K242988 · Genostis AF · HRS · Jun 27, 2025 · Orthopedic
Device Facts
Record ID
K242988
Device Name
Genostis Osteosynthesis System
Applicant
Genostis AF
Product Code
HRS · Orthopedic
Decision Date
Jun 27, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Plate and screw Implants are intended for temporary fixation, correction or stabilization of bones in various anatomical regions. One-Quarter Tubular Plates 2.4/2.7 are indicated for fixation of fractures of the foot, hand, wrist, and forearm (including radius). The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression. Straight Plates 2.4/2.7 are indicated for fixation of fractures, osteotomies, malunions and nonunions of the foot, hand, wrist, and forearm (including radius) and arthrodesis of small bones of the foot and hand. Reconstruction Plates 2.4/2.7 are indicated for fixation of fractures of the foot, hand, wrist, and forearm (including radius) Radius Distal Styloid Plate 2.4/2.7 is indicated for the fixation for fractures of the radial styloid that have metaphyseal comminution or as an adjunct to either volar of dorsal plate fixation when additional support of the radial styloid is required. The Radius Proximal head plates 2.4/2.7 are indicated for extra-articular and intra-articular fractures of the proximal radius and multifragmented radial neck fractures. The Radius Distal Volar Plates 2.4/2.7 are indicated for the fixation for fractures, fusions, or osteotomies of the distal radius. One-Third Tubular Plates 3.5 are indicated for fixation of fractures of the elbow (distal humerus and ulna), and fibula. The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression. Reconstruction Plates 3.5 are indicated for fixation of fractures of the elbow (distal humerus and ulna), wrist, and forearm (including radius), fibula, and calcaneus Straight Plates 3.5 are indicated for fixation of fractures, osteotomies, malunions and nonunions of the elbow (distal humerus and ulna), wrist, forearm (including radius), and fibula. Tibia T-Plates 3.5 are indicated for buttressing of partial articular fractures and bone fragments of the distal tibia. The Clavicle Superior Plates 3.5 are indicated for: - Fractures of the clavicle shaft - Fractures of the lateral clavicle - Malunions of the clavicle - Non-unions of the clavicle The Clavicle Hook Plate 3.5 is indicated for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint. The Humerus Proximal Plates 3.5 are indicated for the fixation of fractures, fracture dislocations, non-unions and osteotomies of the proximal humerus. The Tibia Proximal Lateral Plates 3.5 are indicated for: - Split-type fractures of the lateral tibial plateau - Lateral split fractures with associated depressions - Pure central depression fractures - Split or depression fractures of the medial plateau The Tibia Distal Medial Plates 3.5 are indicated for the fixation of fractures, osteotomies, and non-unions of the distal tibia. The Tibia Distal Anterolateral Plates 3.5 are indicated for the fixation of fractures, osteotomies, and non-unions of the distal tibia. The Humerus Distal Dorsolateral Plates 2.7/3.5 are indicated for - Intraarticular fractures of the distal humerus - Supracondylar fractures of the distal humerus - Non-unions of the distal humerus The Humerus Distal Medial Plates 2.7/3.5 are indicated for - Intraarticular fractures of the distal humerus - Supracondylar fractures of the distal humerus - Non-unions of the distal humerus The Ulna Proximal Olecranon Plates 3.5 are indicated for - Complex extra- and intra-articular olecranon fractures - Non-unions of the proximal ulna - Osteotomies - Simple olecranon fractures Broad Curved Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, femur, and tibia. Broad Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, femur, and tibia. Narrow Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, and distal tibia. T Plate 4.5/5.0 is intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression. T-Buttress Plates 4.5/5.0 are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load sharing purposes, e.g., buttressing, tension banding, neutralization, or compression. LL-Buttress Plates 4.5/5.0 are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load sharing purposes, e.g., buttressing, tension banding, neutralization, or compression. Femur Distal Plates 4.5/5.0 are indicated for distal shaft fractures, supracondylar fractures, intraarticular fractures, and periprosthetic fractures of the distal femur Tibia Proximal Lateral Plates 4.5/5.0 are indicated for proximal shaft fractures, metaphyseal fractures, intra-articular fractures and periprosthetic fractures of the proximal tibia.
Device Story
Genostis Osteosynthesis System consists of metallic bone fixation plates, screws, and surgical instruments; used for temporary bone stabilization, fracture fixation, and osteotomies. Implants feature locking, non-locking, or combination hole interfaces. System utilized by surgeons in clinical settings to achieve bone alignment and stability. Output is physical fixation of bone segments; aids healing by maintaining anatomical position. Benefits include restoration of skeletal integrity and support for bone union. Delivered sterile via radiation.
Clinical Evidence
Bench testing only. Includes biomechanical testing (ASTM F543 for screws, ASTM F382 for plates) and biocompatibility testing (ISO 10993-5, -10, -18, -23). Bacterial endotoxin testing performed. All results satisfactory.
Technological Characteristics
Materials: Titanium alloy (Ti 6Al 4V ELI) per ASTM F136 and ISO 5832-3. Design: Fracture fixation plates and screws with locking/non-locking interfaces. Sterilization: Radiation. Connectivity: None.
Indications for Use
Indicated for temporary fixation, correction, or stabilization of bone fractures, osteotomies, malunions, nonunions, and arthrodesis in various anatomical regions including foot, hand, wrist, forearm, elbow, humerus, femur, tibia, fibula, and clavicle. Specific indications vary by plate type (e.g., buttressing, tension banding, neutralization, compression).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Synthes LCP Low Bend Medial Distal Tibia Plates 3.5mm (K001945)
Synthes 4.5 mm Narrow LCP Plates and 5.00 mm Locking Screw (K000682)
Synthes LCP Olecranon Plate and Sysnthes 3.5 mm Locking Screw (K000684)
Synthes Quarter Tubular Plate 2.7 with Collar (K010321)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 27, 2025
Genostis AF
% Cheryl Wagoner
Principal Consultant/ Owner Wagoner Consulting LLC
Wagoner Consulting LLC
5215 Crosswinds Drive
Wilmington, North Carolina 28409
Re: K242988
Trade/Device Name: Genostis Osteosynthesis System
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Code: HRS, HWC
Dated: May 23, 2025
Received: May 23, 2025
Dear Cheryl Wagoner:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242988 - Cheryl Wagoner
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K242988 - Cheryl Wagoner
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thomas Mcnamara -S
For: Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K242988
Device Name
Genostis Osteosythesis System
Indications for Use (Describe)
Intended Use: Plate and screw Implants are intended for temporary fixation, correction or stabilization of bones in various anatomical regions.
One-Quarter Tubular Plates 2.4/2.7 are indicated for fixation of fractures of the foot, hand, wrist, and forearm (including radius). The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
Straight Plates 2.4/2.7 are indicated for fixation of fractures, osteotomies, malunions and nonunions of the foot, hand, wrist, and forearm (including radius) and arthrodesis of small bones of the foot and hand.
Reconstruction Plates 2.4/2.7 are indicated for fixation of fractures of the foot, hand, wrist, and forearm (including radius)
Radius Distal Styloid Plate 2.4/2.7 is indicated for the fixation for fractures of the radial styloid that have metaphyseal comminution or as an adjunct to either volar of dorsal plate fixation when additional support of the radial styloid is required.
The Radius Proximal head plates 2.4/2.7 are indicated for extra-articular and intra-articular fractures of the proximal radius and multifragmented radial neck fractures.
The Radius Distal Volar Plates 2.4/2.7 are indicated for the fixation for fractures, fusions, or osteotomies of the distal radius.
One-Third Tubular Plates 3.5 are indicated for fixation of fractures of the elbow (distal humerus and ulna), and fibula. The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
Reconstruction Plates 3.5 are indicated for fixation of fractures of the elbow (distal humerus and ulna), wrist, and forearm (including radius), fibula, and calcaneus
Straight Plates 3.5 are indicated for fixation of fractures, osteotomies, malunions and nonunions of the elbow (distal humerus and ulna), wrist, forearm (including radius), and fibula.
Tibia T-Plates 3.5 are indicated for buttressing of partial articular fractures and bone fragments of the distal tibia.
The Clavicle Superior Plates 3.5 are indicated for:
- Fractures of the clavicle shaft
- Fractures of the lateral clavicle
- Malunions of the clavicle
- Non-unions of the clavicle
The Clavicle Hook Plate 3.5 is indicated for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint.
FORM FDA 3881 (8/23)
PTC Publishing Services (301) 443-6740
{4}
The Humerus Proximal Plates 3.5 are indicated for the fixation of fractures, fracture dislocations, non-unions and osteotomies of the proximal humerus.
The Tibia Proximal Lateral Plates 3.5 are indicated for:
- Split-type fractures of the lateral tibial plateau
- Lateral split fractures with associated depressions
- Pure central depression fractures
- Split or depression fractures of the medial plateau
The Tibia Distal Medial Plates 3.5 are indicated for the fixation of fractures, osteotomies, and non-unions of the distal tibia.
The Tibia Distal Anterolateral Plates 3.5 are indicated for the fixation of fractures, osteotomies, and non-unions of the distal tibia.
The Humerus Distal Dorsolateral Plates 2.7/3.5 are indicated for
- Intraarticular fractures of the distal humerus
- Supracondylar fractures of the distal humerus
- Non-unions of the distal humerus
The Humerus Distal Medial Plates 2.7/3.5 are indicated for
- Intraarticular fractures of the distal humerus
- Supracondylar fractures of the distal humerus
- Non-unions of the distal humerus
The Ulna Proximal Olecranon Plates 3.5 are indicated for
- Complex extra- and intra-articular olecranon fractures
- Non-unions of the proximal ulna
- Osteotomies
- Simple olecranon fractures
Broad Curved Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, femur, and tibia.
Broad Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, femur, and tibia.
Narrow Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, and distal tibia.
T Plate 4.5/5.0 is intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
T-Buttress Plates 4.5/5.0 are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia.
The plates should only be used for load sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
LL-Buttress Plates 4.5/5.0 are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia.
The plates should only be used for load sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
FORM FDA 3881 (8/23)
Page 2 of 3
PSC Publishing Services (301) 443-6740
EF
{5}
Femur Distal Plates 4.5/5.0 are indicated for distal shaft fractures, supracondylar fractures, intraarticular fractures, and periprosthetic fractures of the distal femur
Tibia Proximal Lateral Plates 4.5/5.0 are indicated for proximal shaft fractures, metaphyseal fractures, intra-articular fractures and periprosthetic fractures of the proximal tibia.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
PSC Publishing Services (301) 443-6740
Page 164
{6}
510(k) #: K242988
510(k) Summary
Prepared on: 2025-05-22
# Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Genostis AG |
| --- | --- |
| Applicant Address | Kirchbergstrasse 160 Burgdorf 3400 Switzerland |
| Applicant Contact Telephone | +41764559774 |
| Applicant Contact | Dr. Sabine Söhndel |
| Applicant Contact Email | sabine.soehndel@genostis.com |
| Correspondent Name | Wagoner Consulting LLC |
| Correspondent Address | 5215 Crosswinds Drive Wilmington NC 28409 United States |
| Correspondent Contact Telephone | 910-386-9019 |
| Correspondent Contact | Mrs. Cheryl Wagoner |
| Correspondent Contact Email | cheryl@wagonerconsultingllc.com |
# Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | Genostis Osteosynthesis System |
| --- | --- |
| Common Name | Single/multiple component metallic bone fixation appliances and accessories |
| Classification Name | Plate, Fixation, Bone |
| Regulation Number | 888.3030 |
| Product Code(s) | HRS, HWC |
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K001945 | Synthes LCP Low Bend Medial Distal Tibia Plates 3.5mm | HRS |
| K000682 | Synthes 4.5 mm Narrow LCP Plates and 5.00 mm Locking Screw | HRS |
| K000684 | Synthes LCP Olecranon Plate and Sysnthes 3.5 mm Locking Screw | HRS |
| K010321 | Synthes Quarter Tubular Plate 2.7 with Collar | HRS |
| K010766 | LCP L-Buttress Plates 4.5/5.0 | HRS |
| K011335 | Synthes 3.5 mm LCP one-third tubular plate | HRS |
Page 265
{7}
Page 266
| K030597 | Synthes LCP Proximal Tibial Plate 3.5, Lateral | HRS |
| --- | --- | --- |
| K033995 | Synthes LCP Distal Humerus Plate , dorsolateral with support | HRS |
| K040777 | Synthes LCP Proximal Radius Plates 2.4 | HRS |
| K041911 | Synthes 4.5 mm Curved Broad LCP Plates | HRS |
| K052390 | Synthes LCP Proximal Lateral Tibia Plate 4.5/5.0 | HRS |
| K061753 | Synthes LCP Clavicle Hook Plate | HRS |
| K062564 | Synthes LCP Distal Femur Plates | HRS |
| K063049 | Synthes LCP plate 2.7 straight Synthes 2.4/2.7 mm locking screw | HRS |
| K073186 | Synthes LCP Superior Clavicle Plate | HRS |
| K080522 | Synthes 3.5 mm T Plate | HRS |
| K082625 | Synthes PHILOS long-proximal diaphyseal humeral plate 3.5 | HRS |
| K092812 | Synthes LCP Anterolateral Distal Tibia Plate 3.5 | HRS |
| K102998 | Acumed Acu-Loc Plating System | HRS |
| K112583 | Synthes 2.7/3.5 mm Cortex Screw | HWC |
| K161616 | Synthes 4.5 mm Cortex Screw | HWC |
| K071715 | Acumed Acu_LOC 2 Plating System | HRS |
## Device Description Summary
**21 CFR 807.92(a)(4)**
The Genostis Osteosynthesis System comprises of fracture fixation plates, corresponding screws and required surgical instruments. Different shapes and sizes are available to accommodate for different anatomical regions and locations of fractures. The plate screw interface is either locking, non-locking or, for some plates, realized by a combination hole offering both options (locking and non-locking).
All plates and screws are made of titanium alloy (Ti 6Al 4V ELI) according to ASTM F136 and ISO 5832-3.
The implants of the Genostis Osteosynthesis System are sterilized by radiation and are delivered sterile.
## Intended Use/Indications for Use
**21 CFR 807.92(a)(5)**
Intended Use: Plate and screw Implants are intended for temporary fixation, correction or stabilization of bones in various anatomical regions.
One-Quarter Tubular Plates 2.4/2.7 are indicated for fixation of fractures of the foot, hand, wrist, and forearm (including radius). The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
Straight Plates 2.4/2.7 are indicated for fixation of fractures, osteotomies, malunions and nonunions of the foot, hand, wrist, and forearm (including radius) and arthrodesis of small bones of the foot and hand.
Reconstruction Plates 2.4/2.7 are indicated for fixation of fractures of the foot, hand, wrist, and forearm (including radius)
{8}
Radius Distal Styloid Plate 2.4/2.7 is indicated for the fixation for fractures of the radial styloid that have metaphyseal comminution or as an adjunct to either volar of dorsal plate fixation when additional support of the radial styloid is required.
The Radius Proximal head plates 2.4/2.7 are indicated for extra-articular and intra-articular fractures of the proximal radius and multifragmented radial neck fractures.
The Radius Distal Volar Plates 2.4/2.7 are indicated for the fixation for fractures, fusions, or osteotomies of the distal radius.
One-Third Tubular Plates 3.5 are indicated for fixation of fractures of the elbow (distal humerus and ulna), and fibula. The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
Reconstruction Plates 3.5 are indicated for fixation of fractures of the elbow (distal humerus and ulna), wrist, and forearm (including radius), fibula, and calcaneus
Straight Plates 3.5 are indicated for fixation of fractures, osteotomies, malunions and nonunions of the elbow (distal humerus and ulna), wrist, forearm (including radius), and fibula.
Tibia T-Plates 3.5 are indicated for buttressing of partial articular fractures and bone fragments of the distal tibia.
The Clavicle Superior Plates 3.5 are indicated for:
- Fractures of the clavicle shaft
- Fractures of the lateral clavicle
- Malunions of the clavicle
- Non-unions of the clavicle
The Clavicle Hook Plate 3.5 is indicated for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint.
The Humerus Proximal Plates 3.5 are indicated for the fixation of fractures, fracture dislocations, non-unions and osteotomies of the proximal humerus.
The Tibia Proximal Lateral Plates 3.5 are indicated for:
- Split-type fractures of the lateral tibial plateau
- Lateral split fractures with associated depressions
- Pure central depression fractures
- Split or depression fractures of the medial plateau
The Tibia Distal Medial Plates 3.5 are indicated for the fixation of fractures, osteotomies, and non-unions of the distal tibia.
The Tibia Distal Anterolateral Plates 3.5 are indicated for the fixation of fractures, osteotomies, and non-unions of the distal tibia.
The Humerus Distal Dorsolateral Plates 2.7/3.5 are indicated for
- Intraarticular fractures of the distal humerus
- Supracondylar fractures of the distal humerus
- Non-unions of the distal humerus
The Humerus Distal Medial Plates 2.7/3.5 are indicated for
- Intraarticular fractures of the distal humerus
- Supracondylar fractures of the distal humerus
- Non-unions of the distal humerus
The Ulna Proximal Olecranon Plates 3.5 are indicated for
- Complex extra- and intra-articular olecranon fractures
- Non-unions of the proximal ulna
- Osteotomies
- Simple olecranon fractures
Broad Curved Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, femur, and tibia.
Broad Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, femur, and tibia.
Narrow Plates 4.5/5.0 are indicated for fixation of fractures, osteotomies, malunions and nonunions of humerus, and distal tibia.
{9}
Page 268
T Plate 4.5/5.0 is intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load-sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
T-Buttress Plates 4.5/5.0 are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
LL-Buttress Plates 4.5/5.0 are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial plateau and distal tibia. The plates should only be used for load sharing purposes, e.g., buttressing, tension banding, neutralization, or compression.
Femur Distal Plates 4.5/5.0 are indicated for distal shaft fractures, supracondylar fractures, intraarticular fractures, and periprosthetic fractures of the distal femur
Tibia Proximal Lateral Plates 4.5/5.0 are indicated for proximal shaft fractures, metaphyseal fractures, intra-articular fractures and periprosthetic fractures of the proximal tibia
| Indications for Use Comparison | 21 CFR 807.92(a)(5) |
| --- | --- |
| The indications between the Subject device and its predicates are the same with minor wording differences. | |
| Technological Comparison | 21 CFR 807.92(a)(6) |
| The Subject and predicate devices have the same technological characteristics including similar designs, materials, chemical composition and principle of operation. | |
| Non-Clinical and/or Clinical Tests Summary & Conclusions | 21 CFR 807.92(b) |
This application includes biomechanical and biocompatibility testing per appropriate ISO and ASTM standards. All results were satisfactory. All standards are included in the standards section of this application.
Bacterial endotoxin testing was performed. Cytotoxicity (ISO 10993-5), Sensitization (ISO 10993-10), and Irritation/Intracutaneous reactivity (ISO 10993-23) were performed. Chemical characterization (ISO 10993-18) and toxicological risk assessment were also performed as part of the biocompatibility risk assessment. Screw performance testing per ASTM F543 and plate performance testing per ASTM F382 were performed.
All testing indicates that the Subject device is substantially equivalent to the predicates presented in this application. Each test was conducted on final sterile product.
For clarity, the buttress plates include: - One Third Tubular Plates 3.5, One Quarter Tubular Plates 2.4/2.7, Tibia T-Plate 3.5, T Plate 4.5/5.0, T- Buttress Plate 4.5/5.0, L - Buttress Plate 4.5/5.0
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.