K971509 · Novatech SA · NWA · Aug 21, 1997 · General, Plastic Surgery
Device Facts
Record ID
K971509
Device Name
ENDOXANE
Applicant
Novatech SA
Product Code
NWA · General, Plastic Surgery
Decision Date
Aug 21, 1997
Decision
ST
Submission Type
Traditional
Regulation
21 CFR 878.3720
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
INDICATIONS ARE LISTED ACCORDING TO THEIR FREQUENCY : -TRACHEOBRONCIAL TUNORS' -TRACHEAL STENDSIS WITH SCARRING BRONCIAL STANQSIS AFTER SURGICAL ANASTOMOTIC, ANASTOHOSES RESECTIONS OR PULMONARY THANSPLANTATION THE STENDSES HAVE AS A COMMON CHARACTERISTIC THE IMPORTUNIE IN THE EXTENSIC CEMPRESSION OR THE COLLAPSE OF THE CARTILAGENS MALL. INDICNATION ATHARS FOCUSES ON AN ENDOSCORIC THERAPEUTIC ACCIENCY AND Lies a IT CONCERNS A RESECTION (LASER, CRYO, THERMO COGULATION) OR CHATAL. " THE SWENT INSERTEDN IS INDICATED ID THE EWOOSCOPIC RESECTICA 15 15-CONFLETTE IN CARR OF A PERSISTENT EXTRINGIC COMPRESSION OR IF THE LUMEN OF THE ADMANY IS INSUPEICIENT. THE MOST FREQUENT LOCALICAS ARF TRACENAL. THE LEBET MAIN BRONCHUS IS THE SECOND LOCATION FOLLOWED JY THE RIGHT MAIN BRONCHUS.
Device Story
ENDOXANE (Dumon Stent) is a non-armed, medical-grade silicone tube used as a tracheal/bronchial prosthesis. Designed for endoscopic insertion by physicians to maintain airway patency in cases of extrinsic compression or cartilaginous wall collapse. Features external stubs to prevent migration by anchoring between cartilaginous rings; surface coating minimizes mucus adhesion and tissue trauma. Stent regains round shape after insertion. Used in clinical settings following endoscopic resection or dilation. Benefits include improved airway lumen, reduced risk of migration, and facilitated secretion evacuation.
Clinical Evidence
Evidence based on a published study in 'The Journal of Bronchology' (1996) involving 1,500 stent placements. Reported outcomes include no observed cases of compression, even in serious tumors (e.g., osteosarcoma). Stents regained round shape within two days. Complications were reported as uncommon, including migration (9.5%) and obstruction by secretions.
Technological Characteristics
Medical-grade silicone (polysiloxane) construction. Features external stubs for anti-migration. Surface treated with a proprietary coating to reduce roughness, prevent tissue adhesion, and facilitate mucus flow. Available in multiple shapes (standard, pediatric, Y-stent, hourglass). Non-armed design.
Indications for Use
Indicated for patients with tracheobronchial tumors, tracheal stenosis with scarring, or bronchial stenosis following surgical anastomotic resections or pulmonary transplantation. Applicable when endoscopic resection is incomplete, extrinsic compression persists, or airway lumen is insufficient.
Regulatory Classification
Identification
The tracheal prosthesis is a rigid, flexible, or expandable tubular device made of a silicone, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the trachea or trachealbronchial tree. It may be unbranched or contain one or two branches. The metal tracheal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
Predicate Devices
ENDOXANE Stent (Bryan Corporation) (K4894380)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, a staff with two snakes entwined around it, which is a common symbol associated with medicine and healthcare.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 8 2005
Mr. Bruno Ferreyrol Regulatory Affairs Office Novatech S.A. Avenu du Vent des Dames Z1 des Paluds 13685 Aubagne France
Re: K971509
Trade/Device Name: ENDOXANE® Regulation Number: 878.3720 Regulation Name: Prosthesis, tracheal, non-expandable, silicone Regulatory Class: 2 Product Code: NWA Dated: April 23, 1997 Received: April 25, 1997
Dear Mr. Ferreyrol:
This letter corrects our substantially equivalent letter of August 21, 1997.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
## Page 2 - Mr. Bruno Ferreyrol
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmaldsmamain.html.
Sincerely yours,
Sincerely yours,
Mark N Wilkerson
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure: Indication for use statement for K971509.
{2}------------------------------------------------
Page
i(10(k) Number (if known): K971509
Device Name: ENDOXANF
Indic
MBICATIONS ARE LISTED ACCORDING TO THEIR FREQUENCY : -TRACHEOBRONCIAL TUNORS'
-TRACHEAL STENDSIS WITH SCARRING
BRONCIAL STANQSIS AFTER SURGICAL ANASTOMOTIC, ANASTOHOSES RESECTIONS OR PULMONARY THANSPLANTATION
THE STENDSES HAVE AS A COMMON CHARACTERISTIC THE IMPORTUNIE IN THE EXTENSIC CEMPRESSION OR THE COLLAPSE OF THE CARTILAGENS MALL. INDICNATION ATHARS FOCUSES ON AN ENDOSCORIC THERAPEUTIC ACCIENCY AND Lies a IT CONCERNS A RESECTION (LASER, CRYO, THERMO COGULATION) OR CHATAL. " THE SWENT INSERTEDN IS INDICATED ID THE EWOOSCOPIC RESECTICA 15 15-CONFLETTE IN CARR OF A PERSISTENT EXTRINGIC COMPRESSION OR IF THE LUMEN OF THE ADMANY IS INSUPEICIENT. THE MOST FREQUENT LOCALICAS ARF TRACENAL. THE LEBET MAIN BRONCHUS IS THE SECOND LOCATION FOLLOWED JY THE RIGHT MAIN BRONCHUS.
ASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)
Concurrence of CORH, Office of Device Evaluation 100E1
Mark N. Millan
(Division Sign-Off)
Division of General Restorative Devices
| <strong>Labels</strong> | <strong>Values</strong> |
|-------------------------|-------------------------|
| 510(k) Number | K971459 |
OR
7997-37
CZZ6Z3Z77B
Over The-Counter Use
ર દ
(Optional Formal 1-2-96)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Novatech. The logo is in black and white, with the word "NOVATECH" in large, bold letters. Below the word "NOVATECH" is the phrase "NEW BIOTECHNOLOGY FOR LIFE" in smaller letters. A white curved line is below the word "NOVATECH".
ﻌﺎ
K 971509
## 510(k) Summary
Date of Application :
April 23td, 1997
Applicant/Manufacturer Address :
Telephone Number :
Fax Number :
网
Name of Contact Person:
NOVATECH S.A. Avenue du Vent des Dames ZI des Paluds 13685 Aubagne France
(33) 442 82 92 92
(33) 442 82 92 27
Bruno FERREYROL Executive Director
Certification
N 460002 - ISO 9002
NOVATECH S.A. Avenue du Vent des Dames Z.1. les Paluds 13685 AUBAGNE - FRANCE TFL. (33) 04 42 82 92 92 - FAX. (33) 04 42 82 92 27
{4}------------------------------------------------
The proprietary name or trade name of our medical device is ENDOXANE ¯ ¯ he more common name of this device is an endo tracheal and endo bronchial stent, or Dumon Stent.
The Food and Drug Administration considers this device as a Class III - Tracheal Prosthesis.
The legally marketed device that we have cited for comparison is the ENDOXANE ® Stent, the same as ours, which was sold by our sole United States distributor, the Bryan Corporation (510K # K 4894380).
The device description can be found in Section IV Description of Device. In short, the ENDOXANE® is an endoscopy stent made of silicone. This stent, the most fitted in the world, is presented as a silicone tube with stubs allowing to avoid any movement. The non armed polysiloxane is covered with a new coating especially designed in order to keep its' flexibility which facilitates tolerance and evacuation of secretions.
A well-adapted ENDOXANE® does not move and cough reflexes are inhibited by reflex center saturation. The stubs prevent a direct contact between the stent surface and the mucus membrane and distribute the pressures among the small surfaces.
The main characteristics of ENDOXANE ® Stents can be divided into three main categories :
- 1. Histo compatibility : The medical grade silicone used for the manufacturing of ENDOXANE® is perfectly compatible during prolonged contact with trachea bronchial tissues.
- 2. Anti-encrustation and Anti-adhesion : NOVATECH has designed a new coating which guarantees high quality treatment of the surface favoring an easy flow of mucus. The surface is homogeneous and has a very fine grain. The parameter roughness is very weak. Furthermore, the surface treatment used avoids any tissue coloring and facilitates the positioning as well as the removal process. The extremities of the stents are manufactured according to a patented system which considerably reduces the risks of "mechanic" accumulation of mucus. The sides allows being non traumatic as no obstacle of mucus flow can be detected.
- 3. Anti-Migration System : All types of ENDOXANE® stents benefit from an exclusive and patented maintaining system. A perfect adhesion is assured by stubs found on the outer surface giving support to the surface and interpolating between the cartilaginous rings. This configuration prevents possible risks of stent migration if the size is well chosen.
The main indications according to their frequency are :
- 1. tracheobronchial tumors
- 2. Tracheal stenosis with scarring
- 3. Bronchial stenosis after surgical anastomotic, anastomoses resections or pulmonary transplantation.
These stenoses have, as a common characteristic, the importance of the extrinsic compression or the collapse of the cartilingeous wall. The indication always focuses on an endoscopic therapeutic action, whether is concerns a resection or dilation. The stent insertion is indicated if the endoscopic resection is incomplete, in case of a persistent extrinsic compression, or if the lumen is insufficient. The most frequent locations are tracheal. The left main bronchus is the second location, followed by the right main bronchus.
Our medical device has the same technological characteristics as the predicate device being that the two devices are the same. The only technological advancement that differs between the two devices is the new coating that we created in 1996. We have also developed three new shapes of the ENDOXANE Stents, each one with its particular use. The ENDOXANE BB Pediatric Stent was created in 1994 and is equipped with two lines of stubs opposite each other to facilitate the passage through the vocals cords. The ENDOXANE Y Stent is basically the same stent as Bryan Corporation but we added new dimension which are 15-12-12. This model is still used for carina pathologies associated with trachea and bronchus compression. The third new shape is the ENDOXANE ST Stent (hourglass shape) which is perfectly adapted to post intubation benion stenosis.
The conclusion that one can reach concerning our stents is best stated in the study published in "The Journal of Bronchology" (3:6-10, 1996) about the placing of 1500 stents (Cavalière, Diaz, Dumon, Vergnon). Not a single case of compression was observed, even in cases of the most serious tumors (osteosarcoma, for example). In all of the cases of extreme compression, the stent that was oval shaped at the outset regained its round shape in less than two days. "Complications were uncommon, usually easily manage, and rarely life threatening. The main complications were emigration (9.5%)n obstruction by secretions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.