SYNTHES LARGE FRAGMENT DYNAMIC COMPRESSION LOCKING (DCL) SYSTEM
Applicant
Synthes (Usa)
Product Code
KTT · Orthopedic
Decision Date
May 1, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Synthes Large Fragment DCL is intended for fixation of various long bones, such as the humerus, femur and tibia. It is also for use in fixation of osteopenic bone and fixation of non-unions or malunions.
Device Story
The Synthes Large Fragment DCL System is a metallic bone fixation plate and screw system used to treat long bone fractures. The system features limited-contact profile plates with combination holes that allow for both dynamic compression (using standard cortex, cancellous, or cannulated screws) and fixed-angle locking (using 5.0 mm threaded locking screws). The locking mechanism creates a fixed-angle construct by engaging the screw threads with the threaded plate holes, providing stability for unicortical or bicortical purchase. The device is intended for use by surgeons in clinical settings to stabilize fractures, non-unions, or malunions, including in osteopenic bone. The system includes broad and narrow plates of varying lengths and thicknesses. The device is provided sterile or non-sterile (requiring steam sterilization).
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Materials: 316L stainless steel (ASTM F139), commercially pure titanium (ASTM F67), Ti-6Al-7Nb (ASTM F1295). Design: Limited-contact profile plates with combination dynamic compression/locking holes. Locking mechanism: Threaded conical screw heads engaging with threaded plate holes. Dimensions: Broad plates (5.2 mm thick, 17.5 mm wide), narrow plates (4.2 mm thick, 13.5 mm wide). Sterilization: Moist heat (steam) validated to SAL 10^-6 per AAMI guidelines.
Indications for Use
Indicated for fixation of long bones (humerus, femur, tibia), osteopenic bone, and non-unions or malunions in patients requiring orthopedic bone fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Synthes Broad and Narrow Dynamic Compression Plates
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a design inside. The logo appears to be a stylized representation of a medical device or implant. The overall image has a clean and professional look.
### 3.0 Summary of Safety and Effectiveness Information:
# Summary of Safety and Effectiveness Information
| SPONSOR: | Synthes (USA)<br>1690 Russell Road<br>Paoli, PA 19301 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | (610) 647-9700 |
| CONTACT: | Angela Silvestri |
| DEVICE NAME:: | Synthes Large Fragment Dynamic Compression Locking (DCL)<br>System |
| DEVICE CLASSIFICATION: | 21 CFR 888.3030 Single/multiple component metallic bone fixation<br>appliances and accessories and 888.3040 Smooth/threaded metallic<br>bone fixation fastener. |
| PREDICATE DEVICE: | Synthes Broad and Narrow Dynamic Compression Plates<br>Synthes Anatomical Locking Plate System |
| DESCRIPTION OF DEVICE: | The Large Fragment DCL system consists of limited-contact profile<br>plates in broad and narrow sizes, which include combination dynamic<br>compression/locking screw holes. The plates accept 4.5 mm cortex,<br>6.5 mm cancellous, 4.5 mm cannulated, 7.0 mm cannulated, and 5.0<br>mm locking screws. This device is manufactured in either stainless<br>steel or titanium. |
| INDICATIONS: | Synthes Large Fragment DCL is intended for fixation of various long<br>bones, such as the humerus, femur and tibia. It is also for use in<br>fixation of osteopenic bone and fixation of non-unions or malunions. |
:
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a logo that appears to be a stylized image of a person. To the right of the word is a small circle with a "c" inside, indicating a copyright.
#### 4.0 Device Name:
Synthes Large Fragment Dynamic Compression Locking (DCL) System
#### રું... Establishment Registration:
Synthes is registered with the Device Registration and Listing Branch (1719045) of the Food and Drug Administration (FDA). The devices subject to this premarket notification are manufactured by Synthes (USA), 1101 Synthes Avenue, Monument, CO 80132.
#### 6.0 Classification Information:
The classification of the Synthes Large Fragment DCL System in our determination. Class II, as per Title 21 of the Code of Federal Regulations (21 CFR), sections 888.3030: Single/multiple component metallic bone fixation appliances and accessories and 888.3040 Smooth/threaded metallic bone fixation fastener.
#### 7.0 Information Relating to Performance Standards and Special Controls:
Synthes Large Fragment DCL System is manufactured from 316L stainless steel, which adheres to American Society for Testing and Materials (ASTM) standard F 139; commercially pure titanium, which adheres to ASTM standard F67; and Titanium - 6% Aluminum - 7% Niobium (Ti-6Al-7Nb), which adheres to ASTM standard F 1295. Synthes is not aware of any performance standards or special controls established to date.
#### Sterilization Information: 8.0
Synthes will provide the device both sterile. Sterilization information on the sterile device can be found in Attachment I (a). Of course, non-sterile devices must be sterilized prior to use; moist heat sterilization is recommended using the parameters identified in These parameters have been validated using the Association for the Attachment I (b). Advancement of Medical Instrumentation (AAMI) guideline Steam Sterility Validation to provide a sterility assurance level (SAL) of 10t6.
#### 9.0 Description of the Device:
Synthes DCL System is a plate and screw system intended to treat fractures of various long bones. The primary feature of this system is limited-contact profile and the combination dynamic compression/locking screw holes. The standard screws facilitate reduction and create compression between the plate and bone, while the locking screws form a locked, fixed angle construct with the plate.
The DCL System includes both broad and narrow straight plates. The broad plates are 5.2 mm thick, 17.5 mm wide and are available in lengths ranging from 121 mm - 409 mm (6 - 22 screw holes). The narrow plates are 4.2 mm thick, 13.5 mm wide and are available in lengths ranging from 49 mm - 409 mm (2 - 22 screw holes). The screw holes are combined to allow placement of standard 4.5 mm cortex, 6.5 mm cancellous and 4.5 mm and 7.0 mm cannulated screws on one side of each hole, or 5.0 mm threaded conical, locking screws on the opposite side. The holes are
Confidential
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a stylized image inside. To the right of the word is a registered trademark symbol.
uniformly spaced along the length of the plate. The holes are oriented so that the compression component of the holes is always directed towards the middle of the plate.
The DCL System utilizes a locking feature that screw to the plate, enabling unicortical or bicortical screw purchase. This feature consists of conical shaped, threaded locking screw holes in the plates that correspond with the conical- shaped, threaded head of the screw.
The 5.0 mm locking screws feature self-tapping tips, flat head profiles with rounded edges, and are available in lengths from 14mm to 90mm. The threads just below the head of each locking screw engages with the threaded side of the holes in the DCL plate. The engagement of these screws to the plate creates a locked, fixed angle construct.
Confidential engineering drawings may be found in Attachment II.
Synthes 4.5 mm cortex, 6.5 mm cancellous and 7.0 mm cannulated screws that are utilized with this system are commercially available devices.
The following Class I Exempt instruments are utilized with the DCL system: 4.3 mm drill bit and 4.3mm threaded drill guide.
#### 10.0 Proposed Labels / Labeling:
Please see Attachment III for sterile and non-sterile device labeling.
#### 11.0 Commercially Available Device Information:
The following Synthes devices are commercially available: Narrow and Broad Dynamic Compression Plates (pre-amendment status), Anatomical Locking Plate System (a.k.a. Synthes LISS) (K961413), 4.5 mm cortex and 6.5 mm cancellous (pre-amendment status), 4.5 mm Cannulated Screws (K963172) and 7.0 mm Cannulated Screws (K962011). Please see Attachment IV for information on these devices.
## 12.0 Comparison to Commercially Available Device:
A comparison of Synthes' DCL System to Synthes Broad and Narrow Dynamic Compression Plates and Synthes Anatomical Locking Plate System follows:
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized eagle with three wing-like shapes extending from its head.
Public Health Service
MAY = 1 2000
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Angela J. Silvestri Manager, Regulatory Affairs SYNTHES (USA) 1690 Russell Road Post Office Box 1766 Paoli, Pennsylvania 19301-1222
Re: K000682
Trade Name: Large Fragment Dynamic Compression Locking (DCL) System Regulatory Class: II Product Code: KTT Dated: February 25, 2000 Received: February 29, 2000
Dear Ms. Silvestri:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to vour premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{4}------------------------------------------------
Page 2 - Ms. Angela J. Silvestri
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Donna R. Lochner
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Synthes logo. The logo consists of a circular symbol on the left and the word "SYNTHES" in bold, sans-serif font on the right. The circular symbol contains a stylized image of a person or figure. The word "SYNTHES" is in all capital letters and appears to be slightly italicized.
#### 1.0 Indications for Use Statement
Page _________________________________________________________________________________________________________________________________________________________________________
KOOO 682 ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------510(k) Number (if known):
Synthes (USA) Large Fragment Dynamic Compression Locking (DCL) System Device Name: ____________
Indications for use:
Synthes Large Fragment DCL is intended for fixation of various long bones, such as the humerus, femur and tibia. It is also for use in fixation of osteopenic bone and fixation of non-unions or malunions.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
Dans R. Lochner
(Division Sign-Off) (Division Sign-On)
Division of General Restorative Devices Division on Senon Kooo682
Confidential
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.