K243241 · Davol, Inc., A Subsidiary of C.R. Bard, Inc. · OWT · Feb 19, 2025 · General, Plastic Surgery
Device Facts
Record ID
K243241
Device Name
Phasix ST Umbilical Hernia Patch
Applicant
Davol, Inc., A Subsidiary of C.R. Bard, Inc.
Product Code
OWT · General, Plastic Surgery
Decision Date
Feb 19, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Phasix™ ST Umbilical Hernia Patch is indicated for use in the reinforcement of soft tissue, where weakness exists, in procedures involving the repair of umbilical hernias.
Device Story
Sterile, single-use, self-expanding, fully resorbable surgical mesh; used for umbilical hernia repair. Comprised of two layers of poly-4-hydroxybutyrate (P4HB) with posterior side co-knitted with polyglycolic acid (PGA) fibers; coated with resorbable hydrogel (sodium hyaluronate, carboxymethylcellulose, polyethylene glycol) to minimize tissue attachment. Includes SorbaFlex™ Memory Technology (polydioxanone (PDO) monofilament in P4HB sleeve) for stability and placement. Features hybrid positioning straps (P4HB and polypropylene) with blue delineation marker. Operated by surgeons in clinical settings. Mesh allows fibroblastic response for tissue ingrowth; hydrogel resorbs within 30 days; P4HB resorbs in 12-18 months; PDO resorbs in 24-32 weeks. Output is a reinforced tissue site. Benefits include ease of insertion, proper placement, and reduced visceral adhesions.
Clinical Evidence
No clinical data. Evidence based on bench testing (biocompatibility, physical/functional characteristics, design validation) and reliance on historical animal/histological data from reference devices (K142818, K143380) and predicate devices (K101928, K190185). Biocompatibility testing included chemical characterization, cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, and implantation studies.
Technological Characteristics
Materials: Poly-4-hydroxybutyrate (P4HB), polyglycolic acid (PGA), polydioxanone (PDO), polypropylene, and hydrogel (sodium hyaluronate, carboxymethylcellulose, polyethylene glycol). Principle: Self-expanding resorbable mesh with memory ring. Sterilization: Ethylene oxide. Form factor: Mesh patch with positioning straps and memory ring. No software or electronic components.
Indications for Use
Indicated for reinforcement of soft tissue where weakness exists in umbilical hernia repair procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font in blue.
February 19, 2025
Davol, Inc., a subsidiary of C.R. Bard, Inc. Shannon Green Manager, Regulatory Affairs 100 Crossings Blvd Warwick, Rhode Island 02886
Re: K243241
Trade/Device Name: Phasix™ ST Umbilical Hernia Patch Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OWT, OOD, FTL Dated: October 8, 2024 Received: October 10, 2024
Dear Shannon Green:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Tek N. Tek N. Lamichhane -S Lamichhane -S Date: 2025.02.19 22:27:19
22:27:19 Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K243241
Device Name
Phasix ST Umbilical Hernia Patch
Indications for Use (Describe)
Phasix™ ST Umbilical Hernia Patch is indicated for use in the reinforcement of soft tissue, where weakness exists, in procedures involving the repair of umbilical hernias.
Type of Use (Select one or both, as applicable)
> | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Phasix™ ST Umbilical Hernia Patch (K243241) 510(k) Summary
# 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (1)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
### Submitter Information:
Davol, Inc. C.R. Bard, Inc. 100 Crossings Blvd Warwick, RI 02886
Phone: (401) 825-8774 Fax:
Contact Person: Shannon Green Title: Manager, Regulatory Affairs Email: shannon.green2@bd.com
Date Prepared: February 19, 2025
# Subject Device Name:
| Name of Device: | Phasix™ ST Umbilical Hernia Patch | |
|-----------------------|----------------------------------------------|--|
| Common or Usual Name: | Surgical Mesh | |
| Classification Name: | Mesh, Surgical, Absorbable, Abdominal Hernia | |
| Regulatory Class: | Class II | |
| Regulation Number: | 21 CFR 878.3300 | |
| Product Codes: | OWT, OOD, FTL | |
# Primary Predicate Device:
| Name of Device: | Ventralex™ ST Hernia Patch (K101928), cleared on<br>March 23, 2011 |
|-----------------------|--------------------------------------------------------------------|
| Common or Usual Name: | Surgical Mesh |
| Classification Name: | Mesh, Surgical, Polymeric |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 878.3300 |
| Product Code: | FTL |
# Secondary Predicate Device:
{5}------------------------------------------------
| Name of Device: | Phasix™ ST Mesh with Open Positioning System<br>(K190185), cleared on June 12, 2019 |
|-----------------------|-------------------------------------------------------------------------------------|
| Common or Usual Name: | Surgical Mesh |
| Classification Name: | Mesh, Surgical, Absorbable, Abdominal Hernia |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 878.3300 |
| Product Codes: | OWT, OOD, FTL |
#### Reference Devices:
Phasix™ Mesh (K142818), cleared March 31, 2015 Phasix™ ST Mesh (K143380), cleared June 5, 2015
### Device Description:
Phasix™ ST Umbilical Hernia Patch is a sterile, single-use device for prescription use only. It is a self-expanding, fully resorbable mesh with a resorbable hydrogel coating and a positioning pocket and strap. Phasix™ ST Umbilical Hernia Patch is comprised of 2 layers of poly-4hydroxybutyrate (P4HB), with the posterior side being co-knitted with polyglycolic acid (PGA) fibers, identical to the mesh component of the secondary predicate device (Phasix™ ST Mesh with Open Positioning System). P4HB degrades through hydrolysis and a hydrolytic enzymatic digestive process and is essentially completely resorbed in 12-18 months. Phasix™ ST Umbilical Mesh is coated on the PGA surface with a resorbable, chemically modified sodium hyaluronate (HA), carboxymethylcellulose (CMC), and polyethylene glycol (PEG) based hydrogel. The hydrogel is identical to both the primary predicate device (Ventralex™ ST Hernia Patch (K101928) and the secondary predicate device (Phasix™ ST Mesh with Open Positioning System, K190185). The fascial side of the mesh allows for a prompt fibroblastic response through the interstices of the mesh, allowing for complete tissue ingrowth. The visceral side of the mesh is the resorbable hydrogel coating, which separates the mesh from underlying tissues and organ surfaces to help minimize tissue attachment to the mesh. Shortly after hydration in saline, the coating becomes a hydrated gel that is resorbed from the site in less than 30 days. Phasix™ ST Umbilical Hernia Patch contains SorbaFlex™ Memory Technology, which provides memory and stability to the mesh, facilitating ease of initial insertion, proper placement and fixation of the device. The SorbaFlex™ Memory Technology is comprised of an extruded polydioxanone (PDO) resorbable monofilament contained within a knitted P4HB containment sleeve. PDO is resorbed within 24-32 weeks. The PDO ring and hybrid positioning straps (comprised of P4HB and polypropylene materials that are connected by overlapping the materials and sewing them together with clear PP monofilament, with a delineation marker dyed blue with
{6}------------------------------------------------
[phthalocyaninato(2-)] copper), are based on the design of the primary predicate Ventralex™ ST Hernia Patch (K101928). The subject device has the identical intended use as the primary and predicate devices; soft tissue repair/reinforcement.
# Indications for Use of Device:
The Phasix™ ST Umbilical Hernia Patch is indicated for use in the reinforcement of soft tissue, where weakness exists, in procedures involving the repair of umbilical hernias.
# Technological Comparison to Predicate Devices:
The subject device, Phasix™ ST Umbilical Hernia Patch, has the following similarities to the predicate devices:
- . The intended use of the subject device, i.e. for the repair/reinforcement of soft tissue, is identical to Ventralex™ ST Hernia Patch (K10128) and Phasix™ ST Mesh with Open Positioning System (K190185).
- The indications for use, i.e. umbilical hernia repair, is similar to Ventralex™ ST . Hernia Patch (K101928) and Phasix™ ST Mesh with Open Positioning System (K190185).
- The mesh design is similar to Ventralex™ ST Hernia Patch (K101928).
- The mesh sizes are identical to Ventralex™ ST Hernia Patch (K101928). .
- . The principle of operation for hernia repair is similar to Ventralex™ ST Hernia Patch (K101928).
- The poly-4-hydroxybutyrate (P4HB) co-knitted with polyglycolic acid (PGA) fibers . which are coated with chemically modified sodium hyaluronate (HA), carboxymethylcellulose (CMC), and polyethylene glycol (PEG) materials are identical to the materials of Phasix™ ST Mesh with Open Positioning System (K190185).
- The polydioxanone (PDO) ring, as well as the polypropylene and dyed . polypropylene components of the positioning straps are identical to the materials of Ventralex™ ST Hernia Patch.
- The sterilization method (ethylene oxide) is identical to Ventralex™ ST Hernia . Patch (K101928) and Phasix™ ST Mesh with Open Positioning System (K190185).
{7}------------------------------------------------
The subject device, Phasix™ ST Umbilical Hernia Patch, incorporates the following changes as compared to the predicate devices:
- · The positioning strap of the subject device is a hybrid of both resorbable poly-4hydroxybutyrate (P4HB) and non-resorbable polypropylene as compared to the polypropylene only strap of the primary predicate Ventralex™ ST Hernia Patch and includes a delineation marker to identity where to cut excess strap material to ensure no polypropylene is left behind. The marker is dyed blue with [phthalocyaninato (2-)] copper.
- There is no pre-attached, removable positioning system with the subject device as . compared to the secondary predicate Phasix™ ST Mesh with Open Positioning System (K190185).
Phasix™ ST Umbilical Hernia Patch has the identical intended use as both the primary and secondary predicate devices. The subject device has similar indications for use as both the primary and secondary predicate device. The subject device is manufactured from material identical to the mesh component of the secondary predicate device and incorporates the identical PDO ring and polypropylene component of the positioning strap as the primary predicate device. Phasix™ ST Umbilical Patch has identical packaging materials as the primary predicate device. The subject device has the identical sterilization method (ethylene oxide) as compared to the predicate devices and similar principle of operation and manufacturing processes as compared to the predicate devices. Any differences in the technological characteristics were thoroughly tested and the results demonstrate that there are no new questions of safety and effectiveness.
# Performance Data:
The following performance data is provided in support of substantial equivalence determination.
# 1. Performance Standards:
No performance standards have been established for this device under Section 514 of the Federal Food, Drug and Cosmetic Act.
# 2. Biocompatibility Testing:
The biocompatibility evaluation for Phasix™ ST Umbilical Patch was conducted in accordance with the Guidance for Industry and Food and Drug Administration Staff Use
K243241
{8}------------------------------------------------
of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" September 8, 2023, and International Standard ISO 10993-1 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process," as recognized by the FDA.
The subject device mesh materials are identical to the mesh materials utilized in the successful clearance of the primary predicate Ventralex™ ST Hernia Patch (K101928) and the secondary predicate Phasix™ ST Mesh with Open Positioning System (K190185), and therefore considered safe and biocompatible. Utilizing the Biocompatibility Evaluation Flow Chart, Attachment D of the Guidance for Industry and Food and Drug Administration Staff Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (September 8, 2023), it has been concluded that all biocompatibility requirements are met. Thus, the tests specific to the individual materials of Phasix™ ST Umbilical Hernia Patch were not repeated. However, in order to address any risks associated with the interaction of materials that have not previously been combined in a single device, i.e. poly-4-hdyroxybutyrate (P4HB) and polydioxanone (PDO) and the interaction of the subject device with the packaging materials, chemical characterization and biological characterization testing were completed. As all materials of the subject device are utilized in the predicate devices, which are owned by Davol, Inc. and well known and have a greater than ten (10) year history of clinical use in the same anatomical space, comparative testing was not conducted. Refer to Table 1.
| Test Method | Results | |
|------------------------------------------|------------------------------------------------------------------------|------------------------------|
| Material Chemical Characterization | PASS | |
| Exhaustive Extraction at 50°C for 72 hrs | Gravimetric Analysis<br>GCMS<br>LCMS-UV-CAD<br>HS-GCMS<br>ICPMS<br>GPC | |
| Water | | |
| Ethanol | | Gravimetric Analysis<br>GCMS |
| Hexane | | |
# Table 1: Biocompatibility Testing of Phasix™ ST Umbilical Hernia Patch
{9}------------------------------------------------
| | GPC |
|----------------------------------------------------------------------|--------|
| Cytotoxicity - Packaging Materials | |
| MEM Cell Cytotoxicity - Tyvek Envelope | PASS |
| MEM Cell Cytotoxicity - Foil Pouch | PASS |
| Cytotoxicity - Device | |
| MEM Cell Cytotoxicity | PASS |
| Sensitization (Kligman Maximization) | PASS |
| Irritation / Intracutaneous Reactivity<br>(Intracutaneous Injection) | PASS |
| Acute Systemic Toxicity | PASS |
| Material Mediated Pyrogenicty | PASS |
| Limulus Amebocyte Lysate (LAL) | PASS |
| Subacute Toxicity Study in Rats (28 days) | PASS |
| Subchronic Toxicity Study in Rats (13 weeks) | PASS |
| Chronic Toxicity (26 weeks) | PASS |
| Genotoxicity | |
| • AMES Assay | • PASS |
| • Mouse Lymphoma Assay | • PASS |
| Implantation (Intramuscular Implantation Test) | |
| • 4 weeks | • PASS |
| • 8 weeks | • PASS |
| • 13 weeks | • PASS |
| • 26 weeks | • PASS |
| Toxicology (Toxicological Risk Assessment) | PASS |
The biologic and chemical characterization test results demonstrate that the subject device is biocompatible and there are no interactions between the subject device new combination of materials, or the subject device and the packaging materials to affect the previously established safety and effectiveness. Therefore, the subject device is safe and is biocompatible for its intended use.
# 3. Product Testing:
The performance test results demonstrate that the subject device successfully met the established acceptance criteria and is substantially equivalent to Phasix™ ST Mesh with Open Positioning System (K190185) and Ventralex™ ST Hernia Patch (K101928). Completed performance testing on the subject device is listed in Table 2.
# Table 2: Performance Testing
{10}------------------------------------------------
| Performance Test | |
|------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Substantial Equivalency Testing –<br>Physical and Functional Characteristics | Physical Characteristics |
| | - Mesh Weave Characteristics |
| | - Pore Size |
| | - Determination of Device and Mesh<br>Thickness |
| | - Weight per Unit Area (Linear Density) |
| | - Device Stiffness |
| | - Dimensional Measurements |
| | Functional Characteristics |
| | - Ball Burst Strength |
| | - Suture Retention Strength (Machine<br>and Cross directions) |
| | - Tear Strength (Cross and Machine<br>directions) |
| | - PGA Pull-Out Strength |
| | - Three-Tack Pluck Force |
| | - Gel Disruption Analysis |
| Functional Testing of the Subject Device | - 25 Degree Recoil |
| | - Containment Sleeve Puncture |
| | - Strap Attachment Strength |
| | - Pocket Integrity |
| Resorption Profile of the Ring and<br>Containment Sleeve | - <i>In-vitro</i> degradation of PDO with<br>Phasix and Phasix ST |
| | - <i>In-vitro</i> degradation of PDO with<br>Phasix ST (PGA pull-out) |
| Design Validation Usability Testing | - IFU |
| | - Packaging/Labeling |
| | - Insertion |
| | - Positioning/Placement |
| | - Fixation |
| Human Factors / Simulated Testing | - IFU |
| | - Packaging/Labeling |
| | - Device Preparation |
| | - Insertion |
| | - Positioning and Placement |
| | - Fixation |
{11}------------------------------------------------
#### 4. Animal Studies:
Animal studies were not conducted on the subject device. Based on identical materials and substantially equivalent properties measured through bench testing, in-vivo safety and performance of the subject device was evaluated through the animal and histological data presented in the reference devices: Phasix™ Mesh (K142818) and Phasix™ ST Mesh (K143380), which were utilized to support the clearance of the secondary predicate device, Phasix™ ST Mesh with Open Positioning System (K190185).
In addition, Phasix™ ST Mesh with Open Positioning System (K190185) was studied in a comprehensive 4 week GLP study in a porcine model. The subject device, Phasix™ ST Umbilical, is comprised of the same combination of poly-4-hydroxybutyrate (P4HB), polyglycolic acid (PGA) and chemically modified sodium hyaluronate (HA), carboxymethylcellulose (CMC), and polyethylene glycol (PEG). The polydioxanone (PDO) ring in the subject device is identical to the primary predicate Ventralex™ ST Hernia Patch (K101928) and the material was evaluated via a rat model over 1, 2, 4, 8,12,16, 24 and 32 weeks. Based on the studies conducted for the reference devices as well as for both predicate devices, additional testing was not required for the subject device.
### Conclusion:
Phasix™ ST Umbilical Hernia Patch has the identical intended use, materials of construction, packaging materials and sterilization method as the primary and secondary predicate devices. Furthermore, the subject device has similar indications for use, technological characteristics, principle of operation, and manufacturing processes as the primary and secondary predicate devices. Any differences in the technological characteristics between the subject device and the predicate devices were thoroughly assessed and evaluated. All test results support that the subject device's safety, effectiveness and performance are similar to the predicate devices. Therefore, Phasix™ ST Umbilical Hernia Patch is substantially equivalent to Ventralex™ ST Hernia Patch (K101928) and Phasix™ ST Mesh with Open Positioning System (K190185).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.