K242822 · Therapanacea Sas · MUJ · Feb 25, 2025 · Radiology
Device Facts
Record ID
K242822
Device Name
ART-Plan+ (v.3.0.0)
Applicant
Therapanacea Sas
Product Code
MUJ · Radiology
Decision Date
Feb 25, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Organ at risk and lymph node segmentation
Deep learning algorithm
Mean DSC >= 0.8
—
—
—
Quantitative evaluation of autosegmentation of organs at risk: minimum sample size of 24 patients
>1 (medical experts)
Brain metastasis segmentation
Deep learning algorithm
Mean lesion-wise sensitivity >= 0.86; mean lesion-wise precision >= 0.70; mean lesion-wise DSC >= 0.78; mean patient-wise DSC >= 0.83; mean patient-wise FP <= 2.1
—
—
—
Quantitative evaluation of autosegmentation of Brain metastasis: minimum sample size of 51 patients
>1 (medical experts)
Glioblastoma segmentation
Deep learning algorithm
Mean sensitivity >= 0.80; mean DSC >= 0.76
—
—
—
Quantitative evaluation of autosegmentation of Glioblastoma: minimum sample size of 43 patients
Quantitative and qualitative evaluation of automatic treatment plans generations: minimum sample size of 20 patients
>1 (medical experts)
Indications for Use
ART-Plan+'s indicated target population is cancer patients for whom radiotherapy treatment has been prescribed. In this population, any patient for whom relevant modality imaging data is available. ART-Plan+'s includes several modules: - SmartPlan which allows automatic generation of radiotherapy treatment plan that the users import into their own Treatment Planning System (TPS) for the dose calculation, review and approval. This module is available for supported prescriptions for prostate only. - Annotate which allows automatic generation of contours for organs at risk, lymph nodes and tumors, based on medical practices, on medical images such as CT and MR images ART-Plan+ is not intended to be used for patients less than 18 years of age. The indicated users are trained medical professionals including, but not limited to, radiotherapists, radiation oncologists, medical physicists, dosimetrists and medical professionals involved in the radiation therapy process. The indicated use environments include, but are not limited to, hospitals, clinics and any health facility offering radiation therapy.
Device Story
ART-Plan+ is a software platform for radiotherapy planning; inputs include 3D medical images (CT, MR). It utilizes deep-learning algorithms to perform automatic segmentation of organs at risk, lymph nodes, and tumors; it also generates automatic radiotherapy treatment plans for prostate cases. Used in hospitals and clinics by radiation oncologists, medical physicists, and dosimetrists. The software integrates into existing radiotherapy workflows; users import generated contours and plans into their own Treatment Planning System (TPS) for final dose calculation, review, and approval. By automating time-consuming contouring and planning tasks, the device aims to improve efficiency and consistency in the radiation therapy process, ultimately benefiting patients through standardized, high-quality treatment planning.
Clinical Evidence
Bench testing only. Validation included non-regression testing (n=24), qualitative evaluation (n=18), and quantitative evaluation (n=24) of OAR autosegmentation. Brain metastasis (n=51) and glioblastoma (n=43) segmentation were validated against expert contours using DSC, sensitivity, precision, and false-positive metrics. Treatment plan generation (n=20) was validated via quantitative DVH goal comparison (≤5% difference) and qualitative expert review (≥93% clinical acceptability). All tests met pre-defined acceptance criteria.
Technological Characteristics
Software-based radiotherapy planning platform. Utilizes deep-learning algorithms for image segmentation and treatment planning. Compatible with CT and MR imaging modalities. Operates as a standalone software application integrated into clinical workflows. No hardware components. Connectivity via standard clinical network/TPS integration.
Indications for Use
Indicated for cancer patients (≥18 years) prescribed radiotherapy where relevant modality imaging data is available. Used for automatic contouring of organs at risk, lymph nodes, and tumors, and automatic generation of radiotherapy treatment plans (prostate only).
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
February 25, 2025
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Therapanacea SAS Bhairavi Ajachandra Compliance Director 7 bis boulevard Bourdon Paris, 75004 France
Re: K242822
Trade/Device Name: ART-Plan+ (v.3.0.0) Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: MUJ, OKB, LLZ Dated: September 18, 2024 Received: September 18, 2024
Dear Bhairavi Ajachandra:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Locan Weidner
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
#### K242822
Device Name
ART-Plan+ (v.3.0.0)
Indications for Use (Describe)
ART-Plan+'s indicated target population is cancer patients for whom radiotherapy treatment has been prescribed. In this population, any patient for whom relevant modality imaging data is available.
ART-Plan+'s includes several modules:
- SmartPlan which allows automatic generation of radiotherapy treatment plan that the users import into their own Treatment Planning System (TPS) for the dose calculation, review and approval. This module is available for supported prescriptions for prostate only.
- Annotate which allows automatic generation of contours for organs at risk, lymph nodes and tumors, based on medical practices, on medical images such as CT and MR images
ART-Plan+ is not intended to be used for patients less than 18 years of age.
The indicated users are trained medical professionals including, but not limited to, radiotherapists, radiation oncologists, medical physicists, dosimetrists and medical professionals involved in the radiation therapy process.
The indicated use environments include, but are not limited to, hospitals, clinics and any health facility offering radiation therapy.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K242822
### 510(k) Summary
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92.
# Contact details
Applicant Name:
TheraPanacea SAS
#### Applicant Address:
7 bis boulevard Bourdon 75004 Paris France
#### Applicant Contact Telephone:
+33620604982
#### Applicant Contact:
Mrs. Bhairavi AJACHANDRA
#### Applicant Contact Email:
b.ajachandra@therapanacea.eu
### Device Name:
| Device Trade<br>Name | Regulation<br>Number | Common name | Device<br>Class | Product<br>Code(s) | Classification<br>Name |
|-----------------------|----------------------|------------------------------------------------------|-----------------|------------------------------------------------------|-----------------------------------------------------------|
| ART-Plan+<br>(v3.0.0) | 892.5050 | Medical charged-particle<br>radiation therapy system | Class II | MUJ<br>Associated<br>Product<br>Code(s):<br>QKB, LLZ | System,<br>Planning,<br>Radiation<br>Therapy<br>Treatment |
# Legally marketed predicate device
| Predicate # | Predicate trade name<br>(primary predicate is listed<br>first) | Product code |
|-------------|----------------------------------------------------------------|--------------|
| K222728 | Radiation Planning Assistant<br>(RPA) | MUJ |
{5}------------------------------------------------
# Legally marketed reference devices
| Reference device # | Reference device trade name | Product code |
|--------------------|-----------------------------|--------------|
| K213628 | VBrain | QKB |
| K234068 | ART-Plan | MUJ |
### Device Description Summary:
ART-Plan+ is a software platform allowing contour regions of interest on 3D images and to provide an automatic treatment plan. It includes several modules:
-Home: tasks
-Annotate and TumorBox: contouring of regions of interest
-SmartPlan: creation of an automatic treatment plan based on a planning CT and a RTSS
-Administration and settings : preferences management, user account management, etc.
-Institute Management: institute information, including licenses, list of users, etc.
-About: information about the software and its use, as well as contact details.
Annotate, TumorBox and SmartPlan are partially based on a batch mode, which allows the user to launch the operations of autocontouring and autoplanning without having to use the interface or the viewers. In that way, the software is completely integrated into the radiotherapy workflow and offer to the user a maximum of flexibility.
ART-Plan+ offers deep-learning based automatic segmentation of OARs and LNs for the following localizations:
-Head and neck (on CT images)
-Thorax/breast (on CT images)
-Abdomen (on CT and male on MR images)
-Pelvis male (on CT and MR images)
-Pelvis female (on CT images)
-Brain (on CT images and MR images)
ART-Plan+ offers deep-learning based automatic segmentation of targets for the following localizations:
{6}------------------------------------------------
-Brain (on MR images)
### Intended Use/Indication for use:
ART-Plan+'s indicated target population is cancer patients for whom radiotherapy treatment has been prescribed. In this population, any patient for whom relevant modality imaging data is available.
ART-Plan+'s includes several modules:
- SmartPlan which allows automatic generation of radiotherapy treatment plan that the users import into their own Treatment Planning System (TPS) for the dose calculation, review and approval. This module is available for supported prescriptions for prostate only.
- Annotate which allows automatic generation of contours for organs at risk, lymph nodes and tumors, based on medical practices, on medical images such as CT and MR images
ART-Plan+ is not intended to be used for patients less than 18 years of age.
The indicated users are trained medical professionals including, but not limited to, radiotherapists, radiation oncologists, medical physicists, dosimetrists and medical professionals involved in the radiation therapy process.
The indicated use environments include, but are not limited to, hospitals, clinics and any health facility offering radiation therapy.
### Indication for use comparison:
The intended use and indications for use of the proposed device, ART-Plan+ v3.0.0 and the primary predicate are similar as they are both softwares intended to be used in the planning of radiotherapy treatment:
-They allow creation of contours
-Thev allow the planification of external beam irradiation with photon beams using computerized tomography (CT) images.
-They allow the user to import the generated plan into their own Treatment Planning System (TPS) for dose calculation, review and approval.
The intended use and indications for use of the proposed device, ART-Plan+ v3.0.0 and the reference devices are similar as they are also softwares intended to be used in the planning of radiotherapy treatment in order to provide Al based segmentation tools to contour known (diagnosed) tumors, for organs at risk and lymph nodes.
# Technological Comparison:
ART-Plan+ is substantially equivalent to the Radiation Planning Assistant (RPA) (K222728) predicate device in the following aspects:
{7}------------------------------------------------
- . Both devices allow the planification of external beam irradiation with photon beams using computerized tomography (CT) images.
- . Both devices allow the creation of contours and treatment plans that the user imports into their own Treatment Planning System (TPS) for the dose calculation, review and approval.
There are no differences between the proposed device and the predicate that represent an additional claim for the proposed device.
| Technical Information | | | |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Property | Proposed Device<br>ART-Plan v3.0.0 | Primary Predicate<br>Radiation Planning<br>System (RPA) | Comment |
| Delineation<br>Method | AI | AI | The proposed device and primary predicate share an AI delineation method. |
| Segmentation<br>Features | Automatically delineates<br>OARs and lymph nodes and<br>targets.<br><br>Deep learning algorithm.<br><br>Automatic segmentation<br>includes the following<br>localizations:<br>* head and neck (on CT)<br>* thorax/breast (for<br>male/female and on CT)<br>* abdomen (on CT images<br>and MR images)<br>* pelvis male (on CT and<br>MR)<br>* pelvis female (on CT<br>images)<br>* brain (on CT images and<br>MR images) | The RPA provides autocontouring for a range of structures | The proposed device proposes segmentation on the same localizations with the same medical images of different modalities using AI. |
| Regions and<br>Volumes of<br>Interest<br>(ROI) | AI Based autocontouring | AI Based autocontouring, | The proposed device and the primary predicate allow AI automatic contouring. |
# Non-Clinical and/or Clinical Tests Summary & Conclusions
In order to determine the substantial equivalence of ART-Plan + v3.0.0. the following tests have been performed:
Non regression testing of autosegmentation of organs at risk - Validation of the performance of autoseqmentation (on CT and MR images) of already existing structures after retraining or algorithm improvement was performed by comparison of the DSC between contours generated by the previous version of ART-Plan (v2.2.0) and manual contours performed by medical experts and the DSC between contours generated by the new version of ART-Plan + (v3.0.0) and manual contours performed by medical experts. The evaluation was performed on a minimum sample size of 24 patients.
{8}------------------------------------------------
The contours are considered acceptable for clinical use if the following acceptance criterion is achieved:
-Mean DSC should not regress negatively between the current and last validated version of Annotate beyond a maximum tolerance marqin set to -5% relative error.
Qualitative evaluation of autosegmentation of organs at risk - Validation of the performance of autosegmentation of new organs at risk (or organs at risk not passing the non regression testing) on CT images was performed by qualitative evaluation of the contours by medical experts. This evaluation was performed on a minimum sample size of 18 patients.
The contours are considered acceptable for clinical use if the following acceptance criterion is achieved:
-The clinicians' qualitative evaluation of the auto-segmentation is considered acceptable for clinical use without modifications (A) or with minor modifications / corrections (B) with a A+B % above or equal to 85% considering the following scale:
A. the contour is acceptable for a clinical use without any modification
B. the contour would be acceptable for clinical use after minor modifications/corrections
C. the contour requires major modifications (e.g. it would be faster for the expert to manually delineate the structure)"
Quantitative evaluation of autosegmentation of organs at risk - validation of the performance of autosegmentation of new organs at risk on CT images was performed by quantitative evaluation of the contours. This evaluation was performed on a minimum sample size of 24 patients.
The contours are considered acceptable for clinical use if the following acceptance criterion is achieved:
- Mean DSC (annotate) ≥ 0.8
Inter-expert variability evaluation of autosegmentation of organs at risk – validation of the performance of autosegmentation of new organs at risk on CT images was performed by an inter-expert variability evaluation of the contours. The same images were asked to be contoured by two independent medical experts. DSC was calculated between them and with contours generated by the software and the compared. This evaluation was performed on a minimum sample size of 13 patients.
The contours are considered acceptable for clinical use if the following acceptance criterion is achieved:
- Mean DSC (annotate) ≥ Mean DSC (inter-expert) with a tolerance margin of -5% of relative error
{9}------------------------------------------------
Quantitative evaluation of autoseqmentation of Brain metastasis - Validation of the performance of autoseqmentation of brain metastasis on MR images was performed by quantitative evaluation. Contours provided by medical experts were compared with contours generated by the software. This evaluation was performed on a minimum sample size of 51 tients.
The contours are considered acceptable for clinical use if the following acceptance criterion is achieved:
- The lesion-wise sensitivity is equal to or superior to state-of-the-art as benchmark: mean lesion-wise sensitivity ≥ 0.86
#### AND
- The lesion-wise precision is equal to or superior to state-of-the-art as benchmark: mean lesion-wise precision ≥ 0.70
#### AND
- The lesion-wise DSC similarity coefficient (DSC) is equal to or superior to state-of-the-art as benchmark: mean lesion-wise DSC ≥ 0.78
#### AND
- The patient-wise DSC similarity coefficient (DSC) is equal to or superior to state-of-the-art as benchmark: mean patient-wise DSC ≥ 0.83
#### AND
- The patient-wise false positive (FP) is equal to or superior to state-of-the-art as benchmark: mean patient-wise FP ≤ 2.1
Quantitative evaluation of autosegmentation of Glioblastoma - Validation of the performance of autosegmentation of glioblastoma on MR images was performed by quantitative evaluation. Contours provided by medical experts were compared with contours generated by the software. This evaluation was performed on a minimum sample size of 43 patients.
The contours are considered acceptable for clinical use if the following acceptance criterion is achieved:
-The sensitivity is equal to or superior to state-of-the-art as benchmark: mean sensitivity ≥ 0.80
#### AND
-The DSC is equal to or superior to state-of-the-art as benchmark: mean DSC ≥ 0.76
{10}------------------------------------------------
Quantitative and qualitative evaluation of automatic treatment plans generations - validation of the performance of automatic radiotherapy treatment plans generation was performed by both quantitative and qualitative evaluation. The quantitative evaluation consisted of direct comparison with manual plans. The qualitative evaluation consisted of medical experts determining the clinical acceptability of plans. These evaluations were performed on a minimum sample size of 20 patients.
- The treatments plans are considered acceptable for clinical use if the following acceptance criteria are achieved:
Quantitative evaluation: effectiveness difference (%) in DVH achieved goals between manual plans and automatic plans ≤ 5%
AND
Qualitative evaluation: % of clinical acceptable automatic plans ≥ 93% after expert review.
All validation tests were carried out using datasets representative of the worldwide population receiving radiotherapy treatments. Finally, all tests passed their respective acceptance criteria, thus showing ART-Plan + v3.0.0 clinical acceptability.
#### Not applicable
Based on the information presented in these 510(k) premarket notifications the TheraPanacea ART-Plan is considered substantially equivalent. The TheraPanacea ART-Plan + 3.0.0 is as safe and effective as the currently marketed predicate device.
Based on testing and comparison with the predicate device, TheraPanacea ART-Plan indicated no adverse indications or results. It is our determination that the TheraPanacea ART-Plan + 3.0.0 performs within its design specifications and is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.