K242661 · Tri Dental Implants Int. AG · DZE · Jan 8, 2025 · Dental
Device Facts
Record ID
K242661
Device Name
TRI-matrix® X-Force Implants
Applicant
Tri Dental Implants Int. AG
Product Code
DZE · Dental
Decision Date
Jan 8, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TRI-matrix® X-Force implants are intended for placement in the bone of the maxillary or mandibular arch for the rehabilitation of edentulous and partially edentulous patients. TRI-matrix® X-Force implants allow for one and two stage surgical procedures. When a one-stage surgical procedure is applied, the implant may be immediately loaded when good primary stability is achieved and with appropriate occlusal loading. TRI-matrix® X-Force implants with lengths of 18, 20, or 22 mm, when placed in the maxilla, are indicated only for multiple unit restorations in splinted applications that utilize at least two implants.
Device Story
Tissue-level endosseous dental implants; tapered body; double-lead threads; self-cutting flutes. Input: surgical site in maxilla/mandible. Transformation: mechanical integration into bone via threaded design; grit-blasted/acid-etched surface (TRI SBA) for osseointegration; pink anodized transgingival collar. Output: stable foundation for prosthetic restoration (crowns/bridges). Used in dental clinics by dentists/oral surgeons. Clinical decision-making: implant placement restores chewing function and esthetics. Benefits: immediate loading capability; improved stability via aggressive apical flutes; adaptation for immediate placement/restoration.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing (biocompatibility per ISO 10993-5/12, static/dynamic mechanical testing per ISO 14801) and engineering analysis.
Technological Characteristics
Material: Titanium alloy (Ti-6Al-4V ELI) per ASTM F136. Surface: Grit-blasted and acid-etched (TRI SBA) with pink anodized collar. Connection: Internal. Sterilization: Gamma irradiation. Dimensions: Body diameters 3.3-5.7mm; lengths 8-22mm. Form factor: Tissue-level, tapered body, double-lead threads, self-cutting flutes.
Indications for Use
Indicated for edentulous and partially edentulous patients requiring rehabilitation of maxillary or mandibular arches. Suitable for one or two-stage surgical procedures. Immediate loading permitted if primary stability and occlusal loading conditions are met. Implants 18-22mm in maxilla restricted to splinted multiple-unit restorations (min. 2 implants).
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
TRI Dental Implants Int. AG % Floyd Larson President PaxMed International, LLC 12264 El Camino Real Suite 400 San Diego, California 92130
January 8, 2025
Re: K242661
Trade/Device Name: TRI-matrix® X-Force Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: August 30, 2024 Received: December 16, 2024
## Dear Floyd Larson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K242661
Device Name
TRI-matrix® X-Force Implants
#### Indications for Use (Describe)
TRI-matrix® X-Force implants are intended for placement in the bone of the maxillary or mandibular arch for the rehabilitation of edentulous and partially edentulous patients. TRI-matrix® X-Force implants allow for one and two stage surgical procedures. When a one-stage surgical procedure is applied, the implant may be immediately loaded when good primary stability is achieved and with appropriate occlusal loading. TRI-matrix® X-Force implants with lengths of 18, 20, or 22 mm, when placed in the maxilla, are indicated only for multiple unit restorations in splinted applications that utilize at least two implants.
Type of Use (Select one or both, as applicable)
< Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary TRI Dental Implants Int. AG TRI-matrix® X-Force Implants
January 7, 2024
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | TRI Dental Implants Int. AG | |
|---------------------------|--------------------------------------------|--------------------|
| | Boesch 80A/Postfach 419 | |
| | CH-6331, Huenenberg, Switzerland | |
| | Telephone: | +41-32-510-1606 |
| | Fax: | +41-32-510-1606 |
| Official Contact | Sandro Venanzoni, Chief Technology Officer | |
| Representative/Consultant | Floyd G. Larson, MS, MBA | |
| | Kevin Thomas, PhD | |
| | PaxMed International, LLC | |
| | 1925 Palomar Oaks Way, Suite 210 | |
| | Carlsbad, CA 92008 | |
| | Telephone: | +1-858-792-1235 |
| | Fax: | +1-858-792-1236 |
| | Email: | flarson@paxmed.com |
| | | kthomas@paxmed.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | TRI-matrix® X-Force Implants |
|------------------------|---------------------------------------------------------------------------------------------|
| Common Name | Dental implant |
| Regulation Number | 21 CFR 872.3640 |
| Regulation Name | Endosseous dental implant |
| Regulatory Class | Class II |
| Product Code | DZE |
| Classification Panel | Dental |
| Reviewing Office | Office of Health Technology 1 (Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices) |
| Reviewing Division | Division of Health Technology 1 B (Dental Devices) |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device
K203660, TRI-matrix® Implant Line, TRI Dental Implants Int. AG
Reference Devices
K151916, TRI® Dental Implant System, TRI Dental Implants Int. AG
K212785, Blue Sky Bio Implant System, Blue Sky Bio, LLC
- K230108, Straumann® BLC and TLC Implants, Straumann USA, LLC (on behalf of Institut Straumann AG)
- K192221, Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts, Implant Direct Sybron Manufacturing, LLC
{5}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT
TRI-matrix® X-Force implants are intended for placement in the bone of the maxillary or mandibular arch for the rehabilitation of edentulous and partially edentulous patients. TRI-matrix® X-Force implants allow for one and two stage surgical procedures. When a one-stage surgical procedure is applied, the implant may be immediately loaded when good primary stability is achieved and with appropriate occlusal loading. TRI-matrix® X-Force implants with lengths of 18, 20 or 22 mm, when placed in the maxilla, are indicated only for multiple unit restorations in splinted applications that utilize at least two implants.
### SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to obtain marketing clearance for TRI-matrix X-Force implants, a line extension of the TRI-matrix implant line previously cleared in K203660. TRI-matrix X-Force implants are tissue-level endosseous dental implants that are an adaptation of the TRI-matrix Tissue Level implants cleared in K203660, with the same prosthetic interface, but with modified external features. They are intended for use in the mandible or maxilla to restore chewing function. This submission includes only implants, and they are compatible with prosthetic components previously cleared in K203660 as part of the TRI-matrix implant line, including the TRI-matrix Crown Abutment.
TRI-matrix X-Force implants have a tapered body, double-lead threads and self-cutting flutes. A gritblasted and acid-etched surface, named the TRI SBA Surface, is applied to the endosseous portion of the implant. TRI-matrix X-Force implants incorporate a pink anodized, machined, 1.8 mm transgingival collar that is not treated with the TRI SBA Surface. Compared with TRI-matrix implants previously cleared in K203660, TRI-matrix X-Force implants have a modified thread design more suited to immediate placement and restoration, a more tapered apical portion (smaller apical diameter for a given coronal diameter), and more aggressive apical cutting flutes.
TRI-matrix X-Force 3.3 mm diameter implants have a platform diameter of 3.7 mm. TRI-matrix X-Force 3.7 mm and 4.1 mm diameter implants are available in two (2) platform diameters: 3.7 mm and 4.5 mm. TRI-matrix X-Force 4.7 mm diameter and 5.7 mm diameter implants have a platform diameter of 4.5 mm. The combinations of subject device implant body and platform diameters and implant lengths are summarized in Table 1 Summary of Subject Device TRI-matrix X-Force Implant Sizes.
| Body<br>Ø,<br>mm | Platform Ø,<br>(Implant/Abutment<br>connection),<br>mm | Length, mm | | | | | | | |
|------------------|--------------------------------------------------------|------------|----|----|----|----|----|----|----|
| 3.3 | 3.7 | 8 | 10 | 12 | 14 | 16 | | | |
| 3.7 | 3.7, 4.5 | 8 | 10 | 12 | 14 | 16 | 18 | 20 | 22 |
| 4.1 | 3.7, 4.5 | 8 | 10 | 12 | 14 | 16 | 18 | 20 | 22 |
| 4.7 | 4.5 | 8 | 10 | 12 | 14 | 16 | | | |
| 5.7 | 4.5 | 8 | 10 | 12 | 14 | 16 | | | |
Table 1. Summary of Subject Device TRI-matrix X-Force Implant Sizes
{6}------------------------------------------------
#### PERFORMANCE DATA
Non-clinical data submitted to demonstrate substantial equivalence included:
- Referenced from the primary predicate device K203660:
- sterilization validation according to ISO 17665-1 and ISO 17665-2; .
Provided in this submission:
- biocompatibility evaluation according to ISO 10993-5 and ISO 10993-12 ●
- worst-case analysis and static and dynamic testing according to ISO 14801 on subject devices to . demonstrate that static and dynamic testing referenced from the primary predicate device K203660 demonstrates the subject implants do not create a new worst-case as compared to this predicate device with identical connection platform.
No clinical data were included in this submission.
#### EQUIVALENCE TO MARKETED DEVICES
The subject device is substantially equivalent in indications and design principles to the primary predicate device and the reference devices listed above. Provided at the end of this summary are tables comparing the Indications for Use Statements and the technological characteristics of the subject device, the primary predicate device, and reference devices.
The Indications for Use Statement (IFUS) for the subject device is substantially equivalent to that of the primary predicate K203660. Slight differences in the language of the subject device and primary predicate device IFUS do not affect the intended use as an endosseous dental implant for support of a prosthesis to restore chewing function. Minor differences between the IFUS for the subject device and the primary predicate include the specific device names and special requirements for implants with lengths of 18 mm, 20 mm and 22 mm. Substantial equivalence of the latter is supported by the reference device K212785. The minor differences do not impact substantial equivalence both IFUS express equivalent intended use to facilitate dental prosthetic restorations and are expressed equivalently using different specific wording.
The designs of the subject TRI-matrix X-Force implants, the primary predicate implants (K203660) and implants of reference device K151916 also are substantially equivalent in terms of endosseous thread form designs, prosthesis attachment (screw-retained), restorations (single-unit or multi-unit), abutmentimplant interface (internal connection), material, and surface treatment. The ranges of dimensions of body diameter, platform diameter and endosseous length, for the TRI-matrix X-Force implants are within the ranges of the primary predicate device K203660 and the reference device K151916 except that the subject device includes longer implants (18 mm, 20 mm and 22 mm) for Ø 3.7 and Ø 4.1, and one shorter implant (8 mm) for Ø 3.3. In both cases, suitable reference devices have been identified. Long implants are substantially equivalent in diameter and length, and in Indications for Use, to reference device K212785. The range of implant lengths for Ø 3.3 subject device implants is substantially equivalent to that of Straumann TLC implants cleared in reference device K230108.
The subject device implants are manufactured from titanium alloy conforming to ASTM F136, using the identical material and processing used for titanium alloy components cleared in the primary predicate K203660 and the reference device K151916. They are grit-blasted and acid-etched, as are the implants of the primary predicate K203660 and the reference device K151916. The subject device implants have a pink collar that is anodized using a standard anodization process that is identical to the anodization process used on tissue level implants cleared in K203660.
All subject device implants are provided sterile by gamma irradiation, the identical sterilization method used for sterile components in K203660.
Risks associated with the subject device implant designs are mitigated by engineering analysis demonstrating that they are not a new worst-case relative to implants of the primary predicate K203660
{7}------------------------------------------------
510(k) Summary K242661
and by mechanical testing of implants and prosthetic components provided in in K203660. The mechanical testing data and engineering rationale demonstrate that the subject device implants in combination with compatible TRI Dental Implants angled abutments have sufficient strength for their intended use.
Minor differences in the designs, dimensions or sizes among the subject device, the primary predicate device and the reference devices do not affect substantial equivalence. These minor differences do not impact safety or effectiveness because these differences are mitigated by the engineering analysis.
#### CONCLUSION
The subject device, the primary predicate device, and the reference devices have the same intended use, have similar technological characteristics, and are materials. The subject device and the reference devices encompass the same range of physical dimensions, are packaged in the same materials, and are sterilized using the same methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
{8}------------------------------------------------
| | Indications for Use Statement |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device | |
| TRI-matrix® X-Force<br>Implant Line<br>TRI Dental Implants<br>Int. AG | TRI-matrix® X-Force implants are intended for placement in the bone of the maxillary<br>or mandibular arch for the rehabilitation of edentulous and partially edentulous<br>patients. TRI-matrix® X-Force implants allow for one and two stage surgical<br>procedures. When a one-stage surgical procedure is applied, the implant may be<br>immediately loaded when good primary stability is achieved and with appropriate<br>occlusal loading.. TRI-matrix® X-Force implants with lengths of 18, 20, 22, or 24 mm,<br>when placed in the maxilla, are indicated only for multiple unit restorations in splinted<br>applications that utilize at least two implants. |
| Primary Predicate Device | |
| K203660<br>TRI-matrix® Implant<br>Line<br>TRI Dental Implants<br>Int. AG | The TRI-matrix® Implant Line is intended for placement in the bone of the maxillary<br>or mandibular arch for the rehabilitation of edentulous and partially edentulous<br>patients. TRI-matrix® Implant Line allows for one and two-stage surgical procedures.<br>When a one-stage surgical procedure is applied, the implant may be immediately<br>loaded when good primary stability is achieved and with appropriate occlusal loading.<br>TRI-matrix® Implant Line 6.5 mm implants are intended for delayed loading only. |
| | The TRI-matrix® Crown Abutment is intended for use with an endosseous implant to<br>function as a substructure that also serves as the final restoration, in a partially or<br>completely edentulous patient. The abutment screw is intended to secure the TRI-<br>matrix® Crown Abutment to the endosseous implant. |
| Reference Device | |
| K151916<br>TRI® Dental Implant<br>System<br>TRI Dental Implants<br>Int. AG | The TRI® Dental Implant System is intended for placement in the bone of the<br>maxillary or mandibular arch for the rehabilitation of edentulous and partially<br>edentulous patients. TRI® Dental Implant System allows for one and two-stage surgical<br>procedures. When a one-stage surgical procedure is applied, the implant may be<br>immediately loaded when good primary stability is achieved and with appropriate<br>occlusal loading.<br>TRI Dental Implant System 6.5 mm implants are intended for delayed loading only. |
| Reference Device | |
| K212785<br>Blue Sky Bio Implant<br>System<br>Blue Sky Bio, LLC | Blue Sky Bio Long Implant System is intended for surgical placement in the bone of<br>the upper jaw to provide support for prosthetic devices, such as artificial teeth, to<br>restore chewing function. Implants may be used with single-stage or two-stage<br>procedures, and may be loaded immediately when good primary stability is achieved<br>and with appropriate occlusal loading. Blue Sky Bio Long implants can be placed<br>bicortically in cases of reduced bone density. Blue Sky Bio Long implants are only<br>indicated for multiple unit restorations in splinted applications that utilize at least two<br>implants. Blue Sky Bio Long Implant System with a 45° angulation are indicated for<br>surgical installation in the pterygoid region only, in cases of severe jaw resorption, in<br>order to restore patient esthetics and chewing function. |
### Table 2. Substantial Equivalence – Indications for Use Statement
{9}------------------------------------------------
| Reference Device | | | | | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|
| K230108<br>Straumann® BLC<br>and TLC Implants<br><br>Straumann USA,<br>LLC | Straumann® dental implants are indicated for the functional and esthetic oral<br>rehabilitation of the upper or lower jaw of edentulous or partially edentulous patients.<br>They can be used for immediate, early or late implantation following the extraction or<br>loss of natural teeth. The implants can be placed with immediate function for single-<br>tooth and/or multiple-tooth restorations when good primary stability is achieved and<br>with appropriate occlusal loading to restore chewing function. | | | | | | |
| Reference Device | | | | | | | |
| K192221<br>Legacy 2, Legacy3,<br>Legacy4,<br>simplyLegacy2,<br>simplyLegacy3 dental<br>implants;<br>Legacy2, Legacy3,<br>Legacy4 fixture-<br>mounts<br><br>Implant Direct<br>Sybron<br>Manufacturing, LLC | Legacy2, simplyLegacy2, Legacy3, simplyLegacy3, and Legacy4 dental implants are<br>two-piece implants for one-stage or two-stage surgical procedures. These implants are<br>intended for use in partially and fully edentulous upper and lower jaws in support of<br>single or multiple-unit restorations and terminal or intermediate abutment support for<br>fixed bridgework.<br><br>Implants can be indicated for immediate loading when good primary stability has been<br>achieved and with appropriate occlusal loading. Narrow (3.2mmD) implants: Indicated for single-tooth replacement<br>(mandibular central and lateral incisors; maxillary lateral incisors), multiple-<br>tooth replacements or denture stabilization. Short (<10mm) 3.7mm implants: Indicated for single-tooth (mandibular and<br>maxillary central and lateral incisors), multiple tooth replacements or denture<br>stabilization. The Legacy 2, Legacy3, and Legacy4 fixture-mounts are intended for use with the<br>corresponding dental implants (Legacy2, Legacy3, and Legacy4, respectively). The<br>fixture-mounts can function as an abutment. As an abutment, fixture-mounts are<br>intended for use with dental implants in the maxillary and/or mandibular arches to<br>provide support for crowns or bridges for edentulous or partially edentulous patients. Fixture-mounts as an abutment for narrow (3.2mmD) implants: Indicated for<br>single-tooth replacement of mandibular central and lateral incisors and<br>maxillary lateral incisors. Fixture-mounts as an abutment for short (8mm) 3.7mmD implants: Indicated<br>for tooth replacement of mandibular and maxillary central and lateral incisors. Legacy2, simplyLegacy2, Legacy3, simplyLegacy3, and Legacy4 implants are<br>compatible with the following abutments. | | | | | | |
| | Manufacturer Abutment Line Platform Diameter Implant Direct Legacy 3.0mm, 3.5mm, 4.5mm, 5.7mm | | | | | | |
{10}------------------------------------------------
#### 510(k) Summary K242661
#### TRI-matrix® X-Force Implants Page 7 of 8
## Table 3. Substantial Equivalence – Technological Characteristics
| Comparison | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| | | K203660 | K151916 | K212785 | K230108 |
| | TRI-matrix®<br>X-Force Implant Line | TRI-matrix® Implant Line | TRI® Dental Implant System | Blue Sky Bio Implant System | Straumann® BLC and TLC Implants |
| | TRI Dental Implants Int. AG | TRI Dental Implants Int. AG | TRI Dental Implants Int. AG | Blue Sky Bio, LLC | Straumann USA, LLC |
| Reason for Predicate /<br>Reference Device | Not applicable | Implant designs, prosthetic<br>interface | Implant designs | Long implants | Range of Ø 3.3 mm implant lengths |
| Implant Design | | | | | |
| Tissue Level<br>Body Diameters x Endosseous<br>Lengths (mm)<br>(Also see Table 4) | TRI-matrix® X-Force<br>Ø 3.3 x 8, 10, 12, 14, 16<br>Ø 3.7 x 8, 10, 12, 14, 16, 18, 20, 22<br>Ø 4.1 x 8, 10, 12, 14, 16, 18, 20, 22<br>Ø 4.7 x 8, 10, 12, 14, 16<br>Ø 5.7 x 8, 10, 12, 14, 16 | TRI-matrix® Tissue-Level<br>Ø 3.3 x 10, 11.5, 13<br>Ø 3.75 x 8, 10, 11.5, 13<br>Ø 4.1 x 6.5, 8, 10, 11.5, 13<br>Ø 4.7 x 6.5, 8, 10, 11.5, 13 | TRI Octa<br>Ø 3.75 x 8, 10, 11.5, 13<br>Ø 4.1 x 6.5, 8, 10, 11.5, 13<br>Ø 4.7 x 6.5, 8, 10, 11.5, 13 | | Straumann TLC Implants<br>Ø 3.3 x 8, 10, 12, 14, 16, 18 |
| Platform Diameters (mm) or names | 3.7 (for 3.3, 3.7, and 4.1 mm body)<br>4.5 (for 3.7, 4.1, 4.7, and 5.7 mm body) | 3.7 (for 3.3, 3.75, and 4.1 mm body)<br>4.5 (for 3.75, 4.1, and 4.7 mm body) | 4.8 | | NT/RT |
| Bone Level<br>Body Diameters x Lengths (mm) | | TRI-matrix® Bone-Level<br>Ø 3.75 x 8, 10, 11.5, 13, 16<br>Ø 4.1 x 6.5, 8, 10, 11.5, 13, 16<br>Ø 4.7 x 6.5, 8, 10, 11.5, 13, 16 | TRI-Vent, TRI Narrow<br>Ø 3.3 x 10, 11.5, 13, 16<br>(TRI-Narrow)<br>Ø 3.75 x 8, 10, 11.5, 13, 16<br>Ø 4.1 x 6.5, 8, 10, 11.5, 13, 16<br>Ø 4.7 x 6.5, 8, 10, 11.5, 13, 16 | Blue Sky Bio Long Implant<br>System<br>Ø 3.7 x 20, 22.5, 25<br>Ø 4.3 x 20, 22.5, 25<br>Ø 5.0 x 20, 22.5, 25 | |
| Platform Diameters (mm) | | 3.7 (for 3.75 and 4.1 mm body)<br>4.5 (for 4.7 mm body) | 3.2 (for 3.3 mm body TRI-Narrow)<br>3.5 (for 3.75, 4.1, and 4.7 mm body) | 3.5 | |
| Connection | Internal | Internal | Internal | Internal, External | Internal |
| Material | | | | | |
| Implant | Titanium alloy (Ti-6Al-4V ELI) | Titanium alloy (Ti-6Al-4V ELI) | Titanium alloy (Ti-6Al-4V ELI) | Titanium alloy (Ti-6Al-4V ELI) | Titanium-13 Zirconium alloy |
| Surface | Grit-blasted and acid-etched;<br>Pink anodized collar | Grit-blasted and acid-etched;<br>Pink anodized collar | Grit-blasted and acid-etched;<br>Pink anodized collar | Not stated | SLActive® and<br>SLA® |
| How Provided | | | | | |
| Implant | Sterile by gamma irradiation | Sterile by gamma irradiation | Sterile by gamma irradiation | Sterile by gamma irradiation | Sterile by gamma irradiation |
| Usage | Single patient, single use | Single patient, single use | Single patient, single use | Single patient, single use | Single patient, single use |
{11}------------------------------------------------
| Table 3. Substantial Equivalence – Technological Characteristics (continued) | | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison | Subject Device | Reference Device |
| | TRI-matrix®<br>X-Force Implant Line | K192221<br>Legacy2, Legacy3, Legacy4, simplyLegacy2,<br>simplyLegacy3 dental implants; Legacy2,<br>Legacy3, Legacy4 fixture-mounts |
| | TRI Dental Implants Int. AG | Implant Direct Sybron Manufacturing, LLC |
| Reason for Predicate / Reference Device | Not applicable | Implant diameters and lengths |
| Implant Design | | |
| Tissue Level<br>Body Diameters x<br>Endosseous Lengths (mm)<br>(Also see Table 4) | TRI-matrix® X-Force<br>Ø 3.3 x 8, 10, 12, 14, 16<br>Ø 3.7 x 8, 10, 12, 14, 16, 18, 20, 22<br>Ø 4.1 x 8, 10, 12, 14, 16, 18, 20, 22<br>Ø 4.7 x 8, 10, 12, 14, 16<br>Ø 5.7 x 8, 10, 12, 14, 16 | |
| Platform Diameters (mm) or names | 3.7 (for 3.3, 3.7, and 4.1 mm body)<br>4.5 (for 3.7, 4.1, 4.7, and 5.7 mm body) | |
| Bone Level<br>Body Diameters x<br>Lengths (mm) | | 3.2 x 8, 10, 11.5, 13, 16<br>3.7 x 6, 8, 10, 11.5, 13, 16<br>4.2 x 6, 8, 10, 11.5, 13, 16<br>4.7 x 6, 8, 10, 11.5, 13, 16<br>5.2 x 6, 8, 10, 11.5, 13, 16<br>5.7 x 6, 8, 10, 11.5, 13, 16<br>7.0 x 6, 8, 10, 11.5, 13 |
| Platform Diameters (mm) | | 3.0, 3.5, 4.5, 5.7 |
| Connection | Internal | Internal |
| Material | | |
| Implant | Titanium alloy (Ti-6Al-4V ELI) | Titanium alloy (Ti-6Al-4V ELI) |
| Surface | Grit-blasted and acid-etched;<br>Pink anodized collar | Grit-blasted and acid-etched; |
| How Provided | | |
| Implant | Sterile by gamma irradiation | Sterile by gamma irradiation |
| Usage | Single patient, single use | Single patient, single use |
| | | Table 3. Substantial Equivalence - Technological Characteristics (continued) | |
|--|--|------------------------------------------------------------------------------|--|
| | | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.