K190310 · U Deliver Medical, LLC · PIF · May 31, 2019 · Gastroenterology, Urology
Device Facts
Record ID
K190310
Device Name
U Deliver Bolink ENFit Enteral Feeding Sets
Applicant
U Deliver Medical, LLC
Product Code
PIF · Gastroenterology, Urology
Decision Date
May 31, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The U Deliver Bolink ENFit Enteral Feeding Sets are intended for over-the-counter use to deliver liquid nutritional formulas or water to a patient's enteral access device (feeding tube).
Device Story
Gravity-fed enteral feeding sets; connect liquid nutritional reservoirs to patient feeding tubes. Components include PVC tubing, external flow clamp, and distal ENFit connector (ISO 80369-3). Small cap version includes cross-spike; large cap version includes 40mm thread screw cap. Used in home or clinical settings by patients, caregivers, or clinicians. Device facilitates administration of hydration, medication, and nutrition. Flow controlled manually via external clamp. Non-sterile, single-patient use.
Clinical Evidence
No clinical data. Bench testing and usability testing performed. Usability study demonstrated that both laypersons with no prior tube feeding experience and caregivers/tubefeeders could successfully perform essential tasks using provided instructions, supporting OTC use.
Technological Characteristics
Medical grade, biocompatible materials. PVC tubing. ENFit connectors (ISO 80369-3). Safety screw connector (ISO 18250-3). Gravity-fed mechanism. Non-sterile. Assembled in ISO 14644 Class 8 clean room.
Indications for Use
Indicated for delivery of liquid nutritional formulas or water to enteral access devices (feeding tubes) in patients of all ages, including adults, children, infants, and neonates.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
U Deliver Bolink ENFit Enteral Feeding Sets (K170555)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 31, 2019
U Deliver Medical, LLC % Thomas Dold President/ Principal Consultant Vista Regulatory Solutions, LLC 18205 E. Sioux Vista Dr. Jordan, MN 55352
Re: K190310/S001
> Trade/Device Name: U Deliver Bolink ENFit Enteral Feeding Sets Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: Class II Product Code: PIF Dated: February 7, 2019 Received: February 12, 2019
Dear Thomas Dold:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Shanil P. Haugen, Ph.D. Acting Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K190310
Device Name
U Deliver Bolink ENFit Enteral Feeding Sets
Indications for Use (Describe)
The U Deliver Bolink ENFit Enteral Feeding Sets are intended for over-the-counter use to deliver liquid nutritional formulas or water to a patient's enteral access device (feeding tube).
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
PSC Publishing Services (301) 443-6740
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
| Trade Name: | U Deliver Bolink ENFit Enteral Feeding Sets |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of 510(k) submission: | Traditional |
| Date of Submission: | 05 May 2019 |
| 510(k) Owner and Submitter: | U Deliver Medical, LLC<br>7900 Excelsior Blvd, Suite 108<br>Hopkins, MN 55343<br>USA |
| Phone: | 952-200-0270 |
| FDA Registration Number: | 3014016859 |
| Owner/Operator Number: | 10055940 |
| 510(k) Application Correspondent: | Mr. Thomas Dold<br>President / Principal Consultant<br>Vista Regulatory Solutions, LLC<br>Jordan, MN 55352 |
| Phone: | (612) 965-6057 |
| Email: | tdold@vistareg.com |
| FDA Product Code: | PIF |
| FDA Regulation Number: | 21 CFR 876.5980 |
| FDA Classification Name: | Gastrointestinal tube and accessories |
| Classification Panel: | Gastroenterology and Urology |
| Common Name: | Enteral Feeding Sets |
| FDA Device Class: | Class II |
| FDA Identification: | The connector is attached to an enteral tube and is intended to<br>facilitate the administration of hydration, medication, and feed<br>to a patient. |
# Device Description:
The U Deliver Bolink ENFit Enteral Feeding Sets consist of two configurations:
- U Deliver Bolink Small Cap Gravity Feeding Set with Cross Spike and ENFit Connector (ref. BK-2085) .
- U Deliver Bolink Large Cap Gravity Feeding Set with ENFit Connector (ref: BK-2405)
The U Deliver Bolink Small Cap Gravity Feeding Set with Cross Spike and ENFit Connector uses the free flow of gravity to dispense enteral feeding solutions from a distal reservoir (not supplied) which connects via a safety screw connection meeting the requirements of clauses 4 "General requirement" and 5 "Dimensional requirements for enteral feed reservoir connectors" of ISO18250-3, "Connectors for reservoir delivery systems for healthcare applications - Part 3: Enteral applications', to a PVC tube with external clamp, to a distal ENFit connector, meeting the requirements of ISO 80369-3, 'Small-bore connectors for liquids and gases in healthcare applications - Part 3: Connectors for enteral applications'.
{4}------------------------------------------------
The U Deliver Bolink Large Cap Gravity Feeding Set with ENFit Connector uses the free flow of gravity to dispense enteral feeding solutions from a distal reservoir (not supplied) which connects via a screw cap with a 40 mm thread, to a PVC tube with external clamp, to a distal ENFit connector, meeting the requirements of ISO 80369-3
Both sets are constructed from medical grade, biocompatible materials, and are supplied non-sterile for single patient use only.
# Indications for Use/Intended Use
The U Deliver Bolink ENFit Enteral Feeding Sets are intended for over-the-counter use to deliver liquid nutritional formulas or water to a patient's enteral access device (feeding tube).
## Principle of operation, mechanism of action, and interaction with the patient
The U Deliver Bolink ENFit Enteral Feeding Sets operate by providing a means of interconnecting enteral feeding solution reservoirs to a patient's enteral access device (feeding tube), so that patient enteral feeding can take place when 'new generation' enteral end fittings in accordance with ISO 80369-3 ('ENFit' connections) are present on the feeding tube. A clamp is provided on the tubes to allow feeding to be turned on or off. Although contact of components of the Enteral Feeding Sets may accidentally occur with patient's intact skin during use, the Enteral Feeding Sets have a central fluid path through which feeding solutions flow into the patient during the feeding process.
The U Deliver Bolink ENFit Enteral Feeding Sets are intended for over-the-counter use.
U Deliver Bolink ENFit Enteral Feeding Sets are not intended for a specific patient population: they can be used for adults, children, infants and neonates.
## Device Specification
The ENFit connectors on the Enteral Feeding Sets are designed in compliance with FDA-recognized standard ISO 80369-3 "Small-bore connectors for liquids and gases in healthcare applications — Part 3: Connectors for enteral applications" and tested in accordance with FDA-recognized standard ISO 80369-20 "Small-bore connectors for liquids and gases in healthcare applications — Part 20: Common test methods".
The safety screw connector on the U Bolink ENFit Deliver Enteral Set with Safety Screw Connector is designed in accordance with ISO/DIS 18250-3.
The screw cap on the U Deliver Bolink Large Cap Gravity Feeding Set with ENFit Connector is designed with identical major dimensions to the predicate device, U Deliver Bolink ENFit Enteral Feeding Sets (K170555).
## Manufacture
The U Deliver Bolink ENFit Enteral Feeding Sets are assembled from components manufactured by subcontractors/suppliers to Cedic, the contract manufacturer, in environmentally controlled areas (clean room Class 8 according to ISO 14644), together with tubing and adhesive sourced 'off-the-shelf' by Cedic from other suppliers.
## Performance Data
Bench tests were carried out on samples of the U Deliver Bolink ENFit Enteral Feeding Sets and were submitted with the predicate device (K170555). No new device testing was necessary as no changes have been made to the device design, materials or packaging. The testing necessary for this submission was usability testing to demonstrate that both laypersons with no prior tube feeding experience and caregivers/tubefeeders were able to perform the essential tasks using only the proposed instructions for use. The results of usability testing demonstrate that the subject device(s) are appropriate for over-the-counter use.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the text "U Deliver Medical LLC" in blue font. The words "U Deliver" are on the top line and are a darker shade of blue. The words "Medical LLC" are on the second line and are a lighter shade of blue.
# Predicate device
The Predicate Device selected for comparison with the U Deliver Bolink ENFit Enteral Feeding Sets:
| Predicate Device: | U Deliver Bolink ENFit Enteral Feeding Sets |
|--------------------------|---------------------------------------------|
| Sponsor: | U Deliver Medical, LLC |
| 510(k) Number: | K170555 |
| Clearance Date: | 17 August 2017 |
| FDA Product Code: | PIF |
| FDA Classification Name: | Gastrointestinal Tubes and Accessories |
| FDA Regulation No: | 21-CFR-876.5980 |
The Subject Device and the Predicate Device have identical properties and features. The only difference that exists is:
- The indications for use and intended use have changed from Prescription-Use-Only to ● Over-the-Counter Use.
## Conclusion:
Based on the information contained within this submission, it is concluded that the U Deliver Bolink ENFit Enteral Feeding Sets intended for over-the-counter use, are substantially equivalent and are as safe and effective as the identified predicate device already in interstate commerce within the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.