21 CFR 878.4740) |
| Regulatory Class: | Class 2 - Stapler, Surgical | Product Code: | GAG |
#### III. PREDICATE DEVICES
| Predicate Device<br>510(k) Number | Predicate Device Name | Predicate Device Product Codes |
|-----------------------------------|--------------------------------------------------------------|--------------------------------------------------------------------------------------|
| K020779 | PROXIMATE® Linear Cutters with Safety<br>Lockout and Reloads | Devices: TLC55, TLC75, TLC10<br>Reloads: TCR55, TCT55, TCR75,<br>TCT75, TCR10, TCT10 |
These predicates have not been Subjected to a recall related to these design modifications.
{4}------------------------------------------------
| Subject<br>Device<br>Product<br>Code | Product Code<br>Description | Staple<br>Line<br>Length | Cut<br>Line<br>Length | Subject<br>Device<br>Reloads | Predicate Device<br>Product Code | Predicate K<br>number |
|--------------------------------------|------------------------------------|--------------------------|-----------------------|------------------------------|------------------------------------------------------------------|-----------------------|
| GLC60 | ECHELON<br>LINEAR™ 60mm<br>Cutter | 61mm | 55mm | GLCR60B<br>GLCR60G | TLC55<br>Compatible<br>Reloads:<br>TCR55 (Blue)<br>TRT55 (Green) | K020779 |
| GLC80 | ECHELON<br>LINEAR™ 80mm<br>Cutter | 81mm | 75mm | GLCR80B<br>GLCR80G | TLC75<br>Compatible<br>Reloads:<br>TCR75 (Blue)<br>TRT75 (Green) | K020779 |
| GLC100 | ECHELON<br>LINEAR™ 100mm<br>Cutter | 101mm | 95mm | GLCR100B<br>GLCR100G | TLC10<br>Compatible<br>Reloads:<br>TCR10<br>TRT10 | K020779 |
# Table 5-1: ECHELON LINEAR™ Cutters Product Configuration
# Table 5-2: Reload codes and corresponding reload color
| Reloads Codes<br>for<br>60mm | Reload Codes for<br>80mm | Reload Codes for<br>100mm | Corresponding Reload<br>Color |
|------------------------------|--------------------------|---------------------------|-------------------------------|
| GLCR60B | GLCR80B | GLCR100B | Blue |
| GLCR60G | GLCR80G | GLCR100G | Green |
{5}------------------------------------------------
### DEVICE DESCRIPTION
The ECHELON LINEAR™ Cutter delivers two, double-staggered rows of staples while simultaneously dividing the tissue between rows. The instrument has a safety lockout feature that is designed to prevent firing if either no reload or a used reload is installed. An unclamp lockout feature prevents knife exposure by allowing the instrument to open only when the Firing Knob is in the home position. A clamp lockout feature prevents closure of the jaws when the knife is not in the home position. A Staple Retaining Cap on the reload protects the staple leg points during shipping and transportation.
This device may be used on the general population for routine wound closure via stapling.
The 60 mm reload creates a 61 mm staple line and cuts tissue approximately 52 mm beyond the tissue stop.
The 80 mm reload creates an 81 mm staple line and cuts tissue approximately 72 mm beyond the tissue stop.
The 100 mm reload creates a 101 mm staple line and cuts tissue approximately 92 mm beyond the tissue stop.
The instruments are shipped without a reload and must be loaded prior to use. The instrument may be loaded eight times for a maximum of eight firings per instrument during a single procedure.
#### IV. INDICATIONS FOR USE
The ECHELON LINEAR™ Cutters have application in gastrointestinal, thoracic, and pediatric surgery for transection, resection, and the creation of anastomoses.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE V. PREDICATE DEVICE
ECHELON LINEAR™ Cutter is substantially equivalent to PROXIMATE® Linear Cutters with respect to operating principle of device and Intended Use of the device. Both the devices are sterile, single use manual devices. The key technological difference between the Subject and the predicate device is the change to the reloads with the addition of GST technology and a new 3D style anvil pocket.
The Subject device is a next generation Linear Cutter to be used in open procedures requiring a surgical stapler. To improve stapling performance, the Subject device incorporates proven GST Reload Technology, leveraged from Ethicon Endocutters, and new 3-D B-shaped final staple form into the new ECHELON LINEARTM Cutter Reload as shown in illustration 5-1.
{6}------------------------------------------------
# Illustration 5-1: GST Reload Technology and 3-D Final Staple Form
Image /page/6/Picture/2 description: The image contains two separate images with text below each. The image on the left is a close-up of a blue object with a textured surface, and the text below it reads "GST". The image on the right is a 3-D rendering of a metal object, and the text below it reads "3-D".
The Subject and predicate staplers have the following identical features:
- Indication (the compatibility with staple line reinforcement materials has not been evaluated ● for the Subject device and thus has been removed)
- Intended use
- Contraindications ●
- Materials
- MR compatibility
- Sterilization Method
- Operational principles ●
The following differences exist between the Subject and predicate staplers:
- Reloads: Subject device reloads contain GST technology used in Ethicon Endocutters along with a new 3D style anvil pocket. Also, only the newly designed Green and Blue reloads will be compatible with the device.
- New handle half coupling. The two device halves can be joined proximally or separated when maneuvering on tissue.
- . Clamp arm that latches when the device is clamped closed and unlatches by pressing the clamp release button.
- Firing knob consists of dual foldable firing knobs that can be activated from either the left ● side or the right side of the assembled device.
#### VI. PERFORMANCE DATA:
{7}------------------------------------------------
The following performance data demonstrate that the Subject device is substantially equivalent to the predicate device and the differences between the devices were found not to affect safety or performance.
Bench Testing:
- Device Stapling Performance and Comparison to Predicate ●
- Formed Staple Height (FSH) O
- Staple Form Quality (SFO) O
- Staple Line Integrity (SLI) о
- Device Functional Requirement ●
- Force to Close/Clamping Force O
- O Force to Fire
- Force to Couple/Decouple Device Halves о
- Force to Press Clamp Release Button O
- Human Factor Report o
- Product Characterization ●
- o Staple line Strength Test
Animal Testing: In-vivo testing evaluations included
- Acute Hemostasis evaluation study
- Tissue Healing response, Survival Study using both gastrointestinal and pulmonary models.
Clinical studies: The premarket submission did not rely on the assessment of clinical performance data to demonstrate device performance and equivalence.
Biocompatibility: studies were performed and confirmed that the Subject device is biocompatible for the intended patient contact profile.
Electrical Safety and Electromagnetic Compatibility: This device is not powered and therefore this does not apply.
# VII. CONCLUSIONS
The conclusions of the testing criteria demonstrate that the Subject device: ECHELON LINEAR™ CUTTER performs substantially equivalent to the predicate device and does not raise new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.