DSC (mean) >= 0.8 or DSC (mean) >= 0.54 or DSC (mean) >= mean (DSC inter-expert) + 5%; HD95 (mean) <= 5.75 mm; A+B % >= 85%
—
—
—
Validation of autosegmentation on CT, MR, and synthetic-CT from CBCT images: minimum 17 patients; US vs non-US patient data comparison: minimum 17 patients; Qualitative evaluation of new organs at risk: minimum 15 patients
>1 (medical experts)
Contour propagation
—
A+B % >= 85% for deformable registration; A+B % >= 50% for rigid registration
—
—
—
Qualitative evaluation of propagated contours: minimum 15 patients
>1 (medical experts)
Synthetic-CT generation
Deep-learning based synthetic CT-generation
DVH parameters (PTV) < 2%; DVH Individual Pass Rates (PTV) > 76.7%; Median Gamma Index 2%/2mm >= 92%; Median Gamma Index 3%/3mm >= 93.57%; Jacobian Determinant = 1 +/- 5%
—
—
—
Dosimetric and geometric validation: 19 patients per supported anatomy
—
Dose calculation
—
Relative differences on DVH parameters (PTV/OARs) <= 4.4% (Lungs <= 24.4%); Median Gamma Index 2%/2mm >= 86.3%; Median Gamma Index 3%/3mm >= 91.75%
—
—
—
Dosimetric validation: 45 patients per supported anatomy
—
Indications for Use
ART-Plan's indicated target population is cancer patients for whom radiotherapy treatment has been prescribed. In this population, any patient for whom relevant modality imaging data is available. ART-Plan is not intended to be used for patients less than 18 years of age. The indicated users are trained medical professionals including, but not limited to, radiotherapists, radiation oncologists, medical physicists, dosimetrists and medical professionals involved in the radiation therapy process. The indicated use environments are, but not limited to, hospitals, clinics and any health facility involved in radiation therapy.
Device Story
ART-Plan is a software platform for radiotherapy treatment planning and adaptation. It ingests 3D medical images (CT, MR, CBCT) to support clinical workflows. Key modules include: Annotate (manual/automatic contouring, synthetic-CT generation from MR/CBCT); SmartFuse (multi-modal image registration); and AdaptBox (CBCT-based synthetic-CT generation, dose computation, and replanning decision support). Operated by radiation oncologists, physicists, and dosimetrists in clinical settings, the device uses deep-learning algorithms to automate segmentation and synthetic-CT generation. Outputs include contoured images, synthetic-CTs, and dosimetric metrics. Clinicians use these outputs to evaluate treatment plans, assess anatomical changes over the course of treatment, and determine the necessity for replanning, potentially improving treatment accuracy and patient outcomes.
Clinical Evidence
Bench testing only. Validation included non-regression testing (n≥17) and quantitative evaluation of auto-segmentation (n≥17) using DSC (mean ≥0.8) and HD95 (≤5.75mm) metrics. US vs. non-US performance was compared. Qualitative evaluation of contours and propagation was performed by experts (n≥15) with an 85% acceptability threshold. Synthetic-CTs were validated for dosimetric accuracy (n=19/anatomy) using DVH parameters, Gamma Index (2%/2mm ≥92%; 3%/3mm ≥93.57%), and Jacobian Determinant (1±5%). Dose engine validation (n=45/anatomy) compared results against commercial solutions, meeting predefined thresholds for DVH and Gamma Index.
Technological Characteristics
Software-based platform for radiotherapy planning. Features deep-learning algorithms for automatic segmentation and synthetic-CT generation from CT, MR, and CBCT inputs. Supports multi-modal rigid and deformable registration. Performs dose computation for photon beam external beam irradiation. Operates on standard clinical imaging data (DICOM). No specific hardware materials; software-only device.
Indications for Use
Indicated for cancer patients (≥18 years) prescribed radiotherapy where relevant modality imaging data is available.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
April 22, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Therapanacea SAS % Bhairavi Ajachandra QARA Manager 7 bis boulevard Bourdon Paris. 75004 FRANCE
Re: K234068
Trade/Device Name: ART-Plan (v.2.2.0) Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: MUJ Dated: December 22, 2023 Received: December 22, 2023
Dear Bhairavi Ajachandra:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Locon Weidner
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
#### Indications for Use
Submission Number (if known)
K234068
Device Name
ART-Plan (v.2.2.0)
Indications for Use (Describe)
ART-Plan's indicated target population is cancer patients for whom radiotherapy treatment has been prescribed. In this population, any patient for whom relevant modality imaging data is available.
ART-Plan is not intended to be used for patients less than 18 years of age.
The indicated users are trained medical professionals including, but not limited to, radiotherapists, radiation oncologists, medical physicists, dosimetrists and medical professionals involved in the radiation therapy process.
The indicated use environments are, but not limited to, hospitals, clinics and any health facility involved in radiation therapy.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k)#: K234068
# 510(k) Summary
#### Contact Details
Applicant Name: Therapanacea SAS Applicant Address: 7 bis boulevard Bourdon Paris 75004 France Applicant Contact Telephone: +33620604982 Applicant Contact: Mrs. Bhairavi AJACHANDRA Applicant Contact Email: b.ajachandra@therapanacea.eu
## Device Name
Device Trade Name: ART-Plan (v.2.2.0)
Common Name: Medical charged-particle radiation therapy system
Classification Name: System, Planning, Radiation Therapy Treatment
Regulation Number: 892.5050
Product Code: MUJ
## Legally Marketed Predicate Devices
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K232479 | ART-Plan | MUJ |
## Device Description Summary
ART-Plan is a software platform allowing users with an account to contour regions of interest on 3D images, perform multi-model registration of images, and help in the decision for the need for replanning based on contours and doses on daily images. It includes several modules:
- Home,
- Annotate,
- SmartFuse,
- AdaptBox,
- Administration,
- About
{4}------------------------------------------------
ART-Plan asks the user to work by project. It is necessary to create a "project" entity in the Patient page by associating it to a reference volume (preferably the positioning CT or the positioning MR) in order to create the contours on this volume in the Annotate module or to use this image as reference to compare it with CBCT images in the AdaptBox module. It is possible to create several projects for a given patient.
The Home module allows the user to search for a patient already present in the software's database or to import it from the structure's imaging servers or from another external source and to manage the different projects of this patient.
The Annotate module allows the user to contour regions of interest on the reference volume. It also allows generation of pseudo-CTs from MRI images. Users are able to visualize, evaluate and modify the HU values of the associated structures on the pseudo-CT, if needed. After validation, manual and automatic contours can be generated on the pseudo-CT images. Registration can also be performed using the pseudo-CT either as a target or source image. All results can be exported upon approval.
The SmartFuse module allows the user to fuse the primary volume of a project with secondary volumes.
The AdapBox module helps the user to decide if a replanning is necessary. For this purpose, the module allows the user to generate a pseudo-CT from a CBCT image, to auto-delineate regions of interest on the pseudo-CT, to compute the dose on both planning CT and pseudo-CT and then define if there is a need for replanning by comparing volume and dose metrics computed on both images and over the course of the treatment. Those metrics are defined by the user.
The Administration module allows users with specific rights to manage the platform's usage parameters. Some more restricted rights are also accessible in the drop-down menu linked to the user account through the Settings menu.
The About module allows the user to obtain information about the software and its use, as well as to contact TheraPanacea.
{5}------------------------------------------------
ART-Plan offers deep-learning based automatic segmentation for the following localizations:
- head and neck (on CT and synthetic-CT from CBCT images)
- thorax/breast (for male/female and on CT and synthetic-CT from CBCT images)
- abdomen (on CT images and MR images)
- pelvis male(on CT images, on synthetic-CT from CBCT and on MR images)
- . pelvis female (on CT images)
- brain (on CT images and MR images)
ART-Plan offers deep-learning based synthetic CT-generation from MR images for the following localizations:
- pelvis male
- Brain
ART-Plan offers deep-learning based synthetic CT-generation from CBCT images for the following localizations:
- pelvis male
- head and neck
- thorax/breast
# Intended Use/Indications for Use
ART-Plan's indicated target population is cancer patients for whom radiotherapy treatment has been prescribed. In this population, any patient for whom relevant modality imaging data is available.
ART-Plan is not intended to be used for patients less than 18 years of age.
The indicated users are trained medical professionals including, but not limited to, radiotherapists, radiation oncologists, medical physicists, dosimetrists and medical professionals involved in the radiation therapy process.
The indicated use environments are, but not limited to, hospitals, clinics and any health facility involved in radiation therapy.
## Indications for Use Comparison
{6}------------------------------------------------
The proposed device has the same indication for use as the predicate device.
## Technological Comparison
The device has the same technological characteristics.
The similarities between the proposed device and the primary predicate are the following:
Both devices allow multi-modal vizualisation and rigid and deformable registration for the same modalities of images
Both devices allow displaying fused and non-fused images to facilitate the comparison and delineation of image data by the user
Both devices allow manual generation, modification and semi-automatic generation of contours for the regions of interest
Both devices allow automatic segmentation on medical images using Al algorithms
Both devices allow generation of synthetic-CT from CBCT and MR images
Both allow dose computation on CT and/or synthetic-CT images for external beam irradiation with photon beams
Both allow assisted CBCT-based off-line adaptation decision-making for supported anatomies
Both devices allow the import, manipulation, visualisation, generation and the export of DICOM images
- New anatomies for generation of synthetic CT from CBCT : Thorax, Breast, Head & Neck l
- -Addition of new structures to the already existing localizations for the segmentation feature
There are no differences between the proposed device and the primary predicate that represent an additional claim for the primary predicate.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
Within the primary predicate ART-Plan v2.1.0, for an auto segmentation model to be judged acceptable, every organ included in the model must pass at least one acceptance criterion with success across the different testings it has been submitted to. These criteria are as follows:
{7}------------------------------------------------
a) The Dice Similarity Coefficient (DSC) is equal to or superior to the acceptance criteria set by the AAPM: DSC (mean)≥ 0.8.
Or
b) The Dice Similarity Coefficient (DSC) is equal to or superior to inter-expert variability: DSC (mean)≥ 0.54 or DSC (mean) ≥ mean (DSC inter-expert) + 5%.
Or
c) The clinicians' s qualitative evaluation of the auto-segmentation is considered acceptable for clinical use without modifications (A) or with minor modifications / corrections (B) with a A+B % above or equal to 85% considering the following scale:
A: the contour is acceptable for a clinical use without any modification
B: the contour would be acceptable for clinical use after minor modifications/corrections
C: the contour requires major modifications (e.g. it would be faster for the expert to manually delineate the structure)"
Within the proposed device ART-Plan (v2.2.0), the same acceptance criteria were used. Moreover, in order to overcome the limitations of reporting DSC alone and to help establish that proposed device is functioning appropriately across each segmented structure, an additional quantitative metric was used: the Hausdorff Distance (95th percentile – HD95). This metric has been chosen as it is the second most used metric, after DSC. Based on litterature review, the acceptance criteria is set as the following: the Hausdorff distance 95th percentile (HD95) is equal to or inferior to the acceptance criteria set HD95 (mean) ≤ 5.75 mm.
Thus, in order to determine the substantial equivalence of ART-Plan v2.2.0, the following tests have been performed:
Non regression testing of autosegmentation of already existing organs at risk in the primary predicate ART-Plan v2.1.0 (previous version) - Validation of the performance of autosegmentation (on CT and MR images) of already existing structures after retraining or algorithm improvement was performed by comparison of the DSC between contours generated by the previous version of ART-Plan (v2.1.0) and manual contours performed by medical experts and the DSC between contours generated by the new version of ART-Plan (v2.2.0) and manual contours performed by medical experts. The evaluation was performed on a minimum sample size of 17 patients.
The contours are considered acceptable for clinical use if the following acceptance criterion is achieved:
- । Mean DSC should not regress negatively between the current and last validated version of Annotate beyond a maximum tolerance margin set to -5% relative error.
Quantitative evaluation of autosegmentation of all organs at risk – Validation of the performance of autosegmentation of all organs at risk on CT, MR and synthetic-CT from CBCT images by comparison with manual performed by medical experts. The evaluation was performed on a minimum sample size
{8}------------------------------------------------
of 17 patients. The contours are considered acceptable for clinical use if the following acceptance criteria are achieved:
- -The Dice Similarity Coefficient (DSC) is equal to or superior to the acceptance criteria set by the AAPM: DSC (mean)≥ 0.8.
- -The Hausdorff distance 95th percentile (HD95) is equal to or inferior to the acceptance criteria set HD95 (mean) ≤ 5.75 mm.
Quantitative evaluation of autosegmentation of all organs at risk – Validation of the performance of autosegmentation of organs at risk on CT images on US patient data compared to performance on non-US (nUS) patient data. The comparison was assessed by comparing automatic contours with manual contours performed by medical experts. The evaluation was performed on a minimum sample size of 17 patients. The contours are considered acceptable for clinical use if the following acceptance criteria are achieved:
- -The Dice Similarity Coefficient (DSC) is equal to or superior to the acceptance criteria set by the AAPM: DSC (mean)≥ 0.8.
- The Hausdorff distance 95th percentile (HD95) is equal to or inferior to the acceptance criteria set HD95 (mean) ≤ 5.75 mm.
or
- Mean DSC (US) ≥Mean DSC (nUS) -
and/or
- -Mean HD95(US) ≤ Mean HD95 (nUS)
Qualitative evaluation of autosegmentation of organs at risk - Validation of the performance of autosegmentation of all new organs at risk (or organs at risk not passing the non regression testing) on CT, MR and synthetic-CT from CBCT images was performed by qualitative evaluation of the contours by medical experts. This evaluation was performed on a minimum sample size of 15 patients.
The contours are considered acceptable for clinical use if the following acceptance criterion is achieved:
- । The clinicians' qualitative evaluation of the auto-segmentation is considered acceptable for clinical use without modifications (A) or with minor modifications / corrections (B) with a A+B % above or equal to 85% considering the following scale:
- A. the contour is acceptable for a clinical use without any modification
- B. the contour would be acceptable for clinical use after minor modifications/corrections
- ﻥ the contour requires major modifications (e.g. it would be faster for the expert to manually delineate the structure)"
{9}------------------------------------------------
Qualitative evaluation of contours propagation - Validation of the performance of propagation of target and organs at risk contours from the planning CT to the synthetic-CT from CBCT was performed by qualitative evaluation of the propagated contours on the synthetic-CT from CBCT. The evaluation was performed on a minimum sample size of 15 patients.
The propagation of contours is considered acceptable for clinical use if the following acceptance criterion is achieved:
- । The clinicians' s qualitative evaluation of the propagated contours post-registration are considered acceptable for clinical use without modifications (A) or with minor modifications / corrections (B) with a A+B % above or equal to 85% for deformable and above or equal to 50% for rigid registration considering the following scale:
- A. the contour is acceptable for a clinical use without any modification
- B. the contour would be acceptable for clinical after use minor modifications/corrections
- ﻥ the contour requires major modifications (e.g. it would be faster for the expert to manually delineate the structure)
Dosimetric validation of synthetic-CT from CBCT - Validation of synthetic-CTs in terms of dosimetric endpoints was performed by a quantitative evaluation to evaluate the non-inferiority of synthetic-CTs compared to CT images. Treatment plans were optimised on the CT images and recalculated on the synthetic-CT of the same patient. The dose distributions of the CTs and the synthetic-CTs were compared to evaluate non-inferiority of sCTs in terms of dosimetric endpoints. The evaluation was performed on a sample size of 19 patients per supported anatomy.
The synthetic-CTs from CBCT are considered acceptable for clinical use if the following acceptance criteria are achieved:
- DVH parameters (PTV): < 2%
- DVH Individual Pass Rates (PTV): > 76.7%
- -Median Gamma Index 2%/2mm: ≥ 92%
- Median Gamma Index 3%/3mm: ≥ 93.57%
Geometric and anatomic validation of the synthetic-CT from CBCT - Validation of synthetic-CTs in terms of anatomical and geometrical accuracy was performed by a quantitative evaluation. A deformable fusion was performed between the synthetic-CT from CBCT and original CBCT. The Jacobian determinant was calculated in order to determine if the synthetic-CT and the original CBCT are identical, thus meaning that the synthetic-CT reflects the anatomy of the CBCT. The evaluation was performed on a sample size of 19 patients per supported anatomy.
The synthetic-CTs from CBCT are considered acceptable for clinical use if the following acceptance criterion is achieved:
- | Jacobian Determinant = 1 +/- 5%
{10}------------------------------------------------
Dosimetric validation of the dose engine - Validation of the dose engine was performed by quality evaluation based on the dosimetric quality evaluated from clinical dosimetric criteria in comparison with commercially available solutions. The evaluation was performed on a sample size of 45 patients per supported anatomy.
The dose engine is considered acceptable for clinical use if the following acceptance criteria are achieved:
- । Relative differences on DVH parameters (PTV/OARs): ≤ 4.4% (with some superior individual values); for Lungs ≤ 24.4%
- । Median Gamma Index 2%/2mm (with different thresholds): ≥ 86.3%
- । Median Gamma Index 3%/3mm (with different thresholds): ≥ 91.75%
All validation tests were carried out using datasets representative of the worldwide population receiving radiotherapy treatments. Finally, all tests passed their respective acceptance criteria, thus showing ART-Plan v2.2.0 clinical acceptability.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.