K160127 · 3B Medical, Inc. · BZD · Nov 10, 2016 · Anesthesiology
Device Facts
Record ID
K160127
Device Name
iCodeConnect
Applicant
3B Medical, Inc.
Product Code
BZD · Anesthesiology
Decision Date
Nov 10, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Software as a Medical Device, Therapeutic
Indications for Use
iCodeConnect, in conjunction with a Luna CPAP or Auto-CPAP device, is intended for use in treating obstructive sleep apnea (OSA) on adult patients (>30 kg). iCodeConnect is intended to augment the standard follow-up care of patients diagnosed with obstructive sleep apnea by displaying usage and therapeutic information that has been transmitted from the patient's Luna CPAP or Auto-CPAP therapy device located in the home to the clinician or healthcare professional. iCodeConnect also provides remote settings capabilities.
Device Story
iCodeConnect is a web-based patient management system for remote monitoring and configuration of Luna CPAP/Auto-CPAP devices. Input data includes usage and therapeutic information collected by the CPAP device in the home. Data is transmitted to the iCodeConnect platform via manual entry, SD card, or a USB-connected Wi-Fi module. The system stores and displays this data for clinicians to review patient compliance and efficacy. Clinicians use the web interface to remotely update therapy settings (pressure, mode, comfort) on the patient's device. This workflow replaces manual SD card mailing, enabling timely data collection and remote clinical intervention. Benefits include improved follow-up care efficiency and remote management of OSA therapy.
Clinical Evidence
No clinical data. Safety and effectiveness were demonstrated through non-clinical bench testing, including functional verification, software validation, and compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 14971, and FCC 47 CFR Part 15.
Technological Characteristics
Web-based software platform and USB-connected Wi-Fi module. Connectivity via Wi-Fi. Power supplied by host CPAP device. Type BF applied part. IP22 ingress protection. Non-sterile, reusable. Software performs data storage, display, and remote parameter configuration.
Indications for Use
Indicated for adult patients (>30 kg) diagnosed with obstructive sleep apnea (OSA) to augment standard follow-up care via remote monitoring and settings adjustment of Luna CPAP or Auto-CPAP devices.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is simple and clean, and it is easily recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 10, 2016
3B Medical, Inc. Alex Lucio Vice President 799 Overlook Drive Winter Haven, Florida 33884
Re: K160127
Trade/Device Name: iCodeConnect Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: October 10, 2016 Received: October 12, 2016
Dear Alex Lucio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K160127
Device Name iCodeConnect
#### Indications for Use (Describe)
iCodeConnect, in conjunction with a Luna CPAP or Auto-CPAP device, is intended for use in treating obstructive sleep apnea (OSA) on adult patients (>30 kg).
iCodeConnect is intended to augment the standard follow-up care of patients diagnosed with obstructive sleep apnea by displaying usage and therapeutic information that has been transmitted from the patient's Luna CPAP or Auto-CPAP therapy device located in the home to the clinician or healthcare professional. iCodeConnect also provides remote settings capabilities.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| <span> <span style="text-decoration: underline;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PRAStaff@fda.hhs.gov
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
# 510(k) Summary
| Device Trade Name | iCodeConnect |
|--------------------------|--------------------------------------------------------------------------------|
| Common/Usual Name | Noncontinuous ventilator (IPPB) |
| Date Prepared | November 10, 2016 |
| Sponsor Identification | 3B Medical, Inc.<br>799 Overlook Drive<br>Winter Haven, FL 33884 |
| Phone | 863-226-6285 |
| Fax | 863-226-6284 |
| Email | alucio@3bproducts.com |
| Submission Correspondent | Alex Lucio<br>3B Medical, Inc.<br>799 Overlook Drive<br>Winter Haven, FL 33884 |
| Phone | 863-226-6285 |
| Fax | 863-226-6284 |
| Email | alucio@3bproducts.com |
{4}------------------------------------------------
| Establishment Registration | # 3008566132 |
|----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| BMC Medical CO., LTD<br>5/F Main Building No.19Gucheng Street<br>West, Shijingshan,Beijing, CHINA 100043 | |
| Classification | Class II Device |
| Classification Name | Non-continuous ventilator |
| Classification Panel | Medical Device |
| Classification Reference | 21 CFR 878.5905 |
| Products Code | BZD |
| Medical Specialties | Anesthesiology |
| Predicate Device(s) | EasyCare Online (K093684) |
| Reason for Submission: | New Device |
| Intended Use | iCodeConnect, in conjunction with a Luna CPAP<br>or Auto-CPAP device, is intended for use in<br>treating obstructive sleep apnea (OSA) on adult<br>patients (>30 kg).<br>iCodeConnect is intended to augment the standard<br>follow-up care of patients diagnosed with<br>obstructive sleep apnea by displaying usage and<br>therapeutic information that has been transmitted<br>from the patient's Luna CPAP or Auto-CPAP<br>therapy device located in the home to the clinician<br>or healthcare professional. iCodeConnect also<br>provides remote settings capabilities. |
| Device Description | iCodeConnect is a web based patient management<br>system used to store and manage sleep compliance<br>and efficacy data uploaded from a Luna CPAP or<br>Auto-CPAP device.<br>iCodeConnect allows device patient data to be |
{5}------------------------------------------------
uploaded using a variety of methods, including manual entry, SD Card, and wireless transmission. Use of wireless transmission (i.e. WiFi) can provide wireless compliance monitoring and remote settings update functions.
For wireless transmission, iCodeConnect consists of the wireless data transmission module Wi-Fi kit and web based iCodeConnect patient management software. TheWi-Fi Kit is intended to be used in conjunction with the Luna CPAP System (E-20C-H-O), Auto-CPAP System (E-20A-H-O / E-20AJ-H-O), which were cleared in K141770.
1) iCodeConnect Software is an analysis and management platform for the patient's therapy data. This platform, adopting the form of a website, not only realizes the uniform management of the patient's data and the information of the provider, the doctor and the patient, but also allows for the Wi-Fi Kit or other wireless communication module to transmit data and remotely updating therapy device settings, which reduces the inconvenience brought by the traditional mailing of SD card and makes data collection more timely.
Wi-Fi Kit is an accessory module that attaches via USB cable to the data output port of a Luna® CPAP or Auto-CPAP devices. The wireless module Wi-Fi Kit uploads the device's therapy data to the iCodeConnect software with the help of Wi-Fi off-the-shelf technology. After connecting to the internet. the wireless kit shall automatically collect the user information, statistical information and detailed information in the therapy device for uploading.
Non-Clinical Testing Non-clinical testing of the iCodeConnect has been
3
{6}------------------------------------------------
carried out to cover functional verification and device performance. The proposed iCodeConnect has been tested to appropriate standards and other applicable requirements. The device was designed and tested according to:
- IEC 60601-1-2:2014, medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance -Collateral compatibility - Requirements and tests
- IEC 60601-1 Medical Electrical Equipment-Part 1: General requirements for basic safety and essential performance.
- IEC | 60601-1-1 | Medical | | Electrical Equipment- Part 1-11: General requirements for basic safety and essential performance-Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- ISO14971:2007 | Medical | | devices Application of risk management to medical devices
- FCC Regulation 47 CFR Part 15 Radio Frequency Devices
Additional design considerations were applied based on published guidance and draft guidance documents from FDA:
1. FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)
2. FDA Guidance - Design Considerations for Devices Intended for Home Use (Nov24, 2014)
FDA Guidance -Content of of Premarket 3. Submissions for Management of Cybersecurity in Medical Devices (Oct 2, 2014)
{7}------------------------------------------------
4. FDA Guidance - Radio Frequency Wireless Technology in Medical Devices (Aug 14, 2013)
All tests confirmed that iCodeConnect met the predetermined and demonstrated substantial equivalence to the predicate device.
Clinical Testing Clinical testing was not required to demonstrate the safety and effectiveness of the iCodeConnect. The product functionality has been adequately assessed
{8}------------------------------------------------
## Substantial Equivalence
| Features/<br>Function | Proposed Device | Predicate Device | Comments |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|
| | iCodeConnect<br>(510K Number: K160127) | EasyCare Online<br>(510K Number: K093684) | |
| Product Code | BZD | BZD | Same as predicate |
| Regulation<br>Number | 868.5905 | 868.5905 | Same as predicate |
| Regulation<br>Name | Noncontinuous<br>ventilator (IPPB) | Noncontinuous<br>ventilator (IPPB) | Same as predicate |
| Intended Use | iCodeConnect, in conjunction with a Luna<br>CPAP or Auto-CPAP device, is intended for<br>use in treating obstructive sleep apnea (OSA)<br>on adult patients (>30 kg).<br><br>iCodeConnect is intended to augment the<br>standard follow-up care of patients diagnosed<br>with obstructive sleep apnea by displaying<br>usage and therapeutic information that has been<br>transmitted from the patient's Luna CPAP or<br>Auto-CPAP therapy device located in the home<br>to the clinician or healthcare professional.<br>iCodeConnect also provides remote settings<br>capabilities. | EasyCare Online is intended to<br>augment the standard follow-up care<br>of<br>patients diagnosed with<br>obstructive sleep apnea by<br>displaying usage and therapeutic<br>information that has been transmitted<br>from the patient's flow generator<br>located in the home to the care giver.<br>EasyCare Online also provides<br>remote settings capabilities.<br><br>It is intended to be used by<br>Clinicians in conjunction with<br>ResMed compatible flow generators. | |
| Features/<br>Function | Proposed Device | Predicate Device | |
| | iCodeConnect<br>(510K Number: K160127) | EasyCare Online<br>(510K Number: K093684) | Comments |
| Device<br>Composition | A web-based software and a wireless module | A web-based software and a wireless<br>module | Same as predicate |
| Software<br>Function | The iCodeConnect Software is designed to<br>display the data which was uploaded from the<br>CPAP / Auto-CPAP device, store the<br>information in a database and provide a secure<br>interface, allowing the user to query and view<br>patient and treatment information.<br>The iCodeConnect Software supports applying<br>new settings to the CPAP/Auto-CPAP device<br>remotely from the clinician's PC. | The Server System of Easycare<br>Online is designed to display the<br>data which was got from the flow<br>generator, store the information in a<br>database and provide a secure<br>interface, allowing the user to query<br>and view patient and treatment<br>information.<br>The Server System Software<br>supports to apply new settings to the<br>flow generator device remotely from<br>the clinician's PC. | Same as predicate |
| Remoter<br>Settings<br>Change<br>Function | Flow generator settings can be changed,<br>through wireless module can be made from the<br>clinicians' PC by iCodeConnect Software. | Flow generator settings can<br>be<br>changed, through wireless module<br>can be made from the clinicians' PC<br>by Server System. | Same as predicate |
| Data Transfer<br>Technology | Wireless | Wireless | Same as predicate |
| Connectivity | WiFi | GSM | Both<br>the<br>wireless<br>modules<br>are<br>pre-<br>approved. |
| Features/<br>Function | Proposed Device | Predicate Device | |
| | iCodeConnect<br>(510K Number: K160127) | EasyCare Online<br>(510K Number: K093684) | Comments |
| Therapy<br>Setting | Pressure<br>Mode<br>Comfort | Pressure<br>Mode<br>Comfort | Same as predicate |
| Power<br>Requirements | The wireless module is powered via electrical<br>connection form the flow generator. | The wireless module is powered via<br>electrical connection form the flow<br>generator. | Same as predicate |
| Environmental<br>Specifications | Operating temperature:<br>5°C to 35°C<br>Operating humidity :<br>15% ~ 93%, Non-condensing<br>Storage and transport temperature:<br>-25°C ~ 70°C<br>Storage and transport humidity:<br>≤ 93%, Non-condensing | Operating temperature:<br>5°C to 35°C<br>Operating humidity :<br>10 to 95% non-condensing<br>Storage and transport temperature:<br>-20°C to 60°C<br>Storage and transport humidity:<br>10 to 95% non-condensing | Both the devices meet<br>the requirement. |
| Degree of<br>Protection<br>Against<br>Electric Shock | Type BF Vertical Applied Part | Type BF Vertical Applied Part | Same as predicate |
| Degree of<br>Protection<br>Against Ingress<br>of Water | IP22 | IP21 | Both the devices meet<br>the requirement. |
| Sterilization/<br>Reuse | Not provided sterile<br>Reusable with cleaning instructions | Not provided sterile<br>Reusable with cleaning instructions | Same as predicate |
{9}------------------------------------------------
510K) Summary
{10}------------------------------------------------
510K) Summary
{11}------------------------------------------------
510K) Summary
iCodeConnect
## Conclusions
Hardware testing carried out for iCodeConnect indicates it meets design and performance functional requirements. Software verification demonstrates that device features and the system configuration functions are substantially equivalent to the predicate device. This information indicates that the iCodeConnect is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.