The LDS Needle is intended for use with a luer-tip syringe for the administration of drugs. The OcuSafe® LDS Needle is intended for use with a luer-tip syringe for the administration of drugs. The OcuSafe® LDS Needle is indicated for intravitreal use. The SteriCap® LDS Safety Needle is intended for use with a luer-tip syringe for the administration of drugs. The SteriCap Safety Needle has an integrated passive needle cap/shield that covers the needle prior to use and automatically activates to cover the needle immediately after use. A lock mechanism can be manually activated to lock/unlock the needle cap. The locked needle cap minimizes risk of accidental needle stick. The VitreJect® LDS Safety Needle is intended for use with a luer-tip syringe for the administration of drugs. The VitreJect® LDS Safety Needle has an integrated passive needle cap/shield that covers the needle prior to use and automatically activates to cover the needle immediately after use. A lock mechanism can be manually activated to lock/unlock the needle cap. The locked needle cap minimizes risk of accidental needle stick. The VitreJect® LDS Safety Needle is indicated for intravitreal use.
Device Story
Single-lumen needles designed for fluid injection/aspiration; used with luer-tip syringes. Features Low Dead Space (LDS) specification (≤ 5µL) to reduce fluid volume loss. SteriCap® and VitreJect® models include spring-actuated, non-removable sliding safety caps for passive needle shielding post-use; manual lock mechanism included. Used by healthcare professionals in clinical settings. Devices are single-use, sterile (EO), non-pyrogenic. Intravitreal variants (OcuSafe®, VitreJect®) support ophthalmic procedures. Output is manual drug delivery; safety features reduce risk of sharps injury to clinicians.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, hemolysis, ocular/intravitreal irritation). Performance testing included dimensional verification, bond/material strength, needle quality, color coding, luer connector, particulate testing, and sharps injury prevention. Sterilization validated per ISO 11135; shelf-life validated via accelerated aging (ASTM F1980-16).
Technological Characteristics
Materials: Stainless steel needle, polypropylene hub/cap, silicone lubricant, polyacrylate adhesive. Design: Single-lumen, luer-slip/lock connection, lancet bevel. Safety models include stainless steel spring. Standards: ISO 7864, ISO 9626, ISO 80369-7, ISO 6009, ISO 10993 series. Sterilization: Ethylene Oxide (SAL 10^-6). LDS spec: ≤ 5µL.
Indications for Use
Indicated for administration of drugs via luer-tip syringe. OcuSafe® and VitreJect® variants are indicated for intravitreal use. SteriCap® and VitreJect® variants include safety features to minimize accidental needle stick.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 25, 2024
OcuJect, LLC % Julie Stephens President Regulatory Resources Group, Inc. 111 Laurel Ridge Dr Alpharetta, Georgia 30004
Re: K242956
Trade/Device Name: LDS Needle; OcuSafe® LDS Needle; SteriCap® LDS Safety Needle; VitreJect® LDS Safety Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: QNS, FMI, QYM Dated: September 25, 2024 Received: September 25, 2024
Dear Julie Stephens:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Shruti N. Mistry -S
Shruti Mistry Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242956
#### Device Name
LDS Needle; OcuSafe® LDS Needle; SteriCap® LDS Safety Needle; VitreJect® LDS Safety Needle
#### Indications for Use (Describe)
The LDS Needle is intended for use with a luer-tip syringe for the administration of drugs.
The OcuSafe® LDS Needle is intended for use with a luer-tip syringe for the administration of drugs. The OcuSafe® LDS Needle is indicated for intravitreal use.
The SteriCap® LDS Safety Needle is intended for use with a luer-tip syringe for the administration of drugs. The SteriCap Safety Needle has an integrated passive needle cap/shield that covers the needle prior to use and automatically activates to cover the needle immediately after use. A lock mechanism can be manually activated to lock'unlock the needle cap. The locked needle cap minimizes risk of accidental needle stick.
The VitreJect® LDS Safety Needle is intended for use with a luer-tip syringe for the administration of drugs. The VitreJect® LDS Safety Needle has an integrated passive needle cap/shield that covers the needle prior to use and automatically activates to cover the needle immediately after use. A lock mechanism can be manually activated to lock/unlock the needle cap. The locked needle cap minimizes risk of accidental needle stick. The VitreJect® LDS Safety Needle is indicated for intravitreal use.
Type of Use (Select one or both, as applicable)
| <div> <span> <svg class="bi bi-check-square-fill" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2zm10.03 4.97a.75.75 0 0 0-1.08.022L7.477 9.417 5.384 7.323a.75.75 0 0 0-1.06 1.06L6.97 11.03a.75.75 0 0 0 1.079-.02l3.992-4.99a.75.75 0 0 0-.01-1.05z" fill-rule="evenodd"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{4}------------------------------------------------
## 510(k) Summary
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements under 21 CFR 807.92.
Submitted By: OcuJect, LLC 1441 Avocado Ave, Suite 204 Newport Beach, CA 92660 USA Phone: (678) 523-8977
Contact Person:
Julie Stephens, President/Consultant Regulatory Resources Group, Inc. Phone: (678) 513-0693
September 27, 2024 Date Prepared:
Device Name and Classification:
| Trade Name: | LDS Needle<br>OcuSafe® LDS Needle<br>VitreJect® LDS Safety Needle<br>SteriCap® LDS Safety Needle |
|--------------------|--------------------------------------------------------------------------------------------------|
| Regulation Number: | 21 CFR 880.5570 |
| Regulation Name: | Hypodermic Single Lumen Needle |
| Class: | II |
| Product Code(s): | QNS, FMI, QYM |
## Legally Marketed Predicate Device:
Predicate Devices: SteriCap® Safety Needle / VitreJect® Safety Needle, 510(k) # K233343 VitreJect® Needle / OcuSafe® Needle, 510(k) # K230959 SteriCap™ Mini Needle and Standard Needles, 510(k) # K212805 Reference Predicate Device: EZ-Injec LDV Sterile Safety Needle, 510(k) # K210444
## Device Description:
The LDS Needle / OcuSafe® LDS Needle / VitreJect® LDS Safety Needle / SteriCap® LDS Safety Needle are designed to provide a means of fluid injection and aspiration. The devices are single-lumen needles intended for use with a luer-tip syringe. The Low Dead Space (LDS) needle has a decreased internal volume within the luer taper to reduce fluid volume loss. The LDS Needle and OcuSafe® LDS Needle have a removable cap that is removed prior to the needle's use. The SteriCap® LDS Safety Needle and VitreJect® LDS Safety Needle have a spring-actuated, non-removable sliding cap that protects the needle prior and during its use. They are intended for use by health care professionals for administration of drugs. Their operation is manual. The LDS Needle / OcuSafe® LDS Needle / VitreJect® LDS Needle / SteriCap® LDS Needle are single use only, non-toxic, non-pyrogenic, and sterilized by ethylene oxide gas. The VitreJect® LDS Safety Needle and OcuSafe® LDS Needle are suitable for intravitreal use.
{5}------------------------------------------------
| Product Name | Description | Configurations |
|------------------------------|------------------------------------------|------------------------|
| LDS Needle | Low Dead Space | 31G<br>4mm |
| | | 33G<br>8mm |
| | | 29G; 30G; 31G<br>13mm |
| OcuSafe® LDS Needle | Low Dead Space<br>Intravitreal | 31G<br>4mm |
| | | 33G<br>8mm |
| | | 29G; 30G; 31G<br>13mm |
| VitreJect® LDS Safety Needle | Safety<br>Low Dead Space<br>Intravitreal | 31G; 33G<br>4mm |
| | | 30G; 31G; 33G<br>5.5mm |
| | | 30G; 31G; 33G<br>6mm |
| SteriCap® LDS Safety Needle | Safety<br>Low Dead Space | 31G; 33G<br>4mm |
| | | 30G; 31G; 33G<br>5.5mm |
| | | 30G; 31G; 33G<br>6mm |
# 510(k) Summary
## Indications for Use:
The LDS Needle is intended for use with a luer-tip syringe for the administration of drugs.
- The OcuSafe® LDS Needle is intended for use with a luer-tip syringe for the administration of drugs. The OcuSafe® LDS Needle is indicated for intravitreal use.
- The SteriCap® LDS Safety Needle is intended for use with a luer-tip syringe for the administration of drugs. The SteriCap Safety Needle has an integrated passive needle cap/shield that covers the needle prior to use and automatically activates to cover the needle immediately after use. A lock mechanism can be manually activated to lock/unlock the needle cap. The locked needle cap minimizes risk of accidental needle stick.
- The VitreJect® LDS Safety Needle is intended for use with a luer-tip syringe for the administration of drugs. The VitreJect® LDS Safety Needle has an integrated passive needle cap/shield that covers the needle prior to use and automatically activates to cover the needle immediately after use. A lock mechanism can be manually activated to lock/unlock the needle cap. The locked needle cap minimizes risk of accidental needle stick. The VitreJect® LDS Safety Needle is indicated for intravitreal use.
Comparison of Indications of Use to the Predicate Device
The Indications for Use statement for the LDS Needle / OcuSafe® LDS Needle / VitreJect® LDS Safety Needle / SteriCap® LDS Safety Needle [Proposed Devices] did not change from devices under the respective previously cleared 510(k)s [Predicate Devices]. The Low Dead Space (LDS) will be added to the trade name. labels and labeling of these needles going forward. The Low Dead Space (LDS) usage is stated in FDA's Product Classification Database within the 21 CFR 880.5570 regulation with Product Code QNS. The OcuSafe® LDS Needle and the VitreJect® LDS Safety Needle maintain "indicated for intravitreal use" in their Indications for Use. The QYM Product Code for Ophthalmic Needle within the 21 CFR 880.5570 regulation is added as an additional Subsequent Product Code.
{6}------------------------------------------------
#### OcuJect LLC Page 3 of 6 LDS Needle / OcuSafe® LDS Needle / VitreJect® LDS Safety Needle / SteriCap® LDS Safety Needle
# 510(k) Summary
| Proposed Device<br>510(k) #: K242956 | Predicate Devices<br>510(k) #: K233343;<br>K230959; K212805 |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| LDS Needle<br>OcuSafe® LDS Needle<br>SteriCap® LDS Safety Needle<br>VitreJect® LDS Safety Needle | Needle<br>OcuSafe® Needle<br>SteriCap® Safety Needle<br>VitreJect® Safety Needle |
| The LDS Needle is intended for use with a luer-tip syringe for the<br>administration of drugs. | Same<br>No Changes |
| The OcuSafe® LDS Needle is intended for use with a luer-tip syringe for the<br>administration of drugs. The OcuSafe® LDS Needle is indicated for<br>intravitreal use. | |
| The SteriCap® LDS Safety Needle is intended for use with a luer-tip syringe<br>for the administration of drugs. The SteriCap Safety Needle has an<br>integrated passive needle cap/shield that covers the needle prior to use<br>and automatically activates to cover the needle immediately after use. A<br>lock mechanism can be manually activated to lock/unlock the needle<br>cap. The locked needle cap minimizes risk of accidental needle stick. | |
| The VitreJect® LDS Safety Needle is intended for use with a luer-tip syringe<br>for the administration of drugs. The VitreJect® LDS Safety Needle has<br>an integrated passive needle cap/shield that covers the needle prior to<br>use and automatically activates to cover the needle immediately after<br>use. A lock mechanism can be manually activated to lock/unlock the<br>needle cap. The locked needle cap minimizes risk of accidental needle<br>stick. The VitreJect® LDS Safety Needle is indicated for intravitreal use. | |
## Technological Characteristics:
The technological characteristics of the LDS Needle / OcuSafe® LDS Needle / VitreJect® LDS Safety Needle / SteriCap® LDS Safety Needle [Proposed Devices] did not change from devices under the respective previously cleared 510(k)s [Predicate Devices]. The proposed devices and the predicate devices have the identical design, materials, manufacturing process, sterilization method, operation method, and packaging configuration.
A comparison of the technological characteristics between the Proposed Devices and Predicate Devices are illustrated in the following table:
{7}------------------------------------------------
#### OcuJect LLC Page 4 of 6 LDS Needle / OcuSafe® LDS Needle / VitreJect® LDS Safety Needle / SteriCap® LDS Safety Needle
# 510(k) Summary
Comparison of Technological Characteristics to the Predicate Device
| Characteristic | Proposed Device | Predicate Device | Associated Testing Standard | |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Applicant | LDS Needle<br>OcuSafe® LDS Needles<br>SteriCap® LDS Safety Needles<br>VitreJect® LDS Safety Needles | Needle<br>OcuSafe® Needles<br>SteriCap® Safety Needles<br>VitreJect® Safety Needles | | |
| Trade Name(s) | OcuJect, LLC | Identical | - | |
| 510(k) Numbers | Needle, OcuSafe®; SteriCap®; VitreJect®<br>K242956 | Identical<br>K233343; K230959; K212805 | - | |
| Product Code;<br>Regulation; Class | FMI, QNS, QYM; 21 CFR 880.5570;<br>Class II<br>Additional testing per standards to support QYM were performed within 510(k) K230959 and K233343 | FMI; 21 CFR 880.5570;<br>Class II | ISO 10993-10<br>ISO 10993-7<br>USP <789><br>USP <85><br>ISO 7864<br>ISO 9626<br>ISO 80369-7 | |
| Intended Users and<br>Environment | Health care professionals in a clinical setting | Identical | | |
| Principle of<br>Operation | Connects to a Luer type syringe to serve as a conduit for manual movement of fluid | Identical | ISO 80369-7,<br>ISO 7864,<br>ISO 9626 | |
| Device Components | Needle Assembly: needle, needle hub, spring-loaded needle cap (VitreJect and SteriCap)<br>Designed to fit standard 6% Luer fittings | Identical | ISO 80369-7,<br>ISO 7864,<br>ISO 9626 | |
| Material | Needle | Stainless Steel | Identical | ISO 10993-1, |
| | Needle hub | Polypropylene | | ISO 10993-5, |
| | Lubricant | Silicone | | ISO 10993-7, |
| | Spring | Stainless Steel (VitreJect and SteriCap) | | ISO 10993-10, |
| | Adhesive | Polyacrylate | | ISO 10993-11, |
| | Needle cap | Polypropylene | | ISO 10993-12,<br>ISO 7864,<br>ISO 9626,<br>ISO 6009,<br>ISO 80369-7<br>Internal Standards,<br>ISO 23908 |
| Needle Length<br>(Exposed) | VitreJect® LDS: 4 mm / 5.5 mm / 6 mm<br>SteriCap® LDS: 4 mm / 5.5 mm / 6 mm<br>OcuSafe® LDS: 13 mm / 8 mm / 4 mm<br>LDS Needle: 13 mm / 8 mm / 4 mm | Identical | ISO 7864,<br>ISO 9626 | |
| Needle Gauge | VitreJect® LDS: 30G, 31G and 33G.<br>SteriCap® LDS: 30G, 31G and 33G<br>OcuSafe® LDS: 29G, 30G, 31G and 33G<br>LDS Needle: 29G, 30G, 31G and 33G | Identical | ISO 7864,<br>ISO 9626 | |
| Connection | Luer-slip<br>Luer-lock | Identical | ISO 7864,<br>ISO 80369-7 | |
| Needle Color<br>Coding | Conforms to ISO 6009 | Identical | ISO 6009 | |
| Needle tip<br>configuration | Lancet bevel | Identical | ISO 7864,<br>ISO 9626 | |
| Sterilization and<br>Shelf Life | Provided Sterile, Single-Use<br>100 individual blister packed devices<br>packaged into shelf carton<br>Sterilization Method: EO<br>SAL: 10-6<br>Shelf Life: 5 years | Identical | AAMI TIR28,<br>ISO 11135,<br>ISO 10993-7,<br>USP <85>,<br>USP <161>,<br>ANSI/AAMI ST72,<br>ASTM F1980-16, | |
| Characteristic | Proposed Device<br>LDS Needle<br>OcuSafe® LDS Needles<br>SteriCap® LDS Safety Needles<br>VitreJect® LDS Safety Needles | Predicate Device<br>Needle<br>OcuSafe® Needles<br>SteriCap® Safety Needles<br>VitreJect® Safety Needles | Associated<br>Testing Standard | |
| Biocompatibility per<br>ISO 10993-1 | Non-cytotoxic | Identical | ISO 10993-5 | |
| | Non-sensitizer | Identical | ISO 10993-10 | |
| | Non-irritant, Intracutaneous Reactivity | Identical | ISO 10993-10 | |
| | Non-irritant, Ocular | Identical | ISO 10993-10 | |
| | Non-irritant, Intravitreal Injection | Identical | ISO 10993-10 | |
| | Non-irritant, Intraocular | Identical | ISO 10993-10 | |
| | Non-pyrogenic | Identical | ISO 10993-11,<br>USP <151> | |
| | Non-toxic | Identical | ISO 10993-11 | |
| | Non-hemolytic | Identical | ISO 10993-4<br>ASTM F756-17 | |
{8}------------------------------------------------
#### OcuJect LLC Page 5 of 6 LDS Needle / OcuSafe® LDS Needle / VitreJect® LDS Safety Needle / SteriCap® LDS Safety Needle
# 510(k) Summary
The minor difference for the LDS Needle / OcuSafe® LDS Needle / VitreJect® LDS Safety Needle / SteriCap® LDS Safety Needle is that a low dead space specification was established that is less than or equal to 5µL (≤ 0.005 mL)]. The "Low Dead Space (LDS)" specification was confirmed by testing using 58 samples for the 95% confidence and 95% reliability. The "Low Dead Space (LDS)" designation will be added to the trade name, labels and labeling of the LDS Needle / OcuSafe® LDS Needle / VitreJect® LDS Safety Needle / SteriCap® LDS Safety Needle [Proposed Devices].
## Summary of Testing:
The biocompatibility risk assessment was completed as directed by FDA guidance under ISO 10993-1 biocompatibility categorized as an externally communicating device with limited (<24 hours) direct tissue and indirect blood path contact with the testing completed to demonstrate the needle devices are biocompatible for the intended use. Biocompatibility endpoints included Cytotoxicity, Sensitization, Irritation or Intracutaneous reactivity, Acute systemic toxicity, Material mediated pyrogenicity, Indirect hemolysis, Direct hemolysis, Particulates, Occular Irritation, and Intravitreal Injection Irritation. Performance and safety testing included dimensional and physical properties verification, bond and material strength verification, needle quality, color coding, luer connector, particulate testing, sharps injury prevention features. The ethylene oxide sterilization cycle was validated based on the principles of ISO 11135. The residual levels meet the requirements of ISO 10993-7. LAL testing for each lot ensures the devices meet the acceptance criteria of 20 EU/device (for LDS Needle / SteriCap® LDS Safety Needle) or ≤ 0.2 EU/device (for the OcuSafe® LDS Needle / VitreJect® LDS Safety Needle with the intravitreal indication for use). The shelf-life was validated using industry-standard accelerated aging technique and assured functional performance of the device as well as packaging and sterile barrier integrity. The shelflife validation included anticipated stress associated with transportation and handling. Packaging integrity was validated using the bubble leak test method and seal strength. The LDS Needle / OcuSafe® LDS Needle / VitreJect® LDS Safety Needle / SteriCap® LDS Safety Needle IProposed Devices] did not change from devices under the respective previously cleared 510(k)s [Predicate Devices].
{9}------------------------------------------------
#### OcuJect LLC Page 6 of 6 LDS Needle / OcuSafe® LDS Needle / VitreJect® LDS Safety Needle / SteriCap® LDS Safety Needle
# 510(k) Summary
## Substantial Equivalence Conclusions:
The LDS Needle / OcuSafe® LDS Needle / VitreJect® LDS Safety Needle / SteriCap® LDS Safety Needle [Proposed Devices] met the established safety and performance characteristics for a hypodermic single-lumen needle. The addition of the "low dead space (LDS)" specification to the trade name, labels and labeling did not raise any new or different questions of safety, performance or effectiveness when compared to the predicate device. The LDS Needle / OcuSafe® LDS Needle / VitreJect® LDS Safety Needle / SteriCap® LDS Safety Needle are as safe and effective as the respective predicate devices previously cleared under the 510(k)'s [Predicate Devices] and will perform as intended.
The LDS Needle / OcuSafe® LDS Needle / VitreJect® LDS Safety Needle / SteriCap® LDS Safety Needle are substantially equivalent to the Needle / VitreJect® Safety Needle / SteriCap® Safety Needle previously cleared under 510(k) #: K233343; K239059; and K212805.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.