K243126 · Novatech SA · NWA · Oct 29, 2024 · General, Plastic Surgery
Device Facts
Record ID
K243126
Device Name
TRACHEOBRONXANE™ DUMON®
Applicant
Novatech SA
Product Code
NWA · General, Plastic Surgery
Decision Date
Oct 29, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3720
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
For maintaining open the airway, after dilation of the stenosis or resection of the obstruction, in particular in case of: - Tracheobronchial tumors - Tracheal stenoses with scarring - Stenoses following surgical anastomosis, resection or pulmonary transplantation.
Device Story
TRACHEOBRONXANE™ DUMON® stents are flexible, medical-grade silicone tracheobronchial tubes designed to maintain airway patency by exerting radial force. Available in various profiles (TD, TF, BD, Y, ST), sizes, and thicknesses to conform to individual patient anatomy; some models include barium sulfate for radiopacity. Stents feature studs to facilitate anchorage on cartilage rings. Used in operating rooms, treatment rooms, or endoscopy suites by physicians. Stents are implanted to support the trachea or bronchi; they improve respiration capacity and limit mucus obstruction risk. Device is supplied sterile and intended for long-term use. Clinical benefit includes maintaining an open airway in patients with obstructive pathologies.
Clinical Evidence
Bench testing only. No clinical data presented. Performance testing included fatigue testing simulating normal usage, weighing tests for absorption/disposal of substances, compression tests for radial performance, and elongation tests. Biocompatibility testing conducted per ISO 10993-1, including chemical characterization, cytotoxicity, sensitization, irritation, and material-mediated pyrogenicity, demonstrated no adverse biological effects.
Technological Characteristics
Medical-grade silicone with barium sulfate for radiopacity. Radial force support mechanism. Various profiles (TD, TF, BD, Y, ST) with integrated anchorage studs. Supplied sterile via steam sterilization. 3-year shelf life. No energy source. Manual operation. Compliant with ISO 10993-1 for long-term tissue contact.
Indications for Use
Indicated for infants, children, and adults of any gender requiring airway maintenance following dilation or resection of obstructions, including tracheobronchial tumors, scarring, or post-surgical/transplant stenoses.
Regulatory Classification
Identification
The tracheal prosthesis is a rigid, flexible, or expandable tubular device made of a silicone, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the trachea or trachealbronchial tree. It may be unbranched or contain one or two branches. The metal tracheal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 29, 2024
Novatech SA Stuart K. Montgomery President Boston Medical Products, Inc. Z.I. Athélia III -1058, Voie Antiope La Ciotat CEDEX, 13705 France
Re: K243126
Trade/Device Name: Tracheobronxane™ Dumon® Regulation Number: 21 CFR 878.3720 Regulation Name: Tracheal Prosthesis Regulatory Class: Class II Product Code: NWA Dated: September 19, 2024 Received: September 30, 2024
Dear Stuart K. Montgomery:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Ethan L. Nyberg -S
Ethan Nyberg, Ph.D. Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
Device Name
## TRACHEOBRONXANE™ DUMON®
Indications for Use (Describe)
For maintaining open the airway, after dilation of the stenosis or resection of the obstruction, in particular in case of:
- · Tracheobronchial tumors
- · Tracheal stenoses with scarring
- · Stenoses following surgical anastomosis, resektion or pulmonary transplantation
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/7 description: The image shows the logo for Novatech, a biotechnology company. The logo features a blue square on the left side, with a white outline of lungs inside. To the right of the square, the word "Novatech" is written in a bold, blue font. Below "Novatech", the phrase "new biotechnology for life" is written in a smaller, blue font.
## TRACHEOBRONXANE™ DUMON®
## 510(k) Summary
NOVATECH SA Z.I. Athélia III – 1058, Voie Antiope 13705 La Ciotat CEDEX FRANCE
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Novatech. The logo features a blue square on the left side with a white outline of lungs inside. To the right of the blue square, the word "Novatech" is written in a bold, sans-serif font. Below "Novatech" is the tagline "new biotechnology for life" in a smaller, lighter font.
#### 1 Submitter
NOVATECH SA Z.I. Athélia III – 1058, Voie Antiope 13705 La Ciotat CEDEX FRANCE
## Contact Person
Stuart K. Montgomery President Boston Medical Products Inc. 70 Chestnut Street Shrewsbury, MA 01545, USA Ph: 508-898-9300 ext. 240 Fax: 508-898-2373
## Date Prepared
September 19, 2024
#### 2 Device
| Device Name: | TRACHEOBRONXANE™ DUMON® |
|------------------------------------|----------------------------------------|
| Regulation Number and Description: | 878.3720 Tracheal Prosthesis |
| Medical Specialty | GENERAL AND PLASTIC SURGERY<br>DEVICES |
| Product Code: | NWA |
| Device Class: | 2 |
#### Predicate Device 3
| Device Name: | ENDOXANE® |
|---------------------|------------------------------|
| Original Applicant: | NOVATECH SA |
| 510(k) Number: | K971509 |
| Regulation Number: | 878.3720 Tracheal Prosthesis |
| Product Code: | NWA |
| Device Class: | 2 |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a logo for Novatech. The logo is comprised of a blue square on the left side of the image. Inside the blue square is a white outline of a pair of lungs. To the right of the blue square is the word "Novatech" in a large font, and below that is the phrase "new biotechnology for life" in a smaller font.
#### Device Description 4
The TRACHEOBRONXANE™ DUMON® stents are flexible tracheobronchial tubes made of medical grade silicone, designed to improve patient respiration capacity while acting as trachea or bronchi support. These medical devices are designed to conform to individual anatomies.
Each stent has a specific treatment, to improve the evacuation of mucus in order to limit the risk of obstruction.
A large range of shape, thickness and sizes is available depending on the pathology to be treated, patient intrinsic factor and location of the implantation. Indeed, the TRACHEOBRONXANE™ DUMON® range consists of various models which vary from each other in the following characteristics:
- The stents type: TD, TF, BD, Y, ST corresponding to various profiles made of silicone with different hardness;
- . The dimensions: each profile consists of different sizes in diameter and length;
- The radiopacity: stents can be provided radiopaque or not radiopaque; barium sulfate . is added during the manufacturing process in order to make the stent radiopaque which allows obtaining a better visibility by X-rays radiography.
The TRACHEOBRONXANE™ DUMON® stents have been designed with studs in order to facilitate their anchorage in the trachea or bronchi (on cartilage rings). The number of stud rows is depending on the stent profile and stent final destination (bronchi or trachea).
The TRACHEOBRONXANE™ DUMON® stent holds the respiratory tract open by exerting a radial force.
#### 5 Intended Use
For maintaining open the airway, after dilation of the stenosis or resection of the obstruction, in particular in case of:
- Tracheobronchial tumors
- Tracheal stenoses with scarring
- Stenoses following surgical anastomosis, resection or pulmonary transplantation. ●
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Novatech. The logo features a blue square on the left side with a white outline of lungs inside. To the right of the blue square, the word "Novatech" is written in a bold, sans-serif font. Below "Novatech", the phrase "new biotechnology for life" is written in a smaller, lighter font.
#### 6 Comparison of technological characteristics with predicate device
## Table 1: Side-by-Side Comparison of TRACHEOBRONXANE™ DUMON® with the Predicate Device
| Item | Proposed Device:<br>TRACHEOBRONXANE™ DUMON® | Predicate Device:<br>ENDOXANE® | |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|----------------------------------------------------------------------------|
| 510k | TRACHEOBRONXANE™ DUMON® | K971509 | |
| Product Code | NWA | same | |
| Regulation Number | 878.3720 - Tracheal prosthesis | same | |
| Class | 2 | same | |
| Intended Use | The product is intended for use in the trachea<br>and/or bronchus to keep the airway open<br>The product is for long-term use. | same | |
| Indication for Use | For maintaining open the airway, after dilation of<br>the stenosis or resection of the obstruction, in<br>particular in case of:<br>• Tracheobronchial tumors<br>• Tracheal stenoses with scarring<br>• Stenoses following surgical anastomosis,<br>resection or pulmonary transplantation | same | |
| Patient Target Group | • Infants and young children<br>• Children and young people<br>• Adults<br>• Patients of any gender | same | |
| Intended User | Physician | same | |
| Anatomical Sites | trachea and/or bronchus | same | |
| Environment of Use | • Operating room<br>controlled area with access restrictions,<br>strict hygiene concept, controlled airflow and<br>ideal illumination; product to be used under<br>sterile conditions only<br>• Treatment room<br>operating room with lesser requirements but<br>has controlled hygiene concept; product<br>may be used under sterile or non-sterile<br>conditions<br>• Endoscopy suite | same | |
| Energy used and / or<br>delivered | No energy is used for this device | same | |
| Item | Proposed Device:<br>TRACHEOBRONXANE ™ DUMON® | | Predicate Device:<br>ENDOXANE® |
| Design | Stent profile | Scheme | Same, additionally<br>TRACHEOBRONXANE ™<br>DUMON® BB - Ultra Thin<br>Stent |
| | TRACHEOBRONXANE ™<br>DUMON® BD<br>Bronchial Stent | | |
| | TRACHEOBRONXANE ™<br>DUMON® TD<br>Tracheal Stent | | |
| | TRACHEOBRONXANE ™<br>DUMON® TF<br>Thin Tracheal Stent | | |
| | TRACHEOBRONXANE ™<br>DUMON® ST<br>Hourglass Stent | | |
| | TRACHEOBRONXANE ™<br>DUMON® Y<br>Bifurcation Stent | | |
| | | | |
| Materials | medical grade silicone, barium sulfate | | same |
| Principles of<br>Operation | Manually operable | | same |
| Sterilization | Supplied sterile | | Supplied non-sterile |
| Shelf-Life | 3 years | | 5 years |
| Chemical Safety | No hazardous substances | | same |
| Compatibility with<br>environment and<br>other devices | The manufacturer recommends using a delivery<br>system type TONN ™ / NOVATECH | | same |
| Mechanical Safety | Visual examination, weighing test, compression<br>test and elongation test has been performed on<br>TRACHEOBRONXANE ™ DUMON® stents.<br>Additionally, a fatigue test simulated the normal<br>usage of the stents has been performed. | | same |
{8}------------------------------------------------
## Table 1: Side-by-Side Comparison of TRACHEOBRONXANE™ DUMON® with the Predicate Device
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Novatech. The logo consists of a blue square on the left side with a white outline of lungs inside. To the right of the square is the word "Novatech" in a bold, sans-serif font, with the words "new biotechnology for life" underneath in a smaller font.
#### 7 Performance Data
## Biocompatibility Testing
According to its intended purpose and ISO 10993-1, TRACHEOBRONXANE™ DUMON® is categorized as:
| Category | Contact | Contact duration |
|------------------------|---------|------------------------|
| Implant medical device | Tissue | C-Long term (>30 days) |
#### Table 1: Side-by-Side Comparison of TRACHEOBRONXANE™ DUMON® with the Predicate Device
| TRACHEOBRONXANE™ DUMON® | ENDOXANE® |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Applicable standard: ISO 10993-1 | same |
| Chemical characterization<br>Chemical characterization of final products, manufacturing process,<br>packaging, sterilization, and storage revealed no risk of adverse<br>effects on the patient.<br>Chemical testing and subsequent risk mitigating assessment<br>revealed no risk of systemic toxic effects.<br>Therefore, the biological effects as follows can be excluded:<br>systemic toxicity systemic implantation effects genotoxicity carcinogenicity reproductive and developmental toxicity toxicokinetics | same |
| Cytotoxicity<br>in-vitro studies on NuSil medical-grade silicones passed.<br>in-vitro studies on DUMON® final stents passed. | same |
| Sensitization and irritation<br>Chemical testing did not reveal sensitizing substances.<br>In-vivo studies on NuSil medical-grade silicones passed.<br>Long history of safe clinical use. | same |
| Irritation<br>Chemical testing did not reveal irritating substances.<br>In-vivo studies on NuSil medical-grade silicones passed.<br>Long history of safe clinical use. | same |
| Material-mediated pyrogenicity<br>The devices do not contain any of the substances listed in DIN EN<br>ISO 10993-11:2018, Annex G, and are not made of biologically-<br>derived materials.<br>In-vivo studies on NuSil medical-grade silicones revealed no risk for<br>material-mediated pyrogenicity.<br>Long history of safe clinical use. | same |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Novatech. The logo consists of a blue square on the left side with a white outline of lungs inside. To the right of the square is the word "Novatech" in a bold, sans-serif font. Below "Novatech" is the tagline "new biotechnology for life" in a smaller font.
#### Record's title
#### TRACHEOBRONXANE™ DUMON® 510(k) Summary
| TRACHEOBRONXANE™ DUMON® | ENDOXANE® |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|
| Systemic toxicity (acute, subacute, subchronic, chronic) | |
| See chemical characterization.<br>In-vivo studies on NuSil medical-grade silicones passed.<br>Long history of safe clinical use. | same |
| Implantation effects | |
| See chemical characterization.<br>In-vivo studies on NuSil medical-grade silicones passed.<br>Atraumatic design without sharp edges.<br>Removal of device after 12 months recommended.<br>Long history of safe clinical use. | same |
| Genotoxicity | |
| See chemical characterization.<br>In-vitro studies on NuSil medical-grade silicones passed. | same |
| Carcinogenicity | |
| See chemical characterization.<br>Long history of safe clinical use and extensive clinical data. | same |
| Reproductive and developmental toxicity | |
| See chemical characterization.<br>Long history of safe clinical use. | same |
| Degradation | |
| Due to the non-polar nature of silicone (body contacting material) no<br>degradation isexpected in physiological surrounding.<br>Chemical testing on DUMON® stents revealed no degradation<br>products.<br>Conducted shelf-life tests gave no hint for degradation of material.<br>Long history of safe clinical use. | same |
| Toxicokinetics | |
| Chemical testing on DUMON® stents revealed no substances in<br>toxicologically relevant amounts. | same |
| Conclusion: | Conclusion: |
| No adverse biological effects expected. | No adverse biological effects<br>expected. |
According to the available biological information and experimental data, and supported by the long history of safe use of TRACHEOBRONXANE™ DUMON® there is no reason to doubt the biological suitability of this device.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for Novatech. The logo features a blue square on the left side with a white outline of lungs inside. To the right of the square, the word "Novatech" is written in a bold, sans-serif font. Below "Novatech", the phrase "new biotechnology for life" is written in a smaller, less bold font.
## Benchmark Performance Tests
Several performance tests on TRACHEOBRONXANE™ DUMON® stents have been carried out to analyze the device characteristics.
The following tests demonstrate the technical equivalence of aged stents with new manufactured products.
## Fatique
Specific test procedures were performed on brand new stents and stents, which were stressed by fatigue test. This fatigue test simulated the normal usage of the stents. The gathered data of stressed and unstressed stents were compared. Additionally, stents were stored unstressed in a water bath to influence of the water bath in comparison to the influence of the fatigue test.
#### Weighing
A weighing test was conducted on the stents to assess the absorption or disposal on substances out of or into the stent.
The weighing test was performed on the stents treated in the water bath and treated in the fatigue test.
#### Compression
The compression test examines the radial performance of the stents.
Stents from fatigue test are compressed with the use of a defined compression stamp to examine the force required to compress the stents.
#### 8 Shelf Life/Sterilization
| Method of Sterilization: | Steam Sterilization |
|--------------------------|---------------------|
| Shelf-Life: | 3 years |
#### 9 Conclusion
The subject device TRACHEOBRONXANE™ DUMON® is substantially equivalent to predicate device, ENDOXANE®.
ENDOXANE® is the submitter's own legally marketed predicate device.
The only difference between the products is TRACHEOBRONXANE™ DUMON® is supplied sterile and ENDOXANE® is supplied non-sterile.
The indications for use and materials of construction remain unchanged between the subject device and the manufacturer's existing device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.