K143380 · C.R. Bard, Inc. · OWT · Jun 5, 2015 · General, Plastic Surgery
Device Facts
Record ID
K143380
Device Name
Phasix ST Mesh
Applicant
C.R. Bard, Inc.
Product Code
OWT · General, Plastic Surgery
Decision Date
Jun 5, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Phasix™ ST Mesh is indicated for use in the reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as for the repair of hernias.
Device Story
Phasix™ ST Mesh is a sterile, fully resorbable surgical mesh prosthesis for soft tissue reinforcement and hernia repair. It consists of a co-knitted base of poly-4-hydroxybuterate (P4HB) monofilament and polyglycolic acid (PGA) fibers, coated on the visceral side with a resorbable hydrogel (sodium hyaluronate, carboxymethylcellulose, and polyethylene glycol). The P4HB provides mechanical support during the healing period, with minimal resorption through 12-26 weeks and significant degradation by 12-18 months. The hydrogel coating minimizes tissue attachment to the visceral side, while the fascial side allows for fibroblastic response and tissue ingrowth. The device is used by surgeons in clinical settings. It is provided sterile via ethylene oxide. The device benefits patients by providing temporary mechanical support for tissue defects while minimizing post-operative adhesions.
Clinical Evidence
No clinical study was required. Evidence consists of bench testing (mesh weave, stiffness, pore size, burst strength, density, tear resistance, thickness, suture pullout, and in vitro degradation) and in vivo porcine studies (mechanical strength, tissue response, and resorption profile at 4, 12, and 24 weeks). Biocompatibility testing was performed per ISO 10993.
Technological Characteristics
Fully resorbable surgical mesh. Materials: P4HB monofilament, PGA fibers, and HA/CMC/PEG hydrogel coating. Sterilization: Ethylene oxide. Packaging: Tyvek envelope and foil pouch. No electrical or software components.
Indications for Use
Indicated for reinforcement of soft tissue where weakness exists, including hernia repair procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three intertwined strands and three human profiles facing right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 5, 2015
C. R. Bard Incorporated Mr. Tony John Regulatory Affairs Specialist 100 Crossing Boulevard Warwick, Rhode Island 02886
Re: K143380
Trade/Device Name: Phasix™ ST Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: OWT, OOD, FTL Dated: November 24, 2014 Received: November 25, 2015
Dear Mr. John:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K143380
Device Name Phasix™ ST Mesh
Indications for Use (Describe)
The Phasix™ ST Mesh is indicated for use in the reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as for the repair of hernias.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
## This 510(k) Summary is provided per the requirements of section 807.92(c).
## Submitter Information:
| Company Name: | Davol Inc., Subsidiary of C. R. Bard, Inc. |
|------------------------|------------------------------------------------|
| Company Address: | 100 Crossings Boulevard<br>Warwick, RI 02886 |
| Telephone: | (401) 825-8692 |
| Fax: | (401) 825-8765 |
| Submitter's Name: | Tony John, MS<br>Regulatory Affairs Specialist |
| Date Summary Prepared: | June 4, 2015 |
## Device Identification:
| Trade Name: | Phasix TM ST Mesh |
|----------------------|----------------------------------------------|
| Common/Usual Name: | Surgical Mesh |
| Classification Name: | Mesh, Surgical, Absorbable, Abdominal Hernia |
| Device Class: | Class II |
| Regulation Number: | 21 CFR § 878.3300 |
| Product Code: | OWT, OOD, FTL |
## Predicate Device Names:
- . TephaFLEX® Mesh , K070894/K111946/K113723 (Tepha Inc), FDA cleared on: April 13, 2007, September 26, 2011, and February 15, 2012, respectively.
- Ventralight ST Mesh, K101851 (Davol Inc), FDA cleared on : July 15, 2010 ●
PREMARKET NOTIFICATION PHASIX™ ST MESH
{4}------------------------------------------------
Davol Inc. Subsidiary of C.R. Bard, Inc. 100 Crossings Blvd. Warwick. RI 02886 (401) 463-7000
Image /page/4/Picture/1 description: The image shows the logo for Bard Davol Inc. The logo features the word "BARD" in a bold, sans-serif font, with a green outline. Below the word "BARD" is the text "DAVOL INC." in a smaller font. To the right of the logo is the text "K143380 Pat", followed by the words "TECHNOLOGY", "TECHNIQUE", "TRAINING", and "TRUST" stacked vertically.
#### Device Description
The proposed Phasix™ ST Mesh is a fully resorbable mesh with a resorbable hydrogel coating. It is a sterile mesh prosthesis designed for the reinforcement and reconstruction of soft tissue deficiencies. Phasix ™ ST Mesh is co-knitted using poly-4-hydroxybuterate (P4HB) and polyglycolic acid (PGA) fibers. P4HB is produced from a naturally occurring monomer and is processed into monofilament fibers and then knitted into a surgical mesh. P4HB degrades through a process of hydrolysis and a hydrolytic enzymatic digestive process. It has been developed to reinforce areas where weakness exists while minimizing the variability of resorption rate (loss of mass) and strength to provide support throughout the expected healing period. Preclinical implantation studies indicate that resorption of the P4HB fibers is minimal throughout the 12 week expected healing period and up to 26 weeks post implantation. Significant degradation of the mesh fibers observed in preclinical studies within 12 to 18 months indicates loss in mechanical integrity and strength. While fiber segments were observed at 18 months, they continued to degrade. Phasix ™ ST Mesh is coated on the PGA surface with a resorbable, chemically modified sodium hyaluronate (HA), carboxymethylcellulose (CMC) and polyethylene glycol (PEG) based hydrogel. The fascial side of the mesh allows for a prompt fibroblastic response through the interstices of the mesh, allowing for complete tissue ingrowth, similar to P4HB mesh alone. The visceral side of the mesh is a resorbable hydrogel coating, separating the mesh from underlying tissues and organ surfaces to help minimize tissue attachment to the mesh. Shortly after hydration, the biopolymer coating becomes a hydrated gel that is resorbed from the site in less than 30 days.
#### Indications for Use
Phasix™ ST Mesh is indicated for use in the reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as for the repair of hernias.
#### Comparison of Technological Characteristics with the Predicate Device
The proposed Phasix™ ST Mesh has the same materials and design as the predicate TephaFlex® Mesh and Ventralight™ ST Mesh devices.
- The base mesh in the proposed Phasix™ ST Mesh is constructed of the same P4HB . monofilament and knit pattern as the predicate TephaFlex® Mesh. The base mesh component of the predicate Ventralight™ ST Mesh is made of polypropylene.
- Both the proposed Phasix™ ST Mesh and predicate Ventralight™ ST Mesh contain a HA/CMC . PEG based hydrogel. These devices also contain PGA fibers co-knit into the mesh to allow the
## PREMARKET NOTIFICATION PHASIX™ ST MESH
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Bard Davol Inc. logo. The logo is green and features the word "BARD" in a stylized font. Below the word "BARD" is the text "DAVOL INC." On the right side of the image, there are four words stacked vertically: "TECHNOLOGY", "TECHNIQUE", "TRAINING", and "TRUST". The image also includes the text "K143380 Page 3 of 4" at the top.
hydrogel to adhere to the mesh. The purpose of the hydrogel coating is to minimize tissue attachment to the mesh. The TephaFLex® Mesh predicate does not contain a hydrogel coating or PGA fibers.
The proposed and predicate devices have the same intended use and similar indications for use statements. In addition, all three devices are sterilized via ethylene oxide and packaged in a Tyveke envelope and foil pouch. Where minor technological differences exist between the proposed and predicate devices, testing demonstrates that the differences do not adversely affect the performance of the proposed device.
## Performance Data:
The following performance data are provided in support of the substantial equivalence determination.
#### Biocompatibility Testing
The materials used in the construction of the proposed Phasix™ ST Mesh are known to be biocompatible as demonstrated by their safe clinical use in the predicate devices. However, in order to address any risks associated with interactions between the three device components (i.e. P4HB, PGA, and the hydrogel coating), complete biocompatibility testing in accordance with FDA's Blue Book Memorandum #G95-1 issued on May 1, 1995, Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" and FDA Guidance "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management Process". The following studies were completed:
- . Cytotoxicity
- . Genotoxicity
- Sensitization
-
- - Intracutaneous Reactivity .
- Local and Systemic Toxicity (4 and 13 week) ● Local Toxicity (4, 8, 13, and 20 week)
- . Systemic Toxicity (Acute)
- Pyrogenicity
.
#### Electrical safety and electromagnetic compatibility (EMC)
There are no electrical or metal components in the Phasix™ Mesh; therefore the proposed device does not require EMC and Electrical Safety evaluation.
#### Software Verification and Validation Testing
The proposed PhasixTM Mesh does not contain software.
#### PREMARKET NOTIFICATION PHASIX™ ST MESH
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for BARD DAVOL INC. The logo is green and features the word "BARD" in a stylized font. To the right of the logo are the words "TECHNOLOGY", "TECHNIQUE", "TRAINING", and "TRUST" stacked vertically. At the top right of the image is the text "K143380 Page 4 of 4".
## Bench Testing
Bench testing was performed to compare the proposed Phasix™ ST Mesh to the predicate TephaFlex® Mesh and Ventralight™ ST Mesh devices. In accordance with FDA's "Guidance for the Preparation of a Premarket Notification for a Surgical Mesh" published March 2, 1999, the following physical and performance characteristics were evaluated:
- . Mesh weave characteristics
- . Device stiffness
- . Mesh pore size
- Burst strength
- Mesh density
- Tear resistance ●
- . Mesh thickness
- . Suture pullout strength.
Additionally an in vitro degradation study was completed to demonstrate that the degradation of the PGA fibers and hydrogel coating present on the proposed Phasix™ ST Mesh construct do not impact the resorption profile of the P4HB mesh.
## Animal Studies
In vivo porcine studies were performed to characterize the mechanical strength, tissue response, and resorption profile of the proposed Phasix™ ST Mesh device in comparison to the cited predicates at 4, 12, and 24 weeks.
## Clinical Study
No clinical study was required in support of the Phasix™ ST Mesh.
## Conclusion:
All test results provided in this submission demonstrate that the proposed Phasix™ ST Mesh is substantially equivalent to the cited TephaFlex® Mesh and Ventralight™ ST Mesh predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.