K210444 · Poonglim Pharmatech, Inc. · QNS · Feb 16, 2021 · General Hospital
Device Facts
Record ID
K210444
Device Name
EZ-Injec LDV Sterile Safety Needle
Applicant
Poonglim Pharmatech, Inc.
Product Code
QNS · General Hospital
Decision Date
Feb 16, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This product is intended for use to inject fluid into or withdraw fluids from parts of the surface of the skin.
Device Story
Sterile, single-use hypodermic needle; consists of sharpened SUS304 metal tube joined to polypropylene hub; includes protective needle cap and integrated safety guard. Used by clinicians to inject or withdraw fluids from body below skin surface. Safety guard deployed after use to prevent needlestick injuries and reuse. Design features reduced hub inner volume to minimize dead space and fluid loss. Sterilized via EO gas. Device provides physical access for fluid delivery/aspiration; safety guard mitigates sharps injury risk.
Clinical Evidence
Bench testing only. No clinical data. Testing included ISO 9626 (tubing), ISO 7864 (sterile needles), ISO 23908 (sharps protection), USP <788> (particulate matter), and biocompatibility (ISO 10993-1: cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, hemocompatibility). Sterilization validated per ISO 11135:2014; shelf-life validated per ASTM F1980-16.
Technological Characteristics
25G x 25mm needle; SUS304 cannula; polypropylene hub and protector; epoxy adhesive. EO gas sterilization. Dead space ≤0.0054ml. Integrated safety guard for sharps injury protection. Complies with ISO 9626, ISO 7864, ISO 23908, and USP <788>.
Indications for Use
Indicated for injection of fluids into or withdrawal of fluids from parts of the surface of the skin. No specific patient population, age, or gender restrictions are defined.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Poonglim Pharmatech Inc. % Peter Chung President Plus Global 300. Atwood Pittsburgh, Pennsylvania 15213
Re: K210444
Trade/Device Name: EZ-Injec LDV Sterile Safety Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: QNS Dated: February 11, 2021 Received: February 16, 2021
#### Dear Peter Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rumi Young Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210444
Device Name EZ-Injec LDV Sterile Safety Needle
Indications for Use (Describe)
This product is intended for use to inject fluid into or withdraw fluids from parts of the surface of the skin.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| Preparation Date: | February 10, 2021 |
|--------------------|------------------------------------------------------|
| Submitter Name: | POONGLIM Pharmatech Inc. |
| | 21, Jayumuyeok 1-gil, Gunsan-si, Jeollabuk-do, Korea |
| Contact Person: | Peter Chung |
| | President, Plus Global |
| Telephone Number: | 412-512-8802 |
| E-mail Address: | peterchung210@gmail.com |
| Preparation Date: | February 10, 2021 |
| Trade Name: | EZ-Injec LDV Sterile Safety Needle |
| Regulation Name: | Hypodermic Single Lumen Needle |
| Regulation Number: | 21 CFR 880.5570 |
| Product Code: | QNS |
| Device Class: | Class II |
| Predicate Device: | K192222, EZ-Injec Single Use Needle |
# Device Description
Sterile single use needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub). The needle cap covers intended to provide physical protection to the needle tube. This product is packed by sterile paper and sterilized by EO gas.
The safety guard is placed around the needle after using the product for safety and to prevent reuse. EZ-Injec LDV Sterile Safety Needle helps to decrease fluid volume loss because of its smaller hub inner size.
| | Gauge | Length | Wall Type |
|------------------------------------|-------|--------|-----------|
| EZ-Injec LDV Sterile Safety Needle | 25G | 25 mm | TW |
## Indications for Use
EZ-Injec LDV Sterile Safety Needle
This product is intended for use to inject fluids into or withdraw fluids from parts of the surface of the skin.
### Discussions of differences in Indications for Use statement
The indications for use statement for the subject device (EZ-Injec LDV Sterile Safety Needle) is identical to the predicate device.
#### Technological Characteristics
The table below includes a comparison of the technological characteristics between the new device and those of the predicate device:
{4}------------------------------------------------
| Technological<br>Characteristic | Subject Device<br>EZ-Injec LDV Sterile Safety Needle<br>K210444 | Predicate Device<br>EZ-Injec Single Use Needle<br>K192222 | Comments |
|---------------------------------|-----------------------------------------------------------------|------------------------------------------------------------|----------------|
| Gauge | 25G | 18, 19, 21, 22, 23, 25, 26, 27, 29, 30,<br>31, 32, 33, 34G | Same |
| Length | 25mm | 4, 6, 8, 13, 16, 25, 30, 40mm | Same |
| Dead space<br>specification | ≤0.0054ml | ≤0.0544ml (25G-25mm) | See Comment #1 |
| Sharps prevention<br>function | Safety Guard | No Safety Guard | See Comment #2 |
| Hub of needle | Polypropylene (PP) | Polypropylene (PP) | Same |
| Protector | Polypropylene (PP) | Polypropylene (PP) | Same |
| Cannula | SUS304 | SUS304 | Same |
| Adhesive | Epoxy | Epoxy | Same |
| Sterilization<br>method | EO gas | EO gas | Same |
### Discussions of differences in technological characteristics
- Comment #1
EZ-Inject LDV sterile safety needle is a modification to the EZ-Inject single use needle (premarket submission no.:K192222). The modification involves a design change to the inner hub to reduce dead space of the needle. Comparative bench testing and analysis of drug delivery capability was provided to verify and validate this change.
### Comment #2
EZ-Injec LDV sterile safety needle is a modification to the EZ-Injec single use needle (premarket submission no.: K192222). The modifications involve an addition of the safety guard to protect the user from sharps injuries. This change was validated by bench testing and performed in accordance with ISO 23908.
# Performance Testing
The sterile, single lumen hypodermic needles described in this summary were tested and demonstrated to be in conformance with the following FDA recognized standards:
- ISO 9626 Second edition: Stainless steel needle tubing for the manufacture of medical devices -● Requirements and test methods
- ISO 7864 Fourth edition: Sterile hypodermic needles for single use Requirements and test methods ●
- ISO 23908 First edition: Sharps injury protection Requirements and test methods Sharps protection ● features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
Additionally, bench testing demonstrating the low dead space capability of the needle was conducted.
### Biocompatibility
{5}------------------------------------------------
In accordance with ISO 10993-1, the needle is classified as: Externally Communicating Device, Blood Path Indirect, Limited Contact (<24 hours). The following testing was conducted:
- Cytotoxicity
- Sensitization ●
- Irritation ●
- Acute Systemic Toxicity
- Material-Mediated Pyrogenicity ●
- Hemocompatibility ●
Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injections and met the USP acceptance criteria.
## Sterility, Shipping and Shelf-Life
- . Sterilization validation was performed in accordance with ISO11135:2014 to prove that the EO Gas sterilization process has been suitable for the continuous production. Through validation, sterilization process was deemed acceptable.
- Shelf life of 5 years is validated using the FDA recognized standard ASTM F1980-16 Standard Guide ● for Accelerated Aging of Sterile Barrier Systems for Medical Devices
# Conclusions
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The EZ-Injec LDV Sterile Safety Needle (K210444) is substantially equivalent to the EZ-Injec Single Use Needle (K192222) with respect to the indications for use, target populations, treatment method, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.