Respiration Data Management Software (PAP Link PC)
K242935 · BMC Medical Co., Ltd. · BZD · Jan 15, 2025 · Anesthesiology
Device Facts
Record ID
K242935
Device Name
Respiration Data Management Software (PAP Link PC)
Applicant
BMC Medical Co., Ltd.
Product Code
BZD · Anesthesiology
Decision Date
Jan 15, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Respiration Data Management Software is installed in a computer to perform patient management by treatment data viewing, treatment data reporting and remotely adjusting the prescribed compatible BMC CPAP and BPAP devices' settings. It is intended for healthcare professionals in healthcare facilities use only.
Device Story
Respiration Data Management Software is a PC-based application for healthcare professionals to manage patients treated with compatible BMC CPAP and BPAP devices. The software receives therapy data via serial connection, removable media, or wireless (TCP/HTTPS) transmission. It processes and displays clinical parameters including mask leak, AHI, pressure, minute ventilation, and respiratory rate. Clinicians use the software to view and report treatment data, store patient records, and remotely adjust device settings (e.g., initial pressure, set pressure, ramp time, humidifier level). Data can be uploaded to a cloud platform. The software assists in patient follow-up and management by providing centralized access to therapy information and remote parameter modification. It is intended for use in professional healthcare facilities.
Clinical Evidence
No clinical data. Substantial equivalence is supported by software verification and validation testing conducted in accordance with FDA guidance, and cybersecurity evaluation.
Technological Characteristics
PC-based software running on Microsoft Windows. Connectivity via serial, removable media, and wireless (TCP/HTTPS). Supports data storage, report generation, and remote parameter adjustment for compatible BMC CPAP/BPAP devices. Software verification and validation performed per FDA guidance.
Indications for Use
Indicated for healthcare professionals in healthcare facilities to manage patients using compatible BMC CPAP and BPAP devices. Not for patient use.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 15, 2025
BMC Medical Co., Ltd. Amy Wang Regulatory Affairs Specialist Room 10, 17F, Building 4, Huiya Plaza, No.16 Lize Road Fengtai District Beijing, 100073 China
Re: K242935
Trade/Device Name: Respiration Data Management Software Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: December 13, 2024 Received: December 13, 2024
Dear Amy Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K242935 Device Name
Respiration Data Management Software
Indications for Use (Describe)
Respiration Data Management Software is installed in a computer to perform patient management by treatment data viewing, treatment data reporting and remotely adjusting the prescribed compatible BMC CPAP and BPAP devices' settings. It is intended for healthcare professionals in healthcare facilities use only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| Device Trade Name | Respiration Data Management Software |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | PAP Link PC |
| Common/Usual Name | Respiration Data Management Software |
| Date Prepared | September 25, 2024 |
| Sponsor Identification | BMC Medical Co., Ltd.<br>Room 10, 17F, Building 4, Huiya Plaza, No.16<br>Lize Road, Fengtai District, 100073 Beijing,<br>PEOPLE'S REPUBLIC OF CHINA |
| Submission Correspondent | Amy Wang<br>BMC Medical Co., Ltd.<br>Phone 86-18701556244<br>Fax 86-22-82939881<br>Email wangliping@bmc-medical.com |
| Establishment Registration # | 3008566132<br>BMC Medical Co., Ltd.<br>Room 10, 17F, Building 4, Huiya Plaza, No.16 Lize<br>Road, Fengtai District, 100073 Beijing, PEOPLE'S<br>REPUBLIC OF CHINA |
| Classification | Class II Device (21 CFR 868.5905) |
| Classification Name | Noncontinuous ventilator |
| Classification Panel | Medical Device |
| Products Code | BZD, MNS |
| Medical Specialties | Anesthesiology |
| Predicate Device(s) | ResScan (K140054) |
| Reason for Submission: | New Device |
{5}------------------------------------------------
Indications for Use Respiration Data Management Software is installed in a computer to perform patient management by treatment data viewing, treatment data reporting and remotely adjusting the prescribed compatible BMC CPAP and BPAP devices' settings. It is intended for healthcare professionals in healthcare facilities use only. Device Description The Respiration Data Management Software is a softwareonly device, which allows physicians and other clinical staff
to transmit, analyze and review respiration data from the compatible CPAP and BPAP devices (non-life support systems), thus to assist the users in patient management and follow-up work. This also includes remote modification of the treatment parameters of the compatible devices. It can also store patient data, upload data to the cloud platform, generate and print reports.
#### Substantial Equivalence:
The subject and predicate devices are substantially equivalent in the following aspects:
- Same Indications for Use
- Similar operating principle
- Similar technological characteristics
Comparison of Technological Characteristics with the Predicate Device:
{6}------------------------------------------------
| | Subject Device | Predicate Device | |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Respiration Data<br>Management<br>Software (K242935) | ResScan<br>(K140054) | Comparison |
| Classification | | | |
| Device<br>Classification | Class II Device | Class II Device | Equivalent |
| Classification<br>Panel | Anesthesiology | Anesthesiology | Equivalent |
| Regulation<br>Number&<br>Product<br>Code | 21 CFR 868.5905, BZD<br>Non continuous ventilator<br>21 CFR 868.5895, MNS<br>Continuous Ventilator | 21 CFR 868.5905, BZD<br>Non continuous ventilator<br>21 CFR 868.5895, MNS<br>Continuous Ventilator<br>21 CFR 868.5895, MNT<br>Continuous Ventilator<br>21 CFR 868.5895, CBK<br>Ventilator, continuous,<br>facility use | The regulation number &<br>product code depends on the<br>compatible devices. Both the<br>subject device and predicate<br>device are compatible with<br>BZD and MNS devices. The<br>predicate device includes more<br>compatible product codes. This<br>difference does not raise new<br>risks to the subject device. |
| Regulation<br>Name | Noncontinuous ventilator | Noncontinuous ventilator | Equivalent |
| Intended Use and Indications for Use | | | |
| Intended Use of<br>the Device | Respiration<br>Data<br>Management Software is<br>installed in a computer<br>to perform patient<br>management by<br>treatment data viewing,<br>treatment data reporting<br>and remotely adjusting<br>the prescribed<br>compatible BMC CPAP<br>and BPAP devices'<br>settings. It is intended<br>for healthcare<br>professionals in<br>healthcare facilities use<br>only. | ResScan is intended to<br>augment the standard<br>follow-up care of patients<br>by providing transfer of<br>machine and therapeutic<br>information. This includes<br>the ability to remotely<br>change settings in non-life<br>support devices only.<br>It is intended to be used<br>by Clinicians in<br>conjunction with ResMed<br>compatible therapy<br>devices, using ResMed's<br>proprietary<br>communications protocol. | Equivalent |
| Environment of<br>Use | Professional healthcare<br>facilities | Professional healthcare<br>facilities | Equivalent |
| Patient<br>Population | Not for patient use. For<br>healthcare professionals<br>use only. | Not for patient use. For<br>healthcare professionals<br>use only. | Equivalent |
| Performance | | | |
| Application<br>Type | PC-based software<br>program | PC-based software<br>program | Equivalent |
| Operating<br>System | Microsoft Windows | Microsoft Windows | Equivalent |
| Functionality | Display of therapy data<br>Generate reports<br>Settings management<br>Patient management<br>Upload Data | Display of therapy data<br>Generate reports<br>Settings management<br>Patient management | Most functions of the<br>subject device are<br>equivalent to the predicate<br>device. The subject device<br>includes a data upload<br>function than the predicate<br>device. Users can upload<br>the data to the cloud<br>platform (K160127).<br>Software V&V testing and<br>cybersecurity testing<br>demonstrates safety and<br>effectiveness of this<br>function. |
| Compatible flow<br>generators | BMC compatible<br>therapy devices (BZD,<br>MNS) | ResMed compatible<br>therapy devices (BZD,<br>MNS, MNT and CBK) | Both the subject device<br>and predicate device are<br>compatible with the<br>devices with BZD and<br>MNS product codes. The<br>predicate device includes<br>more compatible product<br>codes. This difference<br>does not raise new risks<br>to the subject device. |
| Communication<br>medium | Serial connection<br>Removable medium<br>Wireless medium<br>(TCP/HTTPS) | Serial connection<br>Removable medium | The subject device can<br>wirelessly upload data to<br>the cloud platform<br>(K160127), as well as<br>download the software<br>upgrade package from the<br>cloud platform. Software<br>V&V testing and<br>cybersecurity testing<br>demonstrates safety and<br>effectiveness of these<br>functions. |
| Patient<br>information | Mask Leak<br>AHI<br>Pressure<br>Minute Ventilation<br>Respiratory rate | Mask Leak<br>AHI<br>Pressure<br>Minute Ventilation<br>Respiratory rate | Equivalent |
| Range for<br>patient<br>parameters | Depends on the<br>compatible CPAP devices | Depends on the compatible<br>CPAP devices | Equivalent |
| Changeable<br>settings | Initial Pressure<br>Set Pressure<br>Ramp time<br>Humidifier level | Start Pressure<br>Set Pressure<br>Ramp time<br>Setting time | Equivalent. Remote<br>setting change<br>functionality only applies<br>to non-life support<br>devices. Changeable<br>settings of the subject<br>device and the predicate<br>device are similar.<br>Difference does not raise<br>new risks. |
| Changeable<br>setting range | Depends on the<br>compatible CPAP devices | Depends on the compatible<br>CPAP devices | Equivalent |
{7}------------------------------------------------
### 510(K) Summary
{8}------------------------------------------------
### 510(K) Summary
{9}------------------------------------------------
## Non-Clinical Data
The following non-clinical data were used in support of the substantial equivalence determination.
#### Software Verification and Validation
Software verification and validation were conducted and documentation was provided in accordance with FDA's Guidance "Content of Premarket Submissions for Device Software Functions".
#### Cybersecurity
Cybersecurity was considered and evaluated in accordance with FDA guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions".
# Conclusion
The subject device Respiration Data Management Software has the same intended use and similar technological characteristics as the predicate device. Differences in the technological characteristics between the subject device and predicate device do not raise new/different questions of safety or effectiveness. Thus, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.