The Enteral Drainage System is intended to allow the evacuation of excess gas and drainage of gastrointestinal fluids from patients utilizing an enteral device. The Enteral Medicine Straw is intended for drawing and delivery of medicine to patients through an ENFit syringe.
Device Story
Enteral Drainage System and Enteral Medicine Straw facilitate enteral nutrition management. Drainage System: bag (DEHP-free PVC) with 3μm hydrophobic filter, tubing, and clamp; connects to nasogastric/naso-jejunal tubes to evacuate excess gas/fluids via gravity; used by trained staff in clinical settings. Medicine Straw: thermoplastic polyurethane tubing with male enteral connector; attaches to ENFit syringe to draw/deliver medication. Both devices utilize ENFit connectors to prevent intravenous misconnection. Devices are single-patient use, replaceable every 24 hours. Output allows clinicians to monitor drained fluids and safely administer enteral medications. Benefits include gastric decompression and standardized, secure medication delivery.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1, USP <151>), misconnection, leak, and tensile strength testing. Performance verified against ISO 80369-3:2016, ISO 80369-20:2015, and ISO 20695:2020. Packaging integrity validated per ISO 11607-1:2019 and relevant ASTM standards (F1980, D4332, D4169, F1929).
Technological Characteristics
Materials: DEHP-free PVC, thermoplastic polyurethane, ABS. Design: Vented drainage bag (50-1000ml) with 3μm hydrophobic filter; straw with male enteral connector. Connectivity: ISO 80369-3 compliant ENFit connectors. Energy: Gravity-fed (no active power). Sterilization: Ethylene Oxide (EO).
Indications for Use
Indicated for neonatal, pediatric, and adult patients with compromised or limited gastric motility requiring stomach decompression or medication delivery via enteral access.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Farrell Valve System with Equilibrium Technology and Enfit Enteral (K142012)
U Deliver Bolink ENFit Enteral Feeding Sets (K190310)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 18, 2025
HMC Premedical S.p.a.
Danilo Bosetti
QA RA Manager
Via Bosco, 1/3
Mirandola, MO 41037
Italy
Re: K242917
Trade/Device Name: Enteral Drainage System, Enteral Medicine straw
Regulation Number: 21 CFR 876.5980
Regulation Name: Gastrointestinal Tube and Accessories
Regulatory Class: Class II
Product Code: PIF
Dated: May 13, 2025
Received: May 16, 2025
Dear Danilo Bosetti:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242917 - Danilo Bosetti
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K242917 - Danilo Bosetti
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anthony Lee -S
Anthony C. Lee, Ph.D, M.B.A.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K242917
Device Name
Enteral Drainage System
Enteral Medicine straw
Indications for Use (Describe)
The Enteral Drainage System is intended to allow the evacuation of excess gas and drainage of gastrointestinal fluids from patients utilizing an enteral device.
The Enteral Medicine Straw is intended for drawing and delivery of medicine to patients through an ENFit syringe.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
| 510(k) #: K242917 | 510(k) Summary | Prepared on: 2025-06-17 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | HMC Premedical S.p.a. | |
| Applicant Address | Via Bosco, 1/3 Mirandola MO 41037 Italy | |
| Applicant Contact Telephone | +39 0535 22704 | |
| Applicant Contact | Mr. Danilo Bosetti | |
| Applicant Contact Email | FDA@hmcgroup.it | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Enteral Drainage System, Enteral Medicine straw | |
| Common Name | Gastrointestinal tube and accessories | |
| Classification Name | Gastrointestinal Tubes With Enteral Specific Connectors | |
| Regulation Number | 21 CFR 876.5980 | |
| Product Code(s) | PIF | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K142012 | Farrell Valve System with Equilibrium Technology and Enfit Enteral | PIF |
| K190310 | U Deliver Bolink ENFit Enteral Feeding Sets | PIF |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The Enteral Drainage System and the Enteral Medicine straw are intended to be used during enteral nutrition treatments in combination with enteral feeding tubes, syringes and other compatible enteral medical devices.
The Enteral Drainage System is intended to allow the evacuation of excess gas and drainage of gastrointestinal fluids from patients utilizing an enteral device.
The device consists of:
- a bag in DEHP FREE PVC, equipped with a hydrophobic 3μm filter to allow gas escape;
- two short PVC tubes (5-6 cm) on top and bottom of the bag; the bottom one is equipped with a male enteral connector, and the top one with a female enteral connector that is capped off with an ENFit-compatible cap;
- a clamp to regulate the fluid flow, placed over the bottom tube.
It is available in five sizes, depending on the volume of the bag included in the device:
MED50LU, Enteral drainage bag 50 ml;
MED100LU, Enteral drainage bag 100 ml;
MED250LU, Enteral drainage bag 250 ml;
MED500LU, Enteral drainage bag 500;
MED1000LU, Enteral drainage bag 1000 ml.
The device is connected to the nasogastric or naso-jejunal feeding tube between feeding sessions to allow the drainage of stomach or intestinal fluids and gasses in excess. The transparency of the materials allows for constant monitoring and visualization of drained fluids.
The bag must be placed at a lower level than the patient's stomach and also acts as a venting device. | | |
{5}
The Enteral Medicine straw, coded C005LU, is intended for drawing and delivery of medicine to patients through an ENFit enteral syringe. It consists of a male enteral connector (that allows the connection to a compatible enteral syringe) and a tubing in thermoplastic polyurethane.
Once the fluid is drawn up from the container into the syringe, the cannula must be disconnected, and the syringe can then be connected to another medical device (normally, a feeding tube or a gastrostomy tube) to administer its content to the patient.
All enteral connections of the subject devices are designed to avoid for inadvertent delivery of enteral feedings through the intravenous route. i.e., the devices cannot be connected to devices endowed with a luer connector.
The devices are sterilized with Ethylene Oxide (EO).
These devices are targeted to neonatal, pediatric and adult patients, with compromised or limited gastric motility; they are for single patient use and can be reused multiple times within a twenty-four-hour period, after which they are to be replaced.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The Enteral Drainage System is intended to allow the evacuation of excess gas and drainage of gastrointestinal fluids from patients utilizing an enteral device.
The Enteral Medicine Straw is intended for drawing and delivery of medicine to patients through an ENFit syringe.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
Enteral Drainage System:
The indications for use of the subject device are included in those of the predicate device. Both devices are intended for stomach decompression during enteral feeding procedure. The subject device has a narrower intended use in respect to the predicate device, since it is only intended for drainage: its intended use is included in the wider intended use of the predicate device.
Enteral Medicine Straw:
The subject device and the predicate are both intended to draw fluid solutions and deliver them to a patient through an enteral access device. The indications for use of the predicate device do not expressly mention the use for drawing, but this is implicitly included in the intended use of the device (the device draws the solution from a container and then it delivers it). Thus, their intended use is substantially equivalent.
The predicate device is intended to be directly connected to an ENFit feeding tube, and solution collected from the container is delivered to the patient by gravity; as for the subject devices, the solution is drawn from the container by means of an ENFit enteral syringe and then delivered to the patient's ENFit enteral access. The use of the ENFit syringe allows to measure and fractionate the doses of solution drawn from the container and to be delivered to the patient. This difference in the principle of operation does not change the intended use of the devices and does not have any significant effect on the safety or effectiveness of the subject device, also considering that the subject device is prescription use only (to be used by adequately trained staff), while the predicate device is intended for over-the-counter use.
## Technological Comparison
21 CFR 807.92(a)(6)
Enteral Drainage System:
The subject device has similar technological characteristics and design as the predicate device, as they both include a vented bag, tubing with clamps and male/female enteral connectors. They operate in a similar way, allowing the excess gas and fluids to evacuate in the vented bag.
There is a slight difference in the tube length and in the inner and outer diameter of the tube. The difference is not relevant because the fluid flow is determined by the inner diameter of the enteral connector. Performance tests were conducted on the proposed device and the test results of tube part showed that the performance of the proposed device met the pre-determined acceptance criteria. Therefore, the difference on tube size will not raise new issues of safety and effectiveness of the proposed device.
The proposed device is available in 5 different configurations according to their bag volume, while the predicate device is offered only with a 500 mL bag. The air vent filter is the same for all bag sizes.
The bag is only intended to collect discarded fluids and its different capacity has no effects on the operating modalities of the device. Both the subject and the predicate device are made of PVC and ABS. The results of biocompatibility testing performed on the proposed device show no adverse effects.
Enteral Medicine Straw:
The subject device has similar technological characteristics and design as the predicate device, all including an enteral connector and a
{6}
path through which solutions flow.
The proposed device has a simpler design than the predicate device (which includes a cross spike connector and a clamp) and must be connected to an ENFit enteral syringe - by means of which the solution is drawn from the container and subsequently delivered to the patient's enteral access - whereas the predicate device directly connects the fluid container to the feeding tube.
Moreover, the subject device carries a male enteral connector that matches with the female ENFit connector of the syringe, while the predicate device has a female ENFit connector that matches with the male ENFit connector of the feeding tubes. However, these minor differences do not affect the safety or effectiveness of the proposed devices as shown by the performance tests conducted on the proposed device.
Both the subject and predicate devices are to be connected to enteral end fittings compliant with ISO 80369-3 ('ENFit' connections). The proposed device is manufactured with raw materials which are similar to those used for the predicate device, well known and largely used for similar devices already cleared under the same product code.
The results of biocompatibility testing performed on the proposed device show no adverse effects.
# Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Nonclinical tests were conducted to demonstrate substantial equivalence to the predicate devices. The test results demonstrated that the proposed devices comply with the applicable sections of the standards cited below.
## Biocompatibility:
The materials used to manufacture the subject devices are similar to those used for the predicate devices. All the materials used to manufacture the subject device are largely used for other legally marketed devices under the same product code.
Biocompatibility has been tested according to the requirements of ISO 10993 series.
The subject devices meet requirements for ISO 10993-1 as the predicate devices.
Material-Mediated Pyrogenicity was also tested according to USP <151>.
## Performance tests:
Non-clinical tests were performed to confirm that differences in materials and technological characteristics do not affect the safety or effectiveness of the devices.
Misconnection testing, leak testing and tensile strength testing were performed.
Tests were conducted according to:
- ISO 80369-3:2016 and ISO 80369-20:2015 (at time 0, after accelerated aging and after simulated distribution) on male and female connectors;
- ISO 80369-3:2016 - Point 5 to verify the dimensional requirements for enteral small-bore connectors;
- ISO 20695:2020 for Enteral nutrition
The integrity of the device packaging was tested according to:
ISO 11607-1:2019 "Packaging for terminally sterilized medical devices – Requirements for materials, sterile barrier systems and packaging systems";
ASTM F1980-16 "Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices";
ASTM D4332-14 "Standard Practice for Conditioning Containers, Packages, or Packaging Component for Testing";
ASTM D 4169-16 "Standard Practice for Performance Testing of Shipping Containers and Systems";
ASTM F 1929-15 "Standard Test Method For Detecting Seal Leaks In Porous Medical Packaging By Dye Penetration".
Animal and clinical testing was not applicable to support substantial equivalence
The conclusions drawn from the nonclinical testing on the subject devices demonstrate that the devices are as safe, as effective, and performs as well as the correspondent legally marketed predicate devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.